Ray tracing models for estimating light collection properties of microstructured tapered optical fibers for optical neural interfaces

Open Access Open Access

Abstract

Tapered optical fibers (TFs) were recently employed for depth-resolved monitoring of functional fluorescence in subcortical brain structures, enabling light collection from groups of a few cells through small optical windows located on the taper edge [Pisano et al., Nat. Methods 16, 1185 (2019) [CrossRef]  ]. Here we present a numerical model to estimate light collection properties of microstructured TFs implanted in scattering brain tissue. Ray tracing coupled with the Henyey–Greenstein scattering model enables the estimation of both light collection and fluorescence excitation fields in three dimensions, whose combination is employed to retrieve the volume of tissue probed by the device.

© 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

The possibility to genetically express fluorescent indicators of neural activity in subpopulations of neurons has made it possible to obtain unprecedented insight into signaling in the living brain. A wide set of molecular probes can be employed to monitor intracellular calcium transients, transmembrane potential, or neurotransmitter concentration [14]. These tools can be engineered to target specific cell types, and their overall response in terms of fluorescence dynamic range can be improved by restricting the expression to specific subcellular compartments [3]. However, exciting and collecting fluorescence in subcortical structures of the mouse brain remain a challenge, as advanced multiphoton microscopy techniques are still limited to depths below ${\sim}{{1}}\;{\rm{mm}}$ [5]. Fiber photometry (FP) is the most common method to monitor functional fluorescence in deeper brain regions [6]. In its most conventional implementation, it exploits flat-cleaved optical fibers to excite fluorescence and collect the resulting time-varying optical signal. This method has enabled important applications to the study of functional connectivity [7], and it is compatible with fluorescence lifetime measurements [8], functional magnetic resonance imaging [9], and analysis in freely moving animals and electrophysiology [10]. However, FP also shows important limitations when small cellular groups are targeted, since both excitation light and emitted fluorescence undergo tissue attenuation and scattering, strongly linking the detection volume size to the size of the fiber core [11].

Technologies have been proposed to overcome this limitation, including the use of implanted micro light-emitting diodes (µLEDs) coupled to micro photodetectors (µPDs) [12] and tapered optical fibers (TFs) [1,13]. With µLED/µPD pairs, the lower bound of detection volume is limited by the minimum size of the active elements and by the Lambertian emission profile of µLEDs. Conversely, TFs can be microstructured to tailor the volume over which signals are collected. This is achieved by covering the surface of the taper with a reflective metal to build a metal-confined narrowing waveguide, and then opening small windows, with focused ion beam (FIB) milling [14] or direct laser writing [15], through which light can couple to guided modes. However, the experimental estimation of emission and detection volumes from small detection points in living tissue is difficult, and, due to tissue heterogeneity, most certainly requires averaging across a high number of independent measurements. Conversely, although several works have proposed methods to estimate how light spreads in scattering tissue from implanted devices [16,17], the community would greatly benefit from generalized approaches to assess the collection efficiency of implanted photonic systems. We present a numerical method to evaluate the collection properties of optical windows on microstructured TFs (µTFs), taking into account tissue attenuation and scattering. Collection efficiency maps are first evaluated experimentally in stained brain slices by using a two-photon microscopy system synchronized with light collection. These results are used as a point of comparison to implement a ray tracing model (RT) based on a point source scanning in the volume close to the optical window, giving access to a three-dimensional (3D) collection efficiency map and to the related sensitivity volume. The combination of collection efficiency maps $\eta (x,y,z)$ with emission patterns, obtained with the same ray tracing approach, allows access to the photometry efficiency field $\rho (x,y,z)$, thus estimating the collection properties when the same optical window is used to excite and collect fluorescence.

µTFs were realized by micropatterning an aluminum-coated TF with FIB milling to open small apertures on the taper edge [14,18], with a resolution of ${\sim}{{20}}\;{\rm{nm}}$. A ${\sim}{\rm{600 - nm}}$-thick Al layer was deposited on a TF with numerical aperture (NA) 0.39 and taper angle $\psi = {{4}}^\circ$, and a ${{45}}\;{\rm{\unicode{x00B5}{\rm m}}}\;{{\times}}\;{{45}}\;{\rm{\unicode{x00B5}{\rm m}}}$ optical window was milled at a taper diameter of ${\sim}{{50}}\;{\rm{\unicode{x00B5}{\rm m}}}$ (Fig. 1A). The process was stopped when the entire thickness of the metal coating was removed.

 figure: Fig. 1.

