Interpreting holographic molecular binding assays with effective medium theory

Open Access Open Access

Abstract

Holographic molecular binding assays use holographic video microscopy to directly detect molecules binding to the surfaces of micrometer-scale colloidal beads by monitoring associated changes in the beads’ light-scattering properties. Holograms of individual spheres are analyzed by fitting to a generative model based on the Lorenz-Mie theory of light scattering. Each fit yields an estimate of a probe bead’s diameter and refractive index with sufficient precision to watch a population of beads grow as molecules bind. Rather than modeling the molecular-scale coating, however, these fits use effective medium theory, treating the coated sphere as if it were homogeneous. This effective-sphere analysis is rapid and numerically robust and so is useful for practical implementations of label-free immunoassays. Here, we assess how measured effective-sphere properties reflect the actual properties of molecular-scale coatings by modeling coated spheres with the discrete-dipole approximation and analyzing their holograms with the effective-sphere model.

© 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Holographic particle characterization works by recording in-line holograms of colloidal particles [1] and then analyzing each hologram pixel-by-pixel [2] with a generative model based on the Lorenz-Mie theory of light scattering [35]. This analysis yields each particle’s diameter with nanometer-scale precision and its refractive index to within a part per thousand [2,6], which should be fine enough to detect nanometer-scale molecules binding to the surfaces of micrometer-scale colloidal beads [7,8]. The resulting label-free assay for molecular binding, shown schematically in Fig. 1(a), shares the benefits of industry-standard bead-based immunoassays [9] while eliminating the time, cost and complexity of conventional fluorescence detection. Holographic immunoassays have immediate applications for medical diagnostics, a proof-of-concept study recently having been reported [8], and currently are being developed into diagnostic and serological tests for infection by SARS-CoV-2, the virus responsible for COVID-19.

 figure: Fig. 1.

Fig. 1. (a) Geometry of a bead-based molecular binding assay. A substrate bead with diameter $d_0$ and refractive index $n_0$ is coated with a layer of functional groups that bind target molecules from the surrounding solution. The surface groups are modeled as a dielectric layer of thickness $a_1$ and refractive index $n_1$. The coating of bound target molecules is modeled as a second layer of thickness $a_2$ and refractive index $n_2$. The bead is dispersed in a medium of refractive index $n_m$. The bounding sphere has diameter $d_p = d_0 + 2(a_1 + a_2)$. (b) Synthetic hologram of a coated bead computed with the DDA. Parameters are chosen to mimic a typical molecular binding assay based on a probe bead with a polystyrene core: $d_0 = 1$ µm, $n_0 = 1.60$, and molecular overlayers with $a_1 = a_2 = 10$ nm, $n_1 = n_2 = 1.55$. (c) The effective-sphere model treats the coated sphere as if it were a homogeneous dielectric sphere of effective diameter $d_p^{\ast }$ and refractive index $n_p^{\ast }$. (d) Fitting the coated-sphere’s hologram from (b) with the Lorenz-Mie theory for a homogeneous sphere yields an indistinguishable hologram for $d_p^{\ast } = 1.03$ µm and $n_p^{\ast } = 1.594$.

Download Full Size | PDF

Practical implementations of holographic molecular binding assays rely on the effective-sphere model [7,1014], which treats the scatterer as a homogeneous isotropic sphere regardless of its actual composition and microstructure. Values obtained for the particle’s diameter and refractive index then are interpreted to be characteristics of an effective sphere whose properties represent averages over the particle’s inhomogeneities using Maxwell Garnett effective-medium theory [15].

The effective-sphere model has been validated through studies on mesoporous spheres [14,16] and fractal colloidal clusters [1113,17]. In both cases, the two phases that make up the particle are distributed uniformly throughout the particle’s volume, which is consistent with the assumptions underlying Maxwell Garnett theory. The heterogeneity in bead-based molecular binding assays, by contrast, is restricted to thin surface layers. The present study assesses how properties of coated spheres estimated with the effective-sphere model reflect the presence and composition of such surface layers with the goal of guiding the development of fast and effective holographic molecular binding assays for research applications and medical testing.

We appraise the effective-sphere analysis of coated beads by using it to analyze synthetic holograms computed with the discrete-dipole approximation (DDA) [18]. Direct comparisons between ground truth values and fits to the effective-sphere model demonstrate that effective-sphere analysis usefully characterizes coated spheres, reliably detecting the presence of coatings and offering insights into their properties. The results of this study are consistent with trends in bead diameter and refractive index reported in experimental demonstrations of holographic molecular binding assays [7,8]. This positive outcome furthermore means that molecular binding assays based on holographic particle characterization can benefit from the speed and robustness of effective-sphere analysis.

1.1 Holographic particle characterization

The holograms used for holographic particle characterization are recorded by illuminating a colloidal dispersion with a collimated laser beam [1,2]. Light scattered by a colloidal particle interferes with the remainder of the beam in the focal plane of a microscope that magnifies the interference pattern and relays it to a video camera. Each magnified intensity pattern recorded by the camera is a hologram of the particles in the observation volume and encodes information on their three-dimensional positions, as well as their sizes, shapes and compositions.

