Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group
  • Applied Spectroscopy
  • Vol. 52,
  • Issue 8,
  • pp. 1035-1038
  • (1998)

Use of 0.0 mm Pathlength Gas Cell to Eliminate CO(g) Absorbance from High-CO-Pressure Solid-State IR Spectra of Nonclassical Metal Carbonyls

Not Accessible

Your library or personal account may give you access

Abstract

A transmission-IR gas cell was constructed from 18%-Ni maraging steel so that two parallel 10 mm diameter X 3 mm thick sapphire windows were held approximately 0.1 mm apart. The cell was engineered to withstand static gas pressures up to 150 atm. When 100 atm CO was placed in the cell, the % transmittance was essentially zero between 2200 and 2080 cm-1 because of the intense upsilon (CO) band centered at 2143 cm-1 (this is an important region of interest for the growing subset of metal carbonyl complexes known as nonclassical metal carbonyls). When the cell was assembled with Nujol or Fluorolube mulls of several microcrystalline copper(I) and silver(I) salts between the windows and CO gas was added to the cell, the gas dissolved in the mulling agents and reacted with the metal salts, but no gaseous CO was observed in the spectrum. New spectral bands were observed that are attributed to unusual metal carbonyl species that are stable only under high CO pressure. A relatively weak band due to free CO dissolved in the mulling agents was detected at 2138-2135 cm-1. The new cell was used to demonstrate that the compound Ag(OTeF5) takes up only one equivalent of CO at 100 atm and that the compound Cu(CF3SO3) takes up two equivalents of CO at 136 atm.

PDF Article
More Like This
Absorption coefficients of weakly absorbing solids: theory of a barothermal gas cell

Herbert S. Bennett and Richard A. Forman
Appl. Opt. 15(2) 347-352 (1976)

Augmented sensitivity of an IR-absorption gas sensor employing a metal hole array

Yoshiaki Nishijima, Yuta Adachi, Lorenzo Rosa, and Saulius Juodkazis
Opt. Mater. Express 3(7) 968-976 (2013)

Pulsed quantum-cascade laser-based sensor for trace-gas detection of carbonyl sulfide

Gerard Wysocki, Matt McCurdy, Stephen So, Damien Weidmann, Chad Roller, Robert F. Curl, and Frank K. Tittel
Appl. Opt. 43(32) 6040-6046 (2004)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.