Abstract
A new on-column excitation and detection geometry for multiplexed array capillary electrophoresis is evaluated. The detection limit is improved 100-fold for the same laser power because of more efficient coupling of the laser beam. The illumination is also more uniform over the entire array. Stray light and cross-talk are effectively suppressed by immersing the capillaries in water to roughly match the refractive index. A charge-coupled-device (CCD) camera combined with cutoff filters provides adequate sensitivity and speed for high-speed DNA sequencing based on sieving in a polymer-matrix solution.
PDF Article
More Like This
Fluorescence monitoring of microchip capillary electrophoresis separation with monolithically integrated waveguides
Chaitanya Dongre, Ronald Dekker, Hugo J. W. M. Hoekstra, Markus Pollnau, Rebeca Martinez-Vazquez, Roberto Osellame, Giulio Cerullo, Roberta Ramponi, Rob van Weeghel, Geert A. J. Besselink, and Hans H. van den Vlekkert
Opt. Lett. 33(21) 2503-2505 (2008)
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Login to access Optica Member Subscription