Abstract
Near-infrared (NIR) spectroscopy has been utilized for monitoring the pour point of the lube base oil (LBO) process. Using partial least-squares (PLS) regression, necessary spectral features were successfully extracted and correlated to the reference pour-point data. In a preliminary laboratory feasibility study, it was found that PLS calibration performance was largely governed by the accuracy of the reference pour-point analysis. The same NIR spectroscopic methodology was moved to monitor the pour point continuously in an on-line manner. At this time, the NIR spectrometer was calibrated with the existing on-line pour-point analyzer, which was more accurate. In comparison with the on-line pour-point analyzer, NIR spectroscopy provided faster and more repeatable analytical data. With the help of NIR analysis, LBO process variations during product switch can be detected earlier, and necessary process-control strategy can be applied to improve the process efficiency.
PDF Article
More Like This
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator
or
Login to access Optica Member Subscription