Two-photon imaging through a multimode fiber

Open Access Open Access

Abstract

In this work we demonstrate 3D imaging using two-photon excitation through a 20 cm long multimode optical fiber (MMF) of 350 µm diameter. The imaging principle is similar to single photon fluorescence through a MMF, except that a focused femtosecond pulse is delivered and scanned over the sample. In our approach, focusing and scanning through the fiber is accomplished by digital phase conjugation using mode selection by time gating with an ultra-fast reference pulse. The excited two-photon emission is collected through the same fiber. We demonstrate depth sectioning by scanning the focused pulse in a 3D volume over a sample consisting of fluorescent beads suspended in a polymer. The achieved resolution is 1 µm laterally and 15 µm axially. Scanning is performed over an 80x80 µm field of view. To our knowledge, this is the first demonstration of high-resolution three-dimensional imaging using two-photon fluorescence through a multimode fiber.

© 2015 Optical Society of America

Corrections

14 January 2016: Corrections were made to Eq. (1) and the body text.

1. Introduction

An endoscope allows the visualization of the interior of organs or cavities inside of the body. A widely used endoscopic system for clinical applications is based on multi-core fibers (MCFs). The main benefit of a MCF is that it is 1. thin (0.3 mm for 6,000 cores) and 2. the part entering the body is passive, i.e does not contain active electronic components. However, MCF deliver pixelated images due to the gap between the cores (3-5 µm), which limits the resolution of the system to several micrometers [1–8].

In a similar fashion to scattering media [9–13], wavefront shaping methods using MCFs and multimode fibers (MMFs) have been able to effectively “de-pixelate” images to obtain diffraction limited resolution. Wavefront shaping methods include phase conjugation in the analog or digital domain [14–19], transmission matrix to provide wide field imaging [20–23] and iterative optimization [24–30]. However, these imaging approaches have been demonstrated using monochromatic light sources. Thus, for fluorescence imaging, only one-photon excitation was possible.

Two-photon excited fluorescence (TPEF) is obtained with pulsed light sources in the femtosecond regime. Andersen et al have demonstrated two-photon imaging through a custom MCF with 169 cores [31]. We have previously shown scanning and focusing of pulses in a step index MMF while maintaining a pulse duration of 500 fs through 30 cm fiber propagation [32]. In graded index fibers, spatial focusing of pulses has been shown [33]. However, in these previous studies, TPEF imaging through multimode fibers was not demonstrated.

In this paper, we demonstrate, for the first time, high resolution two-photon excitation imaging through a graded-index multimode optical fiber. We use time-gated digital phase conjugation to focus and scan high intensity femtosecond pulses through the fiber [32]. Sectioned images of fluorescent beads suspended in a PDMS polymer volume are imaged. The high resolution (1µm), small probe diameter (350 µm), large collection efficiency, sectioning capability and ultrashort focused pulse delivery (120 fs) of our system opens the possibility for in-vivo minimally invasive multiphoton endoscopy in areas such as the brain, the ear or the eye.

2. Focusing pulses through the multimode fiber

2.1 Calibration

To generate spatially focused pulses through the multimode fiber, we use time-gated digital phase conjugation, which requires the following calibration step depicted in Fig. 1. The 100 fs pulses from a Ti:sapphire laser are stretched out in time by a prism pair separated by 164 cm as described in Appendix A. This compensates for the group velocity dispersion (GVD) that the pulses suffer as they travel through the fiber. The GVD compensated beam (calibration beam) is then focused in front of the 20 cm length graded index fiber by a 40x microscope objective. We call this the “calibration spot”. As light from the calibration spot enters the MMF it is decomposed into multiple fiber modes, which generate, after propagation to the proximal side of the fiber, a temporally dispersed and spatially scrambled light field. In an off-axis holographic arrangement, a digital hologram of the time-gated light field is recorded on the CCD at the other end of the fiber. The process is repeated at different locations of the calibration spot in a three-dimensional grid. The digital holograms are stored. Recently, a rotational memory effect of multimode fibers has been reported, which can reduce the calibration time [34]. Fiber bending introduces changes in the propagation of the fiber modes. In our set-up, recalibration is required if the fiber is bent. However, this can be solved by incorporating a bending-compensation method via geometric characterization of fiber bending [20] or by measurement of the proximal speckle of a fiber with a distal reflective coating [35].

