Abstract
We demonstrate intravascular optical coherence tomography (OCT) imaging with frame rate up to 3.2 kHz (192,000 rpm scanning). This was achieved by using a custom-built catheter in which the circumferential scanning was actuated by a 1.0 mm diameter synchronous motor. The OCT system, with an imaging depth of 3.7 mm (in air), is based on a Fourier domain mode locked laser operating at an A-line rate of 1.6 MHz. The diameter of the catheter is 1.1 mm at the tip. Ex vivo images of human coronary artery (78.4 mm length) were acquired at a pullback speed of . True 3D volumetric imaging of the entire artery, with dense and isotropic sampling in all dimensions, was performed in second acquisition time.