Abstract
Fibre-optic based time-resolved fluorescence spectroscopy (TRFS) is an advanced spectroscopy technique that generates sample-specific spectral-temporal signature, characterising variations in fluorescence in real-time. As such, it can be used to interrogate tissue autofluorescence. Recent advancements in TRFS technology, including the development of devices that simultaneously measure high-resolution spectral and temporal fluorescence, paired with novel analysis methods extracting information from these multidimensional measurements effectively, provide additional insight into the underlying autofluorescence features of a sample. This study demonstrates, using both simulated data and endogenous fluorophores measured bench-side, that the shape of the spectral fluorescence lifetime, or fluorescence lifetimes estimated over high-resolution spectral channels across a broad range, is influenced by the relative abundance of underlying fluorophores in mixed systems and their respective environment. This study, furthermore, explores the properties of the spectral fluorescence lifetime in paired lung tissue deemed either abnormal or normal by pathologists. We observe that, on average, the shape of the spectral fluorescence lifetime at multiple locations sampled on 14 abnormal lung tissue, compared to multiple locations sampled on the respective paired normal lung tissue, shows more variability; and, while not statistically significant, the average spectral fluorescence lifetime in abnormal tissue is consistently lower over every wavelength than the normal tissue.
Published by Optica Publishing Group under the terms of the Creative Commons Attribution 4.0 License. Further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation, and DOI.
Corrections
28 January 2024: A minor correction was made to an affiliation.
1. Introduction
Lung cancer, the leading cause of cancer-related deaths worldwide [1], has a diagnostic deficit. Only 16.6% of patients are diagnosed with the disease at an early stage with an anticipated 5-year survival rate of 80% [2]. This drops to 10% for diagnoses made at an advanced stage. Thus, we need new diagnostic strategies that enable early and accurate diagnoses. The development of non-surgical radiation-free diagnostic alternatives, such as Fluorescence Lifetime Imaging Microscopy (FLIM) and Time-Resolved Fluorescence Spectroscopy (TRFS) have been shown to discriminate diseased tissue and, additionally, can be available bed-side [3–5]. When used in a label-free setting, these devices produce a detailed molecular profile of tissue autofluorescence (AF), potentially facilitating a more informed and rapid diagnosis [3,6].
AF describes the fluorescence of naturally occurring molecules, i.e., endogenous fluorophores [7]. Endogenous fluorophores, due to their intrinsic molecular structures, have distinct absorption and emission profiles when measured across a suitable wavelength range [8, Fig. 1] [9, Fig. 2]. In addition to the spectral emission, the average time the fluorophore spends in an excited state, known as the fluorescence lifetime, can be measured. This value is fluorophore specific and sensitive to the environment [7]. Therefore, in a given physico-chemical environment (i.e., with a set temperature, pH, viscosity, media dielectric constant, free or enzyme-bound), a single fluorophore will exhibit a specific fluorescence lifetime and emission profile [10].
Interpreting AF in lung tissue presents challenges arising from the overlapping excitation and emission profiles, similarities in lifetimes, and the unknown environments of endogenous fluorophores [10]. Depending on the optical setup, various devices measure the emission spectrum at distinct wavelength locations, also referred to as channels [11]. Each channel has a bandwidth (with smaller bandwidths implying higher resolution) within which photons are accumulated, collectively covering a specific wavelength range. Therefore, both the range and bandwidth dictate which fluorophores the device can measure, and, in the context of tissue delineation, this variability can lead to different conclusions (see Table 1 and Supplement 1). In addition, the systems previously employed for measuring AF collected fluorescence in a limited number of parallel spectral channels. For example, the ms-TRFS device [12], utilised in detecting oral and oropharyngeal cancer [13], assessed fluorescence across 4 spectral channels with resolutions ranging from 14 nm to 26.5 nm. In these setups, the underlying signal is essentially marginalised into distinct, low-resolution spectral bins, thereby limiting the detection of subtle yet potentially significant changes in the tissue being explored. Moreover, lung cancer is often characterised by structural deregulation [14], and consequently, alterations may manifest not only in the physical-chemical changes of lung AF but also in the relative abundance of distinct endogenous fluorophores. This insight could provide additional information about the variations in lifetime between normal and abnormal lung tissue.
Development in TRFS technology has allowed for alternative devices which assess multichannel AF profiles across a large wavelength range and at a higher resolution (i.e., smaller bandwidth for individual channels) [11,24,25]. This allows for the interrogation of both the emission spectrum and spectral fluorescence lifetime (SFL), i.e., the variation of average fluorescence lifetimes in consecutive channels over a broad range. As discussed in [26] and [27, Chapter 3.2.1], given a single fluorophore, we expect SFL to be constant over wavelengths, but in the context of tissue, where multiple fluorophores are excited simultaneously, we expect SFL to vary, depending on the respective emission spectra of the underlying fluorophores and their relative prevalence in the sample (see section 3.1). The relative abundance of underlying fluorophores can also be estimated through fitting multiple exponentials [18] and spectral-temporal un-mixing. This approach, however, requires assuming that the number of underlying fluorophores are known a priori, and we consider SFL to be a proxy for this information. SFL may, therefore, provide additional discriminating information between a diseased and normal tissue sample, than when comparing the fluorescence over a single channel [28,29].