 figure: Fig. 1

Fig. 1 (a) Experimental Setup. Calibration step. A calibration beam, previously pre-chirped for compensation of GVD, is focused and coupled into the multimode fiber. The field emerging from the proximal side of the fiber is interfered with the reference beam. An off-axis digital hologram is recorded. Reconstruction step. The flip mirror is removed. The reference reconstructs the phase conjugated version of the recorded hologram. Light counter propagates through the fiber, generating the initial focus on the distal side. Imaging step. The focused pulses are sequentially scanned over the specimen. The emitted two-photon fluorescence (2PF) is collected through the fiber and measured with the PMT.

Download Full Size | PDF

2.2 Digital scanning of pulsed light

Once the calibration process is completed, the phase conjugated field of each recorded hologram is sequentially reconstructed (the flip mirror is removed) with a spatial light modulator (SLM). The reconstructed field counter propagates through the fiber generating a temporally and spatially sharp focused spot at the distal end. We define the enhancement of the focused pulse as the maximum intensity value in the focused spot divided by the average intensity in the background. As in scattering media, a larger number of controlled modes yields a better focusing capability; hence, greater enhancement [9]. Modes in graded index fibers are confined closer to the core and suffer less modal dispersion than in step-index fibers. However, this results in a drop of the enhancement value when the focus is away from the center of the fiber. Figure 2 shows the experimental results of intensity, enhancement, and spot size of a 40x40 grid of focused pulses generated sequentially 100 μm away from the facet of the fiber. The spots are spaced 2 μm apart. For illustration purposes, Fig. 2(a) shows half of the generated spots (every 4 μm).

 figure: Fig. 2

Fig. 2 Scanning of the focused spot at the fundamental wavelength of 800 nm. (a) Sequential superposition of a focused pulse scanned in a 80x80 µm field of view. Scanning step is 4 µm. (b) Intensity enhancement of the focused pulse as a function of scanning position. (c) FWHM size of the focused pulses as a function of scanning position. (d) Line profile of (b) at the location of the dashed lines. (e) Line profile of (c) at the location of the dashed lines. Scale bars are 10 µm. The dashed orange circle represents the core of the multimode fiber.

Download Full Size | PDF

The FWHM spot size of the focused pulses is inversely proportional to the numerical aperture of the fiber. For an observer located on the distal side, the effective numerical aperture of the fiber depends on the position of observation, being maximum at the center [15]. Therefore, the spot size is minimum at focusing locations near the optical axis as shown in Figs. 2(c) and 2(e). Simulation shows that the spot size across the fiber diameter in Fig. 2(e) is limited by the numerical aperture of the fiber.

2.3 GVD compensation

As demonstrated in [32], the counter-propagation of a time-sampled set of modes excites only fiber modes that travel at approximately the same speed (group velocity). These modes generate a focused pulse which does not suffer from modal dispersion. However, group velocity dispersion GVD still occurs. GVD is a phenomenon by which the spectral components of light travel through a material at different speeds (dn/dλ0) which results in broadening of the propagating pulse in time [36].