Fig. 1. A, scanning electron microscope (SEM) image of µTF with a ${{45}}\;{\rm{\unicode{x00B5}{\rm m}}} \times {{45}}\;{\rm{\unicode{x00B5}{\rm m}}}$ optical window. B, optical setup implemented to measure the light collection properties of µTFs. C, representative data collected by a ${{45}}\;{\rm{\unicode{x00B5}{\rm m}}} \times {{45}}\;{\rm{\unicode{x00B5}{\rm m}}}$ optical window and detected by the fiber PMT, with the related signal on the µscope PMT at different values of ${{z}}$.

Download Full Size | PDF

Light collection properties of µTFs were first measured in a quasi-transparent phosphate saline buffer (PBS):fluorescein solution and then in fluorescein-stained coronal mouse brain slices to estimate the influence of tissue scattering. The optical setup implemented for this purpose is shown in Fig. 1B [11]. A femtosecond-pulsed laser beam at $\lambda = {{920}}\;{\rm{nm}}$ (Chameleon Discovery, Coherent) was sent to a galvanometric mirrors pair (GMs, Cambridge Technology 6215H), a scan and tube lens (focal lengths 50 mm and 200 mm), and an objective lens allowing for a field-of-view (FOV) diameter of ${\sim}{2.0}\;{\rm{mm}}$ (Olympus XLFluor ${{4\times/340a - \rm NA}}\;{ = }\;{0.28}$). The fluorescence spot (in-plane point spread function 3 µm, axial size 32 µm [11]) was raster-scanned close to the optical window (Fig. 1B), with a fraction of the generated fluorescence collected by the window, propagated along the fiber, collimated by a microscope objective (Olympus UIS2 Plan ${{N}}$ ${{40\times}}$), and detected by a photomultiplier tube (fiber PMT, Hamamatsu H7422P-40 SEL). Once synchronized to the GMs’ position, the data acquired by the fiber PMT provide the light collection field $\eta (x,y)$ of the µTF under investigation. Part of the signal is also collected by the microscope objective, and, through an epifluorescence path, it is sent to a second photomultiplier (µscope PMT, Hamamatsu H10770PA-40) to estimate the uniformity of the generated fluorescence intensity in the FOV. Examples of the signals collected by the two PMTs are displayed in Fig. 1C for a µTF featuring a ${{45}} \times {{45}}$ µm window in a 30 µM PBS:fluorescein solution (at different values of ${{z}}$). Standard deviation of intensity detected by the µscope PMT in the scanning region was kept at ${\sim}{{4}}\%$. Similar considerations are valid for data acquired at different positions along $z$, which give access to the 3D collection field $\eta (x,y,z)$, displayed for ${{z}} = {{20}}$, 40, and 60 µm in Fig. 1C.

The same µTF was tested in a 300-µm-thick coronal slice of mouse brain uniformly stained with fluorescein. Slices were obtained from wild-type C57/Bl6 mouse fixed with 4% paraformaldeyde (PFA) and then incubated with Triton TX-100 for 30 min for permeabilization. The staining was performed in 1 mM PBS:fluorescein solution overnight. The 300 µm thickness was chosen as the maximum thickness allowing for an even staining of the slice. The resulting $\eta (x,y)$ at ${\rm{z}} = {{0}}\;{\rm{\unicode{x00B5}{\rm m}}}$ is displayed in Fig. 2A. Due to tissue scattering, light collection is more confined close to the ${{45}}\;{{\times}}\;{{45}}\;{\rm{\unicode{x00B5}{\rm m}}}$ optical window with respect to the measurements performed in quasi-transparent solution. This is confirmed by the comparison between the intensity decay along the collection lobe’s main axis in Fig. 2B (90% decay in brain and in quasi-transparent medium occurs after 90 and 150 µm, respectively).

 figure: Fig. 2.

Fig. 2. A, representative data collected by the fiber PMT. B, comparison between the intensity profile decay in brain slices and in solution.

Download Full Size | PDF

 figure: Fig. 3.

Fig. 3. A (left), ray tracing layout used to estimate $\eta$; A (right), $\eta$ obtained by scanning the point source. B, convergence study as a function of ${{N}}$ (left) and $l$ (right). C, comparison between $\eta (x,y)$ for two values of $l$ (left and center) and their ratio (right). The ratio is computed only where a significative signal was found. D, comparison between experimental and numerical data along the profile lines in the inset.