Previously, we reported on the extensively-parallel (EP-TRFS) device providing a high resolution fluorescence profile of a sample over hundreds of channels [26]. When used in time-correlated single-photon counting (TCSPC) mode, this device constructs a high-resolution histogram of fluorescence concurrently in the temporal and spectral domains. This device excites tissue at 485 nm. Notably, an excitation at 488 nm has been shown to differentiate lung cancer from normal tissue in vivo using an optical endomicroscopy (OEM) setup [30–32]. However, while OEM measures spatial fluorescence without a temporal assessment (thus restricting the fluorescence analysis to emission spectra), our approach evaluates lung AF through high-resolution spectral-temporal profiles.
In addition, a computational tool designed to interrogate the acquired high-resolution histograms, named Multichannel Fluorescence Lifetime Estimation or MuFLE, was suggested to provide detailed insights into fluorescence characteristics over multiple channels [26]. A FLIM setup with an excitation of 475 nm has also exhibited lung cancer discrimination using high-resolution spectral-temporal profiles, however, these lung profiles were assessed without a multi-channel analysis tool [24,33]. Therefore, using these recent developments, we assess the characteristics and applicability of SFL modeled from MuFLE in tissue delineation. Our assumption is that the relative abundances of endogenous fluorophores change in abnormal tissue compared to normal, due to the deregulation of tissue structure in neoplastic samples, thus, changing both the magnitude and shape of SFL, something which low resolution channel devices and conventional analysis tools are unable to capture.
Our overarching goal is to explore the utility of SFL, compared to alternative methods, i.e., Aggregated Fluorescence Lifetime (AFL) estimating fluorescence over marginalised histograms, in investigating mixed systems with multiple underlying fluorophores. Particularly in paired normal and abnormal lung tissue, e.g., in the context of delineating them. The contribution of the paper are as follows: first, we observe that the shape of the SFL is influenced by the relative abundance of underlying fluorophores when assessed in a mixed system. We show this to be true both when altering the relative abundance of simulated fluorophores and endogenous reference fluorophores while the physical-chemical environment (i.e., pH, temperature and viscosity of the solvent) remain unchanged, and despite their emission spectra remaining similar (see section. 3.1 and section. 3.2). Second, we show the shape of SFL, on average, shows more variability in abnormal tissue compared to normal tissue suggesting a more unpredictable alteration of fluorescence changes in suspected cancer tissue (see section. 3.3). However, we observe that there is considerable inter- and intra-sample heterogeneity between and within SFL of both normal and abnormal tissue in both magnitude and shape (see section. 3.3) limiting statistical significance where $n=14$ paired samples are investigated (i.e., 28 tissue sections from lobe resections or 90 spectral histograms in total). Third, we observe that although the AFL and SFL perform similarly in delineating lung tissue, the latter provides additional information that is not preserved when aggregating the fluorescence decays across all channels. To the best of our knowledge, this study is the first to explore the characteristics and utility of SFL in simulated data, bench-side fluorophores, and lung tissue.
2. Data and methods
2.1 EP-TRFS device
The EP-TRFS device is a fibre based setup which was used in time-correlated single photon counting (TCSPC) mode. In addition, the device was comprised of a pulsed laser (laser diode head (LDH-P-C-485, PicoQuant, Germany) and laser driver (PDL 800-D, PicoQuant, Germany) used at a repetition rate of 20 MHz with a complementary metal oxide semi-conductor (CMOS) single photon avalanche diode (SPAD) line sensor, as detailed in [26] (see Fig. 1 A). This allows high resolution histograms of photon arrival times measured at all channels in parallel using $1.5{\times }10^{6}$ exposures, each with a 5 $\mathrm{\mu}\textrm {s}$ exposure time and an optical laser output power of 175 $\mathrm{\mu}\textrm {W}$. Furthermore, we excite the sample such that we remain within the accepted photon detection rate of between 1 and 5% of laser pulses so as to mitigate pile-up risk [34]. Fluorescence spectroscopy measurements in both the temporal and spectral domain, across a broad spectral region between 474 nm-720 nm may be collected. The samples were excited at 485 nm, at this wavelength, as aforementioned, OEM devices have shown through exciting elastin, discrimination between normal and cancerous lungs are visible [30]. Due to the unique fibre-based setup, and spectral-temporal profile of the device, the instrument response function was measured using a quenched form of rose-bengal [35] (see section 4 for more details). Due to the narrower emission profile of rose-bengal, and the emission peaks of the endogenous fluorophores estimated to be excited by a laser of 485 nm, the total fluorescence spectral window the data was analysed in was between 557.13 nm-638.22 nm which consisted 160 individual spectral channels at a wavelength resolution of 0.5 nm.
