Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Real-time measurement of CO2 isotopologue ratios in exhaled breath by a hollow waveguide based mid-infrared gas sensor

Open Access Open Access

Abstract

A hollow waveguide (HWG) based mid-infrared gas sensor using a 2.73 µm distributed feedback (DFB) laser was developed for simultaneously measuring the concentration changes of the three isotopologues 13CO2, 12CO2, and 18OC16O in exhaled breath by direct absorption spectroscopy, and then determining the 13CO2/12CO2 isotope ratio (δ13C) and 18OC16O/12CO2 isotope ratio (δ18O). The HWG sensor showed a fast response time of 3 s. Continuous measurement of δ13C and δ18O in the standard CO2 sample with known isotopic ratios for ∼2 h was performed. Precisions of 2.20‰ and 1.98‰ for δ13C and δ18O respectively at optimal integration time of 734 s were estimated from Allan variance analysis. Accuracy of −0.49‰ and −1.20‰ for δ13C and δ18O, respectively, were obtained with comparison to the values of the reference standard. The Kalman filtering method was employed to improve the precision and accuracy of the HWG sensor while maintaining high time resolution. Precision of 5.45‰ and 4.88‰ and the accuracy of 0.21‰ and −1.13‰ for δ13C and δ18O, respectively, were obtained at the integration time of 0.54 s with the application of Kalman filtering. The concentrations of 12CO2, 13CO2 and 18OC16O in breath cycles were measured and processed by Kalman filtering in real time. The measured values of δ18O and δ13C in exhaled breath were estimated to be −21.35‰ and −33.64‰, respectively, with the integration time of 1 s. This study demonstrates the ability of the HWG sensor to obtain δ13C and δ18O values in breath samples and its potential for immediate respiratory monitoring and disease diagnosis.

© 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Breathing gas analysis is an effective method for disease diagnosis or metabolic status monitoring. Breathing gas diagnostics have the advantage of being non-invasive and highly safe compared with other diagnostic methods. It can even easily collect samples of elderly people or newborns with severe illness and provide real-time results for doctors and patients, which is important for the diagnosis of early diseases. The main components of human exhaled gases include nitrogen (78%), oxygen (15-18%), water vapor (5%), carbon dioxide (4-6%) and argon (1%), hundreds of volatile organic compounds, as well as respiratory gases in the order of one millionth or less [1-2], some of which have been proven to be useful markers for disease diagnosis and metabolic disorders [3]. The changes in the contents of 13CO2, 12CO2, and 18OC16O in respiratory gases are inseparable from the health of the human body, among which 13C is especially often used to diagnose whether a suspected patient is infected with Helicobacter pylori. While a 13C-urea breath test (UBT), patients showing a delta over baseline (DOB) value higher than 4‰ being considered positive for H. pylori infection [4]. Apart from 13C, the 18O isotope in breathing gas can be used as a potential biomarker for diagnosis of diabetes [5-6], and provides a non-invasive diagnostic method for accurately assessment of type 2 diabetes (T2D) and distinguishing type 1 diabetes (T1D) from T2D. Therefore, the real-time online measuring of the δ13C and δ18O in breath CO2 is of great significance.

At present, the methods of isotope abundance detection mainly include mass spectrometry (IRMS) [7] and laser absorption spectroscopy [8-9]. Although mass spectrometry can meet the requirements of measurement accuracy and precision well, it also has some defects, such as sample pretreatment, high price, and difficulty to achieve on-line measurement. In addition, mass spectrometry requires reliable operation and maintenance by professional technicians, which greatly limits the application of this technology. In recent years, laser absorption spectroscopy has been developed to measure stable isotope. This technology has the advantages of high selectivity, high detection sensitivity, fast response, small size and simple operation, and has attracted extensive attention from researchers. Laser absorption spectroscopy for the measurement of 13CO2/12CO2 ratio in respiratory gases has focused on the three main absorption bands of CO2 at 1.6, 2.0, 2,7 and 4.3 µm. In the near-infrared 1.6 µm band, the absorption line strength of CO2 molecules is relatively weak, and the multi-pass cell cannot achieve good measurement precision [10], so the high-finesse optical cavity is usually used to increase the absorption path length, thereby improving measurement precision [11-12]. In the near-infrared 2.0 µm band, the absorption line strength of CO2 molecules is two orders of magnitude higher than that of the 1.6 µm band. It does not require a long effective optical path or a precise cavity operation to achieve a certain absorption depth, so it’s widely employed in the literatures [13-15]; and on this basis, various filtering algorithms (Kalman filtering [16], Savitzky-Golay adaptive filtering [17], empirical mode decomposition (EMD) adaptive algorithm [18], Wiener filtering [19] and wavelet denoising [20], etc.) are used to reduce system noise, thereby improving the detection sensitivity and measurement precision of 13CO2 in breathing gas. Among them, Kalman filtering method is most effective in improving detection precision. In the mid-infrared 4.3 µm band, the CO2 molecular line intensity is three orders of magnitude stronger than that of the 2.0µm band. In recent years, this band has received more and more attention from researchers [21]. In the above research, various traditional gas cells such as multi-pass cells and optical cavities were used as sample cells, but the volume of these conventional gas cells was more than 100 ml, and the gas replacement time was long, resulting in a slow response time of the system. Many respiratory gas measurements even require pre-acquisition of samples and off-line testing. The hollow waveguide has very small sample volume (∼1 ml), which enables rapid update of the breathing gas in the cell, fast response time (∼3 s) and online measurement [22-25]. As shown in Table 1, hollow waveguide (HWG) serving simultaneously as both light guides and gas transmission cell has been considered as a most ideal gas absorption cell for 13CO2/12CO2 isotope ratio measurements [26-31]. The spectral region near 2.7 µm currently covered by commercially available compact DFB lasers with modest cost meets the requirement of CO2 isotopologue ratios measurement to a high degree, though the absorption line intensities of CO2 are hundreds of times weaker than those at 4.3 µm. Meanwhile, no literature has reported measurement of δ18O in respiratory CO2 using HWG based spectroscopic methods, to our best knowledge.

