Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Compact ultrahigh resolution interferometric spectrometer

Open Access Open Access

Abstract

A compact ultrahigh resolving power spectrometer is presented, which combines a Fabry-Perot interferometer (FPI) and a static stepped-mirror interferometer (SMI). The FPI needs to scan N steps, one spectrum per step is obtained from the SMI, and a total of N spectra constitute an ultrahigh resolution spectrum. Compared with Michelson-type interferometers for ultrahigh resolution spectral measurements, the spectrometer is much smaller in physical size and shorter in measurement time. Compared with the combination of a FPI and a Michelson-type interferometer, the spectrometer has much shorter measurement time and higher stability. Preliminary numerical simulations are given by two examples. The spectrometer offers a unique concept that not only provides resolving power higher than 1,000,000 in near-infrared, short-wave infrared or mid-wave infrared region but also achieves short measurement time and small physical size.

© 2019 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Whether for scientific research or industrial applications, military or civil applications, earth remote sensing or space exploration, spectroscopy is of great significance and value. One of the most important performance criteria of spectrometers is spectral resolution (resolving power). The gratings used in either reflection or transmission commonly provide resolving powers from 1000 to 100,000 [1–5], but high spectral resolution gratings are generally large in physical size. For grating spectrometers, the higher the resolution, the larger the physical size and the more expensive the cost. More importantly, resolving power higher than 500,000 can only be provided by interferometric-type spectrometers.

For ultrahigh resolution spectral measurements in the infrared region, Michelson-type interferometers [6–11] still suffer three drawbacks: (1) a very large maximum optical path difference is needed, which means a very large maximum displacement of the moving mirror, which makes the instruments very large in physical size and limits their application where space and mass are at a premium; (2) a large travel of the moving part will reduce the stability against various disturbances; (3) an acceptable solid angle decreases with the increase of optical path difference, which causes the optical throughput to decrease as the spectral resolution increases.

Fabry-Perot interferometers (FPI) [12–21] provide a compact and cost-effective method to get ultrahigh spectral resolution. The FPI faces the problem of overlapping orders, which can be separated either by a fixed narrowband filter [22], a grating [23], or by another FPI [24–30], however, none of these methods can obtain resolving power higher than 1,000,000. A method using a standard Michelson interferometer to separate the FPI overlapping orders [31,32] can provide resolving power higher than 1,000,000 in near-infrared, short-wave infrared or mid-wave infrared region, however, this method still suffers three drawbacks: (1) a large travel of the moving mirror of Michelson interferometer is still needed, which greatly increases spectral measurement time and reduces the stability; (2) the biggest problem associated with the use of a standard Michelson interferometer as a Fourier transform spectrometer (tilt of moving mirror during scanning) is not solved, which greatly reduces the stability; (3) a drive system for moving mirror of Michelson interferometer is needed, which makes the instrument more complex and reduces the stability.

This paper proposes a spectrometer that can achieve ultrahigh spectral resolution, short measurement time and small physical size. After a description of the principle, preliminary numerical simulations are given by two examples with resolving power higher than 1,000,000 in short-wave infrared region. Finally, the conclusion is given.

2. Principle

Figure 1 shows the optics of the compact ultra-high resolution interferometric spectrometer (UHRIS), which consists of a FPI, a static stepped-mirror interferometer (SMI), a collimating lens, a collecting lens, and an area-array detector. The static SMI is suitable for high resolution spectral measurement in narrow spectral bands in the infrared region [33–36]. The SMI comprises one beam splitter and two fixed stepped mirrors with a shape of stairs: the stepped mirror M1 has small steps, the stepped mirror M2 has tall steps, the tall step height corresponds to the sum of the small step heights, the directions of the steps of two stepped mirrors are orthogonal, and so the SMI acquires a spatial array of optical path difference [35]. Namely, the SMI acquires a spatial sampling of the interferogram. In the UHRIS, the spectral resolution of the SMI only needs to be enough to separate the overlapping orders of the FPI. The FPI is first assigned a given spacing and a first spectrum is obtained from the SMI, then the FPI is assigned the next spacing position and the next spectrum is obtained from the SMI, and so on until the full spectral range is covered, finally the multiple spectra obtained from the SMI constitute an ultra-high resolution spectrum.

 figure: Fig. 1

Fig. 1 Optics of the UHRIS combining a FPI and a static SMI: (a) Basic optical layout and (b) Equivalent perspective view.