2.2 Reference fluorophores
Three reference endogenous fluorophores: Flavin Adenine Dinucleotide (Sigma-Aldrich: F8384-100MG), Riboflavin (Sigma-Aldrich: R9504-25G) and Elastin (Sigma-Aldrich: E4527-1G) were measured on the EP-TRFS device. To standardise the environment of the fluorophores, the fluorophores were dissolved in $_{dd}$H$_2$O made up by the addition of buffer to a pH of 7. The fluorophores were collected at room temperature. Elastin was measured at a concentration of 500 $\mathrm{\mu}\textrm {M}$, FAD was measured at a concentration of 100 $\mathrm{\mu}\textrm {M}$ and riboflavin was measured at a concentration of 100 $\mathrm{\mu}\textrm {M}$. To assess the fluorescence profile of mixed endogenous fluorophores in the same environment, 3 samples, referred to as mix 1, mix 2 and mix 3, containing the endogenous fluorophores at varying concentrations, at a pH of 7 were measured (see Supplement 1 Table. S1). To compare changes of SFL in different mixes, to that of different environments, mix 2 was made up in 2 more pHs, referred to as Mix 2a and Mix 2b. Mix 2a at a pH of 4 with the addition of 1 M hydrochloric acid, and Mix 2b at a pH of 9 following the addition of sodium hydroxide. To validate the intensity profiles and excitation wavelengths of all samples, the emission (see section. 3.2) was measured using the bottom read out of a plate reader (biotek, cytation 3 imaging reader).
2.3 Ex vivo lung tissue fragments
Tissue data from ex vivo lobectomy specimens, ranging from 50 mm3 cm in diameter were taken. Samples were obtained from patients undergoing lobe resections between January 2022 to January 2023, for suspected lung cancer (NHS Lothian BioResource, Scotland Research Ethics Service, reference 15/ES/0094). 14 lung samples, paired clinically as non-cancerous and cancerous, which we denote in the paper as normal and abnormal. Samples have been assessed and defined by type, stage and age (see Table 2). The pathological profile of the tissue samples vary. Depending on the size of the tissue sample, 3-6 locations per sample were assessed using our device, with a point sampling approach (fibre core diameter of 32 $\mathrm{\mu}\textrm {m}$). At each location 3 repeated spectroscopy measurements were taken, with the same device setup as detailed in [26]. Data was collected using a 3 axis Cartesian robot [36] (see Fig. 1 B and C) to allow repeated measurements in the same X, Y and Z plane to be collected. The 3 repeated histograms were summed together before further assessment (see Fig. 1 D) to improve signal-to-noise ratio. Moreover, summed histograms with a peak intensity of below 300 were excluded from the analysis due to poor signal-to-noise.
2.4 Analysis
The data from the EP-TRFS device was analysed using two different approaches.
2.4.1 Aggregated fluorescence lifetime
To compare signal measured in the EP-TRFS device with signal from systems measuring autofluorescence in the more traditional low resolution channel method, the 160 channels histogram was aggregated into two distinct channels of 40 nm each (see Supplement 1 sec. 1.B and supplementary Fig. S1), and single lifetime, referred to as AFL, was estimated using least squared fitting. The single exponential fluorescence decay is modelled as follows:
where $\gamma$ and $\tau$ are the average intensity and fluorescence lifetime, $b$ is the bias due to the dark counts of the detector and the fibre background, $h[m]$ is the IRF at the channel, and $*$ denotes convolution. Given $y[m]$ as the observed signal at the $m$-th bin, $\gamma$ and $\tau$ can be estimated by minimising the loss function assuming $h[m]$ is known.2.4.2 Spectral fluorescence lifetime
In the second approach, each channel is analysed either separately with least squares fitting or simultaneously with MuFLE. Given multiple channels, MuFLE, simultaneously estimates the fluorescence intensity and lifetimes over these channels [26]. Here, the fluorescence decay is modelled as
2.4.3 Simulated data
Simulated data was generated to validate the hypothesis that SFL is sensitive to variation in the relative prevalence of fluorophores, even if the emission intensities are very similar to each other, i.e., without distinct peaks. A similar situation arises in the EP-TRFS system and lung tissue since the selected emission wavelength range only captures the tail end of emission spectra of the fluorophores of interest (see Fig. 2). Three hypothetical fluorophores were simulated using the model described in Eq. (3) with different intensity and lifetimes. Exponential and hyperbolic functions were used to model the emission intensities, i.e., $\gamma [p]$, of the individual fluorophores. These functions were chosen to replicate the tail of the emission profile of real fluorophores excited at 485 nm. Across the wavelength range, the lifetime was simulated to be fixed for each fluorophore, i.e., $\tau [m] = \tau$.
