Abstract
We extend the absolute quantum efficiency (QE) calibration method, previously used only for photon counters, to analog detectors with a high dispersion of single-photon responses. Our approach is demonstrated on a reference-free measurement of a photomultiplier tube (PMT) cathode QE with 4% relative uncertainty. It involves not only measuring the biphoton field correlation function, but also a special approximation of the distribution of detector readings in order to determine the average number of photoelectrons and the average charge corresponding to a single photoelectron. Results of the calibration were verified in an independent experiment; the numbers of incident photons detected by the PMT and a single-photon counter are in good agreement.
© 2022 Optica Publishing Group
Full Article | PDF ArticleMore Like This
A. Avella, I. Ruo-Berchera, I. P. Degiovanni, G. Brida, and M. Genovese
Opt. Lett. 41(8) 1841-1844 (2016)
D. A. Safronenkov, N. A. Borshchevskaya, T. I. Novikova, K. G. Katamadze, K. A. Kuznetsov, and G. Kh. Kitaeva
Opt. Express 29(22) 36644-36659 (2021)
Minchol Lee, Donghoon Koo, and Jeongmin Kim
Opt. Lett. 48(1) 5-8 (2023)