Precompensation of 3D field distortions in remote focus two-photon microscopy

Open Access Open Access

Abstract

Remote focusing is widely used in 3D two-photon microscopy and 3D photostimulation because it enables fast axial scanning without moving the objective lens or specimen. However, due to the design constraints of microscope optics, remote focus units are often located in non-telecentric positions in the optical path, leading to significant depth-dependent 3D field distortions in the imaging volume. To address this limitation, we characterized 3D field distortions arising from non-telecentric remote focusing and present a method for distortion precompensation. We demonstrate its applicability for a 3D two-photon microscope that uses an acousto-optic lens (AOL) for remote focusing and scanning. We show that the distortion precompensation method improves the pointing precision of the AOL microscope to < 0.5 µm throughout the 400 × 400 × 400 µm imaging volume.

Published by The Optical Society under the terms of the Creative Commons Attribution 4.0 License. Further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation, and DOI.

Corrections

10 June 2021: Typographical corrections were made to the Data Availability Statement and Refs. 35 and 36.

1. Introduction

Two-photon scanning microscopy is a widely used tool for high resolution functional imaging in life sciences due to its ability to penetrate deep within scattering tissue [1,2]. In neuroscience, two-photon imaging is often used to monitor activity within neurons and across neuronal populations. However, measurements of signals flowing through these 3D structures are restricted with conventional microscopes because the speed of focusing is limited by the inertia of the objective lens. By contrast, rapid remote focus approaches rely on devices that dynamically alter the optical wavefront of the beam prior to the objective, thereby circumventing the need to physically move the objective. Rapid remote focusing has been implemented with electrically tunable lenses (ETLs) [3], liquid crystal spatial light modulators (LC-SLMs) [48], phase-locked ultrasound lenses (tunable acoustic gradient (TAG) lenses) [9], deformable mirrors [10], liquid lenses [11], acousto-optic lenses (AOLs) [1215] and a secondary objective coupled with a piston mirror [1620]. However, remote focus devices cannot always be positioned in a plane conjugate to the back aperture of the objective, due to mechanical space constraints, a limited range of commercially available focal length lenses and uncertainty in the exact position of the back aperture of commercial objectives. This is problematic because it introduces depth-dependent 3D field distortions in the imaging volume [3,2124], which impair the performance of remote focus laser scanning systems. Distortions to the 3D field of view (FOV) lead to missed regions of the specimen and serious mismatches between the 3D z-stacks obtained by mechanically moving the objective and the z-stacks obtained by remote focusing. It also causes erroneous measurements of the dimensions of biological features [21] and makes selection and positioning of regions of interest (ROIs) for selective imaging or photostimulation, within the full FOV inaccurate, which is particularly detrimental when targeting small structures such as neuronal processes.

In a conventional microscope with a collimated input and an infinity corrected objective, non-telecentricity has little effect on the 2D FOV when the focus is changed by mechanically shifting the objective. Although a lateral misalignment of the input beam will result in a tilted point spread function (PSF), changing the focus will not introduce xy plane magnification distortions. However, in a remote focus microscope, where the curvature and tilt of the input beam is modified to change the 3D focal position, a misalignment of the optical input from its ideal telecentric position leads to a distorted 3D FOV. An axially misaligned optical input leads to a z-dependent xy magnification, as well as a varying axial magnification, resulting in uneven z-plane separation. A lateral misalignment of the remote focus unit results in a skew distortion to the xz and/or yz FOV. As the laser is focused in z away from the natural plane (the focal plane for a planar input beam), the FOV progressively drifts laterally. This is problematic as even modest misalignments of the remote focus unit can result in distortions of the dimensions of the FOV and lateral shifts amounting to tens of micrometers over the focus range.

Previous distortion correction approaches [21] have been implemented via post-acquisition image processing. This has several limitations, including loss of information due to cropping of the volumetric data, the time-consuming nature of post-hoc corrections and the use of a complex calibration sample that requires laser manufacturing and which is sensitive to misalignments. Here, we present a fast, convenient, and generic method to measure and precompensate distortions in remote focus microscopes. We demonstrate experimentally that a paraxial remote focus distortion model can be applied to an AOL remote focus microscope to precompensate for the 3D field distortions arising from non-telecentric misalignments and show how similar distortion precompensation solutions could be applied to other remote focus technologies.

2. Model of 3D field distortions in remote focus microscopes

To characterize and correct for the 3D field distortions that arise in a remote focus microscope due to non-telecentric misalignments of the remote focus unit, we developed a paraxial model of a remote focus unit and objective lens. In an ideal telecentric case [Fig. 1(a)] the position of the output of the remote focus unit is located one focal length distance away from the objective (i.e. conjugate to the back aperture) and the optical curvature change produced by the remote focus unit is linearly proportional to the resulting change of z-remote-focus (zRF) in the FOV. Note that zRF is measured with respect to the natural focal plane and is considered a negative distance when located below the natural focal plane (as is the case in Fig. 1). Since the plane conjugate to the output of the remote focus unit is located at infinity, the magnification of the xy FOV is constant with changes in axial focus. An axial misalignment of the remote focus unit, ZERROR, results in a z-dependent xy magnification of the objective FOV, together with a non-uniform z-plane separation for equal changes in curvature of the optical input [Fig. 1(b)]. In this case, the change in position of zRF is not proportional to the change in optical curvature produced by the remote focus device and the resulting z-dependent xy magnification causes tapering of the x FOV with z. A lateral displacement in x, XERROR, of the output of the remote focus unit results in a laterally shifted position of the x-remote-focus (xRF) with z, which we refer to as an xz skew distortion of the imaging volume [Fig. 1(c)]. Similarly, a lateral offset in y would cause a yz skew distortion.

 figure: Fig. 1.