In our experiments, we compensate for the GVD occurring in the 20 cm length of the fiber by passing the beam through a prism pair (shown in Fig. 8) separated by 164 cm, which introduces a total group delay of 7200 fs2. Dispersion compensation enhances the visibility of the fringes at the calibration step, and allows the reconstruction of a 120 fs pulse when used at the reconstruction step, close to the initial 100 fs. In our experimental setup, the compensation of GVD is optional at the reconstruction step. Figure 3 compares the intensity and the pulse length of a line of pulses focused through the fiber when the hologram is reconstructed without GVD compensation and with GVD compensation on the reference. In the absence of GVD compensation, the pulse length is longer, on average 222 fs, due to the temporal broadening produced by GVD in the fiber [Fig. 3(d)]. In contrast, when the pulse is GVD compensated before reconstruction, the spectral dispersion of the excited modes is canceled after propagation through the fiber. All spectral components arrive simultaneously at the distal side of the fiber and produce a short focused pulse, on average 117 fs. A small temporal broadening is observed, from 100 fs to 117 fs which is attributed to the GVD set-up (prism pair in Fig. 8) where some spectral components angularly dispersed by the first prism are not collected by the second prism. The slight variations in pulse width (a standard deviation of 15 fs for no GVD compensation and 17 fs in the other case) are produced by slightly different propagation paths in the fiber, depending on the location of the generated spot and the specific excited modes. Figures 3(e) and 3(f) show the second order interferometric autocorrelation trace and envelope of pulses focused at the center of the fiber. For the case of no GVD compensation [Fig. 3(f)], we can notice an increased non-oscillating intensity at the tails of the pulse, which indicates that the delivered pulse is chirped. For the case of GVD compensation [Fig. 3(e)], the effect is less evident because material dispersion in the fiber is compensated. Figure 3(g) shows the intensity autocorrelation trace of a pulse transmitted through the fiber when wavefront control is not being used (light is focused in the optical axis on the proximal side of the fiber and the pulse length is measured at the distal side). In this case, the pulse is broadened due to GVD and modal dispersion, reaching a pulse length of 745 fs from the initial 100 fs. More details on the pulse width measurements can be found in Appendix A. Regarding the intensity enhancement, it depends on the position of the focused pulse as shown in Fig. 2(b) and 2(d). The enhancement of the generated focused spots [Fig. 3(c)] is on average 252 and 220 within a 20 μm radius from the optical axis, for the case of GVD and no GVD compensation, respectively. Hence, the spectral correction achieved by GVD compensation not only reduces pulse broadening, but also enhances the intensity of the focused spot. Since the cross section of the optical fiber is azimuthally symmetric, for the described results, we can characterize the spots scanned only along a line passing through the optical axis at the center of the core without loss of generality. Taking into account a transmittance of 75% through the GVD compensation set-up (prism pair), the signal to noise ratio of the focused spot is enhanced by a factor of 1.6 when GVD is compensated.

 figure: Fig. 3

Fig. 3 Pulse width characterization. Intensity of the generated focused pulses at 800 nm without (a) and with (b) GVD compensation. (c) Intensity enhancement vs position. (d) Pulse width vs position. (e), (f) Sample interferometric autocorrelation traces of the central focused pulses with and without GVD compensation respectively. (g) Intensity autocorrelation of the delivered pulse when no wavefront control is used. Scale bars are 10 µm. The dashed orange semi-circle in (a) represents the core of the multimode fiber.

Download Full Size | PDF

3. Two-photon imaging through the multimode fiber

Once we are able to generate and scan a grid of focused pulses at different planes through the multimode fiber, the next step is to use the system as a two-photon endoscope. In 2PFE, two photons of long excitation wavelength are simultaneously absorbed by a fluorophore. When the fluorophore returns from the excited to the ground state, a photon of higher energy than any of the two absorbed photons is emitted [37]. We prepared a sample of 1-5 µm fluorescent red microspheres suspended in a transparent PDMS volume of 55 µm thickness. We used an excitation wavelength of 800 nm. The spectral emission of the beads is centered at 575 nm. For more details about the sample preparation, see Appendix A. We scan the focused pulse every 700 nm in a grid of 40x70 points (28x49 µm FOV). We scan the planes from 0 to 50 µm inside the sample. Figure 4 shows the sectioned images. All the frames can be seen in Visualization 1.

 figure: Fig. 4

Fig. 4 Two-photon images of fluorescent beads. (y,x,z) Field of view is 28x49x50 µm. Scale bars are 10 µm. To view all the frames see Visualization 1.

Download Full Size | PDF

The number of fluorescent photons produced in two-photon emission is given by [37]:

nap02δτpfp2(NA22hcλ)2
where na is the number of fluorescent photons, p0 is the average power, τp is the pulse length, fp is the repetition rate of the source, δ is the two-photon absorption cross section and NA is the numerical aperture of the fiber. As described before, the intensity enhancement and spot size is a function of the scanning position at the distal end of the fiber. Hence, we calculate a quantity proportional to na called the normalized two-photon fluorescence (2PF normalized), which is shown in Fig. 5. In a homogeneous fluorescent medium, focused pulses generated near the center of the optical axis produce a larger number of fluorescent photons. Therefore, to reconstruct correctly the original image, we require a compensation of the over-estimation of intensity in the central area, which is performed as a post-acquisition process.

 figure: Fig. 5

Fig. 5 Normalized fluorescent photons as a function of position. Scale bar is 10 µm.