Holographic particle characterization extracts information from recorded holograms by fitting to a generative model for the image-formation process [2]. A standard implementation models the incident beam as a unit-amplitude monochromatic plane wave at frequency $\omega$,

$$\mathbf{E}_0(\mathbf{r}, t) = e^{i k z} e^{-i \omega t} \, \hat{x},$$
that is linearly polarized along $\hat {x}$ and propagates along $\hat {z}$. The wavenumber, $k = 2 \pi n_m/\lambda$, depends on the laser’s vacuum wavelength, $\lambda$, and the refractive index of the medium, $n_m$, and is related to the frequency through the standard dispersion relation, $\omega = c k$, where $c$ is the speed of light in vacuum. This beam illuminates a particle located at $\mathbf {r}_p$ relative to the center of the microscope’s focal plane. The time-averaged intensity pattern recorded by the camera therefore may be modeled as
$$b(\mathbf{r}) = \left\vert { \hat{x} + e^{-i k z_p} \, \mathbf{f}_s(k(\mathbf{r} - \mathbf{r}_p)) } \right\vert^{2},$$
where $\mathbf {f}_s(k\mathbf {r})$ is the Lorenz-Mie scattering function for the particle [35]. In practice, an experimentally recorded hologram is corrected for the dark count of the camera and normalized by the intensity distribution of the illumination to facilitate comparison with Eq. (2).

The scattering function for a homogeneous sphere is parametrized by the sphere’s diameter, $d_p$, and refractive index $n_p$. Fitting a single-particle hologram to Eq. (2) involves optimizing these two parameters plus the particle’s three-dimensional position, $\mathbf {r}_p$. Published implementations [1923] can localize and characterize a sphere in a typical video image in under 50 milliseconds on a desktop workstation.

Our numerical studies are performed with parameters appropriate for the commercial xSight instrument for holographic particle characterization (Spheryx, Inc.) . This platform currently is being used to develop holographic antibody binding assays of the kind depicted in Fig. 1. It operates at a vacuum wavelength of $\lambda = 445$ nm and has an effective system magnification of 120 nm/pixel. We furthermore assume that the medium has the refractive index of water at the imaging wavelength, $n_m = 1.340$. No other calibration constants are required.

Validation experiments on colloidal size standards demonstrate that holographic particle characterization measurements, including those performed with xSight, can resolve the diameter of a micrometer-scale sphere with a precision of 5 nm [24]. Measurements on emulsion droplets demonstrate precision and reproducibility in the refractive index of 0.001 [25]. The former should suffice to detect the formation of a molecular coating through the associated change in the bead’s diameter [7,8], while the latter is useful for distinguishing different types of beads on the basis of their composition [26].

In principle, the hologram of a coated sphere could be analyzed by suitably generalizing the scattering function, $\mathbf {f}_s(k\mathbf {r})$, to account for the thicknesses and refractive indexes of its coatings [20,21]. Such generalizations have been applied successfully to experimental studies of polymer-shelled microbubbles [27] and evaporating aerosol droplets [28]. Introducing these additional adjustable parameters, however, reduces the likelihood that the fits will converge successfully and increases the measurement’s sensitivity to noise and uncorrected interference artifacts in the recorded images. The extracted values for the parameters, moreover, still would reflect effective-medium characterizations of molecular overlayers that could be patchy or incomplete. We therefore seek to understand how coatings influence effective-sphere parameters, which can be obtained rapidly and reliably with instruments that are commercially available.

2. Testing effective-sphere analysis of coated spheres with the discrete-dipole approximation

To assess how effective-sphere parameters reflect the actual properties of a coated sphere, we numerically compute the ideal hologram of a coated sphere using the discrete-dipole approximation (DDA) [18,29,30], and then analyze the hologram using Lorenz-Mie theory for a homogeneous sphere. The DDA treats a scatterer as an ensemble of point-like dipoles. Each elementary dipole scatters the incident plane wave, redirecting a portion to the imaging plane. Some of the scattered light reaches neighboring dipoles, which scatter it a second time. Some of that twice-scattered light also reaches the imaging plane and contributes to the computed image. The first-order DDA truncates the hierarchy after the first neighbor-scattering contribution both to reduce computation time and also to maintain numerical stability.

Our implementation uses the holopy interface [31] to the ADDA library [30]. We model a coated sphere by specifying the properties of the individual dipoles in a discrete three-dimensional square lattice. Dipoles located within the substrate sphere are assigned refractive indexes $n_0$. Those within coatings are assigned $n_1$ or $n_2$, as indicated in Fig. 1(a). The field scattered by this system of dipoles replaces $\mathbf {f}_s(k\mathbf {r})$ in Eq. (2) in computing the ideal hologram. The example in Fig. 1(b) was computed in this way.

Our effective-sphere analysis is performed with the pylorenzmie software suite that automates fits to Eq. (2). The hologram in Fig. 1(d) depicts the effective-sphere model’s best fit to the DDA hologram in Fig. 1(b). Whereas the DDA hologram requires 1 min to compute, the equivalent effective-sphere hologram can be computed in under 5 ms on the same computer hardware. Given that a single nonlinear least-squares fit can require dozens of hologram computations, the speed of the effective-sphere model offers clear advantages provided that its results can be interpreted productively.

Using pylorenzmie to analyze holograms recorded by xSight yields characterization results consistent with those reported by the instrument’s own analytical software [23]. We are confident, therefore, that results of our numerical experiments reflect the performance of the effective-sphere model for real-world measurements.

The standard computational lattice employed by ADDA has $10 n_0/n_m$ dipoles per wavelength, $\lambda$, along each axis. The resulting 37 nm lattice constant for our system is larger than the thickness of a typical molecular-scale coating. Dipoles representing a coating therefore are distributed in a grid across the surface of the model sphere, rather than forming a continuous layer. The spacing between the coating’s dipoles still is smaller than the wavelength of light, which reduces the influence of this discreteness. ADDA further minimizes grid-mediated artifacts through interpolation. As a result, reducing the lattice spacing from the default value to 10 nm does not appreciably change results for the coated spheres’ effective diameters or refractive indexes, although it increases computation time 300-fold.

