Optical and visual quality of real intraocular lenses physically projected on the patient’s eye

Open Access Open Access

Abstract

Visual simulators aim at evaluating vision with ophthalmic corrections prior to prescription or implantation of intraocular lenses (IOLs) in the patient’s eye. In the present study, we present the design, implementation, and validation of a new IOL-in-cuvette channel in an Adaptive Optics visual simulator, which provides an alternative channel for pre-operative simulation of vision with IOLs. The IOL is projected on the pupil’s plane of the subject by using a Rassow system. A second lens, the Rassow lens, compensates for an IOL of 20 D while other powers can be corrected with a Badal system within a 5 D range. The new channel was evaluated by through-focus (TF) optical quality in an artificial eye on bench, and by TF visual acuity in patients, with various IOL designs (monofocal, diffractive trifocal, and refractive extended depth of focus).

© 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Two main physiological changes affect vision as the eye ages: cataracts and presbyopia. A cataract is an opacification of the crystalline lens. The usual treatment for cataract entails the replacement of the natural lens of the eye by an artificial intraocular lens (IOL). In economically developed countries, cataract surgery is performed between 4000 to 11000 times per million population in one year [1]. Another age-related condition, which appears at an earlier stage than cataract, is presbyopia, the loss of accommodation capability in the crystalline lens. While cataract surgery aims at restoring transparency in the lens, it has also become a refractive procedure, as the appropriate IOL power can be selected such that images are projected on the retina, at least for distance. As a result, replacement of the clear crystalline lens with refractive purposes, a procedure known as Refractive Lens Exchange (REL) [2], has become increasingly popular. An alternative to monofocal IOLs in cataract and, even more frequently in REL, are presbyopia-correcting IOLs. The number of multifocal (diffractive or refractive) and extended-depth-of-focus (EDOF) IOLs designs has grown rapidly in the last years [3]. However, it is challenging for surgeons to manage patients’ expectations in cataract surgery or REL, as it is difficult to imagine how one will see with the different types of corrections. Visual simulators allow the patients to experience multifocal vision with different IOL designs before surgery.

Many visual simulators are based on adaptive-optics (AO) [47], where the aberrations of the eye are manipulated with an active element conjugated with the pupil plane that can be programmed. In these systems, the lens design is converted into a spatial phase pattern and mapped either with a Deformable Mirror (DM, [810]), for lenses with smooth varying profiles or with a Spatial Light Modulator (SLM, [6,1113]), well suited to represent zonal refractive lenses or diffractive multifocal lenses. Most AO visual simulators are laboratory prototypes [5,14,15] although some DM-based and SLM-based phoropters or visual simulators have been commercially developed (such as the CRX1 by ImagineEyes [16] or VAO by Voptica [17]). AO Simulators are typically bench- or desk-top, operate by reflection, and visual targets are normally projected on a mini-display inside the system. Alternatively to the AO-visual simulators, simultaneous vision simulators combine images focused at near and at far using two channels provided with Badal systems, and in some configurations a transmission spatial light modulator which helps to split the areas of the pupil devoted for near and far [18,19]. A different way to simulate multifocal lenses is by temporal multiplexing with optotunable lenses to produce simultaneous vision (SimVis). In these simulators, the intraocular lenses are simulated by a temporal pattern, obtained by rapidly changing the power of the optotunable lens [2022]. A commercial product (SimVisGekko, by 2EyesVision) has been developed as a wearable, binocular, see-through, fully programmable, and remotely operated device. Previous works show good correspondence between the pre-operative through-focus visual performance with the simulated IOL with SimVis, and the post-operative through-focus visual performance with the implanted IOL, both on average and at the individual level [23].

Another option to programmable simulators (such as AO-based systems or SimVis) are devices in which the IOL is inserted in a cuvette and projected on the pupil plane of the patient’s eye. With this type of simulation, it is not possible to dynamically flip between different IOLs and, of course, it requires access to the physical lenses (unlike the programmable simulators that allow testing lenses even before manufacturing them). On the other hand, IOL projection systems have the advantage of allowing a direct evaluation of the lens, in principle, without relying on assumptions or limitations of the simulating technology (for example, chromatic aberration in the SLM, resolution of the SLM or DM). In previous studies, we have used phase plates [13], or real 0-Diopter (D) IOLs inserted in a cuvette [4] as a reference to evaluate the accuracy of SLM and SimVis-based simulators. However, often the standard range of powers in commercial IOLs does not include 0 D.

Previous works presented an optical system specifically designed to project an IOL on the eye’s pupil plane while canceling out the base power of the IOL. The system is called the Rassow telescope [24] and it consists of a 4-focal system, with the IOL acting as one of the lenses in the system, and an approximately +20 D achromatic lens acting as the second lens, compensating 20 D of the optical power of the IOL [25]. More recent work described the implementation and use of a Rassow system to evaluate real IOLs, and evaluated on-bench the contrast transfer of six monofocal and four multifocal IOLs [25], although the comparison of performance across IOLs was only attempted in relative, not absolute terms. Two commercial systems have been developed that appear to be based on the Rassow telescope or small variations of it: VirtIOL [26] and ACMIT [27]. In both cases, the IOL from the manufacturer is inserted in a cuvette and projected on the patient’s pupil plane. In the ACMIT instrument, the IOL is inserted into a model eye based on Liou and Brennan's model eye [28].

In our previous work, we compared TF VA pre- (with the simulator) and post-operatively (through the real IOL) in patients that had undergone cataract/RLE surgery with implanted trifocal diffractive IOLs and found high correspondence between simulation (with SimVis and with SLM) and the real implanted IOL [23]. A drawback of that study is that the validation between the simulated IOLs and the real IOLs could only be performed with one IOL (which is implanted) at a time. The IOL-in cuvette projection system allows for the same patient to be tested with multiple IOLs (both real and simulated), converting the new system into a potential “gold-standard” for other, most often programmable, simulators.