Download Full Size | PDF

For the normal incidence and vacuum medium between the FPI plates, the maximum transmission equation of the FPI is given by

2d=mλ.
where d is the plate spacing of the FPI, λ is the wavelength of light, and m is the order that is an integer.

The free spectral range (FSR) in wavenumber of the FPI is given by

FSRσ=12d.

The FSR in wavelength of the FPI is given by

FSRλ=λ22d.

For the normal incidence and vacuum medium between the FPI plates, the transmittance function of the FPI can be written as [12,37]

TFPI(σ,d)=(1R)21+R22Rcos(4πσd).
where σ is the wavenumber (σ=1/λ), R is the reflectance of the FPI plates.

The reflective finesse of the FPI is given by

Fr=πR1R.

For the normal incidence and vacuum medium between the FPI plates, the interferogram of the UHRIS can be expressed as

IUHRIS(σ,d,x)=0B(σ)TFPI(σ,d)[1+cos(2πσx)]dσ=0(1R)2B(σ)[1+cos(2πσx)]1+R22Rcos(4πσd)dσ.
where x is the optical path difference of the SMI, and B(σ) is the input spectral intensity at a wavenumber σ.

First of all, according to the desired spectral resolution δσUHRIS and spectral range σminσσmax, select a particular FPI spacing d0, let free spectral range FSRσ(0)=1/(2d0) be equal to P times the spectral resolution (in wavenumber) δσSMI of the SMI at a center wavenumber σ0, i.e., FSRσ(0)=PδσSMI, so the spectral resolution of the SMI is given by

δσSMI=12Pd0.
where the value of P is determined by making the spectral broadening in the SMI negligible as compared with FSRσ(0)=1/(2d0). In general P3 and P does not need to be an integer.

Secondly, the FPI needs to scan N(P<NFr) steps in order to cover the full spectral range of σminσσmax, and this scanning is generally carried out by using a piezoelectric device. The free spectral range FSRσ(0)=1/(2d0) needs to be scanned N(P<NFr) times with a scanning interval δσUHRIS in wavenumber, so the wavenumber position of center spectral peak moves from σ0FSRσ(0)/2 to σ0+FSRσ(0)/2, namely, the wavenumber position of center spectral peak moves from σ0+(N/2)δσUHRIS to σ0+(N/21)δσUHRIS. Thus, the spectral resolution of the UHRIS can be expressed as

δσUHRIS=FSRσ(0)N=12Nd0.
where P<NπR/(1R) and, for convenience, let N be an even number.

According to Eqs. (7) and (8), it can be obtained that

δσUHRIS=PNδσSMI.

The resolving power of the UHRIS can be written as

RUHRIS=λλλFSRλ/N=mN=2Ndλ.

At the first FPI spacing position d1 (i.e., at the first scanning position of the FPI), the maximum transmission wavenumbers of the FPI are σminσ=m/(2d1)σmax (m is an integer), a first spectrum is obtained from Fourier transform of the first interferogram obtained from the SMI. At the k-th FPI spacing position dk (i.e., at the k-th scanning position of the FPI), where N/2kN/21, the maximum transmission wavenumbers of the FPI are σminσ=m/(2dk)σmax (m is an integer), the k-th spectrum is obtained from Fourier transform of the k-thinterferogram obtained from the SMI. At the N-th FPI spacing position dN (i.e., at the N-th scanning position of the FPI), the maximum transmission wavenumbers of the FPI are σminσ=m/(2dN)σmax (m is an integer), the N-th spectrum is obtained from Fourier transform of the N-th interferogram obtained from the SMI. A total of N spectra are obtained from the SMI, and these N spectra constitute an ultra-high resolution spectrum.

