Abstract
A constant output monochromator has been developed that produces a power of 100 ± 2 μW over a wavelength range from 375 nm to 1400 nm with a spectral bandpass of 20 nm. The output is regulated by controlling the source intensity and the monochromator transmission. A thermopile is incorporated into the instrument for calibrating the output of the open loop system.
© 1968 Optical Society of America
Full Article | PDF ArticleMore Like This
David A. Wayne and Norman G. Dillman
Appl. Opt. 8(3) 603-605 (1969)
Kris M. Rosfjord, Ricardo A. Villalaz, and Thomas K. Gaylord
Appl. Opt. 39(4) 568-572 (2000)
Edward H. Eberhardt
Appl. Opt. 7(10) 2037-2047 (1968)