Typically, in an optical configuration for IOL projection such as the Rassow telescope, the rays impacting the IOL are parallel, in contrast to the converging effect of the rays onto the IOL occurring in real eyes. In real eyes, this results in a magnification between the iris (iris pupil diameter) and its image (entrance pupil diameter) [29]. This difference may be critical in the accuracy of the simulation, as it likely alters the imaging properties of the eye + IOL with respect to the natural (post-operative) case. Another underlying hypothesis is that the wave aberration (representing the IOL) at the entrance pupil and at the iris pupil are equivalent. Solving the effect of magnification and proving the equivalency of both pupil planes will allow us to reproduce the effect of the IOL in the eye, in absolute, and not only relative terms. In fact, except for the presence of the natural crystalline lens, the system that we have implemented in the current study should be set to replicate the absolute performance of the implanted IOL.

In summary, in this study, we describe a new channel that we have implemented in a custom-built AO visual simulator [13,23,30], that incorporates a Rassow telescope for the projection of a real IOL immersed in a cuvette onto the pupil’s plane of the patient. The system corrects for a mismatch in magnification of previous implementations of the Rassow telescope in the literature. We calculated computationally and measured experimentally the optical performance of the system. We measured the through-focus optical performance on bench and the visual performance in patients through three different types of lenses (monofocal, diffractive trifocal, and refractive EDOF). With the incorporation of the new channel in the AO visual simulator, it will be possible, for the first time, to project real IOLs on the patient’s eyes and assess the absolute visual performance with IOLs of different powers and designs. This system holds promise to become a reference for validations of visual validations with other simulators.

2. Methods

We designed and implemented a Rassow system in an Adaptive Optics set-up and tested optical quality on bench and visual quality in patients with monofocal, multifocal, and extended-depth-of-focus IOLs.

2.1 Rassow system configuration

Figure 1 shows the configuration of the implementation. The IOL is immersed in a cuvette composed of metallic support on which the IOL is mounted and filled with distilled water (n=1.33). The Rassow lens (a 60 mm EFL achromatic doublet (Newport PAC16AR.15) that compensates the 20D IOL power) was placed at a distance twice its focal length from the IOL.

 figure: Fig. 1.

Fig. 1. Schematic diagram of the Rassow telescope implemented in the AO system and the Badal lenses (f’R, focal length of Rassow lens; f’B1 and f’B2, focal lengths of Badal lenses 1 and 2 respectively; f’Coll, focal length of collimator lens). Distance between Badal lens 1 and 2 is adjusted. PP and RP represent pupil and retinal conjugated planes, respectively.

Download Full Size | PDF

The capabilities of the Rassow system are complemented with a Badal optometer that allows shifting focus in a ± 4D range around the nominal IOL power (20 D). The Badal system, which was already implemented in the system [30], is composed of two lenses (125 mm, EFL) and two mirrors mounted on a motorized stage that allows the distance between the lenses (and therefore defocus) to be adjusted. The zero-defocus position corresponds to the lens positions such that their separation is equal to the sum of their focal lengths. Higher separations induce positive defocus, and lower separations induce negative defocus.

2.2 Optical computer simulations

Computer simulations of the optical performance of the Rassow and Badal Optometer system (Fig. 1) were performed using Ray Tracing software (Zemax–EE Optical Design Program 2005, Zemax Development Corporation) (Fig. 1), for 4-mm pupil diameter and 555-nm wavelength.

The difference between entrance pupil diameter and iris pupil diameter [29] was theoretically calculated using OpticStudio for corneas of different power. Wavefront maps and the corresponding Modulation Transfer Functions (MTF) of the IOL in the cuvette were obtained on-axis (0 deg) and off-axis at the maximum subtended angle in the AO system (2 deg).

2.3 AO-set up (and implementation of the Rassow system in there)

The Rassow IOL projection system was incorporated in a testing channel in the Viobio Lab AOII, described in previous publications [13,23,30]. The system consists of different channels, the following of which are used in the current study: (1) The psychophysical channel contains a Digital Micro Mirrors Device (DMD, DLP Discovery 4100 0.7 XGA, Texas Instruments Incorporated, USA) on a retinal plane. The DMD Channel was monochromatically illuminated with light coming from a supercontinuum laser source (SCLS, Fianium Ltd, United Kingdom) with a selectable wavelength, 555 nm in this study. (2) The AO-Channel was composed of a Hartmann-Shack wavefront sensor (40×32 microlenses and 3.6-mm effective diameter, HASO 32 OEM, Imagine Eyes, France) and an electromagnetic deformable mirror (52 actuators, 15-mm effective diameter, and 50-µm stroke; MIRAO, Imagine Eyes, France), that was used to measure, correct, and induce aberrations. The illumination source was a SuperLuminescent Diode coupled to an optical fiber (Superlum, Ireland) emitting at 827 nm. (3) The natural pupil monitoring system consists of a camera (DCC1545M, High-Resolution USB2.0 CMOS Camera, Thorlabs GmbH, Germany) conjugated with the eye’s pupil: (6) A testing channel with a conjugate pupil plane (with ×1 magnification) where phase plates or an optotunable lens working under the principle of temporal multiplexing (SimVis) were projected on the subject’s pupil. The Rassow IOL projection system was mounted in a secondary optical breadboard and incorporated in this last channel above the optical set up utilizing two flip-up mirrors so that the IOL in the cuvette was placed in a pupil conjugate plane. Figure 2 shows the schematic diagram of the implementation of the new channel in the Viobio Lab AOII system. The distance between lenses L16 and L17 was modified to control the magnification of the IOL in the Rassow system onto the eye’s pupil.

 figure: Fig. 2.