Tables Icon

Table 1. Summary of various isotope ratio measurements using hollow core wave guides previously reported in the literature and the corresponding key experimental parameters). Where iHWG: substrate-integrated hollow waveguide; DAS: direct absorption spectroscopy; WMS: wavelength modulation spectroscopy; PLS: partial least-squares regression; FTIR: Fourier transform infrared spectroscopy. $\delta {}^{13}C\unicode{x2030} = 1000\left( {\frac{{{}^{13}C/{}^{12}{C_{sample}}}}{{{}^{13}C/{}^{12}{C_{reference}}}} - 1} \right)$, TTR% = 100 (labeled tracer/unlabeled tracee)

In our experiment, a mid-infrared HWG gas sensor for measurement of CO2 isotopologue ratios (δ13C and δ18O) in the breathing gas was developed. The performance of the HWG sensor was evaluated. Kalman filtering was employed to reduce the system noise and improved the detection precision with a high time resolution. The concentrations of 12CO2, 13CO2 and 18OC16O in breath cycles were measured and processed by Kalman filtering, at the same time values of δ18O and δ13C in exhaled breath were determined.

2. Experimental principle

According to Beer-Lambert law, absorption A(v) is related to output light intensity I(v) and incident light intensity I0(v).

$$A(v) = ln({I_0}(v)/I(v)) = c\sigma (v)L$$
where c is the number of molecules absorbed per cubic centimeter (in mol/cm3), σ(v) is the frequency-dependent absorption cross section (in cm2/mol), and L is the optical absorption path length (in cm). The integral absorption AI (cm−1) can be expressed as:
$${A_I} = \int {A(v)dv = \int {ln({I_0}(v)/I(v))dv} } = CL\int {\sigma (v)} dv = CLS(T)/n$$
where S(T) is the line strength of the absorbing molecule at a temperature of T in cm·mol−1, and n is the isotope abundance (ascertained by the HITRAN 2016 database). The gas stable isotope ratio and the isotope delta value [32] can be expressed as:
$$R = \frac{{{}^x{A_I}}}{{{}^a{A_I}}} \times \frac{{{}^aS/{}^an}}{{{}^xS/{}^xn}}$$
$$\delta {}^{13}C = \delta {}^{13}{C_{cal}} + \frac{{{R_{sam}} - {R_{cal}}}}{{{R_{VPDB}}}} \times {10^3}$$
$$\delta {}^{18}O = \delta {}^{18}{O_{cal}} + \frac{{{R_{sam}} - {R_{cal}}}}{{{R_{VPDB - CO2}}}} \times {10^3}$$
wherein x and a are represented as rare isotopic species (13CO2, 18OC16O) and abundant isotopic component (12CO2), respectively. R is the measured ratio of rare/abundant isotopes, Rsam and Rcal are the measured ratio of rare/abundant isotopes for the gas sample with and for the calibration gas, respectively. δ13Ccal and δ18Ocal are known δ-value of the calibration gas. RVPDB is the isotope ratio of the Vienna Peedee Belemnite (VPDB) standard. RVPDB-CO2 is Vienna Peedee Belemnite derived CO2 (VPDB-CO2).

