Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group
  • Applied Spectroscopy
  • Vol. 70,
  • Issue 2,
  • pp. 363-371
  • (2016)

Possible Influences on the Interpretation of Functional Domain (FD) Near-Infrared Spectroscopy (NIRS): An Explorative Study

Not Accessible

Your library or personal account may give you access

Abstract

The influence of subcutaneous adipose tissue (ATT) and oxygen (O2) delivery has been poorly defined in frequency domain (FD) near-infrared spectroscopy (NIRS). Therefore, the aim of this study was to investigate the possible influence of these variables on all FD NIRS responses using a reliable protocol. Moreover, these influences were also investigated when using relative oxy- and deoxyhemoglobin and -myoglobin (oxy[Hb + Mb] and deoxy[Hb + Mb]) values (in %). A regression analysis was carried out for ATT and maximal–minimum oxy[Hb + Mb], deoxy[Hb + Mb], oxygen saturation (SmO2), and total hemoglobin (totHb) amplitudes during an incremental cyclic contraction protocol (ICCP) in a group of 45 participants. Moreover, the same analysis was carried out between subcutaneous ATT and the relative oxy- and deoxy[Hb + Mb] values (in %). In the second part of this study, a regression analysis was performed for peak forearm blood flow (FBF) during ICCP and the absolute and relative NIRS values in a group of 37 participants. Significant exponential correlation coefficients were found between ATT and deoxy[Hb + Mb] (r = 0.53; P < 0.001), oxy[Hb + Mb] (r = 0.57; P < 0.001), and SmO2 amplitudes (r = 0.57; P < 0.001). No significant relations were found between ATT and relative oxy[Hb + Mb] (r = 0.37; P = 0.07) and deoxy[Hb + Mb] (r = 0.09; P = 0.82). Significant positive correlation coefficients were found between force at exhaustion and maximal FBF (r = 0.66; P < 0.001), maximal differences in deoxy[Hb + Mb] (r = 0.353; P = 0.032) and totHb (r = 0.512; P = 0.002) while no significant correlation coefficients were found between these maximal force values and maximal differences in oxy[Hb + Mb] (r = −0.267; P = 0.111) and SmO2 (r = −0.267; P = 0.111). Significant linear correlation coefficients were found between FBF and deoxy[Hb + Mb] (r = 0.51; P = 0.001), oxy[Hb + Mb] (r = −0.50; P = 0.001), SmO2 (r = −0.54; P = 0.001), and totHb amplitude (r = 0.61; P < 0.001). No significant correlations were found when using relative oxy[Hb + Mb] (r = −0.01; P = 0.957) and deoxy[Hb + Mb] (r = −0.02; P = 0.895). Based on these findings, caution is advised when using NIRS values, as subcutaneous ATT and O2 delivery significantly influence NIRS measurements. To eliminate these influences, use of relative deoxy[Hb + Mb] is advised, especially in clinical settings or in people with a higher subcutaneous ATT layer.

© 2015 The Author(s)

PDF Article
More Like This
Hemodynamics of the sternocleidomastoid measured with frequency domain near-infrared spectroscopy towards non-invasive monitoring during mechanical ventilation

Raeef Istfan, Carlos A. Gómez, Matthew Applegate, Dmitry Rozenberg, W. Darlene Reid, and Darren Roblyer
Biomed. Opt. Express 12(7) 4147-4162 (2021)

Hemodynamic responses in rat brain during transcranial direct current stimulation: a functional near-infrared spectroscopy study

Chang-Hee Han, Hyuna Song, Yong-Guk Kang, Beop-Min Kim, and Chang-Hwan Im
Biomed. Opt. Express 5(6) 1812-1821 (2014)

Sustained fatigue assessment during isometric exercises with time-domain near infrared spectroscopy and surface electromyography signals

A. Scano, I. Pirovano, M. E. Manunza, L. Spinelli, D. Contini, A. Torricelli, and R. Re
Biomed. Opt. Express 11(12) 7357-7375 (2020)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.