Fig. 2. (a) Schematic diagram of the Viobio Lab AOII system with the new Rassow-IOL projection system implemented in the testing channel. The inset shows the schematic diagram and a photograph of the Rassow-IOL projection system. Refer to [13,23,30] for a full explanation of the existing system. Lenses and mirrors in the new system are: L11 and L15, 150 mm focal length lens; L16 is the Rassow lens, with 60 mm focal length lens; L17 and L18, 63.5 mm focal length lenses; M10, M11, M12 and M13 are mirrors; PP are conjugated pupil planes. (b) New channel implemented in the system.

Download Full Size | PDF

2.4 Tested IOLs

Three commercial IOLs from PhysIOL were tested in the cuvette and projected at the entrance pupil of an artificial eye or the subject’s eye pupil: Podeye IOL (Monofocal, 20 D), FineVision Trifocal IOL (19.5 D), and Isopure (Isofocal EDOF IOL, 21 D). Four additional monofocal IOLs by Alcon (Acrysof, powers: 18.00, 19.50, 20.50, and 23.00 D) and one monofocal IOL by Johnson & Johnson (Tecnis, power: 22 D) were measured for Badal calibration.

The Podeye is a monofocal biconvex, hydrophobic, aspheric monofocal IOL, aiming at correcting a large proportion of the corneal SA (-0.11 µm SA). The FineVision is a diffractive, trifocal, hydrophilic, and aspheric trifocal IOL [31], with three foci, 0.0 D for far-vision, +1.75 D addition for intermediate-vision, and +3.50 D addition for near-vision. The Isofocal is refractive, hydrophobic, with smooth aspheric surfaces (described by the radius of curvature, and 4 conic constants per surface [32]), aiming at providing optimal vision at far and an extended-depth-of-focus into intermediate distances.

2.5 Subjects

Three subjects participated in the experiment. Subjects were 26, 27, and 28 years old and nearly emmetropic (Spherical refractive error:+ 0.75, +1.50, and -0.75 D; and cylinder: 0, -0.75, and -0.50 D). All measurements were performed under paralyzed accommodation with Tropicamide (2 drops at the beginning of the session, and 1 drop every hour if required).

All protocols met the tenets of the Declaration of Helsinki and had been previously approved by the Spanish National Research Council (CSIC) Bioethical Committee. All participants were acquainted with the nature of the study and provided written informed consent.

2.6 Through-focus on-bench optical quality

An artificial eye consisting of an objective lens (50.8 mm of focal length) and a CCD camera (DCC1240C - High-Sensitivity USB 2.0 CMOS Camera, 1280 × 1024, Global Shutter, Color Sensor; Thorlabs GmbH, Munich, Germany) on the focal plane of the objective was used. Through-focus series of images of an E letter (1.62° angular subtend) displayed in the DMD projector illuminated with green light (555 nm, coherence broken with a holographic diffuser) were taken on the CCD camera (1 Pass), for the three IOLs inserted in the cuvette. The focus was changed in 0.25 D steps using the Badal system. All images were captured with 4.5 mm pupil diameter, with constant laser power (6.90 µW) and similar camera settings, with the DM only correcting the aberrations of the system. The optical quality was assessed as the 2D cross-correlation [33] between each image in the TF series with the image obtained with the monofocal IOL (Podeye) at best focus.

2.7 Through focus visual acuity measurements

Visual Acuity (VA) was measured using a tumbling E letter test with an 8-Alternative Forced Choice (8AFC) procedure and a QUEST (Quick Estimation by Sequential Testing) [34,35] algorithm programmed with the Psychtoolbox package of Matlab [36,37] to calculate the size and orientation of the following presented E letter according to the subject’s response. The E letter was presented for 0.5 seconds, and subjects were asked to indicate the orientation of the E letter. Each VA measurement consisted of 32 trials and 20 reversals. VA was estimated as the mean of the last 10 reversals and its variability, as the standard deviation of that 10 values. VA was measured in a focus range of +2 to -5 D, with a finer sampling was performed near the experimentally found VA peaks.

VA measurements were performed under cycloplegia using a 4.5 mm pupil. The subject’s eye pupil center was aligned with the optical axis of the system with an x-y-z stage and stabilized and a dental impression on a bite bar. Subjects found their best focus with the Badal system while looking at a Maltese cross stimulus prior to performing the VA test for each IOL. The psychophysical stimulus was presented on the DMD, illuminated at 555 nm, and viewed through the real IOL immersed in the cuvette, as well as without the lens. Measurements were performed with the three IOLs in the same session in random order. Subjects rested for 5 min while the corresponding IOL was inserted in the system, with at least 3 breaks. A typical session lasted 120 min.

The multifocal and extended-depth-of-focus IOLs were compared in terms of visual benefit at intermediate (0.60 m) and near (0.40 m). A Visual Benefit metric was calculated as the difference between the logMAR Visual Acuity with a multifocal lens and a monofocal Podeye IOL. Visual Benefit was calculated for both intermediate and near distances [3840]. Visual Imbalance was calculated as the logMAR Visual Acuity standard deviation in a 4.25 D range (from +1 D to -3.25 D).

3. Results

The following results show the optical simulations of the new channel of the system and the on-bench and in vivo measurements.

3.1 Optical simulations

The first simulations aimed at estimating the power of the Rassow lens that compensates for the base power of a 20 D IOL. The Rassow lens was assumed to be a paraxial lens and its specifications were obtained from the closest equivalent lens in Newport’s catalog. After selection of the final focal length of the Rassow lens, the distances between L10-L15 and L15-L16 (Fig. 2) were recalculated to compensate for the difference between the theoretical paraxial lens and the real one. The final focal length of the Rassow lens was 60 mn, and distances L10-L15 and L15-L16 were 213 and 210 mm respectively.