Download Full Size | PDF

Two-photon fluorescence occurs in a smaller volume than single photon absorption. Hence, the point spread function (PSF) of the two-photon imaging fiber device is 2 times smaller than the linear PSF. Additionally, 2PF suppresses the out of focus excitation. This is illustrated in Fig. 6, which shows the linear and 2PF PSF of the fiber imaging system. The intensity after the first zero of the airy disk is significantly reduced by the two-photon process as seen in Fig. 6. In summary, the 2PF images acquired with the proposed system have a lateral resolution smaller than 1.05 µm within a 40 µm diameter circle centered at the optical axis with a two-photon equivalent enhancement larger than 57,000. Resolution is smaller than 1.4 µm in a 60 µm diameter FOV. The axial resolution which is associated to the depth of field is given by: dz=2.22λ/NA2. Near the optical axis, the theoretical single photon axial resolution is 15 µm and the two-photon axial resolution 10 µm (a factor of 2 smaller). This calculated value matches the experimental data shown in the axial line profile of Fig. 4. From the line profile, we obtain a Gaussian fit. Then, we numerically de-convolve the fitted data with a simulated 4 μm size fluorescent bead. The de-convolution results in a two-photon PSF of 10 µm resolution, which matches the theoretical value. The maximum average power delivered to the sample is 22 mW, which corresponds to a spot energy of 0.27 mJ/cm2.

 figure: Fig. 6

Fig. 6 Experimental PSF of the multimode fiber imaging system. (a) Single-photon fluorescence. (b) Two-photon fluorescence. (c) Normalized PSF of (a) and (b) at y = 0.

Download Full Size | PDF

4. Conclusions

We have demonstrated the first implementation of a multimode fiber two-photon endoscope. Our digital scanning method does not require distal mechanical elements, allowing an ultra-thin probe of 350 µm diameter. The large area of the fiber and high numerical aperture (0.34 NA) allows large collection efficiency. The reduced modal dispersion of graded-index fibers allows the simultaneous control of a larger number of modes, generating a brighter focused pulse compared to the case of step-index fibers. Time-gated phase conjugation of a pre-GVD compensated beam prevents pulse broadening due to GVD of the pulse propagating through the fiber, generating an ultrashort pulse of 120 fs duration. The high intensity focused pulse is generated outside of the fiber avoiding nonlinearities in the propagating pulse. Our work sets the basis for in-vivo ultra-thin two-photon endoscopy.

Appendix A materials and methods

The experiments were conducted using a Ti:sapphire oscillator (Coherent Mira 900; central wavelength λc=800 nm; spectral width σλ=10 nm; pulse width=100 fs; repetition rate 76 MHz). The capability of our method to form a sharp focused pulse through the fiber depends on the number of modes than can be controlled simultaneously [32]. In step-index fibers, the pulse spreading in time is given by: στLΔ/(2c1) while in graded-index fibers: στLΔ2/(4c1) [36]. Due to their smaller modal temporal-spreading, we selected a graded-index fiber for our experiments. The fiber has a 200 µm core diameter, 280 µm cladding diameter, 20 cm length, 0.29 NA and supports 26,000 modes at a 800 nm wavelength. The SLM used for time-gated phase conjugation is a Holoeye Pluto, 1920x1080 pixels, phase only.

To increase the resolution of the two-photon imaging system a grin lens (0.5 mm length, 70° view angle, 250 µm diameter) is attached to the distal end of the multimode fiber as shown in Fig. 7. The lens increases the numerical aperture of the fiber from 0.29 to 0.34 (resolution changed from 1.2 µm to 1 µm) but reduces the circular field of view (FOV) of the fiber from 200 µm diameter to 106 µm. Therefore, there is a tradeoff between resolution, FOV and probe diameter. For example, using a fiber with a core diameter equal to 600 µm and a lens of same magnification as the one used in our experiments, the maximum FOV of the system would be 318 µm. The multiphoton imaging principle demonstrated in this work can be extended to lensless and thinner or thicker graded index fibers without loss of generality.

 figure: Fig. 7

Fig. 7 Multimode fiber probe. The white horizontal lines are the reflection of the lamp of the microscope.