Download Full Size | PDF

The RT was built using Zemax OpticStudio (Fig. 3A), with a µTF featuring an optical window and a point source in the position $(x,y)$ emitting at $\lambda = {{520}}\;{\rm{nm}}$. Simulations were run on a workstation with Xeon E5-2630 v4 processor and 64 GB of random access memory. Rays emitted from the point source are scattered in the environment, enter the fiber through the window, and are backpropagated toward a detector placed at the end of straight portion of the fiber of length $l$. To reconstruct $\eta (x,y)$, each source position in the $x {-} y$ plane requires a RT run. A representative output is displayed in Fig. 3A. The taper’s refractive index was set at ${{{n}}_t} = {1.46}$ [19], while the environment at ${{{n}}_e} = {1.36}$ approximating cerebral cortex’s one [11]. The Henyey–Greenstein phase function was used to model scattering in agreement with recent literature, with photon mean free path ${{\rm{\mu}}_{s\:}} = \;{0.04895}\;{\rm{mm}}$, anisotropy factor $g = {0.9254}$, and transmission $T = {0.9989}$ [11,16,20,21], describing the angular distribution of scattered rays,

$$\phi \left(\theta \right) = {\left({4\pi} \right)^{- 1}}\left({1 - {g^2}} \right){\left({1 + {g^2} - 2g\cos \theta} \right)^{- \frac{3}{2}}},$$
in which the anisotropy factor $g$ defines if a scattering event is isotropic or anisotropic $g = \oint_0^\pi 2\pi \phi (\theta)\cos \theta \sin \theta {\rm{d}}\theta .$

The scattering event is assumed anisotropic and highly forward so that ${{g}}$ approaches unity. Apart for the optical window area, the taper was coated by a 1-µm-thick Al layer set as “reflective” surface with complex refractive index ${{{n}}_{\rm Al}} = {{0}.{7 - 7}.{0i}}$. The straight portion of the fiber was modeled in a step-index configuration with core/cladding diameters 200 µm/225 µm and refractive indices of ${{{n}}_{{\rm core}}} = {1.4613}$ and ${{{n}}_{{\rm clad}}} = {1.4079}$, respectively. The length of the straight waveguide $l$ and number of rays $N$ used for each RT run were chosen after a convergence study (Fig. 3B), and set to $l = {{5}}\;{\rm{mm}}$ and $N = {{1}}{{{0}}^5}$ rays, allowing for a reasonable computational time. Convergence of the RT simulation was analyzed by running the model with the source at a fixed position while varying ${{N}}$ or ${{l}}$. The left of Figure 3B shows the variation of $\eta$ and the computational time (${{{T}}_c}$) for ${{N}}$ varying between ${{1}}{{{0}}^5}$ and ${{1}}{{{0}}^6}$ with ${{l}} = {{5}}\;{\rm{mm}}$. $\eta$ remains constant $((3,00\pm 0,01)\times 10^{-2})$, while ${{{T}}_{\rm c}}$ linearly increases. Therefore, $N$ was set to ${{1}}{{{0}}^5}$ in order to minimize computational times. The right of Figure 3B shows the variation of $\eta$ and ${{{T}}_{\rm c}}$ as a function of ${{l}}$ in the range from ${{l}} = {{1}}\;{\rm{mm}}$ to ${{l}} = {{100}}\;{\rm{mm}}$ with ${{N}} = {{1}}{{{0}}^5}$. Both ${{{T}}_{\rm c}}$ and $\eta$ converge to a plateau value starting from ${{l}} = {{50}}\;{\rm{mm}}$, with the single pixel simulation time ${{{T}}_{\rm c}}$ increasing from 5 s to 30 s for from ${{l}} = {{5}}\;{\rm{mm}}$ to ${{l}} = {{10}}\;{\rm{mm}}$. We therefore investigated the differences in the simulation results for these two values of ${{l}}$, reported in Fig. 3C (left and center). The right of Figure 3C shows the ratio between the data for the two lengths: by varying the position of the source within the entire simulation plane, the collected light intensities are proportional, with an average ratio in the simulation domain of ${0.84}\;{{\pm}}\;{0.01}$ (${\rm{mean}}\;{{\pm}}\;{\rm{std}}$ deviation). To contain simulation times without affecting the simulation outcome, in the following, RT results are normalized to the maximum collected signal for ${{l}} = {{5}}\;{\rm{mm}}$. In these conditions, we obtained a total run time of ${\sim}{1.7}\;{\rm{h}}$ in the ${{60}}\;{{\times}}\;{{20}}$ pixel domain (computed as the time required to calculate the result for one pixel multiplied by the number of pixels in the domain), compared to ${\gt}{{10}}\;{\rm{h}}$ required for ${{l}}\; \gt \;{{10}}\;{\rm{mm}}$. The size of the simulation domain was established after preliminary runs to find the adequate dimensions for measuring the entire volume with non-negligible signal collection. The comparison between RT simulation and experimental data for a ${{45}} \times {{45}}\;{\rm{\unicode{x00B5}{\rm m}}}$ window is displayed in Fig. 3D, showing numerical and measured decays along three different axes across the collection lobe, as shown in the insets. The good agreement between experiments and simulations suggests that the proposed model can be employed to estimate collection properties from µTFs. We suggest the small differences between numerical and experimental data at the edge of the window (center and right panels) are due to diffractive effects of the window’s edge not taken into account by the RT model.