2.1 Validating DDA and effective-sphere implementations

We validate the combination of DDA hologram synthesis and Lorenz-Mie analysis by performing numerical experiments on homogeneous spheres for which Eq. (2) should be exact. In this case, fitted values for the diameter, $d_p^{\ast }$, and the refractive index, $n_p^{\ast }$, should agree with the ground-truth values, $d_p$ and $n_p$, used as inputs. Figure 2(a) presents a map of the errors in the estimated diameter, $\epsilon _d(d_p, n_p) = d_p^{\ast } - d_p$, as a function of the ground-truth inputs. Figure 2(b) shows corresponding results for errors in the refractive index, $\epsilon _n(d_p, n_p) = n_p^{\ast } - n_p$. The DDA and Lorenz-Mie formulations agree on the spheres’ diameters to within the ±5 nm precision of holographic characterization measurements over more than half of the selected parameter range, the agreement being better than 1 nm for smaller spheres. Errors in refractive index generally are smaller than ±5 ppt across the entire domain, which also is reasonable for our application. Errors in both diameter and refractive index are consistent with previous reports of the performance of DDA calculations [29,30].

 figure: Fig. 2.

Fig. 2. Holographic images of homogeneous spheres computed with DDA are then fitted to Lorenz-Mie theory. Their deviations from the inputted values are $\epsilon _d(d_p, n_p) = d_p^{\ast } - d_p$ (a), and $\epsilon _n(d_p, n_p) = n_p^{\ast } - n_p$ (b). Errors can be positive or negative for both parameters, but in this region of parameter space, errors in $n_p$ are predominantly negative. Discrete points indicate the properties of polystyrene substrate beads used in published molecular binding assays [7,8] with diameters $d_p = 1$ µm and $d_p = 1.8$ µm. The smaller of these lies in a region of parameter space where agreement between DDA and Lorenz-Mie formulations is particularly good with $\epsilon _d < 5$ nm.

Download Full Size | PDF

The discrete data points in Fig. 2(a) and Fig. 2(b) represent the properties of the micrometer-diameter polystyrene spheres that were used for reported molecular binding assays [7,8]. The computational methods’ errors in size and refractive index are comparable to instrumental uncertainties for the smaller of these particles, but exceed instrumental uncertainties for the larger. We focus our numerical study, therefore, on the characteristics of the smaller commercial probe beads, both for their immediate practical application to medical testing and also to assess the effective-sphere model under conditions where our techniques are most reliable.

2.2 Single coatings

We next examine the effect of adding a single coating of a homogeneous material onto a uniform substrate sphere. This is a model for hologram formation by the probe beads used for holographic molecular binding assays. To facilitate comparison with recent experimental studies [7,8], we focus on the particular case of micrometer-diameter polystyrene spheres with $d_0 = 1$ µm and $n_0 = 1.60$. The coating thickness, $a_1$, and refractive index, $n_1$, are selected at random from the range $a_1 \in $ [5 nm, 20 nm] and $n_1 \in [1.4, 1.7]$. For each set of parameters, we use DDA to compute the coated sphere’s hologram and then fit to the effective-sphere model for $d_p^{\ast }$ and $n_p^{\ast }$.

In the special case that the coating has the same refractive index as the substrate, $n_1 = n_0$, adding a coating is equivalent to increasing the diameter of the bead: $d_p^{\ast } = d_0 + 2 a_1$. Alternatively, setting $n_1 = n_m$ is equivalent to not adding a coating, which should yield $d_p^{\ast } = d_0$.

The results in Fig. 3 confirm that adding a molecular-scale coating increases the apparent diameter of the bead, $\Delta _d = d_p^{\ast } - d_0 > 0$, provided the coating’s refractive index is greater than that of the medium. As expected, the coating’s influence on the effective diameter depends on its refractive index relative to that of the substrate, with low-index coatings increasing $d_p^{\ast }$ by less than their thickness and high-index coatings increasing $d_p^{\ast }$ by more. Indeed, Fig. 4(a) shows that the apparent bead diameter increases nearly linearly with $n_1$ for a fixed layer thickness, $a_1 = 10$ nm. Extrapolating to $n_1 = n_m$ yields $\Delta _d = 0$, as expected. Similarly, setting $n_1 = n_0$ yields $\Delta _d = 2 a_1$. From this, we obtain a phenomenological relationship between the effective diameter and the properties of the coating:

$$d_p^{\ast} = d_0 + 2 a_1 \, \frac{n_1 - n_m}{n_0 - n_m}.$$

 figure: Fig. 3.

Fig. 3. Influence of coating thickness, $a_1$, on effective sphere diameter, $d_p^{\ast }$, depends on the refractive index of the coating, $n_1$, relative to that of the substrate sphere, $n_0$. Data points are computed for a sphere with $d_0 = 1$ µm and $n_0 = 1.60$. The observed increase, $\Delta _d(a_1) = d_p^{\ast } - d_0$, increases monotonically with $n_1$ and agrees with the geometric size, $\Delta _d(a_1) = 2 a_1$. when $n_1 = n_0$, as indicated by the dashed line.

Download Full Size | PDF

 figure: Fig. 4.

Fig. 4. A 10 nm coating with variable refractive index $n_1$ is applied to a $d_0 = 1$ µm polystyrene bead and then is fitted to the effective-sphere model. (a) Shift in effective diameter $\Delta _d (n_1) = d_p^{\ast } - d_0$ is linear with coating index $n_1$, with $n_1 = n_0 = 1.6$ corresponding to a shift of 10 nm. A regression line (dotted) is fitted to the data. (b) Shift in effective refractive index, $\Delta _n (n_1) = n_p^{\ast } - n_0$, is positive for $n_1 > n_0$ and negative for $n_1 < n_0$. The horizontal dashed line indicates the baseline, $\Delta _n = 0$.