Fig. 1. Telecentricity in remote focus microscopes. a) Schematic ray diagram showing xz plane for telecentric output of the wavefront shaping unit located one focal length (f) from the imaging objective (blue lens). Curved wavefronts (curved rainbow-colored lines) represent a converging or diverging optical beam, being scanned (curved arrow). Central ray (black lines) deflected through a beam semi-scan angle θx, with respect to the optical axis (dotted blue line). Each of the three central rays indicate a different time point of a scan. Red asterisk shows the beam focused to an arbitrary xz remote focus within the FOV (gray area) with coordinates (xRF, zRF) indicated. The change in the axial focus of the objective is linearly proportional to the change in curvature produced by the remote focus unit and the central ray in the FOV has no skew distortion. Imaging planes (straight rainbow-colored lines) are thus equally spaced in z for equal steps in beam curvature and the FOV is of constant size. Note that the rainbow color-code matches [Fig. 35]. b) Axial misalignment of the remote focus device output (ZERROR) causes depth-dependent xy FOV magnification and a non-constant z-plane spacing with wavefront curvature, resulting in a change in both xRF and zRF with respect to the telecentric case in (a). Zd is the position of the plane conjugate with the output of the remote focus device. c) Lateral misalignment of the remote focus device output (XERROR) introduces a skew distortion to the FOV, resulting in a skewed local z-axis and a shifted xRF.

Download Full Size | PDF

To quantify the field distortions arising from non-telecentric misalignments of the remote focus unit we derived a set of equations (see Supplement 1, Section S1 and [Fig. S1] for the full derivation). For an optical output of curvature κ, a paraxial objective lens of focal length f, an imaging refractive index n and an (x, z) misalignment of the remote focus unit (XERROR, ZERROR), the (x, z) remote focus with respect to the optical axis, (xRF, zRF) is given by,

$${{{z}}_{{{RF}}}}{{\; = \;}}\frac{{{{ - n\;}}{{{f}}^{{2}}}{{\;\kappa }}}}{{{{\kappa \;}}{{{Z}}_{{{ERROR}}}}{{\; + \;{\textit{1}}}}}}$$
$${{{x}}_{{{RF}}}}{{\; = \;}}\frac{{{{f(}}{{{\theta }}_{{x}}}{{\; - \;\kappa \;}}{{{X}}_{{{ERROR}}}}{{)\;}}}}{{{{\kappa \;}}{{{Z}}_{{{ERROR}}}}{{\; + \;\textit{1}}}}}$$
where θx is the uncompensated x semi-scan angle. Note that Fig. 1 shows the deflection of the central ray of the wavefront at the beginning, middle and end of a scan. For an axially misaligned remote focus unit, the lateral xy plane magnification M, will vary with z and is given by,
$${{M\; = \;}}\frac{{{{n}}{{{f}}^{{2}}}{{\; + \;}}{{{z}}_{{{RF\;}}}}{{{Z}}_{{{ERROR}}}}}}{{{{n}}{{{f}}^{{2}}}}}$$

To precompensate for the distortion, the remote focus unit can be driven to produce a wavefront curvature κCOMP and x semi-scan angle θx COMP, based on the inverse of Eqs. (1) and (2), for the desired corrected z and x focus in the objective FOV, zCORR and xCORR, respectively, where

$${{{\kappa }}_{{{COMP}}}}{{\; = \; - \;}}\frac{{{{{z}}_{{{CORR}}}}}}{{{{\;}}{{{z}}_{{{CORR\;}}}}{{{Z}}_{{{ERROR}}}}{{\; + \;n}}{{{f}}^{{2}}}{{\;}}}}$$
$${{{\theta }}_{{{x\;COMP}}}}{{\; = \;}}{{{\kappa }}_{{{COMP\;}}}}{{{X}}_{{{ERROR}}}}{{ \;+ }}\frac{{{{{x}}_{{{CORR}}}}{{\;(}}{{{\kappa }}_{{{COMP\;}}}}{{{Z}}_{{{ERROR}}}}{{\; + \textit{1})}}}}{{{f}}}\;$$

Thus, field distortions arising from system non-telecentricity can be perfectly compensated, if the wavefront curvature κ and (x, y) semi-scan angles (θx, θy) produced by the remote focus unit are known. The generic nature of this solution suggests that field distortions caused by any paraxially approximated remote focus system can be precompensated by incorporating Eqs. (4) and (5) into the control software.

Magnification distortions are common in widely used ETL-based remote focus microscopes because they are often placed in a non-telecentric position behind the objective [2224]. As a worked example, we show how our solution could be used to precompensate the remote focus distortion reported in an ETL-based non-telecentric remote focus two-photon microscope [3]. Figure 2(a) shows the reported FOV and 2(b) the FOV fitted by Eqs. (1,2). The precompensated FOV calculated by Eqs. (4, 5) is shown in Fig. 2(c). By driving the non-telecentric remote focus unit using Eqs. (4, 5), a perfectly undistorted FOV would be obtained, as predicted in Fig. 2(d).

 figure: Fig. 2.

Fig. 2. Application of the distortion model to an electrically tunable lens remote focus system. a) Experimental xz FOV imaged in fluorescein beneath the objective of a two-photon microscope, remote focused with an electrically tunable lens. Adapted with permission from [3] © The Optical Society. Red lines show extent of x limits of the FOV predicted using eqs. (1,2) with f = 4 mm, ZERROR= 26.6 mm, XERROR = 0.4 mm, n = 1.33 and θx= ± 36 mrad. Natural focal plane (z = 0) indicated by dashed black line. b) Theoretical xz FOV for a z range = [−108, 375] µm with 9 equal steps of curvature, given eqs. (1,2). c) Theoretical xz FOV calculated using precompensated drives κCOMP, Eq. (4) and θx COMP, Eq. (5). d) Accurately precompensated rectangular FOV with equispaced z planes calculated by driving the non-telecentric model of the remote focus system of (a) and (b) with the pre-compensated drive of (c).