The displacement of the k-th FPI plate spacing from the spacing d0 is given by

Δdk=kd0mN+k=kd02d0σ0N+k.
where N/2kN/21.

The spectral resolution of the SMI is also determined by

δσSMI=12xmax.
where xmax is the maximum optical path difference of the SMI.

Based on Eqs. (7) and (12), it can be obtained that

xmax=Pd0.

According to the Nyquist-Shannon sampling criterion [38,39], for a spectral bandwidth Δσ=σmaxσmin, the sampling interval is given by

χ<12Δσ.

The height of the small steps is h1, and the small step mirror M1 has s1 steps. The tall step mirror M2 has s2 steps, and the height of the tall steps is h2=h1×s1. The sampling interval for each of the N interferograms produced by the SMI is

χ=2×h1.

The measurement spectral bandwidth of the SMI is calculated by

Δσ14h1.

The number of sampling points for each of the N interferograms produced by the SMI is

K=s1×s2.

The maximum optical path difference of the SMI is xmax=χ×K+2l0=2×h1×K+2l0, where l0 is the distance between the image of stepped mirror M2 to beam splitter and the stepped mirror M1 as shown in Fig. 1. Let l0=0, it can be obtained that

K=Pd02h1.

Table 1 shows the main advantages and disadvantages of the UHRIS compared with the other two types of spectrometers with resolving power higher than 1,000,000 in short-wave infrared region.

Tables Icon

Table 1. Comparisons of the UHRIS with two other types of spectrometers with resolving power higher than 1,000,000 in near-infrared, short-wave infrared or mid-wave infrared region

It is well known that the throughput (Jacquinot) [40,41] and multiplex (Fellgett) [42,43] advantages are the two basic benefits of interferometric over dispersive spectrometers in the infrared region. The FPI has the throughput advantage as well as the Michelson-type interferometers used for Fourier transform spectrometer in the infrared region [7]. The combination of a FPI and a static SMI makes the UHRIS capable of achieving high optical throughput and therefore achieving high signal-to-noise ratio in the infrared spectral region.

Compared with the combination of a FPI and a fixed narrowband filter, the UHRIS can obtain much higher spectral resolution. Compared with the combination of a FPI and a reflection grating, the UHRIS can obtain higher spectral resolution and higher signal-to-noise ratio in the infrared spectral region [7,44,45].

3. Preliminary numerical simulation with two examples

When the reflectance of the FPI plates is R=0.95, the reflective finesse of the FPI is Fr=πR/(1R)61.2, so the number of scanning steps of the FPI is N61.

3.1. The first example

The first example is that a source spectra covers a central wavenumber σ0=6250cm1 (wavelength 1.6 µm) and a spectral bandwidth Δσ=30cm1 (a wavenumber range from 6235 cm−1 to 6265 cm−1), and the desired spectral resolution is δσUHRIS=0.005cm1. Based on Eqs. (7) and (8), some key parameters of the UHRIS are shown in Table 2. In order to cover the full wavenumber range of 6235 cm−1 to 6265 cm−1, from Eq. (11), the FPI plate spacing is approximately assigned from 1.9999616 cm to 2.00004 cm, namely, the FPI should scan N = 50 steps of approximate 16 nm in step interval. Due to assembling constraints and limitations, the sampling interval cannot be lower than 100 µm and the small step height cannot be lower than 50 µm [35], so the measurement spectral bandwidth of the SMI cannot be greater than 50 cm−1. The main parameters of the SMI in the UHRIS are shown in Table 3. According to Eq. (10), the resolving power of the UHRIS for the wavelength range from 0.8 µm to 2 µm is shown in Fig. 2.

Tables Icon

Table 2. Some key parameters of the UHRIS for the first example

Tables Icon

Table 3. Main parameters of the SMI in the UHRIS for the first example

 figure: Fig. 2

Fig. 2 Resolving power of the UHRIS for the wavelength range from 0.8 µm to 2 µm when d0=2cm and N=50.