Download Full Size | PDF

Figure 4(b) shows the corresponding influence of the coating’s refractive index, $n_1$, on the effective sphere’s refractive index, $n_p^{\ast }$. Unlike the diameter, the change in refractive index, $\Delta _n = n_p^{\ast } - n_0$, does not depend linearly on $n_1$. Generally speaking, a low-index coating with $n_1 < n_0$ reduces the bead’s effective refractive index, $n_p^{\ast } < n_0$, while a high-index coating increases it. The overall magnitude of this effect is less than one part per thousand for the conditions considered, which suggests that trends in the effective refractive index are not likely to provide a practical basis for detecting and characterizing individual molecular overlayers.

2.3 Fractional coverage and partial coatings

Binding sites may not cover the surface of a probe bead uniformly, nor need target molecules to fill all of the available binding sites. Such incomplete coverage is depicted schematically in Fig. 1(a). If target molecules with refractive index $n_1$ fill a fraction, $f$, of the available sites, the remainder of the surface layer is filled with the fluid medium at refractive index $n_m$. The surface layer then has an effective refractive index, $n_1^{\ast }$, intermediate between $n_1$ and $n_m$ that is accounted for by Maxwell Garnett effective-medium theory [15]:

$$L(n_1^{\ast}/n_m) = f L(n_1/n_m),$$
where
$$L(m) = \frac{m^{2} - 1}{m^{2} + 2}$$
is the Lorentz-Lorenz function. From this, we obtain an expression for the effective refractive index of the partial coating
$$\frac{n_1^{\ast}(f)}{n_m} = \sqrt{ \frac{1 + 2 f L(n_1/n_m)}{1 - f L(n_1/n_m)}} $$
$$ \approx 1 + \frac{3}{2} f L(n_1/n_m) \quad\textrm{for}\;fL(n_1/n_m) < 1. $$

The assumed linear dependence of $d_p^{\ast }$ on the refractive index of the coating therefore suggests that the effective diameter of a probe bead depends on the filling factor, $f$, as

$$ d_p^{\ast} = d_0 + 2 a_1 \frac{n_1^{\ast}(f) - n_m}{n_0 - n_m} $$
$$ \approx d_0 + 3 a_1 f \frac{L(n_1/n_m)}{n_0/n_m - 1} . $$
We predict, therefore, that the measured increase in the effective diameter is roughly proportional to the fractional coverage of bound molecules.

Figure 5 presents the representative case of $n_1 = n_0 = 1.6$, with a coating thickness of $a_1 = 10$ nm. This should be a reasonable model for a polystyrene sphere coated with a monolayer of protein. Discrete (red) data points represent results of numerical experiments for randomly selected filling fractions, ranging from a bare sphere at $f = 0$ to a complete coating at $f = 1$. The solid line is a comparison with the linear approximation from Eq. (8), which interestingly agrees better with the data than the dashed curve representing the full expression from Eq. (7).

 figure: Fig. 5.

Fig. 5. Effective shift in diameter from the effective-sphere model, $\Delta _d$, plotted against fraction of surface coverage, $f$ for a sphere of diameter $d_0 = 1$ µm and refractive index $n_0 = 1.60$ coated with a 10 nm-thick layer with $n_1 = 1.60$. Coverage fraction is computed using Maxwell Garnett effective-medium theory, with $f = 0$ corresponding to $n_1^{\ast } = n_m$ and $f = 1$ corresponding to $n_1^{\ast } = n_1$. Dashed and solid curves correspond to predictions of Eq. (7) and Eq. (8), respectively.

Download Full Size | PDF

2.4 Molecular binding assays: double coatings

The foregoing results show that the effective-sphere model reasonably models the light-scattering properties of the probe beads used for holographic molecular binding assays. We next address how those properties change when target molecules occupy the binding sites on the surface of a probe bead to form a second layer, as depicted in Fig. 1(a). We continue to choose $d_0 = 1$ µm and $n_0 = 1.6$ to model the micrometer-diameter polystyrene substrate bead used in experimental studies. Once the coating of binding sites is added, the probe beads have effective diameter $d_0^{\ast }$ and effective refractive index $n_0^{\ast }$, both of which are determined by fitting to the effective-sphere model. Adding a layer of target molecules on top of this constitutes a model for a binding assay with effective properties $d_p^{\ast }$ and $n_p^{\ast }$. The standard assay involves monitoring the difference, $\Delta d_p^{\ast } = d_p^{\ast } - d_0^{\ast }$, in the probe beads’ effective diameter upon binding. We also monitor the change in refractive index, $\Delta n_p^{\ast } = n_p^{\ast } - n_0^{\ast }$.

For concreteness, we choose the two coatings to have the same thickness, $a_1 = a_2 = 10$ nm, while $n_1$ and $n_2$ are chosen at random between 1.4 and 1.7. This range of refractive indexes matches expectations for protein coatings [32] given that the the coatings may not be complete. Performing two fits to the effective-sphere model for each parameter pair $(n_1, n_2)$ yields measurements of $\Delta d_p^{\ast }$ and $\Delta n_p^{\ast }$ that are presented in Fig. 6(a) and Fig. 6(b), respectively.

 figure: Fig. 6.

Fig. 6. Differences in effective sphere parameters between a singly-coated and doubly-coated sphere, varying values of coating effective refractive indices $n_1$ and $n_2$. (a) $\Delta d_p^{\ast }$ is shown to be largely independent of the index of the first coating, $n_1$. (b) $\Delta n_p^{\ast }$ can be either positive or negative depending on coating parameters, but is close to zero when $n_1 = n_0, n_2 = n_0$ (dashed lines), where $n_0 = 1.6$. We observe a larger shift in effective index when the coating indexes are above that of the underlying sphere.

