Spectrally resolved thermal emission of atmospheric gases measured by laser heterodyne spectrometry

Open Access Open Access

Abstract

The demonstration of thermal infrared quantum cascade laser heterodyne spectrometry to resolve local thermodynamic equilibrium molecular emission lines from earth’s atmospheric constituents is presented. The instrument is described, as well as the early steps towards radiometric calibration. Room temperature ethylene emission line measurements carried out in the laboratory are used to validate the instrument. High-resolution (0.02cm1) emission lines from atmospheric carbon dioxide and water vapor are then recorded in a zenith-looking configuration and compared to a radiative transfer model and ideal instrument model expectations.

Published by The Optical Society under the terms of the Creative Commons Attribution 4.0 License. Further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation, and DOI.

High spectral resolution remote sounding of infrared molecular emission provides quantitative characterization of distant molecular samples through the extraction of thermo-physical information contained in spectral lineshapes such as temperature, pressure, local thermodynamic equilibrium (LTE), velocity, and electromagnetic field. This is a versatile and highly selective remote analysis method. When applied to fundamental ro-vibrational molecular bands lying in the long wave and far infrared, atmospheric molecules at LTE and ambient temperatures produce small radiance contrasts that often require large, complex, and cooled spectrometers [1] to be spectrally resolved. Relying on pure molecular emission signals for remote sounding, while challenging, does not require solar illumination, which is an important benefit, especially in the long wave infrared (LWIR), where atmospheric scattering is insignificant. For instance, while ground-based high-resolution spectrometers operating in solar occultation mode deliver high signal-to-noise ratios (SNRs) due to the intense solar background, they require active following of the sun and cannot operate under cloudy conditions or at night.

Spectrally resolved infrared emission sensing is operationally carried out with Fourier transform spectrometers (FTSs). This includes space-borne systems operating in the limb emission sounding mode such as the Envisat/MIPAS [2] and Aura/TES [3]. High-resolution FTS instruments are powerful, but large and complex. Alternatively, when high spectral resolution is needed, laser heterodyne spectro-radiometry (LHR) is an attractive technique. As spectral filtering is done in the radio-frequency (RF) domain [4], optical instrumental background radiation is limited, which is a particular advantage for the small radiance contrasts observed in the LWIR.

High-resolution LHR emission spectroscopy in the LWIR has been reported only in the context of planetary atmosphere remote sounding, predominantly for non-LTE molecular emission, where the high spectral resolution and narrow field-of-view of LHR is exploited. For example, early work demonstrated the resolution of the emission lineshape from carbon dioxide (CO2) at 10μm from which wind velocities in the upper atmosphere of Venus were deduced [5]. CO2 emission in the Mars atmosphere at 10μm [6], ammonia from the Jupiter atmosphere at 11μm [7], and ethane emission from that of Titan at 12μm [8] were also reported using LHRs.

To the best of our knowledge, high spectral resolution emission spectroscopy of the earth’s atmospheric constituents using LWIR LHR has never been reported, despite the usefulness of this approach for ground-based zenith measurements of greenhouse gases, pollutants, and tracers during day and night, and independently of the sun position. In this Letter, we report on the development, demonstration, and assessment of such an LHR for high-resolution LTE molecular thermal emission spectroscopy, and demonstrate a spectrally resolved atmospheric signal from CO2 and water vapor (H2O). First, the implementation of the breadboard instrument is described, as well as the rationale behind the design choices. Secondly, a first iteration of instrumental calibration is sketched out since, in contrast to transmission measurements, emission spectroscopy requires radiometric calibration to infer quantitative information from signals. Thirdly, the instrument demonstrator is validated by measuring room temperature emission spectra of ethylene (C2H4) contained in a gas cell. Lastly, the instrument is coupled to the atmosphere in a zenith-looking configuration to demonstrate clear sky atmospheric emission spectroscopy of H2O and CO2.

The instrument developed for emission spectroscopy is inherited from a quantum cascade laser (QCL) LHR operating in solar occultation mode, which has been fully described elsewhere [9]. However, it has been modified to meet the demands of LTE atmospheric emission spectroscopy.