The temporal resolution was set to 0.05 ns to match the real data and a total of $M = {490}$ time bins were simulated. The number of channels was set to $P={160}$ to match the real data. To simulate mixed samples, the histograms of the individual fluorophores were summed together and convolved with a simulated IRF. The simulated IRF was assumed to be exponential with a lifetime of 0.378 ns, matching the average lifetime of the IRF when fitted with a single exponential decay, recorded in the EP-TRFS device. The intensity at each channel and bin were perturbed using a Poisson noise.
3. Results
3.1 Simulation
It is expected that both the shape and magnitude of the SFL profile is influenced by the relative concentration of the underlying fluorophores with different but constant lifetimes. In the simplest model, we can consider a multi-exponential decay ($y$) of two fluorophores, in the same environment, $a$ and $b$ as
where $\omega$ represents the wavelength, $t$ represents time, and $\tau$ represents lifetime. As the intensities, reflecting the individual emission spectrum of the individual fluorophores, change across the wavelengths, the average lifetime, when approximated by a single exponential will also vary over the wavelengths. When $\gamma _a(\omega ) < \gamma _b(\omega )$, the average lifetime will be closer to $\tau _b$, and when $\gamma _a(\omega ) > \gamma _b(\omega )$, the average lifetime will be closer to $\tau _a$.The 3 individual simulated fluorophores had similar spectral intensity profiles to that of the endogenous fluorophores used, i.e., they drop monotonically at longer wavelengths (see section 2.4.3). In addition, they also had similar fixed lifetime values to that of the endogenous fluorophores. We observe that the shape of SFL reflects the relative abundance of fluorophores. For example, a dip is observed in the SFL in histogram 3 just before 580 nm (see Fig. 3 D), this aligns with the wavelength range where the fluorophore corresponding to the highest lifetime is declining (see Fig. 3 C), in addition, it aligns with the wavelength range where the fluorophore with the lifetime of 2 ns has started to decline, and the fluorophore with a lifetime of 0.8 ns has started to rise (see Fig. 3 C). In both simulated histograms 1 and 2, a similar trend in the SFL shape is also observed that reflects the relative abundance of the fluorophores (see Fig. 3). In particular, the peaks seen in both SFL lifetime ranges between 560 nm-580 nm, reflect the decline in relative abundance of the fluorophore with a lifetime of 0.8 ns and increase in relative abundance of both fluorophores with a lifetime of 5 ns and 2 ns.
3.2 Endogenous fluorophores
We replicate the same experiment in the simulated data using three endogenous fluorophores bench-side. We select three endogenous fluorophores known to be excited at 485 nm in lung tissue, and are therefore also likely to be present in the tissue samples. We first validated the absorption (to confirm the excitation at 485 nm) and emission spectra of 3 endogenous fluorophores (see Fig. 2). The emission profiles from a plate reader were in agreement with the single exponential MuFLE intensity results of the individual fluorophores (see Fig. 2 A, B, C). The magnitude of SFL estimated using MuFLE was also in agreement with the ground truth values reported in the literature while the shape remained broadly constant [37,38] (see Fig. 2 D, E, F).
Following the validation, the mix of these fluorophores were analysed (see section 2.2 and Supplement 1). In all cases, similar intensity profiles of the samples were observed (see Fig. 4 ii-iv). When the environment of the sample was altered, a significant increase in the absolute lifetime value (average of 3.38 ns$\pm$0.18) of Mix 2b (Mix 2 made up to a pH of 9 as mentioned in section. 2.2) was observed, compared to when measured in Mix 2 or Mix 2a (Mix 2 made up to a pH of 4 as mentioned in section. 2.2) (with an average lifetime value of 2.99 ns$\pm$0.15 and 2.95 ns$\pm$0.15 respectively) (see Fig. 4 Ai). When the relative concentration of the individual fluorophores was altered, independent of the environment (see Supplement 1), both the absolute value of the fluorescence lifetime (average lifetime: Mix 1: 2.89 ns$\pm$0.12, Mix 2: 2.99 ns$\pm$0.15 and Mix 3: 3.33 ns$\pm$0.11), and the spectral fluorescence shape of the lifetime across the wavelength changed (see Fig. 4 Bi). The most notable spectral shape change was observed in Mix 3 which reached a plateau across the wavelength between 570 nm-610 nm, instead of a decline in value. The higher lifetime of mix 3 is expected from higher abundance of riboflavin and lower abundance of elastin compared to mix 1 and 2, while the slight lower lifetime of mix 1 compared to mix 2 is expected due to higher abundance of FAD.