Download Full Size | PDF

Binding-induced increases in the effective diameter, $\Delta d_p^{\ast }$, are found to be largely independent of the inner coating’s refractive index, $n_1$. Changes in the apparent size depend much more strongly on the refractive index of the outer coating, $n_2$. This is fortunate for practical molecular binding assays because variations in the density of binding sites might have introduced biases into holographic measurements of $d_p^{\ast }$ arising from optical effects that would have exacerbated their actual influence on the statistics of binding. The absence of such optical biases increases the reliability of measurements of the concentration of target molecules based on holographic measurements of $d_p^{\ast }$.

Binding-induced changes in the bead’s refractive index are far more subtle than changes in the size. The influence of the outer coating on both the apparent size and refractive index increases as the refractive index of the coating becomes more mismatched with the refractive index of the substrate bead. These observations suggest that the sensitivity of molecular binding assays can be improved by reducing the refractive index of the substrate beads. Previous experiments have used commercial polystyrene substrate beads with relatively high refractive indexes, $n_0 = 1.60$. Better choices might include silica with a refractive index around $n_0 = 1.42$, poly(methyl methacrylate) (PMMA) with a refractive index around 1.50, and 3-(trimethoxysilyl)propyl methacrylate (TPM) with a refractive index around $n_0 = 1.51$ [33,34].

3. Comparison with experiments

Insights gained from these numerical studies are useful for interpreting experimental realizations of holographic molecular binding assays [7,8]. The data in Fig. 7 are reproduced from Ref. [8] and show how molecular binding changes effective-sphere properties in two cases: avidin binding to biotinylated polystyrene spheres (Fig. 7(a)) and IgG binding to polystyrene spheres coated with protein A (Fig. 7(b)). Each point in the scatter plot represents the measured diameter and refractive index of a single colloidal particle dispersed in a buffer containing dissolved target molecules. The dark (blue) data points represent the properties of the probe beads, $d_0^{\ast }$ and $n_0^{\ast }$, before incubation. The light (red) points represent properties of the same population of beads after incubation, $d_p^{\ast }$ and $n_p^{\ast }$. To facilitate comparisons, we also plot projected probability distributions, $\rho (d_p^{\ast })$ and $\rho (n_p^{\ast })$, of the measured diameters and refractive indexes, respectively, both before and after incubation. The differences in these distributions, $\Delta \rho (d_p^{\ast }) = \rho (d_p^{\ast }) - \rho (d_0^{\ast })$ and $\Delta \rho (n_p^{\ast }) = \rho (n_p^{\ast }) - \rho (n_0^{\ast })$, emphasize shifts in the distributions due to incubation with target molecules.

 figure: Fig. 7.

Fig. 7. Holographic particle characterization data for (a) avidin binding to biotinylated polystyrene spheres and (b) IgG binding to polystyrene spheres coated with protein A. Data from Ref. [8]. The scatter plots show the distributions of single-particle characterization results. Control data from the probe beads are colored blue. Results after binding are colored red. Each data set also shows the projected distributions of particle diameters, $\rho(d_p^*)$, and refractive indexes, $\rho(n_p^*)$, as well as the differences $\Delta \rho (d_p)$ and $\Delta \rho (n_p)$ in measured diameter and refractive index, respectively, caused by molecular binding. Both cases show an increase in effective diameter and a decrease in effective refractive index after binding. Insets show structures of (a) tetrameric avidin (PDB code: 2AVI) and (b) IgG (PDB code: 1IGT).

Download Full Size | PDF

The biotinylated probe spheres have a mean diameter of $d_0^{\ast } = 1.7935(4)$ µm. This increases to $d_p^{\ast } = 1.7956(5)$ µm after tetrameric avidin binds to the beads. Confidence intervals for these values represent the uncertainty in the mean of more than 15 000 particles in each data set and therefore are much smaller than the uncertainty in a single-bead measurement. The observed shift of $\Delta d_p^{\ast } = 1.1(1)$ nm is smaller than the 5 nm domain size of avidin. Whereas the beads’ diameter increases upon binding, their measured refractive indexes decrease from $n_0^{\ast } = 1.60730(3)$ to $n_p^{\ast } = 1.60693(3)$, a net change of $\Delta n_p^{\ast } = (-0.374 \pm 0.04)$ ppt. The probe beads’ refractive index is consistent with expectations for polystyrene, presumably because biotin is a small molecule; a coating of biotin therefore does not substantially affect the substrate beads’ light-scattering properties.

The protein A-coated spheres have a mean diameter of $d_0^{\ast } = 0.9573(8)$ µm before incubation with the antibody IgG. This increases to $d_p^{\ast } = 0.9622(6)$ µm after 45 min incubation resulting in a shift of $\Delta d_p^{\ast } = 4.9(10)$ nm. This increase is larger than was observed for the avidin-biotin system presumably because IgG is substantially larger than tetrameric avidin. Once again, however, $\Delta d_p^{\ast }$ is much smaller than the size of the target molecule. Uncertainties are larger in this case because the statistical ensemble consists of only 3000 particles per sample.

As for the biotin-avidin system, binding IgG causes the probe beads’ refractive index to shift downward from $n_0^{\ast } = 1.5926(3)$ before binding to $n_p^{\ast } = 1.5897(2)$ after, a decrease of $\Delta n_p^{\ast } = (-2.9 \pm 0.3)$ ppt. The initial refractive index of the protein A coated probe beads is smaller than expectations for polystyrene presumably because of the influence of the protein. Protein A is nearly as large as tetrameric avidin and might be expected to have a comparably sizable influence.

