Two thermoelectrically cooled mid-infrared distributed feedback quantum cascade lasers operated in pulsed mode have been used for the quasi-simultaneous determination of NO and NO<sub>2</sub> in the sub-parts per million meter (sub-ppm-m) range. Using a beam splitter, the beams of the two lasers were combined and sent to a retro-reflector. The returned light was recorded with a thermoelectrically cooled mercury cadmium telluride detector with a rise time of 4 ns. Alternate operation of the lasers with pulse lengths of 300 ns and a repetition rate of 66 kHz allowed quasi-simultaneous measurements. During each pulse the laser temperature increased, causing a thermal chirp of the laser line of up to 1.3 cm<sup>−1</sup>. These laser chirps were sufficient to scan rotational bands of NO centered at 1902 cm<sup>−1</sup> and NO<sub>2</sub> located at 1632 cm<sup>−1</sup>. In that way an absorption spectrum could be recorded from a single laser pulse. Currently achieved limits of detection are 600 parts per billion meter (ppb-m) for NO and 260 ppb-m for NO<sub>2</sub> using signal averaging over 1 min. This work presents the first steps toward a portable stand-off, open-path instrument that uses thermoelectrically cooled detector and lasers.

PDF Article

Cited By

OSA participates in CrossRef's Cited-By Linking service. Citing articles from OSA journals and other participating publishers are listed here.