Download Full Size | PDF

At the calibration step, (Fig. 8, flip mirror 1 and 2 are removed) the calibration beam passes through the prism pair and is coupled into the fiber by a 40x microscope objective (OBJ2). The field emerging from the fiber on the proximal side is imaged by a 20x objective (OBJ1) and a 150 mm lens (L1) onto the CCD, where it is interfered by the pulsed reference beam. The pulsed reference can be delayed and is set at the time in which the visibility of the fringes is maximum [32].

 figure: Fig. 8

Fig. 8 Complete diagram of the experimental setup.

Download Full Size | PDF

At the reconstruction step, flip mirror 3 is removed allowing the reference to reach the SLM and reconstruct the phase conjugate version of the recorded field. Light counter propagated through the fiber and is focused on the distal side imaged on Camera 2 by a 200 mm lens (L2). The hologram can be reconstructed with GVD compensation (flip mirrors 1 and 2 in place) and without GVD compensation (flip mirrors 1 and 2 removed).

At the imaging step, the 2PF is collected through the fiber, reflected by a 650 nm short-reflected dichroic mirror, and focused on the PMT by a 35 mm lens (L3). This ensures the collection of most of the spectral emission of the sample, whose maximum fluorescence is centered at 575 nm.

To measure the pulse width of the generated pulses through the fiber we use second order interferometric autocorrelation by introducing two replicas of the pulses on the reference arm using the interferometer shown in Fig. 8. As a nonlinear element for the second order interferometer, we place a homogeneous sample of PDMS with Rhodamine 6G [38, 39]. The fluorescence is collected through the fiber and measured by the PMT. The background to maximum ratio of the measured second order interferometric traces is 1/8. A sech2 pulse is assumed to calculate the pulse lengths (0.64 times the width of the envelope of the autocorrelation trace). For the pulse width measurement shown in Fig. 3(g), we used an intensity autocorrelator (Carpe APE).

The image sample used in these experiments consists of 1-5 µm fluorescent beads (Cospheric FMR – Red Fluorescent Microspheres) suspended in a transparent PDMS volume. The spectral emission of the beads is centered at 575 nm. The fluorescent beads were mixed with PDMS in a concentration of 1 mg of beads per 100 mg of PDMS with 10 mg of curing agent. A drop of the mixture was placed on a 100 µm thick glass slide and spin coated 60 seconds at 1000 rpm to achieve a flat 55 µm PDMS thickness. Finally, the sample was cured in a hot plate at 80° C for 60 minutes.

To compensate the GVD of the fiber, we use a prism pair in a reflection configuration shown in Fig. 8. For their high angular dispersion and low loss, we selected prisms made of SF10 of 50x50x50 mm. The separation between the prisms was calculated using the following group delay dispersion (GDD) equation for a prism pair [40]:

GDDprism pair=λ32πc2[4l{[d2ndλ2+(2n1n3)(dndλ)2]sinβ2(dndλ)2cosβ}+8D(d2ndλ2)]
where n is the refractive index of SF10, D is the beam diameter and β=2dndλΔλ.

The prisms themselves introduce a GDD of the same sign as the GVD introduced by the fiber. To compensate that, the separation between the prisms is larger than that of a prism pair made of infinitesimally small prisms. For GDD adjustment, the second prism was mounted in a 2D stage that moves parallel and perpendicular to the beam connecting prism 1 (P1) and prism 2 (P2).

Acknowledgments

We would like to thank Dr. Carlos Macias-Romero from the Laboratory for Fundamental BioPhotonics at EPFL, for his helpful discussions about microscopy and nonlinear optics.

