Abstract
We propose a new, to the best of our knowledge, compound technique to measure high-dynamic-range blood flow rate in a large-diameter vessel, which combines the dynamic scattering light (DLS) and the laser speckle contrast imaging (LSCI) methods, possessing the advantages of the high temporal resolution of DLS and the robust property of LSCI. By controlling the second-order spatial correlations of the laser speckle through two imaging systems, the speckle temporal intensity autocorrelation function ${g^2}(t)$ and the decorrelation time ${\tau _c}$ are directly measured using a high-speed camera. It turns out the enhanced spatial second-order correlation helps to measure the blood flow with higher dynamic range and that the measured parameter $\beta$ and the blood flow dynamics $n$ were accurately determined. For further improvement the dynamic range, the modified LSCI method was adopted, and the decorrelation time as a function of blood flow rate was constructed. It reveals the feasibility of measuring the high flow rate in large-diameter vessels and provides significant guidance for the future biomedical study of the myocardial perfusion in coronary artery bypass grafting, ghost imaging, and ghost cytometry.
© 2021 Optical Society of America
Full Article | PDF ArticleMore Like This
Bingxue Liu, Dmitry Postnov, David A. Boas, and Xiaojun Cheng
Biomed. Opt. Express 15(2) 579-593 (2024)
Changyoon Yi, Jaewoo Jung, Jeongmyo Im, Kyung Chul Lee, Euiheon Chung, and Seung Ah Lee
Optica 9(11) 1227-1237 (2022)
Muhammad Mohsin Qureshi, Yan Liu, Khuong Duy Mac, Minsung Kim, Abdul Mohaimen Safi, and Euiheon Chung
Optica 8(8) 1092-1101 (2021)