3. Selection of spectral lines

To simultaneously measure the concentration of CO2 and its isotopes in the breathing gas, the distance between the absorption lines must be within the scanning range of the laser. It is also necessary to consider the interference of other breathing gases on CO2 and its isotope measurements. According to the HITRAN 2016 database [33], CO2 and isotopes have strong absorption lines around 2.73 µm, where the absorption line strength of CO2 molecules is on the same order of magnitude as that in the 2.0 µm band, and does not require a long effective optical path to achieve a high absorbance. Furthermore, the simultaneous measurement of CO2 and isotopes in the breathing gas can be achieved within the laser tuning range, and the line strengths of 12CO2, 18OC16O and 13CO2 are in the same order of magnitude. Therefore, in our experiment, the absorption lines of 12CO2, 18OC16O and 13CO2 were selected as 3661.4948cm−1, 3661.0834cm−1 and 3660.7684cm−1 respectively, the temperature and current of the laser controller were set to 33 °C and 131 mA respectively. The range of laser wavenumber variation during one scanning was 3660.1-3661.8 cm−1, ensuring that these three isotopologues could be detected simultaneously. Human exhaled gases mainly include N2, CO2, O2, H2O, Ar and other trace gases. According to the HITRAN 2016 database, N2, O2 and Ar are not absorbed in the target spectral range, and have no effect on CO2 and its isotopes. The concentration of CO2 and H2O in human exhaled gas is about 5%, and the strongest absorption line strength of H2O in this range is 6.93×10−23 cm−1/(molecule·cm−2), which is stronger than that of the CO2 isotope. Under the conditions of P=33.325 KPa and T=293.15 K, the absorption line in the range of 3658-3668 cm−1with concentration of 5% CO2 and 5% H2O was simulated. It can be seen from Fig. 1 that within this wavenumber range, H2O does not interfere with the measurement of CO2 and its isotopes.

 figure: Fig. 1.

Fig. 1. The simulated absorption lines of 5% H2O, 5% CO2 and the natural abundance of 13CO2 (0.0553%) and 18OC16O (0.0197%) in the range of 3658-3668 cm−1 (a); the inset shows selected absorption lines of CO2 and its isotopes in the range of 3660.6-3661.8 cm−1 (b).

Download Full Size | PDF

Table 2 gives a comparison of the strength, ground state energy and temperature coefficient among the selected three absorption lines. The temperature coefficient of the δ value is to a good approximation equal to the difference of the temperature coefficients of the rare and abundant isotopologue lines. The temperature-induced shift of the isotope ratios is given as −9.6‰/K and −26.7‰/K for the δ13C and δ18O, respectively. The ground state energy of 12CO2 is more than double the ground state energy of the other two molecules. Therefore, the temperature control accuracy of the hollow waveguide must be very high, which increases the complexity of the experimental device. A temperature control device was used to control the temperature of the HWG. The temperature was preset to 25 °C, and maintained constant within ±0.2 °C by the temperature controller.

Tables Icon

Table 2. Comparison of line strength, ground state energy and temperature coefficient among three absorption lines

4. Experimental description

4.1 Experimental set-up

The CO2 and its isotope detection system in the breathing gas based on hollow waveguide are shown in Fig. 2(a). In the optical path section, the laser used was a 2.73 µm tunable DFB laser from Nanoplus. It had a collimating lens inside and a maximum output of 11.2mW. The temperature and current of the DFB laser were controlled using a laser controller (LDC-3724, ILX Lightwave, USA). The peak-to-peak value of the sawtooth wave signal generated by the signal generator was set to be 0.8V and its frequency was set to be 20Hz. The sawtooth wave signal was used to modulate the laser controller to control the laser output wavenumber, so that the laser scanning range could cover the absorption lines of 12CO2, 13CO2 and 18OC16O. The gas cell module was mainly composed of a hollow waveguide (HWEA 10001600) with a length of 1 m, an inner diameter of 1mm and an outer diameter of 1.6mm, and a volume of about 0.78 cm3. The light emitted by the laser passed through the collimating lens to form paraxial parallel laser beam, which was then focused to the HWG via a focusing lens. After entering the hollow waveguide, the light was reflected back and forth on the inner wall, and the emitted light was received by a photodetector (PVI-4TE-10.6, VIGO system S.A., USA). The absorption signal from the detector was digitized with a laptop using a 14-bit analogue/digital data acquisition card controlled with a Labwindows program. A 1.2 L vacuum pump (DIVAC, Germany) was used to evacuate the hollow waveguide. The mass flow controller (GV50A, MKS, USA) and the pressure controller (640, MKS, USA) were used to control the gas flow rate in the gas chamber at 4.33mL/s and the pressure at 200torr, and the pressure dynamic equilibrium was achieved. The hollow waveguide connector is shown in Fig. 2(b). In order to minimize interference from CO2 and its isotopes in the environment, the laser and photodetector was kept at a relatively close distance from the hollow waveguide. The HWG was coiled on the surface of a plastic ring made by 3D printing with a radius of 80mm for compactness, and then covered with a heater band. The HWG temperature was monitored with calibrated platinum resistors (Pt100). A PID temperature controller was used to control the temperature of the HWG. The HWG sensor with all components was packaged in a box. No temperature gradient along the fiber axis was observed.

 figure: Fig. 2.