References and links

1. B. A. Flusberg, E. D. Cocker, W. Piyawattanametha, J. C. Jung, E. L. M. Cheung, and M. J. Schnitzer, “Fiber-optic fluorescence imaging,” Nat. Methods 2(12), 941–950 (2005). [CrossRef]   [PubMed]  

2. K. Carlson, M. Chidley, K.-B. Sung, M. Descour, A. Gillenwater, M. Follen, and R. Richards-Kortum, “In vivo fiber-optic confocal reflectance microscope with an injection-molded plastic miniature objective lens,” Appl. Opt. 44(10), 1792–1797 (2005). [CrossRef]   [PubMed]  

3. C. Liang, K.-B. Sung, R. R. Richards-Kortum, and M. R. Descour, “Design of a high-numerical-aperture miniature microscope objective for an endoscopic fiber confocal reflectance microscope,” Appl. Opt. 41(22), 4603–4610 (2002). [CrossRef]   [PubMed]  

4. A. R. Rouse, A. Kano, J. A. Udovich, S. M. Kroto, and A. F. Gmitro, “Design and demonstration of a miniature catheter for a confocal microendoscope,” Appl. Opt. 43(31), 5763–5771 (2004). [CrossRef]   [PubMed]  

5. W. Göbel, J. N. D. Kerr, A. Nimmerjahn, and F. Helmchen, “Miniaturized two-photon microscope based on a flexible coherent fiber bundle and a gradient-index lens objective,” Opt. Lett. 29(21), 2521–2523 (2004). [CrossRef]   [PubMed]  

6. P. M. Lane, A. L. P. Dlugan, R. Richards-Kortum, and C. E. Macaulay, “Fiber-optic confocal microscopy using a spatial light modulator,” Opt. Lett. 25(24), 1780–1782 (2000). [CrossRef]   [PubMed]  

7. A. F. Gmitro and D. Aziz, “Confocal microscopy through a fiber-optic imaging bundle,” Opt. Lett. 18(8), 565–567 (1993). [CrossRef]   [PubMed]  

8. A. R. Rouse and A. F. Gmitro, “Multispectral imaging with a confocal microendoscope,” Opt. Lett. 25(23), 1708–1710 (2000). [CrossRef]   [PubMed]  

9. I. M. Vellekoop and A. P. Mosk, “Focusing coherent light through opaque strongly scattering media,” Opt. Lett. 32(16), 2309–2311 (2007). [CrossRef]   [PubMed]  

10. M. Kim, Y. Choi, C. Yoon, W. Choi, J. Kim, Q. H. Park, and W. Choi, “Maximal energy transport through disordered media with the implementation of transmission eigenchannels,” Nat. Photonics 6(9), 583–585 (2012). [CrossRef]  

11. D. B. Conkey, A. M. Caravaca-Aguirre, and R. Piestun, “High-speed scattering medium characterization with application to focusing light through turbid media,” Opt. Express 20(2), 1733–1740 (2012). [CrossRef]   [PubMed]  

12. S. M. Popoff, G. Lerosey, R. Carminati, M. Fink, A. C. Boccara, and S. Gigan, “Measuring the Transmission Matrix in Optics: An Approach to the Study and Control of Light Propagation in Disordered Media,” Phys. Rev. Lett. 104(10), 100601 (2010). [CrossRef]   [PubMed]  

13. I. M. Vellekoop and A. P. Mosk, “Universal Optimal Transmission of Light Through Disordered Materials,” Phys. Rev. Lett. 101(12), 120601 (2008). [CrossRef]   [PubMed]  

14. I. N. Papadopoulos, S. Farahi, C. Moser, and D. Psaltis, “Focusing and scanning light through a multimode optical fiber using digital phase conjugation,” Opt. Express 20(10), 10583–10590 (2012). [CrossRef]   [PubMed]  

15. I. N. Papadopoulos, S. Farahi, C. Moser, and D. Psaltis, “High-resolution, lensless endoscope based on digital scanning through a multimode optical fiber,” Biomed. Opt. Express 4(2), 260–270 (2013). [CrossRef]   [PubMed]  

16. A. Yariv, “Three‐dimensional pictorial transmission in optical fibers,” Appl. Phys. Lett. 28(2), 88–89 (1976). [CrossRef]  

17. P. R. Forman, F. C. Jahoda, and B. L. Mason, “Multimode Fiber Interferometry with and without Phase Conjugation,” Appl. Opt. 30(13), 1629–1632 (1991). [CrossRef]   [PubMed]  

18. I. McMichael, P. Yeh, and P. Beckwith, “Correction of Polarization and Modal Scrambling in Multimode Fibers by Phase Conjugation,” Opt. Lett. 12(7), 507–509 (1987). [CrossRef]   [PubMed]  

19. Y. Pu, X. Yang, I. Papadoupoulos, S. Farahi, C.-L. Hsieh, C. A. Ong, C. Moser, D. Psaltis, and K. Stankovic, “Imaging of the Mouse Cochlea with Two-photon Microscopy and Multimode Fiber-based Microendoscopy,” in Biomedical Optics 2014, OSA Technical Digest (online) (Optical Society of America, 2014), BT4A.3.

