Abstract
Here we demonstrate an inexpensive, simple, and ultra-sensitive refractive index sensor based on a tapered tip optical fiber combined with a straightforward image analysis method. The output profile of this fiber exhibits circular fringe patterns whose intensity distribution dramatically changes even with ultra-small refractive index variations in the surrounding medium. The sensitivity of the fiber sensor is measured using different concentrations of saline solutions with a transmission setup consisting of a single wavelength light source, a cuvette, an objective lens, and a camera. By analyzing the areal changes in the center of the fringe patterns for each saline solution, we obtain an unprecedented sensitivity value of 24,160 dB/RIU (refractive index unit), which is the highest value reported so far among intensity-modulated fiber refractometers. The resolution of the sensor is calculated to be 6.9 ×10−9. Moreover, we measure the sensitivity of the fiber tip in the backreflection mode using salt-water solutions and obtained a sensitivity value of 620 dB/RIU. This sensor is ultra-sensitive, simple, easy to fabricate, and low-cost, which makes it a promising tool for on-site measurements and point-of-care applications.
© 2022 Optica Publishing Group
Full Article | PDF ArticleMore Like This
Yi-Hsin Tai and Pei-Kuen Wei
Opt. Lett. 35(7) 944-946 (2010)
Dejun Liu, Arun Kumar Mallik, Jinhui Yuan, Chongxiu Yu, Gerald Farrell, Yuliya Semenova, and Qiang Wu
Opt. Lett. 40(17) 4166-4169 (2015)
M. S. Aruna Gandhi, Yuanfang Zhao, Chenxingyu Huang, Yuan Zhang, H. Y. Fu, and Qian Li
Opt. Lett. 47(7) 1697-1700 (2022)