Abstract

It seems to be generally conceded that the results obtained with a photoelectric colorimeter are significant only when the spectral characteristics of the sample and standard are similar. When this is the case, the selected ordinate method of integrating spectrophotometric data can be greatly abridged, a single selected ordinate being adequate under many circumstances. This means that an abridged spectrophotometer can be used for the dual purpose of production control and, when necessary, the determination of tristimulus values.

© 1949 Optical Society of America

Full Article  |  PDF Article

References

  • View by:
  • |

  1. Handbook of Colorimetry (The Technology Press, Cambridge, 1936).

Other (1)

Handbook of Colorimetry (The Technology Press, Cambridge, 1936).

Cited By

OSA participates in Crossref's Cited-By Linking service. Citing articles from OSA journals and other participating publishers are listed here.

Alert me when this article is cited.


Tables (1)

Tables Icon

Table I

Equations (13)

Equations on this page are rendered with MathJax. Learn more.

X = 0 R E x ¯ d λ , Y = 0 R E y ¯ d λ , Z = 0 R E z ¯ d λ .
X = 0 R E x ¯ Δ λ .
X = 0 R E x ¯ d λ , = 0 R E x ¯ Δ λ , = 0 R E x ¯ Δ λ .
X = 63,319 / 1042.7 ( 0.9804 ) = 59.54 .
X = ( 0.9475 ) X = ( 0.9475 ) ( 59.54 ) = 56.41 .
0 R E x ¯ Δ λ = 0 505 R E x ¯ Δ λ + 505 R E x ¯ Δ λ .
X = [ ( 1539 ) ( 1.0362 ) ( 63 , 319 ) + ( 61 , 780 ) ( 0.9474 ) ( 63 , 319 ) ] 59.54 , = 56.54 .
X = 0 T c E x ¯ d λ and X = 0 T c + Δ c E x ¯ d λ .
X = 0 [ 1 + ( Δ c / c ) ln T c + 1 / n ! ( Δ c / c ) n ( ln T c ) n + ] T c E x ¯ d λ , = X 0 + ( Δ c / c ) X 1 + ( Δ c / c ) n X n + ,
X n = 1 n ! 0 ( ln T c ) n T c E x ¯ d λ .
X = X + Δ c / c 0 ( ln T c ) T c E x ¯ d λ .
E c x ¯
0 λ + 5 R E c x ¯ Δ λ