The optical layout is shown in Fig. 1. The local oscillator (LO) optical path is depicted in red. The LO is a QCL operating at 953.1cm1, whose optical radiation is collimated (CL), attenuated (ATT), expanded, and re-imaged onto a liquid N2-cooled, fast (1 GHz) HgCdTe photodiode (PD1). PD1 has not been optimized for the specific wavelength used in this demonstration, and its heterodyne efficiency η has been estimated to be only 26%. Laser light transmitted through the beam splitter (BS) goes through a 3″ long etalon towards a second photodiode (PD2) for relative LO frequency calibration. The LO can be continuously tuned from 952 to 956cm1 by both temperature and current, though in normal operation the laser temperature is kept constant, and the frequency is tuned over 1cm1 by a current ramp applied on top of the fixed injection current. The current tuning also modulates the LO optical power, which needs accounting for in the calibration.

 figure: Fig. 1.

Fig. 1. Schematic of the LHR optical layout for emission sounding. The optical path of the LO (QCL) is shown in red; the clear sky radiation is shown in blue, and the reference (BB1), background (BB2), and calibration blackbodies (BB3) are shown in green. PD1, photomixer; PD2, etalon photodiode; BS, beam splitter; MOD, modulation plane; ATT, attenuator; CL, collimating lens; OAPM/OAEM, off-axis paraboloidal/elliptical mirror; EFL, effective focal length; FM, flip mirror.

Download Full Size | PDF

The signal path is depicted in blue. In normal operation, the radiation comes from the atmosphere and is focused onto the plane of a chopper (MOD) and re-imaged onto PD1 to be superimposed onto the LO field.

Compared to the LHR developed for atmospheric transmission spectroscopy in solar occultation, new additions included are depicted in green. A blackbody (BB1) provides a cold reference against which the signals are measured. BB1 is a small cavity mounted onto the cold finger of a miniature Stirling cooler (Ricor K508). The temperature is measured by Pt100 platinum resistances to be Tb6070K, which ensures negligible radiance emission compared to the instrument noise floor. The image of the BB1 cavity is superimposed to that of the source at the plane of the gold-coated reflective chopper. In this way, the instrument measures the brightness contrast between BB1 and the scene using phase-sensitive detection.

A gas cell path for controlled LTE molecular emission spectroscopy was also added. To simulate a “deep space” background representative of a transparent atmosphere, BB2, identical to BB1, has been added behind the gas cell.

Lastly, the calibration target (BB3) is a further PID-controlled blackbody (Isotech Hyperion R), operated at temperatures spanning 263.15–293.15 K, hence covering a range representative of the lower troposphere.

Throughout the instrument, off-axis paraboloidal and elliptical gold-coated mirrors (OAPM and OAEM) have been used and aligned using the back-propagated LO methodology. The system optical efficiency κ from BB2 or BB3 to PD1 is estimated as 67% from the optics specifications.

The RF processing chain after PD1 passes the intermediate frequency (IF) signal through an RF filter (50–350 MHz bandpass), which defines the double sideband bandwidth B and spectral resolution as 600 MHz. The RF power is measured by a Shottky diode, whose output voltage is fed into a lock-in amplifier, which demodulates the heterodyne voltage signal Uhet.

To inform the instrument calibration, a theoretical model of the LHR was considered [10,11] and adapted to the present case. The IF signal is produced by PD1; it passes through a low-noise pre-amplifier (input impedance Z=50Ω, gain G1=36dB), followed by a second-stage amplification (gain G2=35dB). The IF signal power is filtered by the passband filter and converted into voltage by the Shottky diode (responsivity of ksd1200V/W). The heterodyne voltage measured is expressed by Eq. (1), where e is the elementary charge, h is Planck’s constant, ν is the photon frequency, and Ps and PLO are the optical powers contributing to the heterodyne mixing, respectively, coming from the source and the LO:

Uhet=2ksd10G1/1010G2/10Z·(ηehν)2PSPLO.
PLO is indirectly measured by the DC voltage component Udc from PD1, which has been calibrated using a power meter such that PLO=(4.94±0.03)104×Udc(3.3±0.2)105. The radiant contrast flux PS received by the detector is expressed by Eq. (2), where integration over the receiver coherent etendue and bandwidth has been accounted for:
PS=12λ2B·(κtgt·Lν(Ttgt)κref·Lν(Tref)).
Lν is the spectral radiance at frequency ν,Tref refers to the cavity temperature of BB1, and Ttgt refers to that of the target blackbody in the field-of-view. κtgt and κref refer to the optical efficiencies of the target and reference optical arm, respectively. Here, for all practical purposes, the second term on the RHS of Eq. (2) is negligible.