The optical quality of the Rassow system was evaluated computationally using the MTF, the spot diagram, and the wave aberrations of the system. Figure 3 shows the estimated MTFs (a) and wave aberrations (b) at 0° and 2° in vertical for a monofocal IOL immersed in the cuvette. Small amounts of spherical aberration were found at 0° (0.02µm), and also of astigmatism (0.05µm) and coma (0.02µm) at 2°.

 figure: Fig. 3.

Fig. 3. Computer simulations from Zemax Ray Tracing software of the optical quality in the image surface with 4.5 mm pupil size for 0° and 2° in terms of MTF (a) and wavefront maps (b).

Download Full Size | PDF

We calculated the magnification factor needed to reproduce the decrease in effective diameter at the IOL plane in the eye by the effect of rays’ convergence at the cornea. Calculations were performed assuming different corneal powers (39, 43, and 47 D), anterior chamber depths (3.2, 4.5, and 5.5 mm), and entrance pupil diameters (3 and 4.5 mm). The resulting iris pupil diameter and the correspondent magnification are shown in Table 1. A magnification of ×0.87, which corresponds to an eye model with a 43 D cornea and an anterior chamber depth of 3.2 mm was implemented in the system.

Tables Icon

Table 1. Difference of magnification (%) in the IOL plane with respect to the entrance beam for different corneal powers, anterior chamber depths, and entrance beam sizes.

3.2 Experimental performance of Rassow system

We tested experimentally the magnification of the Rassow system for different positions of the Badal system, the experimental optical performance with monofocal IOLs of different powers, and the spherical aberration of the system.

Figure 4 shows the experimental measurements of magnification in the cuvette plane, for different positions in the Badal optometer. The average measured magnification was 0.88. The average standard deviation of measured magnification over repeated measurements was 0.02. The standard deviation of measured magnification across defocus positions in the Badal system was 0.01.

 figure: Fig. 4.

Fig. 4. Magnification in the cuvette plane concerning the pupil plane. Error bars indicate standard deviation along 5 measurements of the same condition.

Download Full Size | PDF

The optical performance of the Rassow system was tested using five monofocal IOLs of different powers (Fig. 5), using on bench images of an E-letter, and a cross-correlation optical quality metric (using an E letter image with no lens in the cuvette, with the flip-up mirrors up, as a reference). The Badal optometer allowed spherical compensation for other powers, linearly (R2=0.99) in a 5 D range. The measured optical quality metric varied less than 0.02 across IOLs.

 figure: Fig. 5.

Fig. 5. On-bench images through monofocal IOLs of 18.00D, 19.50D, 20.50D, 22D, and 23.00D and the needed number of diopters to compensate with the Badal system.

Download Full Size | PDF

The spherical aberration induced by the Rassow system was measured with no IOL in the cuvette and defocus corrected with a 20 D trial lens placed in a conjugate pupil plane. In comparison with the spherical aberration measured in the cuvette with monofocal spherical IOLs (-0.07 um and -0.015 um at 4.5-mm pupils, which is consistent with the negative spherical aberrations of these IOLs) measurements of the spherical aberration of the cuvette alone showed very small amounts of positive spherical aberration (<0.02 μm). The experimental results of spherical aberration measurements are in good agreement with predictions from computer simulations of the Rassow system with and without IOLs, as shown in Fig. 6, with the measured values within 75% of the predicted values. RMS for 3rd order aberrations was lower than 0.04 μm and all values of aberrations of the cuvette with and without IOL were lower than 0.15 μm.

 figure: Fig. 6.

Fig. 6. Spherical aberration of the Rassow system without (cuvette alone) IOL in the cuvette and with (IOL in cuvette). The tested IOLs are monofocal IOLs Podeye (monofocal 1) and Acrysoft (monofocal 2). Experimental measurements (solid bars) are compared with predictions from computer simulations (patterned bars). Error bars indicate the standard deviation across three repeated measurements. Data are for 4.5 mm pupil diameter.

Download Full Size | PDF

3.3 Through-focus on-bench measurements

Figure 7(a) shows the TF on-bench images of an E letter obtained for three IOLs: Monofocal Isofocal and Trifocal inserted in the cuvette. The 0 D reference was shifted to the focus position that produced the highest image quality and was shifted by 0, +0.5 and -1.25 D in the monofocal (20 D), trifocal (19.5), and Isofocal (21 D) IOLs, respectively.

 figure: Fig. 7.

Fig. 7. (a) TF on-bench images of an E-stimulus in the artificial eye through the monofocal (top panel), Trifocal (middle panel), and Isofocal (bottom panel) and IOLs. (b) TF image correlation metric for the Monofocal (blue line), Trifocal (green line), and Isofocal (solid orange lines) IOLs. Data are for 4-mm pupil diameters.

Download Full Size | PDF

Qualitatively, the higher quality corresponds to the monofocal IOL at Far; however, the image quality decreases rapidly out of focus. The trifocal IOL shows two distinctive broader areas of higher quality at far and intermediate-near distance. The Isofocal IOL shows a wider area of good focus, consistently with its extended depth of focus nature. Figure 7(b) shows the corresponding TF optical quality correlation metric, using the image of the monofocal IOL at best focus as a reference. For IOLs projected on an artificial eye, the optical quality at far does not decrease bellow 0.95 (relative to monofocal at a far distance as a reference) in neither trifocal and Isofocal IOLs, reflecting high optical quality at distance. The Visual Benefit at intermediate (-1.5 D) is 0.03 for the trifocal and 0.07 for the Isofocal, and at near (-3 D) is 0.14 for the trifocal and 0.01 for the Isofocal. The Visual Imbalance was 0.1, 0.04, and 0.07 for the monofocal, trifocal, and Isofocal, respectively.