Fig. 2. Schematic diagram of (a) detection system of breathing gas based on hollow waveguide. (b) layout of the custom-made HWG gas cell.

Download Full Size | PDF

4.2 Spectral line fitting

In order to obtain the integral absorption AI, the acquired absorption spectrum was fitted to a Voigt profile using an L-M multi-line fitting algorithm. During the fitting process, the baseline of the laser power as a function of current was represented by a 5-level polynomial. Figure 3 shows the experimental data of the isotope absorption line of carbon dioxide near 2.73 µm, which was measured at 296.15 K at a pressure of 33.325 KPa, the average of 100 laser scans. It can be seen from Fig. 3 that the absorption line obtained by the experimental measurement matches well with the Voigt fitting data, and the fitting correlation is as high as 0.999. The noise in fitting residual is mainly caused by the optical stripe and random noise superimposed on the absorption signal. The corresponding concentrations of 13CO2, 18OC16O and 12CO2 are 0.0465% (45 ppm), 0.017% (170 ppm) and 4.24% (42400 ppm) respectively, and the corresponding δ13C and δ18O values are −24.05‰ and −30.22‰ determined by calculation using Eqs. (4) and (5) respectively.

 figure: Fig. 3.

Fig. 3. Result of CO2 isotope absorption spectra fitted to Voigt profile and fitting residuals

Download Full Size | PDF

4.3 Response time

Standard CO2 gases at concentrations of 0%, 3%, 7%, 11%, and 15% were prepared by dynamically mixing purity CO2 with N2 using two mass flow controllers at P=33.325 KPa and T=296.15 K. The absorption signals were measured with different concentrations of CO2 gas, and the corresponding concentration values were obtained by fitting. The relationship between the concentration and the change in time at different 13CO2 concentrations is shown in Fig. 4. It can be seen from Fig. 4 that when the CO2 gas channel is open, concentration of 13CO2 rises rapidly after a rise delay time τd (0-10% of the concentration levels in rising or falling process) of 0.54 s, and the concentration of 13CO2 is stabilized at 0.033% after 3 s response time τr (10-90% of the concentration levels in rising or falling process); when the CO2 gas channel is closed, 13CO2 concentration drops rapidly after 0.54 s of falling delay time τd, and the concentration of 13CO2 decreases to zero after 2.7 s response time τr. The response time of the system is mainly determined by the structure of the gas chamber, the gas flow rate through the gas chamber, and the processing time required by the data acquisition system.

 figure: Fig. 4.

Fig. 4. Comparison between measured gas concentration and standard gas concentration.

Download Full Size | PDF

4.4 Long-term performance of measurement of CO2 isotopologue ratios

The Allan Variance [34] is a method commonly used to assess system stability and detection limits. Under the condition of P=33.325 KPa and T=296.15 K, the reference samples of CO2 isotope (δ13C=-21.34‰, δ18O=-30.89‰) calibrated by isotope mass spectrometry (IRMS) were measured, and an absorption spectrum was collected every 0.54 s during 1.77 h. By using the least squares algorithm to fit the measured absorption line, a continuously measured concentration value could be obtained. The corresponding δ values were calculated by Eq. (4). Allan variance analysis has been performed to determine the optimum integration time and the precision. It can be seen from Fig. 5 that the measurement precision of δ13C and δ18O reaches 2.20‰ and 1.98‰ respectively with the best integration time of 734 s. Table 3 shows the averaged values of δ13C and δ18O are −20.85‰ and −32.19‰ respectively, and the absolute deviations of the measured values are −0.49‰ and −1.20‰ respectively. After Kalman filtering, the measured values of δ13C and δ18O in the 0.54 s are −21.13‰ and −32.02‰ respectively, the absolute deviations of the measured values are 0.21‰ and −1.13‰ respectively, and the measurement precision is 5.45‰ and 4.88‰ respectively with a 0.54s time resolution. Though the demonstrated precision cannot meet the requirement of clinical goals, like diagnosis of infection with Helicobacter pylori and assessment of diabetes demanded a precision better than 1‰ [35], methods for further improvement of measurement precision include: (1) increasing the length of the HWG to increase the absorption path length in the HWG; (2) using wavelength modulation spectroscopy techniques to reduce system noise; (3) controlling temperature of the HWG within ±0.1 °C. to further reduce temperature-induced δ-value drift.

 figure: Fig. 5.

Fig. 5. Measurement results of the reference samples of CO2 isotopes. The upper four panels show raw measurement of δ13C and δ18O (black lines) and the corresponding Kalman-filter output (blue lines). The Allan variances plotted in the lower panel show an optimal integration time of 734 s for the present system.