In both cases, binding with target molecules leads to an increase in the holographically measured particle diameter that is smaller than the physical size of the target molecules and a decrease in the measured refractive index. Looking to the results in Fig. 6, these trends can be explained if the coatings of bound molecules have lower refractive indexes than the effective refractive index of the substrate beads. Specifically, we interpret these results to show that both avidin and IgG have refractive indexes substantially smaller than 1.6 at the densities of the experimentally obtained coatings. Choosing substrate beads with lower refractive indexes therefore should increase the apparent shift in bead diameter upon binding thereby increasing the target molecules’ influence on holographically measurable properties and improving the sensitivity of the assay.

4. Conclusions

We have used the discrete-dipole approximation to model label-free bead-based molecular binding assays performed with holographic particle characterization in the effective-sphere approximation. Our computational study confirms that interpreting the holograms of coated spheres with the Lorenz-Mie theory for homogeneous spheres yields valuable information on the presence and characteristics of the coatings while retaining the demonstrated speed and robustness of standard holographic particle characterization. Our study validates previous experimental reports of holographic molecular binding assays [7,8] and explains trends in those measurements as arising from the mismatch in refractive index between the substrate beads and the molecular coatings. Because this mismatch depends on the filling factor, $f$, of bound molecules on the beads’ surfaces, the particles’ effective diameters and refractive indexes change continuously as binding proceeds. We have shown that changes in the diameter scale linearly with $f$ to a very good approximation. This means that trends in the holographically measured diameter can be mapped directly onto trends in the fraction of filled binding sites.

Our results furthermore provide guidance for optimizing holographic molecular binding assays. Most notably, the sensitivity of such assays to variations in the filling factor can be increased by reducing the refractive index of the substrate beads. The polystyrene substrates used for many standard bead-based assays are not the best choice for this application, therefore, because their refractive index is quite high. Other materials such as silica, PMMA and TPM offer attractive alternatives for such assays. Ultra-low-index fluorinated PMMA spheres might be a particularly good choice [35]. The choice of substrate for holographic molecular binding assays therefore can be optimized both for optical properties and also for physical properties such as buoyancy to facilitate processing of tests. Beads of different sizes and compositions can be functionalized with different binding sites and combined into a multiplexed assay. The individual tests can be monitored in parallel through the unique ability of holographic particle characterization to differentiate particles by both size and refractive index. These considerations should be particularly useful for designing serological and diagnostic tests for viral infection, with immediate urgency being placed on addressing the ongoing COVID-19 pandemic.

The open-source software used for this study is available at https://github.com/laltman2/DDA_binding_assays.

Funding

National Science Foundation (DMR-1420073, DMR-2027013).

Acknowledgments

This work was supported primarily by the MRSEC program of the National Science Foundation through Award No. DMR-1420073. Additional funding was provided by the RAPID program of the National Science Foundation through Award No. DMR-2027013.

Disclosures

DGG: Spheryx, Inc. (F,I,P,S).

References

1. J. Sheng, E. Malkiel, and J. Katz, “Digital holographic microscope for measuring three-dimensional particle distributions and motions,” Appl. Opt. 45(16), 3893–3901 (2006). [CrossRef]  

2. S.-H. Lee, Y. Roichman, G.-R. Yi, S.-H. Kim, S.-M. Yang, A. van Blaaderen, P. van Oostrum, and D. G. Grier, “Characterizing and tracking single colloidal particles with video holographic microscopy,” Opt. Express 15(26), 18275–18282 (2007). [CrossRef]  

3. C. F. Bohren and D. R. Huffman, Absorption and Scattering of Light by Small Particles (Wiley, 1983).

4. M. I. Mishchenko, L. D. Travis, and A. A. Lacis, Scattering, Absorption, and Emission of Light by Small Particles (Cambridge University Press, 2002).

5. G. Gouesbet and G. Gréhan, Generalized Lorenz-Mie Theories (Springer, 2011).

6. B. J. Krishnatreya, A. Colen-Landy, P. Hasebe, B. A. Bell, J. R. Jones, A. Sunda-Meya, and D. G. Grier, “Measuring Boltzmann’s constant through holographic video microscopy of a single sphere,” Am. J. Phys. 82(1), 23–31 (2014). [CrossRef]  

7. F. C. Cheong, B. Sun, R. Dreyfus, J. Amato-Grill, K. Xiao, L. Dixon, and D. G. Grier, “Flow visualization and flow cytometry with holographic video microscopy,” Opt. Express 17(15), 13071–13079 (2009). [CrossRef]  

8. Y. Zagzag, M. F. Soddu, A. D. Hollingsworth, and D. G. Grier, “Holographic molecular binding assays,” Sci. Rep. 10(1), 1932 (2020). [CrossRef]  

9. J. M. Singer and C. M. Plotz, “The latex fixation test: I. application to the serologic diagnosis of rheumatoid arthritis,” Am. J. Med. 21(6), 888–892 (1956). [CrossRef]  

10. M. Hannel, C. Middleton, and D. G. Grier, “Holographic characterization of imperfect colloidal spheres,” Appl. Phys. Lett. 107(14), 141905 (2015). [CrossRef]  

11. C. Wang, F. C. Cheong, D. B. Ruffner, X. Zhong, M. D. Ward, and D. G. Grier, “Holographic characterization of colloidal fractal aggregates,” Soft Matter 12(42), 8774–8780 (2016). [CrossRef]  

12. C. Wang, X. Zhong, D. B. Ruffner, A. Stutt, L. A. Philips, M. D. Ward, and D. G. Grier, “Holographic characterization of protein aggregates,” J. Pharm. Sci. 105(3), 1074–1085 (2016). [CrossRef]  

13. J. Fung and S. Hoang, “Computational assessment of an effective-sphere model for characterizing colloidal fractal aggregates with holographic microscopy,” J. Quant. Spectrosc. Radiat. Transfer 236, 106591 (2019). [CrossRef]  