Extending the scan of the point source to the $z$ direction allows assessment of the 3D collection field $\eta (x,y,z)$ for the window and evaluation of the overall collection volume $V$ as a function of window size and position along the taper (Fig. 4A, left). The volumetric domain was set to ${{200}} \times {{100}} \times {{120}}\;{\rm{\unicode{x00B5}{\rm m}}}$ with a source point scan step of 10 µm, and extended to ${{200}} \times {{100}} \times {{200}}\;{\rm{\unicode{x00B5}{\rm m}}}$ for the largest window size (${{65}} \times {{65}}\;{\rm{\unicode{x00B5}{\rm m}}}$). After each simulation, $\eta (x,y,z)$ maps were processed to estimate the volume enclosed by the iso-intensity surfaces at 10%, 20%, 40%, 60%, and 80% of the maximum collection intensity. This was done for three different square windows with side ${{W}} = {{25}}$, 45, and 65 µm placed at three different locations (${{L}} = {{300}}$, 750, and 1200 µm from the tip) along a taper with angle 4°. The minimum ${{L}}$ was chosen to allow for a limited roughness of the milled surface, while the minimum ${{W}}$ was chosen to have a comparable size with cellular size. As shown in Fig. 4B, ${{V}}$ increases as a function of ${{W}}$ for all the investigated ${{L}}$, following the increase of window’s size at fixed distance from the taper’s tip.

 figure: Fig. 4.

Fig. 4. A, schematic representation of the 3D scanning (left) and reconstructed iso-$\eta$ surfaces used to estimate collection volumes. B and C, volumes contained within the iso-$\eta$ surfaces as a function of ${{W}}$ and ${{L}}$.

Download Full Size | PDF

Instead, Fig. 4C displays the estimated volumes as a function of ${{L}}$ for the different values of ${{W}}$. For the two smallest windows, the collection volumes remain mostly unchanged along the taper, whereas for ${{W}} = {{65}}\;{\rm{\unicode{x00B5}{\rm m}}}$, an increase of about one order of magnitude was recorded for ${{L}}$ increasing from 300 to 1200 µm. Although this approach allows estimation of the collection fields from optical windows, in a typical FP experiment, functional fluorescence is excited through the same device. Therefore, the intensity of the photometry signal depends on the geometrical distribution of excitation light. As reported by previous works, this can be described by the photometry efficiency ($\rho$) parameter, obtained by multiplying the collection field $\eta (x,y,z)$ with the emission field normalized at its maximum $\beta (x,y,z)$ [11,22]. A RT model was implemented to model the light emission field from an optical window placed along the taper, exploiting a stack of detectors close to the window to record irradiance and obtain the related emission field $\beta (x,y,z)$ (Fig. 5A). A representative result of $\beta (x,y,z)$ for a window with ${{W}} = {{45}}\;{{\unicode{x00B5}{\rm m}}}$ and ${{L}} = {{750}}\;{\rm{\unicode{x00B5}{\rm m}}}$ is displayed in Fig. 5B, together with the related collection field $\eta (x,y,z)$ and the resulting photometry efficiency field $\rho (x,y,z)$. Figure 5C reports the volumes enclosed in the iso-surfaces at 10, 20, 40, 60, and 80% of the maximum signal for $\eta$ (${{x}},{{y}},{{z}}$) and $\rho\! ({{x}},{{y}},{{z}})$. Figure 5C reports the volumes enclosed in the iso-surfaces at 10, 20, 40, 60, and 80% of the maximum signal for $\eta$ (${{x}},{{y}},{{z}}$) and $\rho\! ({{x}},{{y}},{{z}})$, highlighting the combined effect of light delivery and collection efficiencies.

 figure: Fig. 5.