20. M. Ploschner, T. Tyc, and T. Cizmar, “Seeing through chaos in multimode fibres,” Nat. Photonics 9(8), 529–535 (2015). [CrossRef]  

21. Y. Choi, C. Yoon, M. Kim, T. D. Yang, C. Fang-Yen, R. R. Dasari, K. J. Lee, and W. Choi, “Scanner-free and wide-field endoscopic imaging by using a single multimode optical fiber,” Phys. Rev. Lett. 109(20), 203901 (2012). [CrossRef]   [PubMed]  

22. D. Loterie, S. Farahi, I. Papadopoulos, A. Goy, D. Psaltis, and C. Moser, “Digital confocal microscopy through a multimode fiber,” Opt. Express 23(18), 23845–23858 (2015). [CrossRef]   [PubMed]  

23. D. Kim, J. Moon, M. Kim, T. D. Yang, J. Kim, E. Chung, and W. Choi, “Toward a miniature endomicroscope: pixelation-free and diffraction-limited imaging through a fiber bundle,” Opt. Lett. 39(7), 1921–1924 (2014). [CrossRef]   [PubMed]  

24. R. Di Leonardo and S. Bianchi, “Hologram transmission through multi-mode optical fibers,” Opt. Express 19(1), 247–254 (2011). [CrossRef]   [PubMed]  

25. T. Cižmár and K. Dholakia, “Exploiting multimode waveguides for pure fibre-based imaging,” Nat. Commun. 3, 1027 (2012). [CrossRef]   [PubMed]  

26. T. Čižmár and K. Dholakia, “Shaping the light transmission through a multimode optical fibre: complex transformation analysis and applications in biophotonics,” Opt. Express 19(20), 18871–18884 (2011). [CrossRef]   [PubMed]  

27. S. Bianchi and R. Di Leonardo, “A multi-mode fiber probe for holographic micromanipulation and microscopy,” Lab Chip 12(3), 635–639 (2012). [CrossRef]   [PubMed]  

28. M. Plöschner, B. Straka, K. Dholakia, and T. Čižmár, “GPU accelerated toolbox for real-time beam-shaping in multimode fibres,” Opt. Express 22(3), 2933–2947 (2014). [CrossRef]   [PubMed]  

29. A. M. Caravaca-Aguirre, E. Niv, and R. Piestun, “High-Speed Phase Modulation for Multimode Fiber Endoscope,” in Imaging and Applied Optics 2014, OSA Technical Digest (online) (Optical Society of America, 2014), ITh3C.1.

30. R. Y. Gu, R. N. Mahalati, and J. M. Kahn, “Noise-reduction algorithms for optimization-based imaging through multi-mode fiber,” Opt. Express 22(12), 15118–15132 (2014). [CrossRef]   [PubMed]  

31. E. R. Andresen, G. Bouwmans, S. Monneret, and H. Rigneault, “Two-photon lensless endoscope,” Opt. Express 21(18), 20713–20721 (2013). [CrossRef]   [PubMed]  

32. E. E. Morales-Delgado, S. Farahi, I. N. Papadopoulos, D. Psaltis, and C. Moser, “Delivery of focused short pulses through a multimode fiber,” Opt. Express 23(7), 9109–9120 (2015). [CrossRef]   [PubMed]  

33. D. G. Shamir Rosen, O. Katz, and Y. Silberberg, Focusing and Scanning through Flexible Multimode Fibers without Access to the Distal End,” http://arxiv.org/abs/1506.08586 (2015).