Download Full Size | PDF

Suppose that a source spectrum contains only wavenumber 6249.745 cm−1, 6249.75 cm−1, 6249.755 cm−1, 6249.995 cm−1, 6250 cm−1 (1.6 µm), 6250.005 cm−1, 6250.245 cm−1, 6250.25 cm−1 and 6250.255 cm−1. Thus, the FPI needs to scan three steps with approximate 16 nm step interval, and three interferograms produced by the SMI of the UHRIS are shown in Fig. 3. The first interferogram (d=1.9999984cm) contains only wavenumber 6249.755 cm−1, 6250.005 cm−1 and 6250.255 cm−1. The second interferogram (d=2cm) contains only wavenumber 6249.75 cm−1, 6250 cm−1 and 6250.25 cm−1. The third interferogram (d=2.0000016cm) contains only wavenumber 6249.745 cm−1, 6249.995 cm−1 and 6250.245 cm−1. Figure 4 shows the spectrum obtained from Fourier transform of the three interferograms in Fig. 3, and the spectral resolution is 0.005 cm−1.

 figure: Fig. 3

Fig. 3 Three interferograms produced by the SMI of the UHRIS with a spectral resolution 0.005 cm−1.

Download Full Size | PDF

 figure: Fig. 4

Fig. 4 Spectrum obtained from Fourier transform of the three interferograms in Fig. 3.

Download Full Size | PDF

Table 4 shows some parameters of both the UHRIS and a standard Michelson interferometer for the same spectral resolution 0.005 cm−1.

Tables Icon

Table 4. Some parameters of both the UHRIS and a standard Michelson interferometer for the same spectral resolution 0.005 cm−1

3.2. The second example

The second example is that a source spectra covers a central wavenumber σ0=4000cm1 (wavelength 2.5 µm) and a spectral bandwidth Δσ=30cm1 (a wavenumber range from 3985 cm−1 to 4015 cm−1), and the desired spectral resolution is δσUHRIS=0.004cm1. Some key parameters of the UHRIS for the second example are shown in Table 5. In order to cover the full wavenumber range from 3985 cm−1 to 4015 cm−1, the FPI plate spacing is approximately assigned from 2.4999400 cm to 2.5000625 cm, namely, the FPI should scan N = 50 steps of approximate 25 nm in step interval. Let P=4, the spectral resolution of the SMI of the UHRIS for the second example is still δσSMI=0.05cm1, and therefore the main parameters of the SMI in the UHRIS for the second example are still shown in Table 3.

Tables Icon

Table 5. Some key parameters of the UHRIS for the second example

Suppose that a source spectrum contains only wavenumber 3999.796 cm−1, 3999.8 cm−1, 3999.804 cm−1, 3999.996 cm−1, 4000 cm−1 (wavelength 2.5 µm), 4000.004 cm−1, 4000.196 cm−1, 4000.2 cm−1 and 4000.204 cm−1. The FPI needs to scan three steps with approximate 25 nm step interval, and three interferograms produced by the SMI of the UHRIS for the second example are shown in Fig. 5. The first interferogram (d=2.4999975cm) contains only wavenumber 3999.804 cm−1, 4000.004 cm−1 and 4000.204 cm−1. The second interferogram (d=2.5cm) contains only wavenumber 3999.8 cm−1, 4000 cm−1 (2.5 µm) and 4000.2 cm−1. The third interferogram (d=2.5000025cm) contains only wavenumber 3999.796 cm−1, 3999.996 cm−1 and 4000.196 cm−1. Figure 6 shows the spectrum obtained from Fourier transform of the three interferograms in Fig. 5, and the spectral resolution is 0.004 cm−1.

 figure: Fig. 5

Fig. 5 Three interferograms produced by the SMI of the UHRIS with a spectral resolution 0.004 cm−1 (the second example).

Download Full Size | PDF

 figure: Fig. 6

Fig. 6 Spectrum obtained from Fourier transform of the three interferograms in Fig. 5.