14. M. A. Odete, F. C. Cheong, A. Winters, J. J. Elliott, L. A. Philips, and D. G. Grier, “The role of the medium in the effective-sphere interpretation of holographic particle characterization data,” Soft Matter 16(4), 891–898 (2020). [CrossRef]  

15. V. A. Markel, “Introduction to the Maxwell Garnett approximation: tutorial,” J. Opt. Soc. Am. A 33(7), 1244–1256 (2016). [CrossRef]  

16. F. C. Cheong, K. Xiao, D. J. Pine, and D. G. Grier, “Holographic characterization of individual colloidal spheres’ porosities,” Soft Matter 7(15), 6816–6819 (2011). [CrossRef]  

17. F. C. Cheong, P. Kasimbeg, D. B. Ruffner, E. H. Hlaing, J. M. Blusewicz, L. A. Philips, and D. G. Grier, “Holographic characterization of colloidal particles in turbid media,” Appl. Phys. Lett. 111(15), 153702 (2017). [CrossRef]  

18. B. T. Draine and P. J. Flatau, “Discrete-Dipole Approximation for scattering calculations,” J. Opt. Soc. Am. A 11(4), 1491–1499 (1994). [CrossRef]  

19. W. J. Wiscombe, “Improved Mie scattering algorithms,” Appl. Opt. 19(9), 1505–1509 (1980). [CrossRef]  

20. W. Yang, “Improved recursive algorithm for light scattering by a multilayered sphere,” Appl. Opt. 42(9), 1710–1720 (2003). [CrossRef]  

21. O. Peña and U. Pal, “Scattering of electromagnetic radiation by a multilayered sphere,” Comput. Phys. Commun. 180(11), 2348–2354 (2009). [CrossRef]  

22. B. J. Krishnatreya and D. G. Grier, “Fast feature identification for holographic tracking: The orientation alignment transform,” Opt. Express 22(11), 12773–12778 (2014). [CrossRef]  

23. L. E. Altman and D. G. Grier, “CATCH: Characterizing and tracking colloids holographically using deep neural networks,” J. Phys. Chem. B 124(9), 1602–1610 (2020). [CrossRef]  

24. S.-H. Lee and D. G. Grier, “Holographic microscopy of holographically trapped three-dimensional structures,” Opt. Express 15(4), 1505–1512 (2007). [CrossRef]  

25. H. Shpaisman, B. Jyoti Krishnatreya, and D. G. Grier, “Holographic microrefractometer,” Appl. Phys. Lett. 101(9), 091102 (2012). [CrossRef]  

26. A. Yevick, M. Hannel, and D. G. Grier, “Machine-learning approach to holographic particle characterization,” Opt. Express 22(22), 26884 (2014). [CrossRef]  

27. F. Saglimbeni, S. Bianchi, G. Bolognesi, G. Paradossi, and R. Di Leonardo, “Optical characterization of an individual polymer-shelled microbubble structure via digital holography,” Soft Matter 8(34), 8822–8825 (2012). [CrossRef]  

28. L. Méès, N. Grosjean, D. Chareyron, J.-L. Marié, M. Seifi, and C. Fournier, “Evaporating droplet hologram simulation for digital in-line holography setup with divergent beam,” J. Opt. Soc. Am. A 30(10), 2021–2028 (2013). [CrossRef]  

29. M. A. Yurkin and A. G. Hoekstra, “The discrete dipole approximation: An overview and recent developments,” J. Quant. Spectrosc. Radiat. Transfer 106(1-3), 558–589 (2007). [CrossRef]  

30. M. A. Yurkin and A. G. Hoekstra, “The discrete-dipole-approximation code ADDA: Capabilities and known limitations,” J. Quant. Spectrosc. Radiat. Transfer 112(13), 2234–2247 (2011). [CrossRef]  

31. S. Barkley, T. Dimiduk, J. Fung, D. Kaz, V. N. Manoharan, R. McGorty, R. Perry, and A. Wang, “Holographic microscopy with Python and HoloPy,” Comput. Sci. Eng. pp. 1 (2019).

32. T. L. McMeekin, M. L. Groves, and N. J. Hipp, “Refractive indices of amino acids, proteins, and related substances,” in Amino Acids and Serum Proteins, vol. 44 of Advances in Chemistry (American Chemical Society, 1964), pp. 54–66.

33. C. van der Wel, R. K. Bhan, R. W. Verweij, H. C. Frijters, Z. Gong, A. D. Hollingsworth, S. Sacanna, and D. J. Kraft, “Preparation of colloidal organosilica spheres through spontaneous emulsification,” Langmuir 33(33), 8174–8180 (2017). [CrossRef]  

34. C. Middleton, M. D. Hannel, A. D. Hollingsworth, D. J. Pine, and D. G. Grier, “Optimizing the synthesis of monodisperse colloidal spheres using holographic particle characterization,” Langmuir 35(20), 6602–6609 (2019). [CrossRef]  

35. S. Sacanna, G. H. Koenderink, and A. P. Philipse, “Microemulsion synthesis of fluorinated latex spheres,” Langmuir 20(19), 8398–8400 (2004). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (7)