3.4 Through-focus VA results

Figure 8 shows the through-focus VA in three patients with the three IOLs projected on the pupil plane with the Rassow-IOL system. The VA with no lens in the cuvette was also measured at a far distance using the former psychophysical channel, with the flip-up mirrors up. The wave aberrations of the eye (along with the corresponding RMS for astigmatism and higher-order aberrations) are shown for each subject. The optical quality of the three subjects measured is shown in terms of RMS wavefront error. The average VA at far with no lens was -0.08 ± 0.07. VA at far with the monofocal IOL and no IOL differed by less than 0 ± 0.03 logMAR on average. Trifocal IOLs increase VA at intermediate, but most predominantly at near, with all subjects showing a peak in VA at around 3.5 D (0.24 ± 0.04 logMAR on average), while the Isofocal IOL provides the highest VA at intermediate distances (0.21 ± 0.02 logMAR on average). The lowest values of VA (both with monofocal and multifocal IOLs) were found in the most aberrated subject (S#2).

 figure: Fig. 8.

Fig. 8. TF logMAR VA for 3 (a, b and c) subjects with no lens (black symbol) and through the monofocal (blue lines), Isofocal (orange lines), and Trifocal (green lines) IOLs projected with the Rassow system for a pupil diameter of 4.5 mm. (d) Wave aberration and RMS at zero defocus.

Download Full Size | PDF

Figure 9 shows the visual benefit at intermediate (9(a)) and near (9(b)) distances for the trifocal and Isofocal IOLs relative to the monofocal IOL, and the visual imbalance (9(c)) for the three IOLs (for each subject and averaged across subjects). In vivo measurements, in terms of TF VA, replicate, on average, on-bench findings. We found that both curves, correlation metric for TF images and TF VA on subjects, are correlated across subjects, especially for the monofocal IOL (slope = -2.57, r = 0.83; slope = -1.85, r = 0.52; and slope = -1.39, r = 0.43, on average for monofocal, trifocal and Isofocal IOLs respectively).

 figure: Fig. 9.

Fig. 9. Visual benefit at intermediate, 0.6 m (a) and near 0.4 m (b) distance (VA MIOL @ int – VA Monofocal @ int; VA MIOL near – VA Monofocal near). (c) Visual imbalance on a range of 4.25 D for monofocal (blue), trifocal (green), and Isofocal EDOF (orange) IOLs for each subject and on average (solid bars). Error bars stand for standard deviation across subjects.

Download Full Size | PDF

4. Discussion

In this study, we introduced a system to project physical IOLs on the pupil's eyes through a Rassow telescope. This represents a new opportunity to validate the experimental simulations of vision with IOLs mapped in an Adaptive Optics element (spatial phase maps) or SimVis simulators (temporal multiplexing), as the physical IOLs can be used as a reference. Previous work used 0 D IOLs in a cuvette [4], which were projected on the eye as if they were phase maps. To our knowledge, this is the first time that a Rassow – IOL projection system has been incorporated into a channel of an Adaptive Optics system, allowing projecting IOLs of standard base power and visual testing.

The use of a Rassow telescope was proposed in the ‘90s, and to a large extent is the basis for systems that project IOLs on the eye. A recent study compared the Contrast Sensitivity Function through physical IOLs in a laboratory-based Rassow system and the commercial prototype VirtIOL [41]. Measurements were done through three IOLs (monofocal, multifocal, and EDOF) on twenty-one subjects. There was no attempt to assess absolute performance values for each lens, and the study was rather limited to relative comparisons.

Our Adaptive Optics and IOL SimVis visual simulators are based on dynamic elements and therefore fully programmable. This allows a rapid switch between corrections. In contrast, in the Rassow system, the IOLs need to be interchanged to compare vision through different designs allowing only static simulations. However, the incorporation of the Rassow IOL projection system into the channels of an Adaptive Optics system provides a reference for validation of the IOL designs with standard powers mapped in the active elements.

To our knowledge, this study asses for the first time the aberrations of the Rassow system (both computationally and experimentally). Since some of the IOLs (for example monofocal aspheric IOLs or the Isofocal IOL) modulate the spherical aberration of the eye, guaranteeing that the Rassow system does not induce higher-order aberrations is critical. We found that the induced spherical aberration is negligible. In addition, in our system, residual aberrations were compensated using the deformable mirror. Nevertheless, our findings confirm that the Rassow system can be used as a stand-alone simulator for comparing IOLs, which can be built using off-the-shelf optical components without affecting the optics of the IOL to be projected.

To the best of our knowledge, a novelty of this work is the adjustment of the system magnification to compensate the effect of the convergence of the rays by the cornea onto the eye’s pupil plane, which effectively decreases the iris pupil diameter that is illuminated [29]. In contrast, the standard 4-focal configuration of the published Rassow systems projects the IOL on the eye assuming that the entrance pupil and the iris pupil diameter are equal. We reduced the effective pupil diameter of the IOL plane by a factor of 0.87 and we calculated that this factor was affected by less than 10% by changes in corneal power, ACD, or entrance pupil diameter using realistic values. While the effect of magnification may be relatively small with certain IOLs that are less dependent on pupil diameter (i.e monofocal or diffractive IOLs), it is expected to play a relevant role with IOLs that modulate higher-order aberrations, the spherical aberration in particular, which are pupil-size dependent.

