Abstract
We report a confocal total-internal-reflection fluorescence (TIRF) microscope that generates a detection volume for analyte molecules of less than 5 al at a water–glass interface. Compared with conventional confocal microscopy, this represents a reduction of almost 2 orders of magnitude, which is important in isolating individual molecules at high analyte concentrations, where many biologically relevant processes occur. Diffraction-limited supercritical focusing and fluorescence collection is accomplished by a parabolic mirror objective. The system delivers TIRF images with excellent spatial resolution and detects single molecules with a high signal-to-background ratio.
© 2004 Optical Society of America
Full Article | PDF ArticleMore Like This
Thomas Ruckstuhl and Stefan Seeger
Appl. Opt. 42(16) 3277-3283 (2003)
Thomas P. Burghardt, Andrew D. Hipp, and Katalin Ajtai
Appl. Opt. 48(32) 6120-6131 (2009)
Kai Hassler, Marcel Leutenegger, Per Rigler, Ramachandra Rao, Rudolf Rigler, Michael Gösch, and Theo Lasser
Opt. Express 13(19) 7415-7423 (2005)