Download Full Size | PDF

3. Experimental evaluation of the distortion precompensation scheme

To test the accuracy of distortion precompensation we developed a simple strategy [Fig. 3(a)] to measure field distortions and employed an acousto-optic lens (AOL) remote focus microscope [Fig. S2] (Supplement 1 Section S2), since AOL technology enables high precision focusing and scanning of a laser beam [25,26]. To quantify distortions caused by non-telecentric remote focusing in our AOL 3D two-photon microscope, we compared a reference image obtained at the natural plane of the objective [Fig. 3(b)], where non-telecentric misalignment does not cause distortion of the FOV, with images of the same region obtained using increasingly strong remote focusing. The resulting change in focus was systematically counteracted by displacing the objective by an equal amount in the opposite direction [Fig. 3(a), inset]. The precision of the mechanical objective focus, which is critical to the precision of the calibration scheme, was determined to be <0.5 µm, with a mechanical alignment gauge.

 figure: Fig. 3.

Fig. 3. Characterization of non-telecentric remote focus distortions in an acousto-optic lens microscope. a) Experimental setup, showing a slide covered with a thin layer of 5 µm green beads. As the axial position of the objective is incrementally shifted away from the sample using motorized z-movement, the beads are kept in focus using remote focusing with an acousto-optic lens (AOL) microscope. Black solid line shows optical beam focusing at the natural plane (gray objective). Pink dotted line shows converging optical beam enabling remote focused imaging with objective displaced (pink outlined objective). Inset summarizes the experimental procedure used to generate panel (c). b) Reference image of 5 µm fluorescent beads obtained at the natural focus of the objective in a 262 µm xy FOV. c) Color-coded intensity projection of the imaged beads at various remote foci (± 200 µm in 5 µm increments, color indicates z-plane location). Color-coded lines indicate depth-dependent deviation of the xy bead location relative to the natural plane image. d) Bead trajectories from (c) in scanned (y) and non-scanned (x) direction tracked using ImageJ TrackMate plugin (black). Bead trajectories predicted from the model at the same location as the experiments (red) and for positions on a grid pattern (pink). Orange and gray asterisks indicate location of beads for +200 µm remote focusing and natural plane imaging, respectively.

Download Full Size | PDF

To perform the calibration, we used a microscope slide with a thin layer of 5 µm fluorescent beads suspended in agar (Phosphorex, Degradex), although any small, bright sample would work (see Supplement 1 Section S3 and [Fig. S3]). In a perfectly telecentric optical path, the mechanical objective focus and AOL remote focus would be in perfect registration and the images of the beads would remain superimposed wherever they were acquired across the entire remote focusing z-range. However, when non-telecentric misalignment is present, field distortions are revealed as displacements in the locations of the beads in the images across the z-range. Figure 3(c) shows a z-intensity projection of the images of the calibration z-stack (C-z-stack) covering a remote focus range of ± 200 µm, without correction for the distortion arising from the misalignments present in our AOL 3D two-photon microscope. Displacement of the individual beads across the C-z-stack were measured using the TrackMate plugin in ImageJ [27]. When plotting the position of the beads for each frame of the C-z-stack, the trajectory of the displacement of each bead over the remote focus z-range enables a measure of the distortion field [Fig. 3(d)]. The XERROR and YERROR present in our AOL microscope led to a measured absolute shift in xRF and yRF of 28 µm and 14 µm, respectively, with respect to the undistorted xRF and yRF at z = 0, over a remote focus range of ± 200 µm. The ZERROR present in our AOL microscope caused the reference 267 × 267 µm xy FOV to vary between 252 × 252 µm and 283 × 283 µm over a ± 200 µm remote focus range, respectively.

To relate bead trajectories to non-telecentric field distortion in a quantitative manner we developed a ray-based model of the AOL 3D microscope. The non-telecentric remote focus distortion predicted by the AOL ray model was found to match the distortion predicted by the paraxial distortion model [Fig. S4-S6]. The ray model was then used to estimate the values of non-telecentric misalignment in the physical system that give rise to the observed distortion, by finding a match between the experimentally measured distortion and the distortion predicted by the model. For the experimental data in Fig. 3(d), the distortion predicted by the model is likely to be caused by a misalignment of the AOL scanner of XERROR, YERROR, ZERROR of 2.3, 1.3, -69 mm, respectively [Fig. S7]. But note that the same experimental field distortion could be caused by a wide range of optical component misalignments in the relay(s) between the remote focus device and the final objective. Nevertheless, the predictions of Eqs. (1) and (2) will still be valid. We then used Eqs. (4,5) to precompensate for the distortion, by modifying the amount of tilt and curvature imparted by the AOL to the optical beam. The distortion precompensation equations were incorporated into our MATLAB-based AOL microscope control software and had little computational overhead. The values for XERROR, YERROR and ZERROR were adjusted to obtain a perfect registration between the remote focus system and the mechanical focus system.

To test the performance of precompensation for magnification distortions arising from axial misalignments of the AOL scanner, the amount of effective axial misalignment, ZERROR, was varied by changing the lenses within the relay, which altered the magnification and introduced an effective axial misalignment of 152 mm or −9 mm (i.e. an addition of −83 mm or +60 mm to the pre-existing axial misalignment of −69 mm in our AOL setup, respectively) [Fig. S8]. Figure 4(a-f) shows the projection before and after the correction of these effective axial misalignments, together with model predictions, confirming the capacity of the distortion precompensation scheme for larger axial misalignments of the remote focus unit. In addition, the z focal plane position was measured before and after distortion precompensation [Fig. 4(g,h)], using a thin layer of 0.2 µm beads suspended in polymer. The AOL was used to remote focus at z = [−200, 0, 200] µm, and the objective was mechanically moved in steps of 1 µm around these positions to measure the real amount of remote focusing. Before precompensation, the out-of-focus error was measured to be −9.5 µm and 13.5 µm for a target remote focus of +200 µm and 200 µm, respectively, but was reduced to only −1.5 µm and 1.5 µm, after precompensation. This confirmed that our precompensation for axial misalignment also corrected for the non-constant z-spacing introduced by the pre-existing axial misalignment of −69 mm.

 figure: Fig. 4.

