Abstract
An improved theory for data reduction of absolute reflectance measurements using the third Taylor method in the α°/d configuration is presented. A brief description is given of an absolute reflectometer operating in the 0.8–2.5-μm region. The reflectometer is operated according to the improved theory. Experimental data for some widely used samples are given, as well as data showing agreement between the current measurements and those made by the U.S. National Institute of Standards and Technology.
© 1991 Optical Society of America
Full Article | PDF ArticleMore Like This
Dan Sheffer, Uri P. Oppenheim, and Adam D. Devir
Appl. Opt. 29(1) 129-132 (1990)
James B. Gillespie and James D. Lindberg
Appl. Opt. 31(7) 955-959 (1992)
Thomas A. Blake, Timothy J. Johnson, Russell G. Tonkyn, Brenda M. Forland, Tanya L. Myers, Carolyn S. Brauer, Yin-Fong Su, Bruce E. Bernacki, Leonard Hanssen, and Gerardo Gonzalez