Fig. 5. A, schematic representation of the ray tracing model used to estimate emission field. B, representative 2D emission field $\beta (x,y)$ acquired at ${{z}} = {{0}}\;{{\unicode{x00B5}{\rm m}}}$ above the optical window. C, $\eta (x,y)$ on the same plane as the data in panel B. D, photometry efficiency $\rho (x,y)$ resulting from data in panels B and C. E, volumes contained in the iso-$\eta$ and iso-$\rho$ surfaces at different percentages of the maximum value.

Download Full Size | PDF

In summary, we implement a RT method that estimates the collection properties of µTFs in living tissue, taking into account tissue attenuation and scattering. Our approach enables estimation of the collection volume of optical windows with different sizes placed at different positions along the taper. In addition, the method can take into account the excitation efficiency resulting from the simulated emission pattern, in order to retrieve the photometry efficiency field, a foremost parameter in the design of FP experiments. Owing to the vibrant research surrounding the development of implantable neural probes [23], we view our method as a valuable resource for the community.

Funding

National Institutes of Health (1UF1NS108177-01, U01NS094190); European Research Council (677683, 692943); EU Horizon 2020 Framework Programme (828972).

Disclosures

L. S., B. S., and F. Pisanello are founders and hold private equity in Optogenix, a company that develops, produces, and sells technologies to deliver light into the brain. Tapered fibers commercially available from Optogenix were used as tools in the research.

REFERENCES

1. F. Pisano, M. Pisanello, S. Y. Lee, J. Lee, E. Maglie, A. Balena, L. Sileo, B. Spagnolo, M. Bianco, M. Hyun, M. De Vittorio, B. Sabatini, and F. Pisanello, Nat. Methods 16, 1185 (2019). [CrossRef]  

2. Y. Yang, N. Liu, Y. He, Y. Liu, L. Ge, L. Zou, S. Song, W. Xiong, and X. Liu, Nat. Commun. 9, 1504 (2018). [CrossRef]  

3. K. D. Piatkevich, S. Bensussen, H. Tseng, S. N. Shroff, V. G. Lopez-Huerta, D. Park, E. E. Jung, O. A. Shemesh, C. Straub, H. J. Gritton, M. F. Romano, E. Costa, B. L. Sabatini, Z. Fu, E. S. Boyden, and X. Han, Nature 574, 413 (2019). [CrossRef]  

4. T. Patriarchi, J. R. Cho, K. Merten, M. W. Howe, A. Marley, W. H. Xiong, R. W. Folk, G. J. Broussard, R. Liang, M. J. Jang, H. Zhong, D. Dombeck, M. von Zastrow, A. Nimmerjahn, V. Gradinaru, J. T. Williams, and L. Tian, Science 360, eaat4422 (2018). [CrossRef]  

5. T. Wang, C. Wu, D. G. Ouzounov, W. Gu, F. Xia, M. Kim, X. Yang, M. R. Warden, and C. Xu, eLife 9, e53205 (2020). [CrossRef]  

6. H. Adelsberger, O. Garaschuk, and A. Konnerth, Nat. Neurosci. 8, 988 (2005). [CrossRef]  

7. Y. Sych, M. Chernysheva, L. T. Sumanovski, and F. Helmchen, Nat. Methods 16, 553 (2019). [CrossRef]  

8. S. J. Lee, Y. Chen, B. Lodder, and B. L. Sabatini, Front. Neurosci. 13, 766 (2019). [CrossRef]  

9. F. Schmid, L. Wachsmuth, M. Schwalm, P.-H. Prouvot, E. R. Jubal, C. Fois, G. Pramanik, C. Zimmer, C. Faber, and A. Stroh, J. Cereb. Blood Flow Metab. 36, 1885 (2016). [CrossRef]  

10. A. A. Patel, N. McAlinden, K. Mathieson, and S. Sakata, Front. Neurosci. 14, 148 (2020). [CrossRef]  

11. M. Pisanello, F. Pisano, M. Hyun, E. Maglie, A. Balena, M. De Vittorio, B. L. Sabatini, and F. Pisanello, Front. Neurosci. 13, 82 (2019). [CrossRef]  

12. L. Lu, P. Gutruf, L. Xia, D. L. Bhatti, X. Wang, A. Vazquez-Guardado, X. Ning, X. Shen, T. Sang, R. Ma, G. Pakeltis, G. Sobczak, H. Zhang, D.-O. Seo, M. Xue, L. Yin, D. Chanda, X. Sheng, M. R. Bruchas, and J. A. Rogers, Proc. Natl. Acad. Sci. USA 115, E1373 (2018). [CrossRef]  