Equations (1) and (2) indicate that at a given wavelength, Uhet is a bilinear function of the source spectral radiance and LO power. Therefore, the effect of the LO power modulation introduced by its continuous frequency tuning also needs to be calibrated.

Radiometric calibration is performed with the dedicated optical arm using BB3 and retaining the ideal physical model from Eq. (1). Heterodyne voltages are recorded as a function of PLO and PS. A subset of the calibration data is shown in Fig. 2, obtained while continuously tuning the LO for four different BB3 temperatures. For each measurement, 10 scans with an integration time per point of 50 ms were recorded, and the in-phase and quadrature components were averaged separately before calculating the signal magnitude. The 1cm1 frequency tuning of the LO does not yield significant spectral radiance variation, so Lν is assumed to be constant across the LO scan.

 figure: Fig. 2.

Fig. 2. Extract from the radiometric calibration data showing the dependence of the heterodyne signal on LO power for different source brightness temperatures. The low LO power data were disregarded as biased. Two linear fitting models were tested: with (solid lines) and without ordinate intercepts (dashed lines).

Download Full Size | PDF

The linear fits to the experimental data are also shown in Fig. 2. For these, the least reliable data where PLO<90μW have been discarded, since they are hardly discernable from the noise floor, especially at the lowest BB3 temperature. The theoretical ideal SNR can be calculated using equations provided in Ref. [11]. The values are 5 and 9 at Ttgt of 263.15 and 293.15 K, respectively, at the experimental conditions of Fig. 2. Using a hot blackbody (1323.15 K), the instrument was established to operate to within two times the ultimate shot noise limit at optimum PLO levels (200μW). The use of magnitude signals from the lock-in amplifier also produces a bias on near-zero signals, as negative signal fluctuations are rendered positive when the magnitude is calculated.

The averaged bilinear proportionality constant between Uhet and the product PS×PLO derived from the calibration measurements is found to be (6.5±0.3)×1012V/W2. In comparison, the theoretical value calculated from Eq. (1) is found to be (7.3±1.3)×1012V/W2. Uncertainties on the calculated value are driven by the large uncertainties on amplifier gains and heterodyne efficiency.

For this first demonstration, the calibration model has been kept ideal. A more sophisticated model informed by further experimental analysis would be needed to fully describe the Uhet dependence and improve the calibration accuracy. An updated model would include effects such as background RF power in the signal or residual emission contributions from optics, their variations with LO power, and non-linearities. Evidence of additional effects can be seen in Fig. 2 for low levels of signal power, where data are better represented when adding a constant parameter to the linear dependence (green dashed lines). This is the scope for follow-on detailed investigations.

The ability of the LHR to spectrally resolve LTE molecular emission was first analyzed in the laboratory using gas cell analogs. Through modeling, low pressure C2H4 was found to be the most suitable test molecule within the LO tuning range. A glass cell of 10 cm length was filled with C2H4 at a total pressure of 20.0±2.0mbar. The following experiments were conducted:

  • – The cell was placed in the LO path to record a high SNR absorption spectrum using direct tunable laser absorption spectroscopy (TLAS) for reference. An averaged spectrum is shown in Fig. 3(a).
  • – The cell was placed as indicated in Fig. 1, and BB2 was set to a high temperature (1323.15 K) to produce a strong background against which heterodyne transmission spectroscopy could be done. This produces good SNR spectra showing the influence of the instrument lineshape of the spectrometer, as shown in Fig. 3(b). Ten spectra acquired over 90 s each were averaged.
  • – Lastly, BB2 was replaced to provide a cold background (80 K) against which the LTE emission of ethylene at room temperature could be measured. This spectrum appears in Fig. 3(c). Ninety-nine spectra were averaged for a total acquisition time of 149 min.

 figure: Fig. 3.

Fig. 3. Experimental C2H4 gas cell spectra (black crosses) with model fitting (red lines) and residuals (black crosses, in separate panels). From top to bottom: (a) direct TLAS, (b) heterodyne transmission spectroscopy against a 1323K blackbody, (c) LTE emission heterodyne spectroscopy against an 80 K blackbody.

