Abstract
When a zone-plate interferometer is used, a bright spot appears at the center of the image plane. The spot makes it difficult to analyze the interference fringes. A simple technique that is based on the principle of fringe-intensity reversal is proposed to analyze the fringes efficiently. A zone plate with a phase fraction of π/2 or 3π/2 is used in this technique to diminish the bright spot. Unlike the masking technique, no part of the data on the fringes is lost. The fringes can, therefore, be analyzed completely. The technique is described in detail, and the results of an experiment in which the shape error of a concave mirror was measured with the proposed zone plate is presented. The experimental results agree well with the results obtained with the Fizeau interferometer.
© 1995 Optical Society of America
Full Article | PDF ArticleMore Like This
Raymond N. Smartt
Appl. Opt. 13(5) 1093-1099 (1974)
Jenn-An Lin, W. T. Yeh, and S. W. Hsu
Appl. Opt. 29(34) 5151-5158 (1990)
Chris S. Anderson
Appl. Opt. 34(32) 7474-7485 (1995)