Abstract
The scanning-beam technique for measuring the response of a detector to an irradiance is analyzed. With this method the irradiance responsivity is determined by integration of the spatial responsivity. Since in practice the integration is approximated by a summation over steps with a finite step size, errors are introduced. It is shown both theoretically and experimentally that the error vanishes when the reciprocal step size lies beyond the diffraction limit. Furthermore, comparison shows that experiment and theory are in good agreement.
© 2000 Optical Society of America