Fig. 4. Experimental tests of distortion correction scheme for axial AOL misalignments: a) Mean intensity projection of imaged beads at various remote foci (± 200 µm in 5 µm increments, color indicates z-plane location). Color smearing indicates depth-dependent deviation of the bead location relative to the natural plane image. The −83 mm of additional effective axial misalignment (see Fig. S8(b)) caused z-dependent magnification of the FOV (lateral misalignment was already pre-compensated). b) Same as (a) with axial distortion precompensation enabled. c) Same as (a), with introduction of +60 mm of additional axial AOL misalignment (See Fig S8(c)). d) same as (c) with distortion precompensation enabled. e) Bead position trajectories tracked using ImageJ TrackMate plugin (black) and bead positions predicted by the ray model at the same locations (red) together with model predictions for a grid pattern (pink) for (b), with ZERROR = +60 mm. Orange and gray asterisks indicate values for +200 µm remote focusing and natural plane imaging, respectively. f) same as (e), with ZERROR = −83 mm. g) Measure of the z-focus error in our AOL system when remote focusing at +200 and +200 µm, using 0.2 µm beads. Out-of-focus error was −9.5 and +13.5 µm, respectively. h) Correction for axial misalignment also corrected non-constant spacing of the z planes. Residual out-of-focus error is −1.5 and +1.5 µm.

Download Full Size | PDF

The performance of the distortion precompensation scheme was tested for a range of effective lateral misalignments of the AOL by deflecting the beam with a pair of Risley prisms (PS810-B, Thorlabs Inc.). This allowed us to introduce a lateral AOL misalignment in a controlled manner, at a high precision, without having to physically move the AOL [Fig. S9]. The effective lateral misalignment of the AOL, XERROR and YERROR could be calculated from the lateral displacement of the optical beam by the Risley prisms and the magnification of the optical telecentric relay. To assess the performance of the distortion correction scheme, the xz and yz skew distortion of the FOV was measured by taking the inverse tangent of the slope of the trajectory of the beads through the C-z-stacks, with and without the distortion correction scheme enabled, for a range of lateral beam displacements. Figure 5(a,b) shows the maximum intensity z-projections of the C-z-stacks acquired without and with distortion precompensation, for an effective lateral AOL misalignment of −5.9 mm in y. Despite the strong distortion introduced by the y misalignment (−18.7 ± 1.3 degree yz skew angle over a ± 200 µm remote focus range), the distortion precompensation reduced the yz skew angle to −0.3 ± 1.3 degrees. Risley prism-based lateral beam displacement across a range of values showed that the distortion compensation scheme was effective at correcting skew distortions arising from effective lateral displacements of the remote focus unit of up to ± 6.8 mm in y [Fig. 5(c)]. Similar results were obtained for x displacement.

 figure: Fig. 5.

Fig. 5. Lateral distortion precompensation: a) Color-coded maximal intensity projection of fluorescent beads for a ±200 µm remote focusing range with 5 µm z-increments, without correction and with the beam incident on the back aperture of the objective misaligned using the Risley prisms (equivalent to a −5.9 mm YERROR). b) Same as (a) with distortion precompensation. c) YZ skew distortion angle in the FOV of the objective (arctan of the yz slope of the central bead) as a function of the effective lateral AOL misalignment in y (red symbols). The distortion due to the inherent axial misalignment of the AOL was precompensated using the correction scheme to isolate the distortion arising from the Risley prism beam deflection (black symbols). Theoretical skew distortion angle measurements (hollow black circles). Light blue area indicates the experimental point shown in (b, d). d) Trajectories of the bead positions (black) with both axial and lateral precompensation. Orange and gray asterisks indicate location of beads values for +200 µm remote focusing and natural plane imaging, respectively. Note that markers and lines are superimposed as residual error is only 0.3 ± 0.2 µm.

Download Full Size | PDF

To quantify the precision of AOL-based laser scanning throughout the 400 × 400 × 400 µm imaging volume after distortion precompensation, we measured the largest displacement of the bead from its position when imaged at the natural focal plane of the objective (z = 0). Figure 5(d) provides an example of the bead displacement measured over a C-z-stack, after correction for the inherent lateral and axial misalignments of the AOL. This revealed a maximum lateral displacement of 0.33 µm ± 0.12 µm over the ± 200 µm remote focus range.

An unexpected benefit of the high precision of the distortion precompensation scheme for non-telecentric misalignments enabled us to identify and correct for small residual distortions in our AOL microscope that arose from weak divergence of the input beam [Fig. S10]. This higher order distortion, which resulted in a non-uniform distortion of the field with a scan-time dependency, could be explained by the ray model (Supplement 1 Section S6), enabling a full correction. This was possible only after precompensation for the larger first order distortions due to system non-telecentricity.

4. Discussion and conclusion

Wavefront-shaping devices are increasingly being used in two-photon microscopy to provide the rapid focusing required for 3D functional imaging and 3D photostimulation. Here we derive a set of equations to quantify and correct for the substantial field distortions introduced by remote focus devices that are not perfectly telecentric. By quantifying distortions and implementing distortion precompensation into the control software of an AOL-based remote focus 3D two-photon microscope, we demonstrated submicron (< 0.5 µm) precision over the 400 × 400 × 400 µm FOV, for a non-telecentric arrangement. The generic nature of the precompensation solution makes it suitable for a range of remote focus devices, enabling non-telecentric designs with improved optical performance.