Download Full Size | PDF

All spectra were frequency calibrated using the transmission signals from the 3″ long germanium etalon. The absolute frequency calibration was provided by the ethylene frequency transitions taken from HITRAN 2012. A gas cell transmission model was fitted to the TLAS spectra using optimal estimation inversion. In Fig. 3(a), the model trace is shown in red. The residual from the fit is shown in the lower panel. The total ethylene pressure retrieved was 22.80±0.05mbar. The quoted errors on total pressure only account for random instrument noise propagation. However, there is also a clear model error as the model fails to reproduce all the molecular transitions appearing in the experimental spectrum: either ethylene transitions are missing in the HITRAN2012 database or interfering molecules were present in the cylinder mixture. Given the mixture specifications, the low pressure and short absorption path length, the latter explanation is less likely.

The heterodyne transmission spectrum in Fig. 3(b) was fitted similarly to the TLAS one, but with the addition of the instrument lineshape convolution. The unaccounted for lines create broader mismatch structures in the residual. The returned ethylene total pressure was 25.5±0.02mbar.

The emission spectrum in Fig. 3(c) was calibrated in spectral radiance and fitted using a simple isothermal radiative transfer model given in Eq. (3):

Lν=Lν,bkgeα(ν)lcellAbb+Lν,cell·(1eα(ν)lcellAbb),
where Lν,bkg is the Planck spectral radiance from BB2 (T=80K), and Lν,cell is that of the cell (Tcell293K); lcell is the cell length; and α(ν) is the spectral absorption coefficient derived from HITRAN2012 molecular data. Abb is an additional broadband absorption term found to be necessary to account for system losses not specifically captured by the model and for an additional baseline bias induced by laser excess noise. When operating in emission mode to capture molecular LTE signals, the contribution from in-quadrature laser excess noise becomes significant against emission signals. As a result, it adds a bias to the demodulated magnitude. This bias was also found to be slowly varying in time and will need to be better characterized and accounted for in subsequent developments of the instrument.

The fitted emission data returned a pressure of 19.5 ± 0.2 mbar and a broadband absorption contribution of 0.208 ± 0.002. The pressure retrieved is comparable to the experimental value, despite the systematic, slightly varying bias to the baseline.

The emission measurements performed on the C2H4 gas cell analog validated the tunable QCL LHR in the laboratory, as well as the overall approach to signal calibration. Laser noise related background issues were also identified as needing further investigation. The next step was to couple the instrument field-of-view to the clear sky and demonstrate the first measurement of spectrally resolved atmospheric emission lines of CO2 and H2O from the ground, in a zenith viewing configuration.

With the same operating conditions, a set of spectra was collected over almost 1 h in mid-latitude summertime clear sky conditions. Each spectrum was acquired over 90 s, and Fig. 4 shows the averaged set. The heterodyne signal was calibrated into spectral radiance using Eqs. (1) and (2). A significant residual second-order polynomial baseline was subtracted, and the constant offset of the resulting spectrum was adjusted. The experimental data are shown as black crosses, a synthetic observation emulating measurement conditions as a blue line, and the same model with noise four times above the shot noise limit as red crosses. The model was produced using an observing system simulator [9] with a constant CO2 volume mixing ratio profile of 400 ppmv, and H2O, temperature and pressure profiles from the closest available radiosonde station (Larkhill, no 3743, sonde released 23/08/2016 09:00 UTC).

 figure: Fig. 4.

Fig. 4. Atmospheric emission spectra, measured with zenith viewing on 23/08/2016 between 13:12 and 14:08 at the Rutherford Appleton Laboratory (51.57°N, 01.31°W). Experimental spectral radiance data (black crosses) were calibrated, and a baseline was compensated for. A simulated observation is shown in blue, and a model with noise exceeding the shot noise by a factor of 4 is shown as red dots.

Download Full Size | PDF

The qualitative agreement between the traces is fair. The varying background observed during laboratory validation measurements [Fig. 3(c)] also shows in the atmospheric measurements. With noise levels 10 times above the shot noise limit, the instrument performs within acceptable limits, especially since the LO was observed to exhibit significant excess noise, compared to other QCL-based LHRs tested and operated in the mid-infrared.

