Abstract
We have developed a three-channel imaging Fabry–Perot interferometer with which to measure atmospheric wind and temperature in the mesosphere and thermosphere through nocturnal airglow emissions. The interferometer measures two-dimensional wind and temperature for wavelengths of 630.0 nm (OI, altitude, 200–300 km), 557.7 nm (OI, 96 km), and 839.9 nm (OH, 86 km) simultaneously with a time resolution of 20 min, using three cooled CCD detectors with liquid-N2 Dewars. Because we found that the CCD sensor moves as a result of changes in the level of liquid N2 in the Dewars, the cooling system has been replaced by thermoelectric coolers. The fringe drift that is due to changes in temperature of the etalon is monitored with a frequency-stabilized He–Ne laser. We also describe a data-reduction scheme for calculating wind and temperature from the observed fringes. The system is fully automated and has been in operation since June 1999 at the Shigaraki Observatory (34.8N, 136.1E), Shiga, Japan.
© 2001 Optical Society of America
Full Article | PDF ArticleMore Like This
Hideaki Nakajima, Shoichi Okano, Hiroshi Fukunishi, and Takayuki Ono
Appl. Opt. 34(36) 8382-8395 (1995)
Brian J. Harding, Thomas W. Gehrels, and Jonathan J. Makela
Appl. Opt. 53(4) 666-673 (2014)
Yiyi Huang, Jonathan J. Makela, and Gary R. Swenson
Appl. Opt. 51(17) 3787-3800 (2012)