Addition of remote focusing devices to conventional microscopes is complicated by the requirement for telecentricity within the optical train [28], which is often not possible due to mechanical space constraints, lens availability and a lack of information on the optical design of commercial microscopes and objectives. The substantial field distortions introduced by non-telecentric arrangements [3,2124] compromise the imaging performance of such 3D microscopes. Our solution to this problem, which enables full precompensation of distortion over the entire 3D FOV, builds on earlier work that quantified remote focus distortions using a sophisticated 3D plastic reference calibration specimen and an image processing based on post-hoc correction [21]. Our quantification method and precompensation strategy has several advantages over such an approach. The simple C-z-stack calibration method using fluorescent beads dispenses with the need for a custom made 3D calibration sample, and thus any issues associated with the polymer having a different refractive index to tissue and hence a different distortion (as noted by the authors, [21]). By contrast, our calibration and precompensation can be carried out in water or even in the target biological tissue itself. This allows any optical element located between the lens and the sample, such as a coverslip, to be integrated into the calibration process. Moreover, our approach directly calibrates the microscope against the xyz stage mechanics and thus provides a single coordinate system for the 3D FOV of the specimen when zooming and/or imaging in any arbitrary plane. Lastly, our precompensation solution restores the full telecentric imaging volume, so that remote focusing and mechanical focusing are in perfect register and laser scanning in any arbitrary direction maps accurately onto a Euclidean 3D space. This prevents regions being missed and eliminates the necessity of post processing the 3D images.

Quantification and precompensation of field distortions, represents a significant step towards achieving aberration-free non-telecentric remote focusing. Compensation for field distortions does not, however, correct for the (largely spherical) aberrations introduced by the objective lens when remote focusing. These have recently been characterized in detail and compensated for in a LC-SLM-remote focus microscope [29]. Combining field and lens-based aberration corrections would enable non-telecentric remote focus designs to approach the near optimal optical performance achieved with telecentric dual objective 3D microscopes [16,20].

Precompensation of remote focus field distortion improves the performance of AOL-based random-access 3D microscopy in several ways. Firstly, it ensures that the location and extent of the FOV is the same as with a perfectly telecentric optical arrangement. The sub 0.5 µm precision achievable with AOL-based laser scanning is critical for selective high-speed functional imaging of fine biological structures such as the synapses, axons and dendrites of neurons. High precision line scanning within an undistorted imaging volume is also critical for the correction of tissue movement in real time, since this involves tracking the movement of an object such as a fluorescent bead, in one location and applying a rigid xyz translation to the imaged volume, to counteract any 3D brain movement [30]. Lastly, an undistorted remote focus FOV enables highly precise alignment of neighboring AOL-based z-stacks for imaging large biological structures and brain regions that extend far beyond a single imaging volume.

As our modelling of ETL remote focusing shows, our approach for correcting distortions in non-telecentric remote focus devices is applicable to other types of 3D microscopes and is expected to improve their performance. For remote focus technologies in which the wavefront shaping control is less precise than that of an AOL or ETL, some post-hoc image processing may still be required. However, if the non-telecentric remote focus distortions are largely precompensated, the loss in volumetric data will be significantly reduced and any residual errors may be corrected with post-processing of the images. Alternatively, the estimate of the non-telecentric misalignments provided by the distortion model can be used to correct the misalignment of optical components in the remote focus path. The advantage of this is that it will also correct misalignment-induced aberration of the PSF.

Precompensation for field distortions is also likely to be particularly important for 3D photostimulation, since this requires accurate focusing to one or more precise locations within the imaging volume. Indeed, 3D two-photon imaging and optogenetic photostimulation methods [7,3134] require accurate LC-SLM-remote focusing of multiple illumination beams to specific locations within the imaging volume, which is typically acquired with galvanometer-based 2D scanning and mechanical focusing. Precise precompensation of the imaging field will be critical for extending these technologies to combine full 3D random access functional imaging and 3D photostimulation and for making photostimulation more precise, so that smaller structures such as synapses (∼1µm) can be selectively photoactivated.

Funding

Engineering and Physical Sciences Research Council (EP/L015455/1); ; National Institute of Neurological Disorders and Stroke (U01NS113273); Wellcome Trust (203048).

Acknowledgments

This research was funded in part by the Wellcome Trust (203048; R.A.S. is in receipt of a Wellcome Trust Principal Research Fellowship in Basic Biomedical Science) and in part by the National Institute of Neurological Disorders and Stroke of the National Institutes of Health under award number U01NS113273 (to R.A.S.). The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. FN was supported by the UK Engineering and Physical Sciences Research Council (EPSRC) studentship. We thank Victoria Griffiths, K. M. Naga Srinivas Nadella, Tomás Fernández-Alfonso for comments on the manuscript.

Disclosures

A.M.V., F.C.N., J.E.M declare no competing interests. P.A.K. and R.A.S. are named inventors on patents owned by UCL Business relating to linear and nonlinear AOL 3D laser scanning technology. P.A.K. has a financial interest in Agile Diffraction Ltd., which aims to commercialize the AOL technology.

Data availability

The SilverLab AOL Imaging Software and scripts for models are available on GitHub [35]. Data underlying the results presented in this paper are available on FigShare [36].

Supplemental document

See Supplement 1 for supporting content.