Download Full Size | PDF

Tables Icon

Table 3. Comparison between δ values measured by hollow waveguide-based isotope abundance measurement system and Isotope Mass Spectrometer (IRMS)

4.5 Real-time online measurement of respiratory cycle

To verify the practicality and effectiveness of the 12CO2, 13CO2, and 18OC16O measurement systems for breathing, a sequence diagram of real-time changes in CO2 and isotope concentrations during a tidal breathing was measured. The pressure in the hollow waveguide was controlled to be 33.325 KPa, and the breathing gas was introduced into the HWG cell by breath tube according to the direction of gas passage shown in Fig. 2 (a). The volume of each exhaled gas in the human body was about 460 ml [36], so the average 10 spectra were selected as one data point. The actual acquisition time was 312 s with time interval of 1.08 s. The respiratory gas circulation process was measured in real time. Figure 6 shows that the real-time CO2 and its isotope concentration changes during a tidal breathing and the results of Kalman filtering. The measured values of δ13C and δ18O in the exhaled gas after Kalman filtering were −21.34‰ and −33.64‰ respectively, and the measurement precision was 27.32‰ and 28.27‰ respectively. The reasons for the fluctuation of the measurement results are as follows: (1) The concentration of CO2 and isotopes in the tidal breathing process continuously rises and changes; (2) The interference effect in the hollow waveguide, fluctuations in temperature, unstable flow during human breathing, and differences in the temperature of the breathing gas itself and the temperature in the gas cell can also affect the measurement precision; (4) In terms of data processing, the fitting algorithm also brings a certain deviation. Although the deviation is small, it is still a non-negligible factor in the calculation of isotope abundance.

 figure: Fig. 6.

Fig. 6. Sequence diagram of the changes of 12CO2, 18OC16O and 13CO2 concentrations during tidal breathing process of a volunteer and the results of the Kalman filter.

Download Full Size | PDF

5. Conclusion

This paper reports the development of a mid-infrared gas sensor for measuring CO2 and its isotope concentration in breathing gas based on hollow waveguide. The sensor has the advantages of non-intrusion, small volume, fast response time, high sensitivity and real-time online measurement. In our experiment, a DFB laser with a center wavelength of 2.73 µm was used as the detection source. The hollow waveguide combined with direct absorption spectroscopy technology was used to achieve high precision measurement of CO2 isotope abundance under high concentration conditions and long-term stability performance of the measurement system. The changes in CO2 and isotope concentrations in the breathing gas from dynamic measurements are analyzed. The system stability and measurement precision are discussed in detail. It can be seen from the experiment results that under the condition of optimal integration time of 734 s, the absolute deviations of δ13C and δ18O measurements were −0.49‰ and −1.20‰ respectively, and the measurement precision could reach 2.20‰ and 1.98‰ respectively. After Kalman filtering, the absolute deviations of δ13C and δ18O in the integration time of 0.54 s were 0.21‰ and −1.13‰ respectively, and the measurement precision was 5.45‰ and 4.88‰ respectively. The concentrations of 12CO2, 13CO2 and 18OC16O in breath cycles were measured and processed by Kalman filtering in real time. The measured values of δ18O and δ13C in exhaled breath were estimated to be −21.35‰ and −33.64‰ respectively with the integration time of 1s. This study demonstrates the ability of the HWG sensor to obtain δ13C and δ18O values in breath samples, and its potential for immediate respiratory monitoring and disease diagnosis.

Funding

Key Research and Development Program of Jiangxi Province (20192BBH80019); Postgraduate Innovation Foundation of Nanchang Hangkong University (YC2018052).

Disclosures

The authors declare no conflicts of interest.

References

1. F. K. Tittel and T. H. Risby, “Current status of midinfrared quantum and interband cascade lasers for clinical breath analysis,” Opt. Eng. 49(11), 111123 (2010). [CrossRef]  

2. J. Wojtas, Z. Bielecki, T. Stacewicz, J. Mikolajczyk, and M. Nowakowski, “Ultrasensitive laser spectroscopy for breath analysis,” Opto-Electron. Rev. 20(1), 26–39 (2012). [CrossRef]  

3. T. Stacewicz, Z. Bielecki, J. Wojtas, P. Magryta, J. Mikolajczyk, and D. Szabra, “Detection of disease markers in human breath with laser absorption spectroscopy,” Opto-Electron. Rev. 24(2), 82–94 (2016). [CrossRef]  

4. L. G. V. Coelho, M. Reber, M. C. F. Passos, R. O. A. Aguiar, P. E. Casaes, M. L. Bueno, F. R. Yazaki, F. J. Castro, W. L. S. Vieira, J. M. M. Franco, and L. P. Castro, “Application of isotopeselective non-dispersive infrared spectrometry (IRIS) for the evaluation of 13C urea breath test: comparison with three concordant methods,” Braz. J. Med. Biol. Res. 32(12), 1493–1497 (1999). [CrossRef]  

