Abstract
A vacuum echelle spectrometer for the near and medium infrared is described together with its prism-order selecting and detector optical systems. The high-dispersion part of the optical system uses a 10.5 cm × 20.2 cm 30-line/mm echelle in a 2.5-m Littrow optical system having 7.5-cm long slits. The limit of resolution at 15 μ using a thermocouple detector is 0.08 cm−1, at 2.5 μ using a cooled PbS cell 0.03 cm−1 can be resolved. Wavelengths can be measured with an accuracy of several thousandths of a wavenumber.
© 1965 Optical Society of America
Full Article | PDF ArticleMore Like This
K. P. Yates and R. F. Buhl
J. Opt. Soc. Am. 45(3) 192-201 (1955)
T. K. McCubbin, R. P. Grosso, and J. D. Mangus
Appl. Opt. 1(4) 431-436 (1962)
C. S. C. Tarbet and E. F. Daly