High-accuracy time-domain jitter measurement of ultrafast lasers

Not Accessible

Your library or personal account may give you access

Abstract

We describe a new method for measuring the timing jitter of modelocked lasers directly in the time-domain. By using cross-correlation between two modelocked lasers, very high accuracy is obtained. We compare the jitter results obtained with three different analysis methods, and we investigate the accuracy of the cross-correlation method. In the presence of typical noise levels, the peak fitting errors are found to vary inversely with the pulse area, giving an accuracy of less than 5 fs for the first few pulses in a cross-correlation scan.

© 1999 Optical Society of America

PDF Article
More Like This
Time-domain Jitter measurement of ultrafast lasers

G. Sucha, M. E. Fermann, and D. J. Harter
CTuO1 Conference on Lasers and Electro-Optics (CLEO:S&I) 1998

Simultaneous time-domain detection and suppression of laser timing Jitter

Jia-Min Shieh, Liang-Iao Chang, C.S. Wang, and Ci-Ling Pan
CTuK6 Conference on Lasers and Electro-Optics (CLEO:S&I) 1999

Sub-10 femtosecond timing jitter of a 10-GHz harmonically mode-locked fiber laser

T. R. Clark, T. F. Carruthers, I. N. Duling, and P. J. Matthews
PD24 Optical Fiber Communication Conference (OFC) 1999