5. C. Ghosh, G. D. Banik, A. Maity, S. Som, A. Chakraborty, C. Selvan, S. Ghosh, S. Chowdhury, and M. Pradhan, “Oxygen-18 isotope of breath CO2 linking to erythrocytes carbonic anhydrase activity: a biomarker for pre-diabetes and type 2 diabetes,” Sci. Rep. 5(1), 8137 (2015). [CrossRef]  

6. C. Ghosh, S. Mandal, G. D. Banik, A. Maity, P. Mukhopadhyay, S. Ghosh, and M. Pradhan, “Targeting erythrocyte carbonic anhydrase and 18O-isotope of breath CO2 for sorting out type 1 and type 2 diabetes,” Sci. Rep. 6(1), 35836 (2016). [CrossRef]  

7. E. S. F. Berman, N. E. Levin, A. Landais, S. Li, and T. Owano, “Measurement of δ18O, δ17O, and 17O-excess in water by off-Axis integrated cavity output spectroscopy and isotope ratio mass spectrometry,” Anal. Chem. 85(21), 10392–10398 (2013). [CrossRef]  

8. M. B. Esler, D. W. T. Griffith, S. R. Wilson, and L. P. Steele, “Precision trace gas analysis by FT-IR spectroscopy. 2. The 13C/12C isotope ratio of CO2,” Anal. Chem. 72(1), 216–221 (2000). [CrossRef]  

9. J. F. Becker, T. B. Sauke, and M. Loewenstein, “Stable isotope analysis using tunable diode laser spectroscopy,” Appl. Opt. 31(12), 1921–1927 (1992). [CrossRef]  

10. D. E. Cooper, R. U. Martinelli, C. B. Carlisle, H. Riris, D. B. Bour, and R. J. Menna, “Measurement of 12CO2/13CO2 ratios for medical diagnostics with 1.6-µm distributed-feedback semiconductor diode lasers,” Appl. Opt. 32(33), 6727–6731 (1993). [CrossRef]  

11. E. R. Crosson, K. N. Ricci, B. A. Richman, F. C. Chilese, T. G. Owano, R. A. Provencal, M. W. Todd, J. Glasser, A. A. Kachanov, and B. A. Paldus, “Stable isotope ratios using cavity ring-down spectroscopy: determination of 13C/12C for carbon dioxide in human breath,” Anal. Chem. 74(9), 2003–2007 (2002). [CrossRef]  

12. V. L. Kasyutich, P. A. Martin, and R. J. Holdsworth, “An off-axis cavity-enhanced absorption spectrometer at 1605 nm for the 12CO2/13CO2 measurement,” Appl. Phys. B 85(2-3), 413–420 (2006). [CrossRef]  

13. S. N. Andreev, E. S. Mironchuk, I. V. Nikolaev, V. N. Ochkin, M. V. Spiridonov, and S. N. Tskhai, “High precision measurements of the 13CO2/12CO2 isotope ratio at atmospheric pressure in human breath using a 2µm diode laser,” Appl. Phys. B 104(1), 73–79 (2011). [CrossRef]  

14. S. V. Kireev, S. L. Shnyrev, and A. A. Kondrashov, “Development of laser noninvasive on-line diagnostics of oncological diseases based on the absorption method in the 4860–4880 cm−1 spectral range,” Laser Phys. 26(7), 075601 (2016). [CrossRef]  

15. S. V. Kireev, A. A. Kondrashov, S. L. Shnyrev, and A. P. Safagaraev, “Use of pump current modulation of diode laser for increased sensitivity of detection of 13СO2 in human exhaled breath,” Laser Phys. Lett. 15(3), 035704 (2018). [CrossRef]  

16. S. V. Kireev, A. A. Kondrashov, S. L. Shnyrev, and N. V. Frolov, “Kalman’s method to improve accuracy of online 13С16O2 measurement in the exhaled human breath using tunable diode laser absorption spectroscopy,” Laser Phys. Lett. 15(9), 095701 (2018). [CrossRef]  

17. S. V. Kireev, A. A. Kondrashov, and S. L. Shnyrev, “Improving the accuracy and sensitivity of 13C online detection in expiratory air using the TDLAS method in the spectral range of 4860–4880 cm−1,” Laser Phys. Lett. 15(10), 105701 (2018). [CrossRef]  

18. S. V. Kireev, A. A. Kondrashov, S. L. Shnyrev, and I. G. Simanovsky, “Using empirical mode decomposition to increase the sensitivity of diode laser-based detection of 13CO2 in human exhaled breath,” Laser Phys. Lett. 15(8), 085704 (2018). [CrossRef]  

19. S. V. Kireev, A. A. Kondrashov, and S. L. Shnyrev, “Implementation of the adaptive Wiener filtering algorithm in the problem of measuring the 13CO2 content in expiratory air using the tunable diode laser absorption spectroscopy technique,” Laser Phys. Lett. 16(4), 045701 (2019). [CrossRef]  

20. M. Sun, H. Ma, Q. Liu, Z. Cao, G. Wang, K. Liu, Y. Huang, X. Gao, and R. Rao, “Precise 13CO2/12CO2 isotopic ratio measurements for breath diagnosis with a 2-µm diode laser,” Acta Phys. Sin. 68(21), 162–171 (2018).