Optical quality and vision with the projected IOLs match the expectations from these IOLs, as well as prior, defocus curves in patients implanted with these IOLs. We found higher quality at far with the monofocal IOL, followed by Isofocal and Multifocal. The Isofocal IOL outperforms the monofocal and multifocal at intermediate, and the Multifocal IOL outperforms the monofocal and Isofocal at near. Defocus curves with the physically projected IOLs are in good agreement with those obtained by simulating the same IOLs with Spatial Light Modulator or SimVis, in terms of the shape of the curve, with peaks and depth of focus as expected [4]. The combination of these simulation strategies and the Rassow system in the same instrument will allow a direct comparison of adaptive optics and static simulations in phakic patients and therefore direct validations of the visual simulators for different lenses. Simulations (with either physical IOL projected on the eye, or the IOL represented as phase maps or temporal patterns on SLM or SimVis) are generally performed on phakic patients, and therefore all share the assumption that the impact of the native aberrations of the crystalline lens in visual performance is secondary to that of the IOL design. This assumption has proved valid, at least with multifocal IOLs, as shown in patients before and after implantation [23] and in a study that concluded that the influence of the crystalline lens’ aberrations is negligible in terms of VA for medium and small pupil diameters [42]. With monofocal IOLs of magnitudes of spherical aberration comparable to those of the ocular components, the presence of spherical aberration from the crystalline lens (negative spherical aberration in young eyes [43] and shifting towards positive in older eyes [44]) likely modulates the spherical aberration in different ways than in the pseudophakic eye. The lack of compensation of spherical aberration of the lens in Rassow-IOL-in cuvette stand-alone system could be a limitation of the system. In our AO system, spherical aberration could be compensated with the deformable mirror, if deemed necessary in certain applications.

In summary, we have presented and validated an optical system for the projection of real IOLs (immersed in a cuvette) on the subject’s eye. The system does not induce any large additional aberrations and replicates the effective diameter of the IOL when it is implanted in the eye. We have demonstrated the system on bench (with an artificial eye) and in real subjects (VA) through three different types of lenses (refractive monofocal, diffractive trifocal, and refractive Isofocal) allowing patients to experience vision and evaluate through-focus visual performance without implanting the IOL. The through-focus visual performance with three different IOLs projected on the eye using the Rassow system show differences across individuals (likely arising from differences across patients with natural aberrations) and across lens designs, with presbyopia-correcting IOLs producing a clear visual benefit at intermediate (Isofocal IOL) and near (trifocal IOL). The incorporation of a validated channel for projecting IOLs on the eye adds a reference in Adaptive Optics visual simulators, in which the IOLs are mapped (spatially or temporally) using active elements.

Funding

Spanish Government ((FIS2017-84753R)).

Disclosures

MV: 2EyesVision (I), CML: 2EyesVision (E), CD: 2EyesVision (I, P, E), SM: 2EyesVision (I, P).

Data availability

Data underlying the results presented in this paper are available in Ref. [45].

References

1. “Cataract surgical rates,” Community eye Heal. 30(100), 88–89 (2017).

2. J. L. Alió, A. Grzybowski, and D. Romaniuk, “Refractive lens exchange in modern practice: when and when not to do it?” Eye and Vis 1(1), 10 (2014). [CrossRef]  

3. R. Rampat and D. Gatinel, “Multifocal and extended depth-of-focus intraocular lenses in 2020,” Ophthalmology (2020).

4. M. Vinas, C. Benedi-Garcia, S. Aissati, D. Pascual, V. Akondi, C. Dorronsoro, and S. Marcos, “Visual simulators replicate vision with multifocal lenses,” Sci. Rep. 9(1), 1539 (2019). [CrossRef]  

5. S. Marcos, J. S. Werner, S. A. Burns, W. H. Merigan, P. Artal, D. A. Atchison, K. M. Hampson, R. Legras, L. Lundstrom, G. Yoon, J. Carroll, S. S. Choi, N. Doble, A. M. Dubis, A. Dubra, A. Elsner, R. Jonnal, D. T. Miller, M. Paques, H. E. Smithson, L. K. Young, Y. Zhang, M. Campbell, J. Hunter, A. Metha, G. Palczewska, J. Schallek, and L. C. Sincich, “Vision science and adaptive optics, the state of the field,” Vision Res. 132, 3–33 (2017). [CrossRef]  

6. M. Vinas, C. Dorronsoro, V. Gonzalez, D. Cortes, A. Radhakrishnan, and S. Marcos, “Testing vision with angular and radial multifocal designs using adaptive optics,” Vision Res. 132, 85–96 (2017). [CrossRef]  

7. E. J. Fernández, P. M. Prieto, and P. Artal, “Binocular adaptive optics visual simulator,” Opt. Lett. 34(17), 2628 (2009). [CrossRef]  

8. R. Sabesan, K. Ahmad, and G. Yoon, “Correcting highly aberrated eyes using large-stroke Adaptive Optics,” J Refract Surg 23(9), 947–952 (2007). [CrossRef]  

9. C. Schwarz, S. Manzanera, P. M. Prieto, P. Piers, and P. Artal, “Binocular visual acuity for the correction of spherical aberration in polychromatic and monochromatic light,” Journal of Vision 14(2), 1–11 (2014). [CrossRef]  

10. S. Manzanera and P. Artal, “Minimum change in spherical aberration that can be perceived,” Biomed. Opt. Express 7(9), 3471–3477 (2016). [CrossRef]  

11. L. Zhao, N. Bai, X. Li, L. S. Ong, Z. P. Fang, and A. K. Asundi, “Efficient implementation of a spatial light modulator as a diffractive optical microlens array in a digital Shack – Hartmann wavefront sensor,” Appl. Opt. 45(1), 90 (2006). [CrossRef]  

12. J. Tabernero, C. Schwarz, and E. J. Ferna, “Binocular visual simulation of a corneal inlay to increase depth of focus,” Invest. Ophthalmol. Vis. Sci. 52(8), 5273–5277 (2011). [CrossRef]  

13. M. Vinas, C. Dorronsoro, A. Radhakrishnan, C. Benedi-Garcia, E. A. LaVilla, J. Schwiegerling, and S. Marcos, “Comparison of vision through surface modulated and spatial light modulated multifocal optics,” Biomed. Opt. Express 8(4), 2055 (2017). [CrossRef]  