3.3 Ex vivo lung tissue autofluorescence
The tissue samples were analysed both in terms of AFL and SFL. We assessed their characteristics both in normal and abnormal tissue samples in individual patients separately (indv), and normal and abnormal tissue samples pooled together across all individuals (pool).
3.3.1 AFL:
3.3.1.1 pool
Despite no significant difference, a slight decline is observed between the average lifetime of both channels from all normal samples (shorter wavelength channel: average lifetime of 1.49 ns$\pm$0.47, longer wavelength channel: average lifetime of 1.35 ns$\pm$0.41) compared to the abnormal samples (shorter wavelength channel: average 1.37 ns$\pm$0.48, longer wavelength channel: average of 1.30 ns$\pm$0.40) (see Supplement 1).
3.3.1.2 indv
In the shorter wavelength channel (see Fig. 5 A) no significant difference was observed between any paired sample. In the longer wavelength channel (see Fig. 5 B), a significantly higher lifetime value of the abnormal lifetime compared to the normal lifetime between one sample (sample number 12, p-value 0.03) was observed.
3.3.2 SFL: magnitude
3.3.2.1 pool
A consistent decrease in the average SFL is observed in abnormal samples compared to normal samples at each channel (see Fig. 6 A). A t-test at each channel between pooled normal and pooled abnormal samples reveal that the drop in lifetime is not significant. When considering the average SFL of all samples, the most prominent difference is observed in the lower wavelength region (between 560 nm - 580 nm) (see Fig. 6 A). Furthermore, a declining trend in the value of lifetime across the wavelength range is observed, independent of sample type (see Fig. 6 A). These observations align with some reported in the literature, i.e. [23].
3.3.2.2 indv
To assess sample specific SFL changes, first, a paired t-test (paired at 160 channels) between the average normal and abnormal SFL was conducted. In 8 of the samples, a consistent decrease in fluorescence lifetime between the average normal and average abnormal sample was observed. In 4 of the samples a consistent increase was observed, and in 2 no significant change was observed, however, in these samples abnormal SFL is higher than normal SFL towards longer wavelengths (see Fig. 6 B). To present some sample specific SFL, three samples were visualised (see Fig. 6 C, D and E). These samples had the largest increase, largest decrease and no significant change between the SFL.
In all samples, overlap in the absolute lifetime value in at least one tissue location is observed (see Supplement 1). Furthermore, the overlap in the lifetime values does not appear constant, with some wavelength ranges having a greater amount of overlap (i.e. in Fig. 6 F and in Fig. 6 H, one location overlaps in the lower wavelength range (560 nm - 570 nm) more prominently). Furthermore, in the sample which shows the greatest decrease in SFL (see Fig. 6 E and H), two of the abnormal locations appear lower than the normal range of between 1.2 ns and 0.7 ns, however, one of the abnormal locations appears in the higher region of the normal locations. T-tests assessed at each channel did not reveal significant change in lifetime except sample 12 where a p-value of less than 0.05 was observed in most wavelength ranges as also demonstrated by AFL.
3.3.3 SFL: shape
3.3.3.1 pool:
To assess the SFL shape, independent of the absolute lifetime value, the Procrustes disparity between all samples was calculated. The median disparity between all normal SFL was 0.076, whereas the median disparity between all abnormal SFL was 0.159, suggesting a larger variety within the SFL shape in the abnormal than the normal. The pairwise distances (measured using the Procrustes distance metric) between all samples was assessed and plotted using tSNE dimensionality reduction (see Supplement 1). A slight difference between sample type can be observed (see Supplement 1), where, irrespective of the sample being normal or abnormal, the squamous cell samples and adenocarcinoma samples separate (see Supplement 1).
3.3.3.2 indv:
The median disparity between the sample specific normal and abnormal SFL was assessed (see Fig. 7). Two samples had noticeably higher median disparity within the normal, compared to the abnormal. However, the other samples either had a combined negligible median disparity suggesting very similar shapes (i.e. less than 0.1), or had a markedly higher disparity within the abnormal sample, compared to the normal sample. Most notable in samples number 2 and 8 which had a median disparity of 0.99 and 0.77 in their abnormal, compared to 0.122 and 0.058 in their normal (see Fig. 7).
4. Discussion
Previous systems used to measure tissue AF are limited in collecting fluorescence in low resolution spectral channels (i.e., [12]). Innovations into both high-resolution spectral devices, and analysis techniques to interrogate these signals allows the total temporal spectral histogram to be assessed in detail. The EP-TRFS device and MuFLE algorithm enable high resolution fluorescence intensity and SFL estimation spanning across an entire emission spectrum from 557.13 nm-638.22 nm providing unforeseen details of tissue AF [26].
