Abstract
Excitation temperatures and electron densities are reported for a magnetron rotating direct-current arc plasma with graphite furnace sample introduction. Temperatures obtained from emission line ratios of the Ar support gas are compared with values obtained for two analyte thermometric species, Zn(I) and Cu(I). At the anode surface, the plasma temperature is about 6500 K. The addition of an easily ionized element does not effect the Ar(I) temperature; however, a temperature increase of up to 1000 K was observed for the analyte thermometric species. At the anode surface, the electron density is about 1.0 × 10<sup>15</sup> cm<sup>−3</sup>. The addition of an easily ionized element increases the electron density by about a factor of two. Possible explanations are given for the increases in temperature and electron density.
PDF Article
More Like This
Translational, rotational, vibrational and electron temperatures of a gliding arc discharge
Jiajian Zhu, Andreas Ehn, Jinlong Gao, Chengdong Kong, Marcus Aldén, Mirko Salewski, Frank Leipold, Yukihiro Kusano, and Zhongshan Li
Opt. Express 25(17) 20243-20257 (2017)
Cited By
You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.
Contact your librarian or system administrator