14. S. Marcos, C. Bened, S. Aissati, A. M. G. Ramos, C. M. Lago, A. Radhkrishnan, M. Romero, S. Vedhakrishnan, and L. Sawides, “VioBio lab adaptive optics: technology and applications by women vision scientists,” Ophthalmic Physiol Opt 40, 75–87 (2020). [CrossRef]  

15. R. Rosén, L. Lundström, and P. Unsbo, “Adaptive optics for peripheral vision,” J. Mod. Opt. 59(12), 1064–1070 (2012). [CrossRef]  

16. J. Ruiz-Alcocer, D. Madrid-Costa, S. García-Lázaro, C. Albarrán-Diego, and T. Ferrer-Blasco, “Visual simulation through an aspheric aberration-correcting intraocular lens in subjects with different corneal profiles using adaptive optics,” Clin. Exp. Optom. 96(4), 379–384 (2013). [CrossRef]  

17. L. Hervella, E. A. Villegas, C. Robles, and P. Artal, “Spherical aberration customization to extend the depth of focus with a clinical adaptive optics visual simulator,” J Refract Surg 36(4), 223–229 (2020). [CrossRef]  

18. P. de Gracia, C. Dorronsoro, Á. Sánchez-González, L. Sawides, and S. Marcos, “Experimental simulation of simultaneous vision,” Investig. Ophthalmol. Vis. Sci. 54(1), 415–422 (2013). [CrossRef]  

19. A. Radhakrishnan, C. Dorronsoro, L. Sawides, and S. Marcos, “Short-term neural adaptation to simultaneous bifocal images,” PLoS One 9(3), e93089 (2014). [CrossRef]  

20. C. Dorronsoro, A. Radhakrishnan, J. R. Alonso-Sanz, D. Pascual, M. Velasco-Ocana, P. Perez-Merino, and S. Marcos, “Portable simultaneous vision device to simulate multifocal corrections,” Optica 3(8), 918 (2016). [CrossRef]  

21. V. Akondi, C. Dorronsoro, E. Gambra, and S. Marcos, “Temporal multiplexing to simulate multifocal intraocular lenses: theoretical considerations,” Biomed. Opt. Express 8(7), 3410 (2017). [CrossRef]  

22. C. Dorronsoro, X. Barcala, E. Gambra, V. Akondi, L. Sawides, Y. Marrakchi, V. Rodriguez-Lopez, C. Benedi-Garcia, M. Vinas, E. Lage, and S. Marcos, “Tunable lenses: dynamic characterization and fine-tuned control for high-speed applications,” Opt. Express 27(3), 2085 (2019). [CrossRef]  

23. M. Vinas, S. Aissati, M. Romero, C. Benedi-Garcia, N. Garzon, F. Poyales, C. Dorronsoro, and S. Marcos, “Pre-operative simulation of post-operative multifocal vision,” Biomed. Opt. Express 10(11), 5801–5817 (2019). [CrossRef]  

24. R. Kusel and B. Rassow, “Präoperative Abschätzung des mit Intraokularlinsen erreichbaren Sehvermögens,” Klin Monatsbl Augenheilkd 215(08), 127–131 (1999). [CrossRef]  

25. F. Schaeffel and H. Kaymak, “A rapid and convenient procedure to evaluate optical performance of intraocular lenses,” Photonics 1(3), 267–282 (2014). [CrossRef]  

26. J. Pujol, M. Aldaba, A. Giner, S. Arasa, and J. Luque, “Visual performance evaluation of a new multifocal intraocular lens design before surgery,” Investig. Ophthalmol. Vis. Sci.55(13), 3752 (2014).

27. W. Brezna, K. Lux, N. Dragostinoff, C. Krutzler, N. Plank, R. Tobisch, A. Boltz, G. Garhöfer, R. Told, K. Witkowska, and L. Schmetterer, “Psychophysical vision simulation of diffractive bifocal and trifocal intraocular lenses,” Transl. Vis. Sci. Technol. 5(5), 13 (2016). [CrossRef]  

28. H.-L. Liou and N. A. Brennan, “Anatomically accurate, finite model eye for optical modeling,” J. Opt. Soc. Am. A 14(8), 1684 (1997). [CrossRef]  

29. S. Norrby, P. Piers, C. Campbell, and M. Van Der Mooren, “Model eyes for evaluation of intraocular lenses,” Appl. Opt. 46(26), 6595–6605 (2007). [CrossRef]  

30. M. Vinas, C. Dorronsoro, D. Cortes, D. Pascual, and S. Marcos, “Longitudinal chromatic aberration of the human eye in the visible and near infrared from wavefront sensing, double-pass and psychophysics,” Biomed. Opt. Express 23(4), 513–522 (2015). [CrossRef]  

31. D. Gatinel, C. Pagnoulle, Y. Houbrechts, and L. Gobin, “Design and qualification of a diffractive trifocal optical profile for intraocular lenses,” J. Cataract Refract. Surg. 37(11), 2060–2067 (2011). [CrossRef]  

32. D. Fernández, S. Barbero, C. Dorronsoro, and S. Marcos, “Multifocal intraocular lens providing optimized through-focus performance,” Opt. Lett. 38(24), 5303–5306 (2013). [CrossRef]  

33. Gonzalez and Woods, “Digital Image Processing,” in 3rd ed. (Pearson / Prentice Hall, 2008).

34. Z.-L. Lu and B. Dosher, Visual Psychophysics. From Laboratory to Theory (MIT books, 2014).

35. A. B. Watson and D. G. Pelli, “Quest: A Bayesian adaptive psychometric method,” Percept. Psychophys. 33(2), 113–120 (1983). [CrossRef]  

36. M. Borgo, A. Soranzo, and M. Grassi, MATLAB for Psychologist (Springer US, 2013), 53(9).

37. D. H. Brainard, “The psychophysics toolbox,” Spatial Vis 10(4), 433–436 (1997). [CrossRef]  

38. International Organization for Standardization, “Ophthalmic optics—Visual acuity testing—Standard and clinical optotypes and their presentation—Amendment 1,” U.S. patent ISO 8596:2017/AMD 1:2019 (2019).

