Abstract
The principle and the alignment aberration functions are described for an axicon-type test interferometer for measuring cylindrical mantle surfaces. Additionally, we show that the derived systematic alignment functions fulfill for reasonably small misalignments the requirements for measurements in the range of approximately 1/100 of a fringe. We verify this with optical path-length calculations, using ray tracing.
© 1997 Optical Society of America
Full Article | PDF ArticleMore Like This
Martin Šarbort, Šimon Řeřucha, Miroslava Holá, Zdeněk Buchta, and Josef Lazar
Appl. Opt. 54(33) 9930-9938 (2015)
Junzheng Peng, Dongbao Ge, Yingjie Yu, Kesheng Wang, and Mingyi Chen
Appl. Opt. 52(30) 7311-7323 (2013)
Johannes Pfund, Norbert Lindlein, and Johannes Schwider
Appl. Opt. 40(4) 439-446 (2001)