Abstract

Coherence-restored pulse interferometry (CRPI) is a recently developed method for optical detection of ultrasound that achieves shot-noise-limited sensitivity and high dynamic range. In principle, the wideband source employed in CRPI may enable the interrogation of multiple detectors by using wavelength multiplexing. However, the noise-reduction scheme in CRPI has not been shown to be compatible with wideband operation. In this work, we introduce a new scheme for CRPI that relies on a free-space Fabry-Pérot filter for noise reduction and a pulse stretcher for reducing nonlinear effects. Using our scheme, we demonstrate that shot-noise-limited detection may be achieved for a spectral band of 80 nm and powers of up to 5 mW.

© 2019 SPIE/OSA

PDF Article
More Like This
Shot-noise-limited optical sensing of ultrasound using pulse interferometry with a free-space Fabry-Pérot

Oleg Volodarsky, Yoav Hazan, and Amir Rosenthal
SW3L.1 CLEO: Science and Innovations (CLEO_SI) 2019

Optical phase shifted pulse interferometry for parallel multichannel ultrasound detection

Yoav Hazan and Amir Rosenthal
11077_17 European Conference on Biomedical Optics (ECBO) 2019

Phase modulated pulse interferometry for simultaneous multi-channel ultrasound detection

Yoav Hazan and Amir Rosenthal
SW3L.2 CLEO: Science and Innovations (CLEO_SI) 2019

References

You do not have subscription access to this journal. Citation lists with outbound citation links are available to subscribers only. You may subscribe either as an OSA member, or as an authorized user of your institution.