We demonstrate that the magnitude and shape of SFL is influenced by the relative abundance of underlying fluorophores and their respective lifetimes (see Fig. 4). We observe this to occur while changing the relative abundance of both simulated fluorophores and endogenous fluorophores in samples matched to the expected fluorophores emitting in tissue with the same optical setup. We also observe that while changing the environment of the endogenous fluorophores, the SFL shape remains relatively similar. This provides a potential explanation to the variation we observe in the SFL tissue data where the magnitude and shape vary considerably. We might expect the magnitude and shape of SFL to not change if the relative abundance of the underlying fluorophores and their environment remain homogeneous e.g., in healthy, organised tissue. We might expect the magnitude to change, while shape remaining similar, if the lifetime of individual fluorophores changes without the respective emission spectra changing, and we might expect both magnitude and shape to vary if the relative abundance of the fluorophores change while the respective lifetimes remain fixed, or if both the relative abundance and lifetimes change, e.g., in heterogeneous, disorganised abnormal tissue, like cancer. We show that aggregating the spectral channels into channels with large bandwidths prevents these subtle spectral information between normal and abnormal tissue to be assessed. Although these observations align with our expectation and our results broadly align with results reported in existing studies, a complete validation of the hypothesis is beyond the scope of this paper. Nonetheless, we observe SFL to provide a detailed overview of tissue characteristics, which may provide potential for improved and informed diagnostics.
A limitation to the EP-TRFS device used in this study is that it produces fluorescence profiles alone. Therefore, specific structures observed within the sample which may account for a unique fluorescence profile, such as a fibrotic area, are not recorded alongside the fluorescence profile. So, without direct microscopic images of structure, intra-heterogenous data will arise from mixed tissue components. Furthermore, inter-heterogeneity may arise from differences in patient history i.e. exposure to smoking, pollution and other environmental factors which may alter the inflammatory state of the lung, limiting how normal the normal counterpart is. Additional limitations of our study include the limited patient sample numbers, and the assumption that every location of a tissue sample we take measurements from is biologically the same. Therefore, the lack of significance can be attributed to both the smaller number of patients included in the study and smaller number of locations sampled for each tissue sample (resulting in a smaller number of point-based histogram samples accumulated per tissue area being assessed) as well as patient demographics and tissue environment.
Although the SFL magnitude and AFL do not show statistical significance in our analysis, the SFL shape show significant variability in abnormal (and often normal) lung tissue indicating the underlying complex molecular dynamics of the structure. Although the significance of the statistical test can be improved through additional data (assuming the difference of lifetime being 50 ps and the population standard deviation being 200 ps, 253 samples will lead to a significance test with level 0.05 and power 0.8), the underlying heterogeneity should be understood in more detail. E.g., if it is due to underlying physical-chemical processes that can both contribute to increase and decrease of lifetime [16], or due to patient demographics such as age or potentially smoking status, or due to the locality of the abnormality, e.g., if it is smaller than the field of view (i.e., <32 $\mathrm{\mu}\textrm {m}$). A better understanding will lead to more informed diagnosis in precision medicine with SFL providing more granular information than AFL.
A major limitation within the spectral range utilised in the analysis, due to the IRF measured, remains. Since we use a fibre-based EP-TRFS system, and our device contains bandpass filters, traditional methods for measuring the IRF remain ineffective. For example, shining light directly onto the sensor does not account for the time delay of the optical signal travelling back and forth through the fibre. Similarly, reflecting excitation light back through the fibre is not appropriate since a bandpass filter blocks light less than 520 nm from entering the detector. A proper alignment of IRF and the observed spectra remain crucial for the least squares fitting routine, and therefore, to measure the IRF, we opt for using a fluorophore that can be excited at 485 nm with as wide a spectral range as possible and with a decay rate shorter than the pulsed light source. Quenched Rose-Bengal has been used in this manner [35]. Variations of fluorescein sodium was also tested (data not shown) at various concentrations, however, we found that the decay rate of this modified fluorophore remained wider than the laser pulse.
An excitation of 485 nm in previous studies using OEM systems has been shown to discriminate normal from abnormal structures, therefore, as aforementioned, the same setup was applied with this method. However, the excitation of the specific endogenous fluorophores with this device, coupled with the IRF limitation prevent the emission peaks of either fluorophore to be measured. Therefore, the optical setup misses the emission peak of the endogenous fluorophores which are expected to be excited using a 485 nm excitation wavelength. However, we show, using both simulated data and data generated bench-side that, regardless of the presence of an emission peak, the ratio of the individual fluorescence compartments influence the SFL shape. A potential discrimination enhancement of this application is to alter the device setup. More specifically, to investigate this emission at a shorter excitation wavelength, i.e., 350 nm, so as to increase the amount of endogenous fluorophores being excited. However, this is beyond the scope of this paper.
