Abstract
As x-ray microscopy is pushed into the nanoscale with the advent of more bright and coherent x-ray sources, associated improvement in spatial resolution becomes highly vulnerable to geometrical errors and uncertainties during data collection. We address a form of error in tomography experiments, namely, the drift between projections during the tomographic scan. Our proposed method can simultaneously recover the drift, while tomographically reconstructing the specimen based on a joint iterative optimization scheme. This approach utilizes the correlation provided from different view angles and different signals. While generally applicable, we demonstrate our method on x-ray fluorescence tomography from a tissue specimen and compare the reconstruction quality with conventional methods.
© 2019 Optical Society of America
Full Article | PDF ArticleMore Like This
Esther H. R. Tsai, Federica Marone, and Manuel Guizar-Sicairos
Opt. Lett. 44(9) 2181-2184 (2019)
K. Pande, J. J. Donatelli, D. Y. Parkinson, H. Yan, and J. A. Sethian
Opt. Express 30(6) 8898-8916 (2022)
Daniel Ching, Selin Aslan, Viktor Nikitin, and Doğa Gürsoy
Opt. Lett. 44(11) 2803-2806 (2019)