Download Full Size | PDF

The resolving power of the UHRIS for the wavelength range from 0.8 µm to 2.5 µm for the second example is shown in Fig. 7.

 figure: Fig. 7

Fig. 7 Resolving power of the UHRIS for the wavelength range from 0.8 µm to 2.5 µm when d0=2.5cm and N=50 (the second example).

Download Full Size | PDF

Table 6 shows some parameters of both the UHRIS and a standard Michelson interferometer for the same spectral resolution 0.004 cm−1.

Tables Icon

Table 6. Some parameters of both the UHRIS and a standard Michelson interferometer for the same spectral resolution 0.004 cm−1

According to Eq. (10), by adjusting the values of d (the plate spacing of the FPI) and N (the number of scanning steps of the FPI), the UHRIS can be scaled to achieve resolving power higher than 1,000,000 in near-infrared (NIR) or mid-wave infrared (MWIR) region.

4. Conclusion

The principle of a compact UHRIS is described in detail, the results of preliminary numerical simulation with two examples are shown, and the comparisons of the UHRIS with two other types of ultra-high resolution spectrometers are given (see Tables 1, 4 and 6). Compared with Michelson-type interferometers used for ultrahigh resolution spectral measurements, the UHRIS has much smaller physical size, higher stability and shorter measurement time. Since the SMI of the UHRIS is static, does not require a drive system and acquires a spatial sampling of the interferogram, compared with the combination of a FPI and a Michelson-type interferometer, the UHRIS has much shorter measurement time, higher stability and compactness. Due to the use of the SMI, there is also a tradeoff for the UHRIS, i.e., the spectral range is narrow. The UHRIS is a unique concept to provide resolving power higher than 1,000,000 in near-infrared, short-wave infrared or mid-wave infrared region while achieving short measurement time and small physical size.

Funding

National Natural Science Foundation of China (NSFC) (61605151).

References

1. C. Palmer and E. Loewen, Diffraction Grating Handbook (Newport Corporation, 2005).

2. Q. Yang, “Compact high-resolution Littrow conical diffraction spectrometer,” Appl. Opt. 55(18), 4801–4807 (2016). [CrossRef]   [PubMed]  

3. W. M. Kunkel and J. R. Leger, “Gain dependent self-phasing in a two-core coherently combined fiber laser,” Opt. Express 26(8), 9373–9388 (2018). [CrossRef]   [PubMed]  

4. L. Yu, H. Xue, and J. X. Chen, “Dual concave grating anastigmatic spectrometer with high spectral resolution for remote sensing,” Appl. Opt. 57(33), 9789–9796 (2018). [CrossRef]   [PubMed]  

5. N. Danz, B. Höfer, E. Förster, T. Flügel-Paul, T. Harzendorf, P. Dannberg, R. Leitel, S. Kleinle, and R. Brunner, “Miniature integrated micro-spectrometer array for snap shot multispectral sensing,” Opt. Express 27(4), 5719–5728 (2019). [CrossRef]   [PubMed]  

6. W. H. Steel, Interferometry (Cambridge University, 1983).

7. P. R. Griffiths and J. A. de Haseth, Fourier Transform Infrared Spectrometry (Wiley-Interscience, 2007).

8. J. Kauppinen and V. M. Horneman, “Large aperture cube corner interferometer with a resolution of 0.001 cm(-1).,” Appl. Opt. 30(18), 2575–2578 (1991). [CrossRef]   [PubMed]  

9. G. Durry and G. Guelachvili, “High-information time-resolved step-scan Fourier interferometer,” Appl. Opt. 34(12), 1971–1981 (1995). [CrossRef]   [PubMed]  

10. Q. Yang, “Moving corner-cube mirror interferometer and reflection characteristic of corner-cube mirror,” Appl. Opt. 49(21), 4088–4095 (2010). [CrossRef]   [PubMed]  