5. Conclusion
Our study demonstrates that the relative concentration of individual fluorophores contributes significantly to the shape and magnitude of SFL, both in simulated data and reference data while this information is not visible in AFL. This emphasises the importance of utilising high-resolution spectral channel data when evaluating tissue AF. Specifically focusing on lung tissue, we observed that, in most cases, the SFL consistently drops in magnitude across all channels in abnormal tissue compared to normal ex vivo lung tissue, although this difference is not statistically significant. Additionally, we noted that the shape of the SFL, on average, displays greater variability in abnormal ex vivo lung tissue compared to normal tissue. Furthermore, our observations revealed significant inter- and intra-heterogeneity among the patient samples measured. While this limits the immediate translation of this approach to tissue delineation, SFL provides unprecedented details on the tissue AF that can help better investigate the underlying tissue heterogeneity.
Funding
Medical Research Council (MR/N13 166/1).
Acknowledgments
We are grateful to the patients who consented to this study, and all the staff at the Department of Thoracic Surgery and Pathology at the Royal Infirmary of Edinburgh.
Disclosures
The authors declare no conflicts of interest.
Data availability
Data presented here is available upon request.
Supplemental document
See Supplement 1 for supporting content.
References
1. J. Ferlay, I. Soerjomataram, R. Dikshit, et al., “Cancer incidence and mortality worldwide: sources, methods and major patterns in globocan 2012,” Int. J. Cancer 136(5), E359–E386 (2015). [CrossRef]
2. F. C. Detterbeck, D. J. Boffa, A. W. Kim, et al., “The eighth edition lung cancer stage classification,” Chest 151(1), 193–203 (2017). [CrossRef]
3. Y. Ouyang, Y. Liu, Z. M. Wang, et al., “Flim as a promising tool for cancer diagnosis and treatment monitoring,” Nano-Micro Lett. 13(1), 133 (2021). [CrossRef]
4. W. Becker, “Fluorescence lifetime imaging–techniques and applications,” J. Microsc. 247(2), 119–136 (2012). [CrossRef]
5. L. Marcu, “Fluorescence Lifetime Techniques in Medical Applications,” Ann Biomed Eng. 40(2), 304–331 (2012). [CrossRef]
6. D. S. Kittle, F. Vasefi, C. G. Patil, et al., “Real time optical biopsy: Time-resolved fluorescence spectroscopy instrumentation and validation,” Sci. Rep. 6(1), 38190 (2016). [CrossRef]
7. J. R. Lakowicz, Principles of Fluorescence Spectroscopy (SSBM, 2013).
8. G. A. Wagnieres, W. M. Star, and B. C. Wilson, “In vivo fluorescence spectroscopy and imaging for oncological applications,” Photochem. Photobiol. 68(5), 603–632 (1998). [CrossRef]
9. A. C. Croce and G. Bottiroli, “Autofluorescence spectroscopy and imaging: a tool for biomedical research and diagnosis,” Eur. J. Histochem. 58(4), 1 (2014). [CrossRef]
10. M. Y. Berezin and S. Achilefu, “Fluorescence lifetime measurements and biological imaging,” Chem. Rev. 110(5), 2641–2684 (2010). [CrossRef]
11. D. Chorvat Jr and A. Chorvatova, “Multi-wavelength fluorescence lifetime spectroscopy: a new approach to the study of endogenous fluorescence in living cells and tissues,” Laser Phys. Lett. 6(3), 175–193 (2009). [CrossRef]
12. D. R. Yankelevich, D. Ma, J. Liu, et al., “Design and evaluation of a device for fast multispectral time-resolved fluorescence spectroscopy and imaging,” Rev. Sci. Instrum. 85(3), 034303 (2014). [CrossRef]
13. M. Marsden, T. Fukazawa, Y. Deng, et al., “Flimbrush: dynamic visualization of intraoperative free-hand fiber-based fluorescence lifetime imaging,” Biomed. Opt. Express 11(9), 5166–5180 (2020). [CrossRef]
14. G. Burgstaller, B. Oehrle, M. Gerckens, et al., “The instructive extracellular matrix of the lung: basic composition and alterations in chronic lung disease,” Eur. Respir. J. 50(1), 1601805 (2017). [CrossRef]
15. P. Uehlinger, T. Gabrecht, T. Glanzmann, et al., “In vivo time-resolved spectroscopy of the human bronchial early cancer autofluorescence,” J. Biomed. Opt. 14(2), 024011 (2009). [CrossRef]
16. M. Wang, F. Tang, X. Pan, et al., “Rapid diagnosis and intraoperative margin assessment of human lung cancer with fluorescence lifetime imaging microscopy,” BBA Clinical 8, 7–13 (2017). [CrossRef]
17. S. Fernandes, G. Williams, E. Williams, et al., “Fluorescence-lifetime imaging: a novel diagnostic tool for suspected lung cancer,” (2021).
