Abstract
We describe basic principles of wavelength-tuning interferometry and demonstrate its application in ophthalmology. The advantage of this technique compared with conventional low-coherence interferometry ranging is the simultaneous measurement of the object structure without the need for a moving reference mirror. Shifting the wavelength of an external-cavity tunable laser diode causes intensity oscillations in the interference pattern of light beams remitted from the intraocular structure. A Fourier transform of the corresponding wave-number-dependent photodetector signal yields the distribution of the scattering potential along the light beam illuminating the eye. We use an external interferometer to linearize the wave-number axis. We obtain high resolution in a model eye by slow tuning over a wide wavelength range. With lower resolution we demonstrate the simultaneous measurement of anterior segment length, vitreous chamber depth, and axial eye length in human eyes in vivo with data-acquisition times in the millisecond range.
© 1997 Optical Society of America
Full Article | PDF ArticleMore Like This
G. F. Schmid, B. L. Petrig, C. E. Riva, K. H. Shin, R. A. Stone, M. J. Mendel, and A. M. Laties
Appl. Opt. 35(19) 3358-3361 (1996)
F. Bien, M. Camac, H. J. Caulfield, and S. Ezekiel
Appl. Opt. 20(3) 400-403 (1981)
B. Golubovic, B. E. Bouma, G. J. Tearney, and J. G. Fujimoto
Opt. Lett. 22(22) 1704-1706 (1997)