34. L. V. Amitonova, A. P. Mosk, and P. W. H. Pinkse, “The Rotational memory effect of a multimode fiber,” Opt. Express 23(16), 20569–20575 (2015). [CrossRef]   [PubMed]  

35. R. Y. Gu, R. N. Mahalati, and J. M. Kahn, “Design of flexible multi-mode fiber endoscope,” Opt. Express 23(21), 26905–26918 (2015). [CrossRef]   [PubMed]  

36. B. E. A. Saleh and M. C. Teich, Fundamentals of Photonics, 2nd ed. (Wiley, 2007).

37. W. Denk, J. H. Strickler, and W. W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science 248(4951), 73–76 (1990). [CrossRef]   [PubMed]  

38. M. Müller, J. Squier, and G. J. Brakenhoff, “Measurement of femtosecond pulses in the focal point of a high-numerical-aperture lens by two-photon absorption,” Opt. Lett. 20(9), 1038–1040 (1995). [CrossRef]   [PubMed]  

39. O. Lammel and A. Penzkofer, “Femtosecond pulse duration measurement by two-photon fluorescence detection,” Opt. Quantum Electron. 32(10), 1147–1160 (2000). [CrossRef]  

40. R. L. Fork, O. E. Martinez, and J. P. Gordon, “Negative dispersion using pairs of prisms,” Opt. Lett. 9(5), 150–152 (1984). [CrossRef]   [PubMed]  

Supplementary Material (1)

NameDescription
Visualization 1: AVI (923 KB)      Optical sectioning

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (8)

Fig. 1
Fig. 1 (a) Experimental Setup. Calibration step. A calibration beam, previously pre-chirped for compensation of GVD, is focused and coupled into the multimode fiber. The field emerging from the proximal side of the fiber is interfered with the reference beam. An off-axis digital hologram is recorded. Reconstruction step. The flip mirror is removed. The reference reconstructs the phase conjugated version of the recorded hologram. Light counter propagates through the fiber, generating the initial focus on the distal side. Imaging step. The focused pulses are sequentially scanned over the specimen. The emitted two-photon fluorescence (2PF) is collected through the fiber and measured with the PMT.
Fig. 2
Fig. 2 Scanning of the focused spot at the fundamental wavelength of 800 nm. (a) Sequential superposition of a focused pulse scanned in a 80x80 µm field of view. Scanning step is 4 µm. (b) Intensity enhancement of the focused pulse as a function of scanning position. (c) FWHM size of the focused pulses as a function of scanning position. (d) Line profile of (b) at the location of the dashed lines. (e) Line profile of (c) at the location of the dashed lines. Scale bars are 10 µm. The dashed orange circle represents the core of the multimode fiber.
Fig. 3
Fig. 3 Pulse width characterization. Intensity of the generated focused pulses at 800 nm without (a) and with (b) GVD compensation. (c) Intensity enhancement vs position. (d) Pulse width vs position. (e), (f) Sample interferometric autocorrelation traces of the central focused pulses with and without GVD compensation respectively. (g) Intensity autocorrelation of the delivered pulse when no wavefront control is used. Scale bars are 10 µm. The dashed orange semi-circle in (a) represents the core of the multimode fiber.
Fig. 4
Fig. 4 Two-photon images of fluorescent beads. (y,x,z) Field of view is 28x49x50 µm. Scale bars are 10 µm. To view all the frames see Visualization 1.
Fig. 5
Fig. 5 Normalized fluorescent photons as a function of position. Scale bar is 10 µm.
Fig. 6
Fig. 6 Experimental PSF of the multimode fiber imaging system. (a) Single-photon fluorescence. (b) Two-photon fluorescence. (c) Normalized PSF of (a) and (b) at y = 0.
Fig. 7
Fig. 7 Multimode fiber probe. The white horizontal lines are the reflection of the lamp of the microscope.
Fig. 8
Fig. 8 Complete diagram of the experimental setup.

Equations (2)

Equations on this page are rendered with MathJax. Learn more.

n a p 0 2 δ τ p f p 2 ( N A 2 2 h c λ ) 2
G D D p r i s m   p a i r = λ 3 2 π c 2 [ 4 l { [ d 2 n d λ 2 + ( 2 n 1 n 3 ) ( d n d λ ) 2 ] sin β 2 ( d n d λ ) 2 cos β } + 8 D ( d 2 n d λ 2 ) ]
Select as filters