21. Z. Wang, Q. Wang, J. Y. L. Ching, J. C. Y. Wu, G. Zhang, and W. Ren, “A portable low-power QEPAS-based CO2 isotope sensor using a fiber-coupled interband cascade laser,” Sens. Actuators, B 246, 710–715 (2017). [CrossRef]  

22. C. Yao, S. Gao, Y. Wang, P. Wang, W. Jin, and W. Ren, “Silica hollow-core negative curvature fibers enable ultrasensitive mid-infrared absorption spectroscopy,” J. Lightwave Technol. (to be published).

23. L. Liu, B. Xiong, Y. Yan, J. Li, and Z. Du, “Hollow waveguide-enhanced mid-infrared sensor for real-time exhaled methane detection,” IEEE Photonics Technol. Lett. 28(15), 1613–1616 (2016). [CrossRef]  

24. J. Chen, A. Hangauer, R. Strzoda, M. Fleischer, and M. C. Amann, “Low-level and ultralow-volume hollow waveguide based carbon monoxide sensor,” Opt. Lett. 35(21), 3577–3579 (2010). [CrossRef]  

25. B. Xiong, Z. Du, L. Liu, Z. Zhang, J. Li, and Q. Cai, “Hollow-waveguide-based carbon dioxide sensor for capnography,” Chin. Opt. Lett. 13(11), 111201 (2015). [CrossRef]  

26. I. Robinson, H. L. Butcher, N. A. Macleod, and D. Weidmann, “Hollow waveguide integrated laser spectrometer for 13CO2/12CO2 analysis,” Opt. Express 27(24), 35670–35688 (2019). [CrossRef]  

27. J. F. Kelly, R. L. Sams, T. A. Blake, M. Newburn, J. Moran, M. L. Alexander, and H. Kreuzer, “A capillary absorption spectrometer for stable carbon isotope ratio (13C/12C) analysis in very small samples,” Rev. Sci. Instrum. 83(2), 023101 (2012). [CrossRef]  

28. E. Tütüncü, M. Nägele, S. Becker, M. Fischer, J. Koeth, C. Wolf, S. Köstler, V. Ribitsch, A. Teuber, M. Gröger, S. Kress, M. Wepler, U. Wachter, J. Vogt, P. Radermacher, and B. Mizaikoff, “Advanced Photonic Sensors Based on Interband Cascade Lasers for Real-Time Mouse Breath Analysis,” ACS Sens. 3(9), 1743–1749 (2018). [CrossRef]  

29. K. Wörle, F. Seichter, A. Wilk, C. Armacost, T. Day, M. Godejohann, U. Wachter, J. Vogt, P. Radermacher, and B. Mizaikoff, “Breath analysis with broadly tunable quantum cascade lasers,” Anal. Chem. 85(5), 2697–2702 (2013). [CrossRef]  

30. A. Wilk, F. Seichter, S. S. Kim, E. Tütüncü, B. Mizaikoff, J. A. Vogt, U. Wachter, and P. Radermacher, “Toward the quantification of the 13CO2/12CO2 ratio in exhaled mouse breath with mid-infrared hollow waveguide gas sensors,” Anal. Bioanal. Chem. 402(1), 397–404 (2012). [CrossRef]  

31. F. Seichter, A. Wilk, K. Wörle, S. S. Kim, J. A. Vogt, U. Wachter, P. Radermacher, and B. Mizaikoff, “Multivariate determination of 13CO2/12CO2 ratios in exhaled mouse breath with mid-infrared hollow waveguide gas sensors,” Anal. Bioanal. Chem. 405(14), 4945–4951 (2013). [CrossRef]  

32. V. L. Kasyutich and P. A. Martin, “13CO2/12CO2 isotopic ratio measurements with a continuous-wave quantum cascade laser in exhaled breath,” Infrared Phys. Technol. 55(1), 60–66 (2012). [CrossRef]  