39. International Organization for Standardization, “Ophthalmic optics—Visual acuity testing—Standard and clinical optotypes and their presentation,” U.S. patent ISO 8596:2017 (2017).

40. X. Barcala, M. Vinas, M. Romero, E. Gambra, J. L. Mendez-Gonzalez, S. Marcos, and C. Dorronsoro, “Multifocal acceptance score to evaluate vision: MAS-2EV,” Sci. Rep. 11(1), 1397 (2021). [CrossRef]  

41. S. Wahl, C. Song, and A. Ohlendorf, “Comparison of two devices to simulate vision with intraocular lenses,” Clin. Ophthalmol. 13, 123–130 (2019). [CrossRef]  

42. E. A. Villegas, S. Manzanera, C. M. Lago, L. Hervella, L. Sawides, and P. Artal, “Effect of crystalline lens aberrations on adaptive optics simulation of intraocular lenses,” J Refract Surg 35(2), 126–131 (2019). [CrossRef]  

43. G. Smith, M. J. Cox, R. Calver, and L. F. Garner, “The spherical aberration of the crystalline lens of the human eye,” Vision Res. 41(2), 235–243 (2001). [CrossRef]  

44. A. C. Kingston and I. G. Cox, “Population spherical aberration: Associations with ametropia, age, corneal curvature, and image quality,” Clin. Ophthalmol. 7, 933–938 (2013). [CrossRef]  

45. C. Benedi, M. Viñas, C. M. Lago, S. Aissati, A. de Castro, C. Dorronsoro, and S. Marcos, Data for the paper “Optical and visual quality of real intraocular lenses physically projected on the patient’s eye,” CSIC, 2021, http://dx.doi.org/10.20350/digitalCSIC/13974.

Data availability

Data underlying the results presented in this paper are available in Ref. [45].

45. C. Benedi, M. Viñas, C. M. Lago, S. Aissati, A. de Castro, C. Dorronsoro, and S. Marcos, Data for the paper “Optical and visual quality of real intraocular lenses physically projected on the patient’s eye,” CSIC, 2021, http://dx.doi.org/10.20350/digitalCSIC/13974.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (9)

Fig. 1.
Fig. 1. Schematic diagram of the Rassow telescope implemented in the AO system and the Badal lenses (f’R, focal length of Rassow lens; f’B1 and f’B2, focal lengths of Badal lenses 1 and 2 respectively; f’Coll, focal length of collimator lens). Distance between Badal lens 1 and 2 is adjusted. PP and RP represent pupil and retinal conjugated planes, respectively.
Fig. 2.
Fig. 2. (a) Schematic diagram of the Viobio Lab AOII system with the new Rassow-IOL projection system implemented in the testing channel. The inset shows the schematic diagram and a photograph of the Rassow-IOL projection system. Refer to [13,23,30] for a full explanation of the existing system. Lenses and mirrors in the new system are: L11 and L15, 150 mm focal length lens; L16 is the Rassow lens, with 60 mm focal length lens; L17 and L18, 63.5 mm focal length lenses; M10, M11, M12 and M13 are mirrors; PP are conjugated pupil planes. (b) New channel implemented in the system.
Fig. 3.
Fig. 3. Computer simulations from Zemax Ray Tracing software of the optical quality in the image surface with 4.5 mm pupil size for 0° and 2° in terms of MTF (a) and wavefront maps (b).
Fig. 4.
Fig. 4. Magnification in the cuvette plane concerning the pupil plane. Error bars indicate standard deviation along 5 measurements of the same condition.
Fig. 5.
Fig. 5. On-bench images through monofocal IOLs of 18.00D, 19.50D, 20.50D, 22D, and 23.00D and the needed number of diopters to compensate with the Badal system.
Fig. 6.
Fig. 6. Spherical aberration of the Rassow system without (cuvette alone) IOL in the cuvette and with (IOL in cuvette). The tested IOLs are monofocal IOLs Podeye (monofocal 1) and Acrysoft (monofocal 2). Experimental measurements (solid bars) are compared with predictions from computer simulations (patterned bars). Error bars indicate the standard deviation across three repeated measurements. Data are for 4.5 mm pupil diameter.
Fig. 7.
Fig. 7. (a) TF on-bench images of an E-stimulus in the artificial eye through the monofocal (top panel), Trifocal (middle panel), and Isofocal (bottom panel) and IOLs. (b) TF image correlation metric for the Monofocal (blue line), Trifocal (green line), and Isofocal (solid orange lines) IOLs. Data are for 4-mm pupil diameters.
Fig. 8.
Fig. 8. TF logMAR VA for 3 (a, b and c) subjects with no lens (black symbol) and through the monofocal (blue lines), Isofocal (orange lines), and Trifocal (green lines) IOLs projected with the Rassow system for a pupil diameter of 4.5 mm. (d) Wave aberration and RMS at zero defocus.
Fig. 9.
Fig. 9. Visual benefit at intermediate, 0.6 m (a) and near 0.4 m (b) distance (VA MIOL @ int – VA Monofocal @ int; VA MIOL near – VA Monofocal near). (c) Visual imbalance on a range of 4.25 D for monofocal (blue), trifocal (green), and Isofocal EDOF (orange) IOLs for each subject and on average (solid bars). Error bars stand for standard deviation across subjects.