Abstract
We demonstrate a fast, direct wavefront-sensing method for dynamic in vivo adaptive optical two-photon microscopy. By using a Shack–Hartmann wavefront sensor and open-loop control, the system provides high-speed wavefront measurement and correction. To measure the wavefront in the middle of a Drosophila embryo at early stages, autofluorescence from endogenous fluorophores in the yolk were used as reference guide stars. The method was tested through live imaging of a Drosophila embryo. The aberration in the middle of the embryo was measured directly for the first time. After correction, the contrast and signal intensity of the structure in the middle of the embryo was improved.
© 2013 Optical Society of America
Full Article | PDF ArticleMore Like This
Xiaodong Tao, Justin Crest, Shaila Kotadia, Oscar Azucena, Diana C. Chen, William Sullivan, and Joel Kubby
Opt. Express 20(14) 15969-15982 (2012)
Xiaodong Tao, Oscar Azucena, Min Fu, Yi Zuo, Diana C. Chen, and Joel Kubby
Opt. Lett. 36(17) 3389-3391 (2011)
Antoine Hubert, Fabrice Harms, Rémy Juvénal, Pauline Treimany, Xavier Levecq, Vincent Loriette, Georges Farkouh, François Rouyer, and Alexandra Fragola
Opt. Lett. 44(10) 2514-2517 (2019)