33. I. E. Gordon, L. S. Rothman, C. Hill, R. V. Kochanov, Y. Tan, P. F. Bernath, M. Birk, V. Boudon, A. Campargue, K. V. Chance, B. J. Drouin, J.-M. Flaud, R. R. Gamache, J. T. Hodges, D. Jacquemart, V. I. Perevalov, A. Perrin, K. P. Shine, M.-A. H. Smith, J. Tennyson, G. C. Toon, H. Tran, V. G. Tyuterev, A. Barbe, A. G. Császár, V. M. Devi, T. Furtenbacher, J. J. Harrison, J.-M. Hartmann, A. Jolly, T. J. Johnson, T. Karman, I. Kleiner, A. A. Kyuberis, J. Loos, O. M. Lyulin, S. T. Massie, S. N. Mikhailenko, N. Moazzen-Ahmadi, H. S. P. Müller, O. V. Naumenko, A. V. Nikitin, O. L. Polyansky, M. Rey, M. Rotger, S. W. Sharpe, K. Sung, E. Starikova, S. A. Tashkun, J. V. Auwera, G. Wagner, J. Wilzewski, P. Wcisło, S. Yu, and E. J. Zak, “The HITRAN2016 molecular spectroscopic database,” J. Quant. Spectrosc. Radiat. Transfer 203, 3–69 (2017). [CrossRef]  

34. P. Werle, R. Mücke, and F. Slemr, “The limits of signal averaging in atmospheric trace-gas monitoring by tunable diode-laser absorption spectroscopy (TDLAS),” Appl. Phys. B 57(2), 131–139 (1993). [CrossRef]  

35. E. R. Crosson, K. N. Ricci, B. A. Richman, F. C. Chilese, T. G. Owano, R. A. Provencal, M. W. Todd, J. Glasser, A. A. Kachanov, B. A. Paldus, T. G. Spence, and R. N. Zare, “Stable Isotope Ratios Using Cavity Ring-Down Spectroscopy: Determination of 13C/12C for Carbon Dioxide in Human Breath,” Anal. Chem. 74(9), 2003–2007 (2002). [CrossRef]  

36. D. D. Arslanov, K. Swinkels, S. M. Cristescu, and F. J. M. Harren, “Real-time, subsecond, multicomponent breath analysis by optical parametric oscillator based off-axis Integrated cavity output spectroscopy,” Opt. Express 19(24), 24078–24089 (2011). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (6)

Fig. 1.
Fig. 1. The simulated absorption lines of 5% H2O, 5% CO2 and the natural abundance of 13CO2 (0.0553%) and 18OC16O (0.0197%) in the range of 3658-3668 cm−1 (a); the inset shows selected absorption lines of CO2 and its isotopes in the range of 3660.6-3661.8 cm−1 (b).
Fig. 2.
Fig. 2. Schematic diagram of (a) detection system of breathing gas based on hollow waveguide. (b) layout of the custom-made HWG gas cell.
Fig. 3.
Fig. 3. Result of CO2 isotope absorption spectra fitted to Voigt profile and fitting residuals
Fig. 4.
Fig. 4. Comparison between measured gas concentration and standard gas concentration.
Fig. 5.
Fig. 5. Measurement results of the reference samples of CO2 isotopes. The upper four panels show raw measurement of δ13C and δ18O (black lines) and the corresponding Kalman-filter output (blue lines). The Allan variances plotted in the lower panel show an optimal integration time of 734 s for the present system.
Fig. 6.
Fig. 6. Sequence diagram of the changes of 12CO2, 18OC16O and 13CO2 concentrations during tidal breathing process of a volunteer and the results of the Kalman filter.

Tables (3)

Tables Icon

Table 1. Summary of various isotope ratio measurements using hollow core wave guides previously reported in the literature and the corresponding key experimental parameters). Where iHWG: substrate-integrated hollow waveguide; DAS: direct absorption spectroscopy; WMS: wavelength modulation spectroscopy; PLS: partial least-squares regression; FTIR: Fourier transform infrared spectroscopy. δ 13 C = 1000 ( 13 C / 12 C s a m p l e 13 C / 12 C r e f e r e n c e 1 ) , TTR% = 100 (labeled tracer/unlabeled tracee)

Tables Icon

Table 2. Comparison of line strength, ground state energy and temperature coefficient among three absorption lines

Tables Icon

Table 3. Comparison between δ values measured by hollow waveguide-based isotope abundance measurement system and Isotope Mass Spectrometer (IRMS)

Equations (5)

Equations on this page are rendered with MathJax. Learn more.

A ( v ) = l n ( I 0 ( v ) / I ( v ) ) = c σ ( v ) L
A I = A ( v ) d v = l n ( I 0 ( v ) / I ( v ) ) d v = C L σ ( v ) d v = C L S ( T ) / n
R = x A I a A I × a S / a n x S / x n
δ 13 C = δ 13 C c a l + R s a m R c a l R V P D B × 10 3
δ 18 O = δ 18 O c a l + R s a m R c a l R V P D B C O 2 × 10 3
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.