13. F. Pisanello, G. Mandelbaum, M. Pisanello, I. A. Oldenburg, L. Sileo, J. E. Markowitz, R. E. Peterson, A. Della Patria, T. M. Haynes, M. S. Emara, B. Spagnolo, S. R. Datta, M. De Vittorio, and B. L. Sabatini, Nat. Neurosci. 20, 1180 (2017). [CrossRef]  

14. F. Pisano, M. Pisanello, L. Sileo, A. Qualtieri, B. L. Sabatini, M. De Vittorio, and F. Pisanello, Microelectron. Eng. 195, 41 (2018). [CrossRef]  

15. A. Rizzo, E. D. Lemma, F. Pisano, M. Pisanello, L. Sileo, M. De Vittorio, and F. Pisanello, Microelectron. Eng. 192, 88 (2018). [CrossRef]  

16. G. Yona, N. Meitav, I. Kahn, and S. Shoham, eNeuro 3, 420 (2016). [CrossRef]  

17. J. M. Stujenske, T. Spellman, and J. A. Gordon, Cell Rep. 12, 525 (2015). [CrossRef]  

18. F. Pisanello, L. Sileo, I. A. Oldenburg, M. Pisanello, L. Martiradonna, J. A. Assad, B. L. Sabatini, and M. De Vittorio, Neuron 82, 1245 (2014). [CrossRef]  

19. M. Pisanello, F. Pisano, L. Sileo, E. Maglie, E. Bellistri, B. Spagnolo, G. Mandelbaum, B. L. Sabatini, M. De Vittorio, and F. Pisanello, Sci. Rep. 8, 4467 (2018). [CrossRef]  

20. J. P. Zinter and M. J. Levene, Opt. Express 19, 15348 (2011). [CrossRef]  

21. E. Maglie, M. Pisanello, F. Pisano, A. Balena, B. Spagnolo, B. L. Sabatini, M. De Vittorio, and F. Pisanello, in 21st International Conference on Transparent Optical Networks (ICTON) (2019), pp. 1–7.

22. P. R. Bargo, S. A. Prahl, and S. L. Jacques, Appl. Opt. 42, 3187 (2003). [CrossRef]  

23. F. Pisanello, Microelectron. Eng. 215, 110979 (2019). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (5)

Fig. 1.
Fig. 1. A, scanning electron microscope (SEM) image of µTF with a ${{45}}\;{\rm{\unicode{x00B5}{\rm m}}} \times {{45}}\;{\rm{\unicode{x00B5}{\rm m}}}$ optical window. B, optical setup implemented to measure the light collection properties of µTFs. C, representative data collected by a ${{45}}\;{\rm{\unicode{x00B5}{\rm m}}} \times {{45}}\;{\rm{\unicode{x00B5}{\rm m}}}$ optical window and detected by the fiber PMT, with the related signal on the µscope PMT at different values of ${{z}}$ .
Fig. 2.
Fig. 2. A, representative data collected by the fiber PMT. B, comparison between the intensity profile decay in brain slices and in solution.
Fig. 3.
Fig. 3. A (left), ray tracing layout used to estimate $\eta$ ; A (right), $\eta$ obtained by scanning the point source. B, convergence study as a function of ${{N}}$ (left) and $l$ (right). C, comparison between $\eta (x,y)$ for two values of $l$ (left and center) and their ratio (right). The ratio is computed only where a significative signal was found. D, comparison between experimental and numerical data along the profile lines in the inset.
Fig. 4.
Fig. 4. A, schematic representation of the 3D scanning (left) and reconstructed iso- $\eta$ surfaces used to estimate collection volumes. B and C, volumes contained within the iso- $\eta$ surfaces as a function of ${{W}}$ and ${{L}}$ .
Fig. 5.
Fig. 5. A, schematic representation of the ray tracing model used to estimate emission field. B, representative 2D emission field $\beta (x,y)$ acquired at ${{z}} = {{0}}\;{{\unicode{x00B5}{\rm m}}}$ above the optical window. C, $\eta (x,y)$ on the same plane as the data in panel B. D, photometry efficiency $\rho (x,y)$ resulting from data in panels B and C. E, volumes contained in the iso- $\eta$ and iso- $\rho$ surfaces at different percentages of the maximum value.

Equations (1)

Equations on this page are rendered with MathJax. Learn more.