References

1. W. Denk, J. Strickler, and W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science 248(4951), 73–76 (1990). [CrossRef]  

2. F. Helmchen and W. Denk, “Deep tissue two-photon microscopy,” Nat. Methods 2(12), 932–940 (2005). [CrossRef]  

3. B. F. Grewe, F. F. Voigt, M. van ’t Hoff, and F. Helmchen, “Fast two-layer two-photon imaging of neuronal cell populations using an electrically tunable lens,” Biomed. Opt. Express 2(7), 2035–2046 (2011). [CrossRef]  

4. A. Cheng, J. T. Gonçalves, P. Golshani, K. Arisaka, and C. Portera-Cailliau, “Simultaneous two-photon calcium imaging at different depths with spatiotemporal multiplexing,” Nat. Methods 8(2), 139–142 (2011). [CrossRef]  

5. M. D. Maschio, A. M. D. Stasi, F. Benfenati, and T. Fellin, “Three-dimensional in vivo scanning microscopy with inertia-free focus control,” Opt. Lett. 36(17), 3503 (2011). [CrossRef]  

6. R. Liu, N. Ball, J. Brockill, L. Kuan, D. Millman, C. White, A. Leon, D. Williams, S. Nishiwaki, S. de Vries, J. Larkin, D. Sullivan, C. Slaughterbeck, C. Farrell, and P. Saggau, “Aberration-free multi-plane imaging of neural activity from the mammalian brain using a fast-switching liquid crystal spatial light modulator,” Biomed. Opt. Express 10(10), 5059 (2019). [CrossRef]  

7. F. Anselmi, C. Ventalon, A. Bègue, D. Ogden, and V. Emiliani, “Three-dimensional imaging and photostimulation by remote-focusing and holographic light patterning,” Proc National Acad Sci 108(49), 19504–19509 (2011). [CrossRef]  

8. S. Quirin, D. S. Peterka, and R. Yuste, “Instantaneous three-dimensional sensing using spatial light modulator illumination with extended depth of field imaging,” Opt. Express 21(13), 16007 (2013). [CrossRef]  

9. L. Kong, J. Tang, J. P. Little, Y. Yu, T. Lämmermann, C. P. Lin, R. N. Germain, and M. Cui, “Continuous volumetric imaging via an optical phase-locked ultrasound lens,” Nat. Methods 12(8), 759–762 (2015). [CrossRef]  

10. M. Žurauskas, O. Barnstedt, M. Frade-Rodriguez, S. Waddell, and M. J. Booth, “Rapid adaptive remote focusing microscope for sensing of volumetric neural activity,” Biomed. Opt. Express 8(10), 4369 (2017). [CrossRef]  

11. K. F. Tehrani, C. V. Latchoumane, W. M. Southern, E. G. Pendleton, A. Maslesa, L. Karumbaiah, J. A. Call, and L. J. Mortensen, “Five-dimensional two-photon volumetric microscopy of in-vivo dynamic activities using liquid lens remote focusing,” Biomed. Opt. Express 10(7), 3591 (2019). [CrossRef]  

12. G. D. Reddy, K. Kelleher, R. Fink, and P. Saggau, “Three-dimensional random access multiphoton microscopy for functional imaging of neuronal activity,” Nat. Neurosci. 11(6), 713–720 (2008). [CrossRef]  

13. P. A. Kirkby, K. M. N. S. Nadella, and R. A. Silver, “A compact acousto-optic lens for 2D and 3D femtosecond based 2-photon microscopy,” Opt. Express 18(13), 13720 (2010). [CrossRef]  

14. W. Akemann, J.-F. Léger, C. Ventalon, B. Mathieu, S. Dieudonné, and L. Bourdieu, “Fast spatial beam shaping by acousto-optic diffraction for 3D non-linear microscopy,” Opt. Express 23(22), 28191–28205 (2015). [CrossRef]  

15. G. Katona, G. Szalay, P. Maák, A. Kaszás, M. Veress, D. Hillier, B. Chiovini, E. S. Vizi, B. Roska, and B. Rózsa, “Fast two-photon in vivo imaging with three-dimensional random-access scanning in large tissue volumes,” Nat. Methods 9(2), 201–208 (2012). [CrossRef]  

16. E. J. Botcherby, C. W. Smith, M. M. Kohl, D. Débarre, M. J. Booth, R. Juškaitis, O. Paulsen, and T. Wilson, “Aberration-free three-dimensional multiphoton imaging of neuronal activity at kHz rates,” Proc National Acad Sci 109(8), 2919–2924 (2012). [CrossRef]  

17. Y. Qi, M. Lei, Y. Yang, B. Yao, D. Dan, X. Yu, S. Yan, and T. Ye, “Remote-focusing microscopy with long working distance objective lenses,” Appl. Opt. 53(16), 3473 (2014). [CrossRef]  

18. E. Z. Chong, M. Panniello, I. Barreiros, M. M. Kohl, and M. J. Booth, “Quasi-simultaneous multiplane calcium imaging of neuronal circuits,” Biomed. Opt. Express 10(1), 267 (2019). [CrossRef]  

19. K. T. Takasaki, D. Tsyboulski, and J. Waters, “Dual-plane 3-photon microscopy with remote focusing,” Biomed. Opt. Express 10(11), 5585 (2019). [CrossRef]  

20. N. J. Sofroniew, D. Flickinger, J. King, and K. Svoboda, “A large field of view two-photon mesoscope with subcellular resolution for in vivo imaging,” Elife 5, e14472 (2016). [CrossRef]  

21. A. D. Corbett, R. A. B. Burton, G. Bub, P. S. Salter, S. Tuohy, M. J. Booth, and T. Wilson, “Quantifying distortions in two-photon remote focussing microscope images using a volumetric calibration specimen,” Front. Physiol. 5, 384 (2014). [CrossRef]  

