Abstract
Recent advances in model observers that predict human perceptual performance now make it possible to optimize medical imaging systems for human task performance. We illustrate the procedure by considering the design of a lens for use in an optically coupled digital mammography system. The channelized Hotelling observer is used to model human performance, and the channels chosen are differences of Gaussians. The task performed by the model observer is detection of a lesion at a random but known location in a clustered lumpy background mimicking breast tissue. The entire system is simulated with a Monte Carlo application according to physics principles, and the main system component under study is the imaging lens that couples a fluorescent screen to a CCD detector. The signal-to-noise ratio (SNR) of the channelized Hotelling observer is used to quantify this detectability of the simulated lesion (signal) on the simulated mammographic background. Plots of channelized Hotelling SNR versus signal location for various lens apertures, various working distances, and various focusing places are presented. These plots thus illustrate the trade-off between coupling efficiency and blur in a task-based manner. In this way, the channelized Hotelling SNR is used as a merit function for lens design.
© 2005 Optical Society of America
Full Article | PDF ArticleMore Like This
Subok Park, Eric Clarkson, Matthew A. Kupinski, and Harrison H. Barrett
J. Opt. Soc. Am. A 22(1) 3-16 (2005)
Cyril Castella, Craig K. Abbey, Miguel P. Eckstein, Francis R. Verdun, Karen Kinkel, and François O. Bochud
J. Opt. Soc. Am. A 24(12) B1-B12 (2007)
J. P. Rolland and H. H. Barrett
J. Opt. Soc. Am. A 9(5) 649-658 (1992)