Abstract
The dependence of the intensities of the 1849- and 2537-Å resonance lines on current density, in the range 0.2–2 A/cm2, Hg pressure 10−3–0.5 torr, and discharge-tube diameter 10–25 mm, have been measured. The intensities nearly saturate at about 2 A/cm2. Absolute output at 1849 Å, measured photochemically, is consistent with the results of B. T. Barnes. The dependence of this output on the measured parameters make this discharge a good absolute flux-density standard at 1849 Å.
© 1971 Optical Society of America
Full Article | PDF ArticleMore Like This
T. J. Hammond and C. F. Gallo
Appl. Opt. 10(1) 58-64 (1971)
W. M. Vaidya and V. D. Dandawate
J. Opt. Soc. Am. 56(12) 1693-1697 (1966)
F. W. Crawford and H. R. Pagels
J. Opt. Soc. Am. 53(6) 734-740 (1963)