22. J. Jiang, D. Zhang, S. Walker, C. Gu, Y. Ke, W. H. Yung, and S. Chen, “Fast 3-D temporal focusing microscopy using an electrically tunable lens,” Opt. Express 23(19), 24362–24368 (2015). [CrossRef]  

23. Y. Nakai, M. Ozeki, T. Hiraiwa, R. Tanimoto, A. Funahashi, N. Hiroi, A. Taniguchi, S. Nonaka, V. Boilot, R. Shrestha, J. Clark, N. Tamura, V. M. Draviam, and H. Oku, “High-speed microscopy with an electrically tunable lens to image the dynamics of in vivo molecular complexes,” Rev Sci Instrum 86(1), 013707 (2015). [CrossRef]  

24. R. Shi, C. Jin, H. Xie, Y. Zhang, X. Li, Q. Dai, and L. Kong, “Multi-plane, wide-field fluorescent microscopy for biodynamic imaging in vivo,” Biomed. Opt. Express 10(12), 6625 (2019). [CrossRef]  

25. G. Konstantinou, P. A. Kirkby, G. J. Evans, K. M. N. S. Nadella, V. A. Griffiths, J. E. Mitchell, and R. A. Silver, “Dynamic wavefront shaping with an acousto-optic lens for laser scanning microscopy,” Opt. Express 24(6), 6283–6299 (2016). [CrossRef]  

26. G. J. Evans, P. A. Kirkby, K. M. N. S. Nadella, B. Marin, and R. A. Silver, “Development and application of a ray-based model of light propagation through a spherical acousto-optic lens,” Opt. Express 23(18), 23493–23510 (2015). [CrossRef]  

27. J.-Y. Tinevez, N. Perry, J. Schindelin, G. M. Hoopes, G. D. Reynolds, E. Laplantine, S. Y. Bednarek, S. L. Shorte, and K. W. Eliceiri, “TrackMate: An open and extensible platform for single-particle tracking,” Methods 115, 80–90 (2017). [CrossRef]  

28. H. Jiang, C. Wang, B. Wei, W. Gan, D. Cai, and M. Cui, “Long-range remote focusing by image-plane aberration correction,” Opt. Express 28(23), 34008 (2020). [CrossRef]  

29. Y. Yang, W. Chen, J. L. Fan, and N. Ji, “Adaptive optics enables aberration-free single-objective remote focusing for two-photon fluorescence microscopy,” Biomed. Opt. Express 12(1), 354 (2021). [CrossRef]  

30. V. A. Griffiths, A. M. Valera, J. Y. N. Lau, H. Roš, T. J. Younts, B. Marin, C. Baragli, D. Coyle, G. J. Evans, G. Konstantinou, T. Koimtzis, K. M. N. S. Nadella, S. A. Punde, P. A. Kirkby, I. H. Bianco, and R. A. Silver, “Real-time 3D movement correction for two-photon imaging in behaving animals,” Nat. Methods 17(7), 741–748 (2020). [CrossRef]  

31. W. Yang, L. Carrillo-Reid, Y. Bando, D. S. Peterka, and R. Yuste, “Simultaneous two-photon imaging and two-photon optogenetics of cortical circuits in three dimensions,” Elife 7, e32671 (2018). [CrossRef]  

32. A. R. Mardinly, I. A. Oldenburg, N. C. Pégard, S. Sridharan, E. H. Lyall, K. Chesnov, S. G. Brohawn, L. Waller, and H. Adesnik, “Precise multimodal optical control of neural ensemble activity,” Nat. Neurosci. 21(6), 881–893 (2018). [CrossRef]  

33. H. W. P. Dalgleish, L. E. Russell, A. M. Packer, A. Roth, O. M. Gauld, F. Greenstreet, E. J. Thompson, and M. Häusser, “How many neurons are sufficient for perception of cortical activity?” Elife 9, e58889 (2020). [CrossRef]  

34. L. Carrillo-Reid, W. Yang, Y. Bando, D. S. Peterka, and R. Yuste, “Imprinting and recalling cortical ensembles,” Science 353(6300), 691–694 (2016). [CrossRef]  

35. F. Neufeldt, P. Kirkby, J. Mitchell, and A. Silver, “AOL Ray Model,” Github (2021), https://github.com/SilverLabUCL/SilverLab-Microscope.

36. F. Neufeldt, P. Kirkby, J. Mitchell, and A. Silver, “Precompensation of 3D field distortions in remote focus two-photon microscopy,” Figshare (2021), https://doi.org/10.5522/04/14614071.

Supplementary Material (1)

NameDescription
Supplement 1       Supplemental methods and figures

Data availability

The SilverLab AOL Imaging Software and scripts for models are available on GitHub [35]. Data underlying the results presented in this paper are available on FigShare [36].

35. F. Neufeldt, P. Kirkby, J. Mitchell, and A. Silver, “AOL Ray Model,” Github (2021), https://github.com/SilverLabUCL/SilverLab-Microscope.

36. F. Neufeldt, P. Kirkby, J. Mitchell, and A. Silver, “Precompensation of 3D field distortions in remote focus two-photon microscopy,” Figshare (2021), https://doi.org/10.5522/04/14614071.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (5)

