Abstract
An instrument for the measurement of circular dichroism (CD) between 5000 cm−1 and 750 cm−1 is described. This instrument also can perform high precision linear dichroism measurements in the same wavelength range. A germanium photoelastic modulator having a large linear and angular aperture is used to vary the polarization of monochromatic ir light between left- and right-circularly polarized at 8 kHz. This polarization-modulated light traverses the sample and falls onto a cooled InSb or HgCdTe detector. Phase-sensitive demodulation yields the CD of the sample to a precision of 4 × 10−6 o.d. units near 3000 cm−1. The instrument has been used to study ir circular dichroism arising from molecular vibrations in cholesteric liquid crystals, crystals, and fluids. CD arising from low-lying electronic transitions of metal complexes and linear dichroism of polymer films have also been studied.
© 1975 Optical Society of America
Full Article | PDF ArticleMore Like This
J. C. Cheng, L. A. Nafie, and P. J. Stephens
J. Opt. Soc. Am. 65(9) 1031-1035 (1975)
Toshio Ishii
Appl. Opt. 14(9) 2143-2145 (1975)
F. A. Modine and R. W. Major
Appl. Opt. 14(3) 761-764 (1975)