18. M. W. Conklin, P. P. Provenzano, K. W. Eliceiri, et al., “Fluorescence lifetime imaging of endogenous fluorophores in histopathology sections reveals differences between normal and tumor epithelium in carcinoma in situ of the breast,” Cell Biochem. Biophys. 53(3), 145–157 (2009). [CrossRef]
19. P. J. Tadrous, J. Siegel, P. M. French, et al., “Fluorescence lifetime imaging of unstained tissues: early results in human breast cancer,” The J. Pathol. A J. Pathol. Soc. Gt. Br. Irel. 199(3), 309–317 (2003). [CrossRef]
20. Y. Wang, C. Song, M. Wang, et al., “Rapid, label-free, and highly sensitive detection of cervical cancer with fluorescence lifetime imaging microscopy,” IEEE J. Sel. Top. Quantum Electron. 22(3), 228–234 (2016). [CrossRef]
21. M. Brandao, R. Iwakura, F. Basilio, et al., “Fluorescence lifetime of normal, benign, and malignant thyroid tissues,” J. Biomed. Opt. 20(6), 067003 (2015). [CrossRef]
22. J. McGinty, N. P. Galletly, C. Dunsby, et al., “Wide-field fluorescence lifetime imaging of cancer,” Biomed. Opt. Express 1(2), 627–640 (2010). [CrossRef]
23. P. De Beule, C. Dunsby, N. Galletly, et al., “A hyperspectral fluorescence lifetime probe for skin cancer diagnosis,” Rev. Sci. Instrum. 78(12), 123101 (2007). [CrossRef]
24. G. O. Williams, E. Williams, N. Finlayson, et al., “Full spectrum fluorescence lifetime imaging with 0.5 nm spectral and 50 ps temporal resolution,” Nat. Commun. 12(1), 6616 (2021). [CrossRef]
25. Q. Wang, S. Fernandes, G. O. Williams, et al., “Deep learning-assisted co-registration of full-spectral autofluorescence lifetime microscopic images with h&e-stained histology images,” Commun. Biol. 5(1), 1119 (2022). [CrossRef]
26. A. Adams, A. Kufcsak, K. Ehrlich, et al., “Simultaneous spectral temporal modelling for a time-resolved fluorescence emission spectrum,” IEEE Transactions on Biomedical Engineering (2023).
27. B. Valeur and M. N. Berberan-Santos, Molecular fluorescence: principles and applications (John Wiley & Sons, 2012).
28. W. Becker, A. Bergmann, and C. Biskup, “Multispectral fluorescence lifetime imaging by tcspc,” Microsc. Res. Tech. 70(5), 403–409 (2007). [CrossRef]
29. J. M. Beechem and L. Brand, “Time-resolved fluorescence of proteins,” Annu. Rev. Biochem. 54(1), 43–71 (1985). [CrossRef]
30. L. Thiberville, S. Moreno-Swirc, T. Vercauteren, et al., “In vivo imaging of the bronchial wall microstructure using fibered confocal fluorescence microscopy,” Am. J. Respir. Crit. Care Med. 175(1), 22–31 (2007). [CrossRef]
31. S. Tian, H. Huang, Y. Zhang, et al., “The role of confocal laser endomicroscopy in pulmonary medicine,” Eur. Respir. Rev. 32(167), 220185 (2023). [CrossRef]
32. L. Thiberville, M. Salaün, S. Lachkar, et al., “Human in vivo fluorescence microimaging of the alveolar ducts and sacs during bronchoscopy,” Eur. Respir. J. 33(5), 974–985 (2009). [CrossRef]
33. Q. Wang, M. Vallejo, and J. Hopgood, “Fluorescence lifetime endomicroscopic image-based ex-vivo human lung cancer differentiation using machine learning,” Pre-print (2020).
34. W. Becker, A. Bergmann, G. L. Biscotti, et al., “Advanced time-correlated single photon counting techniques for spectroscopy and imaging in biomedical systems,” in Commercial and Biomedical Applications of Ultrafast Lasers IV, vol. 5340 (SPIE, 2004), pp. 104–112.
35. M. Szabelski, R. Luchowski, Z. Gryczynski, et al., “Evaluation of instrument response functions for lifetime imaging detectors using quenched rose bengal solutions,” Chem. Phys. Lett. 471(1-3), 153–159 (2009). [CrossRef]
36. T. M. Khan, M. Arshad, and M. A. Choudhry, “Modeling and control of cartesian robot manipulator,” in 2005 Pakistan Section Multitopic Conference, (IEEE, 2005), pp. 1–4.
37. L. Marcu, P. M. French, and D. S. Elson, Fluorescence lifetime Spectroscopy and Imaging: Principles and Applications In Biomedical Diagnostics (CRC Press, 2014).
38. Y. Sun, Y. Sun, D. Stephens, et al., “Dynamic tissue analysis using time-and wavelength-resolved fluorescence spectroscopy for atherosclerosis diagnosis,” Opt. Express 19(5), 3890–3901 (2011). [CrossRef]