The results demonstrate well the first measurement of spectrally resolved LTE emission from atmospheric constituents using a thermal infrared QCL-based LHR, as well as the first steps in producing calibrated spectral radiance data. These results are not yet sufficiently robust quantitatively to obtain meaningful geophysical data, and more work is required to develop this early proof-of-principle study to operational readiness. In particular, given the long integration times, both instrument and calibration stabilities require improvement.

Vertical sensitivity analysis suggests that such an instrument can provide information on the lower troposphere for CO2 and H2O, and up to two independent pieces of information can be resolved (degrees of freedom for signal). This demonstration opens the path to using thermal infrared LHR for the characterization of earth’s atmosphere when direct sun occultation is not practical, for instance, for night time sounding.

The LHR performance can be further improved through spectral multiplexing, as used in astronomy [12], rather than sequential channel acquisition (LO scanning), as demonstrated with this early prototype. Implementing spatial multiplexing to obviate the limit of the single spatial mode field-of-view is another way to improve the instrument SNR, and both enhancement methods will be considered in the future following this demonstration Letter.

Funding

UK Space Agency (RP10G0348B201); Natural Environment Research Council (NERC) (NE/P003230/1).

REFERENCES

1. P. F. Bernath, Phys. Chem. 96, 177 (2000). [CrossRef]  

2. H. Fischer, Atmos. Chem. Phys. 8, 2151 (2008). [CrossRef]  

3. R. Beer, Appl. Opt. 40, 2356 (2001). [CrossRef]  

4. D. Weidmann, Rev. Sci. Instrum. 78, 073107 (2007). [CrossRef]  

5. E. C. Betz, Astrophys. J. 208, L141 (1976). [CrossRef]  

6. M. Mumma, Science 212, 45 (1981). [CrossRef]  

7. K. Fast, Icarus 156, 485 (2002). [CrossRef]  

8. T. Kostiuk, Geophys. Res. Lett. 32, L22205 (2005). [CrossRef]  

9. A. Hoffmann, Atmos. Meas. Tech. 9, 5975 (2016). [CrossRef]  

10. T. G. Blaney, Space Sci. Rev. 17, 691 (1975). [CrossRef]  

11. D. J. Wilson, Infrared Phys. 31, 109 (1991). [CrossRef]  

12. T. Kostiuk, Infrared Phys. Technol. 35, 243 (1994). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (4)

Fig. 1.
Fig. 1. Schematic of the LHR optical layout for emission sounding. The optical path of the LO (QCL) is shown in red; the clear sky radiation is shown in blue, and the reference (BB1), background (BB2), and calibration blackbodies (BB3) are shown in green. PD1, photomixer; PD2, etalon photodiode; BS, beam splitter; MOD, modulation plane; ATT, attenuator; CL, collimating lens; OAPM/OAEM, off-axis paraboloidal/elliptical mirror; EFL, effective focal length; FM, flip mirror.
Fig. 2.
Fig. 2. Extract from the radiometric calibration data showing the dependence of the heterodyne signal on LO power for different source brightness temperatures. The low LO power data were disregarded as biased. Two linear fitting models were tested: with (solid lines) and without ordinate intercepts (dashed lines).
Fig. 3.
Fig. 3. Experimental C 2 H 4 gas cell spectra (black crosses) with model fitting (red lines) and residuals (black crosses, in separate panels). From top to bottom: (a) direct TLAS, (b) heterodyne transmission spectroscopy against a 1323 K blackbody, (c) LTE emission heterodyne spectroscopy against an 80 K blackbody.
Fig. 4.
Fig. 4. Atmospheric emission spectra, measured with zenith viewing on 23/08/2016 between 13:12 and 14:08 at the Rutherford Appleton Laboratory (51.57°N, 01.31°W). Experimental spectral radiance data (black crosses) were calibrated, and a baseline was compensated for. A simulated observation is shown in blue, and a model with noise exceeding the shot noise by a factor of 4 is shown as red dots.

Equations (3)

Equations on this page are rendered with MathJax. Learn more.

U het = 2 k sd 10 G 1 / 10 10 G 2 / 10 Z · ( η e h ν ) 2 P S P LO .
P S = 1 2 λ 2 B · ( κ tgt · L ν ( T tgt ) κ ref · L ν ( T ref ) ) .
L ν = L ν , b k g e α ( ν ) l cell A b b + L ν , cell · ( 1 e α ( ν ) l cell A b b ) ,