Fig. 1.
Fig. 1. Telecentricity in remote focus microscopes. a) Schematic ray diagram showing xz plane for telecentric output of the wavefront shaping unit located one focal length (f) from the imaging objective (blue lens). Curved wavefronts (curved rainbow-colored lines) represent a converging or diverging optical beam, being scanned (curved arrow). Central ray (black lines) deflected through a beam semi-scan angle θx, with respect to the optical axis (dotted blue line). Each of the three central rays indicate a different time point of a scan. Red asterisk shows the beam focused to an arbitrary xz remote focus within the FOV (gray area) with coordinates (xRF, zRF) indicated. The change in the axial focus of the objective is linearly proportional to the change in curvature produced by the remote focus unit and the central ray in the FOV has no skew distortion. Imaging planes (straight rainbow-colored lines) are thus equally spaced in z for equal steps in beam curvature and the FOV is of constant size. Note that the rainbow color-code matches [Fig. 35]. b) Axial misalignment of the remote focus device output (ZERROR) causes depth-dependent xy FOV magnification and a non-constant z-plane spacing with wavefront curvature, resulting in a change in both xRF and zRF with respect to the telecentric case in (a). Zd is the position of the plane conjugate with the output of the remote focus device. c) Lateral misalignment of the remote focus device output (XERROR) introduces a skew distortion to the FOV, resulting in a skewed local z-axis and a shifted xRF.
Fig. 2.
Fig. 2. Application of the distortion model to an electrically tunable lens remote focus system. a) Experimental xz FOV imaged in fluorescein beneath the objective of a two-photon microscope, remote focused with an electrically tunable lens. Adapted with permission from [3] © The Optical Society. Red lines show extent of x limits of the FOV predicted using eqs. (1,2) with f = 4 mm, ZERROR= 26.6 mm, XERROR = 0.4 mm, n = 1.33 and θx= ± 36 mrad. Natural focal plane (z = 0) indicated by dashed black line. b) Theoretical xz FOV for a z range = [−108, 375] µm with 9 equal steps of curvature, given eqs. (1,2). c) Theoretical xz FOV calculated using precompensated drives κCOMP, Eq. (4) and θx COMP, Eq. (5). d) Accurately precompensated rectangular FOV with equispaced z planes calculated by driving the non-telecentric model of the remote focus system of (a) and (b) with the pre-compensated drive of (c).
Fig. 3.
Fig. 3. Characterization of non-telecentric remote focus distortions in an acousto-optic lens microscope. a) Experimental setup, showing a slide covered with a thin layer of 5 µm green beads. As the axial position of the objective is incrementally shifted away from the sample using motorized z-movement, the beads are kept in focus using remote focusing with an acousto-optic lens (AOL) microscope. Black solid line shows optical beam focusing at the natural plane (gray objective). Pink dotted line shows converging optical beam enabling remote focused imaging with objective displaced (pink outlined objective). Inset summarizes the experimental procedure used to generate panel (c). b) Reference image of 5 µm fluorescent beads obtained at the natural focus of the objective in a 262 µm xy FOV. c) Color-coded intensity projection of the imaged beads at various remote foci (± 200 µm in 5 µm increments, color indicates z-plane location). Color-coded lines indicate depth-dependent deviation of the xy bead location relative to the natural plane image. d) Bead trajectories from (c) in scanned (y) and non-scanned (x) direction tracked using ImageJ TrackMate plugin (black). Bead trajectories predicted from the model at the same location as the experiments (red) and for positions on a grid pattern (pink). Orange and gray asterisks indicate location of beads for +200 µm remote focusing and natural plane imaging, respectively.
Fig. 4.
Fig. 4. Experimental tests of distortion correction scheme for axial AOL misalignments: a) Mean intensity projection of imaged beads at various remote foci (± 200 µm in 5 µm increments, color indicates z-plane location). Color smearing indicates depth-dependent deviation of the bead location relative to the natural plane image. The −83 mm of additional effective axial misalignment (see Fig. S8(b)) caused z-dependent magnification of the FOV (lateral misalignment was already pre-compensated). b) Same as (a) with axial distortion precompensation enabled. c) Same as (a), with introduction of +60 mm of additional axial AOL misalignment (See Fig S8(c)). d) same as (c) with distortion precompensation enabled. e) Bead position trajectories tracked using ImageJ TrackMate plugin (black) and bead positions predicted by the ray model at the same locations (red) together with model predictions for a grid pattern (pink) for (b), with ZERROR = +60 mm. Orange and gray asterisks indicate values for +200 µm remote focusing and natural plane imaging, respectively. f) same as (e), with ZERROR = −83 mm. g) Measure of the z-focus error in our AOL system when remote focusing at +200 and +200 µm, using 0.2 µm beads. Out-of-focus error was −9.5 and +13.5 µm, respectively. h) Correction for axial misalignment also corrected non-constant spacing of the z planes. Residual out-of-focus error is −1.5 and +1.5 µm.
Fig. 5.
Fig. 5. Lateral distortion precompensation: a) Color-coded maximal intensity projection of fluorescent beads for a ±200 µm remote focusing range with 5 µm z-increments, without correction and with the beam incident on the back aperture of the objective misaligned using the Risley prisms (equivalent to a −5.9 mm YERROR). b) Same as (a) with distortion precompensation. c) YZ skew distortion angle in the FOV of the objective (arctan of the yz slope of the central bead) as a function of the effective lateral AOL misalignment in y (red symbols). The distortion due to the inherent axial misalignment of the AOL was precompensated using the correction scheme to isolate the distortion arising from the Risley prism beam deflection (black symbols). Theoretical skew distortion angle measurements (hollow black circles). Light blue area indicates the experimental point shown in (b, d). d) Trajectories of the bead positions (black) with both axial and lateral precompensation. Orange and gray asterisks indicate location of beads values for +200 µm remote focusing and natural plane imaging, respectively. Note that markers and lines are superimposed as residual error is only 0.3 ± 0.2 µm.

Equations (5)

Equations on this page are rendered with MathJax. Learn more.

z R F = n f 2 κ κ Z E R R O R + 1
x R F = f ( θ x κ X E R R O R ) κ Z E R R O R + 1
M = n f 2 + z R F Z E R R O R n f 2
κ C O M P = z C O R R z C O R R Z E R R O R + n f 2
θ x C O M P = κ C O M P X E R R O R + x C O R R ( κ C O M P Z E R R O R + 1 ) f
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.