Fig. 1.
Fig. 1. (a) Geometry of a bead-based molecular binding assay. A substrate bead with diameter $d_0$ and refractive index $n_0$ is coated with a layer of functional groups that bind target molecules from the surrounding solution. The surface groups are modeled as a dielectric layer of thickness $a_1$ and refractive index $n_1$. The coating of bound target molecules is modeled as a second layer of thickness $a_2$ and refractive index $n_2$. The bead is dispersed in a medium of refractive index $n_m$. The bounding sphere has diameter $d_p = d_0 + 2(a_1 + a_2)$. (b) Synthetic hologram of a coated bead computed with the DDA. Parameters are chosen to mimic a typical molecular binding assay based on a probe bead with a polystyrene core: $d_0 = 1$ µm, $n_0 = 1.60$, and molecular overlayers with $a_1 = a_2 = 10$ nm, $n_1 = n_2 = 1.55$. (c) The effective-sphere model treats the coated sphere as if it were a homogeneous dielectric sphere of effective diameter $d_p^{\ast }$ and refractive index $n_p^{\ast }$. (d) Fitting the coated-sphere’s hologram from (b) with the Lorenz-Mie theory for a homogeneous sphere yields an indistinguishable hologram for $d_p^{\ast } = 1.03$ µm and $n_p^{\ast } = 1.594$.
Fig. 2.
Fig. 2. Holographic images of homogeneous spheres computed with DDA are then fitted to Lorenz-Mie theory. Their deviations from the inputted values are $\epsilon _d(d_p, n_p) = d_p^{\ast } - d_p$ (a), and $\epsilon _n(d_p, n_p) = n_p^{\ast } - n_p$ (b). Errors can be positive or negative for both parameters, but in this region of parameter space, errors in $n_p$ are predominantly negative. Discrete points indicate the properties of polystyrene substrate beads used in published molecular binding assays [7,8] with diameters $d_p = 1$ µm and $d_p = 1.8$ µm. The smaller of these lies in a region of parameter space where agreement between DDA and Lorenz-Mie formulations is particularly good with $\epsilon _d < 5$ nm.
Fig. 3.
Fig. 3. Influence of coating thickness, $a_1$, on effective sphere diameter, $d_p^{\ast }$, depends on the refractive index of the coating, $n_1$, relative to that of the substrate sphere, $n_0$. Data points are computed for a sphere with $d_0 = 1$ µm and $n_0 = 1.60$. The observed increase, $\Delta _d(a_1) = d_p^{\ast } - d_0$, increases monotonically with $n_1$ and agrees with the geometric size, $\Delta _d(a_1) = 2 a_1$. when $n_1 = n_0$, as indicated by the dashed line.
Fig. 4.
Fig. 4. A 10 nm coating with variable refractive index $n_1$ is applied to a $d_0 = 1$ µm polystyrene bead and then is fitted to the effective-sphere model. (a) Shift in effective diameter $\Delta _d (n_1) = d_p^{\ast } - d_0$ is linear with coating index $n_1$, with $n_1 = n_0 = 1.6$ corresponding to a shift of 10 nm. A regression line (dotted) is fitted to the data. (b) Shift in effective refractive index, $\Delta _n (n_1) = n_p^{\ast } - n_0$, is positive for $n_1 > n_0$ and negative for $n_1 < n_0$. The horizontal dashed line indicates the baseline, $\Delta _n = 0$.
Fig. 5.
Fig. 5. Effective shift in diameter from the effective-sphere model, $\Delta _d$, plotted against fraction of surface coverage, $f$ for a sphere of diameter $d_0 = 1$ µm and refractive index $n_0 = 1.60$ coated with a 10 nm-thick layer with $n_1 = 1.60$. Coverage fraction is computed using Maxwell Garnett effective-medium theory, with $f = 0$ corresponding to $n_1^{\ast } = n_m$ and $f = 1$ corresponding to $n_1^{\ast } = n_1$. Dashed and solid curves correspond to predictions of Eq. (7) and Eq. (8), respectively.
Fig. 6.
Fig. 6. Differences in effective sphere parameters between a singly-coated and doubly-coated sphere, varying values of coating effective refractive indices $n_1$ and $n_2$. (a) $\Delta d_p^{\ast }$ is shown to be largely independent of the index of the first coating, $n_1$. (b) $\Delta n_p^{\ast }$ can be either positive or negative depending on coating parameters, but is close to zero when $n_1 = n_0, n_2 = n_0$ (dashed lines), where $n_0 = 1.6$. We observe a larger shift in effective index when the coating indexes are above that of the underlying sphere.
Fig. 7.
Fig. 7. Holographic particle characterization data for (a) avidin binding to biotinylated polystyrene spheres and (b) IgG binding to polystyrene spheres coated with protein A. Data from Ref. [8]. The scatter plots show the distributions of single-particle characterization results. Control data from the probe beads are colored blue. Results after binding are colored red. Each data set also shows the projected distributions of particle diameters, $\rho(d_p^*)$, and refractive indexes, $\rho(n_p^*)$, as well as the differences $\Delta \rho (d_p)$ and $\Delta \rho (n_p)$ in measured diameter and refractive index, respectively, caused by molecular binding. Both cases show an increase in effective diameter and a decrease in effective refractive index after binding. Insets show structures of (a) tetrameric avidin (PDB code: 2AVI) and (b) IgG (PDB code: 1IGT).

Equations (9)

Equations on this page are rendered with MathJax. Learn more.

E 0 ( r , t ) = e i k z e i ω t x ^ ,
b ( r ) = | x ^ + e i k z p f s ( k ( r r p ) ) | 2 ,
d p = d 0 + 2 a 1 n 1 n m n 0 n m .
L ( n 1 / n m ) = f L ( n 1 / n m ) ,
L ( m ) = m 2 1 m 2 + 2
n 1 ( f ) n m = 1 + 2 f L ( n 1 / n m ) 1 f L ( n 1 / n m )
1 + 3 2 f L ( n 1 / n m ) for f L ( n 1 / n m ) < 1.
d p = d 0 + 2 a 1 n 1 ( f ) n m n 0 n m
d 0 + 3 a 1 f L ( n 1 / n m ) n 0 / n m 1 .