11. Q. Yang, L. Liu, and P. Lv, “Principle of a two-output-difference interferometer for removing the most important interference distortions,” J. Mod. Opt. 65(19), 2234–2242 (2018). [CrossRef]  

12. G. Hernandez, Fabry-Perot Interferometers (Cambridge University, 1986).

13. P. B. Hays and H. E. Snell, “Multiplex Fabry-Perot interferometer,” Appl. Opt. 30(22), 3108–3113 (1991). [CrossRef]   [PubMed]  

14. G. Hernandez and K. C. Clark, “Electro-optic high-resolution Fabry-Perot spectrometer,” Appl. Opt. 33(10), 1989–1992 (1994). [CrossRef]   [PubMed]  

15. M. Pisani and M. Zucco, “Compact imaging spectrometer combining Fourier transform spectroscopy with a Fabry-Perot interferometer,” Opt. Express 17(10), 8319–8331 (2009). [CrossRef]   [PubMed]  

16. M. Zucco, M. Pisani, V. Caricato, and A. Egidi, “A hyperspectral imager based on a Fabry-Perot interferometer with dielectric mirrors,” Opt. Express 22(2), 1824–1834 (2014). [CrossRef]   [PubMed]  

17. T. A. Al-Saeed and D. A. Khalil, “Fourier transform spectrometer based on Fabry-Perot interferometer,” Appl. Opt. 55(20), 5322–5331 (2016). [CrossRef]   [PubMed]  

18. P. Metz, H. Block, C. Behnke, M. Krantz, M. Gerken, and J. Adam, “Tunable elastomer-based virtually imaged phased array,” Opt. Express 21(3), 3324–3335 (2013). [CrossRef]   [PubMed]  

19. P. Metz, J. Adam, M. Gerken, and B. Jalali, “Compact, transmissive two-dimensional spatial disperser design with application in simultaneous endoscopic imaging,” Appl. Opt. 53(3), 376–382 (2014). [CrossRef]   [PubMed]  

20. W. Bao, Z. Ding, P. Li, Z. Chen, Y. Shen, and C. Wang, “Orthogonal dispersive spectral-domain optical coherence tomography,” Opt. Express 22(8), 10081–10090 (2014). [CrossRef]   [PubMed]  

21. W. Bao, Y. Shen, T. Chen, P. Li, and Z. Ding, “High-speed high-precision and ultralong-range complex spectral domain dimensional metrology,” Opt. Express 23(9), 11013–11022 (2015). [CrossRef]   [PubMed]  

22. P. D. Atherton, K. Taylor, C. D. Pike, C. F. W. Harmer, N. M. Parker, and R. N. Hook, “TAURUS: a wide-field imaging Fabry-Perot spectrometer for astronomy,” Mon. Not. R. Astron. Soc. 201(3), 661–696 (1982). [CrossRef]  

23. G. R. Davis, I. Furniss, W. A. Towlson, P. A. R. Ade, R. J. Emery, W. M. Glencross, D. A. Naylor, T. J. Patrick, R. C. Sidey, and B. M. Swinyard, “Design and performance of cryogenic, scanning Fabry-Perot interferometers for the Long-Wavelength Spectrometer on the Infrared Space Observatory,” Appl. Opt. 34(1), 92–107 (1995). [CrossRef]   [PubMed]  

24. S. M. Lindsay, M. W. Anderson, and J. R. Sandercock, “Construction and alignment of a high performance multipass vernier tandem Fabry-Perot interferometer,” Rev. Sci. Instrum. 52(10), 1478–1486 (1981). [CrossRef]  

25. G. Hernandez, “Afocal coupled etalons. DEM: a high-resolution doubleetalon modulator spectrometer,” Appl. Opt. 26(22), 4857–4869 (1987). [CrossRef]   [PubMed]  

26. G. Hernandez and F. G. McCormac, “Afocal coupled etalons: experimental confirmation of a high-resolution double-etalon modulator spectrometer,” Appl. Opt. 27(16), 3492–3495 (1988). [CrossRef]   [PubMed]  

27. P. B. Hays, V. J. Abreu, M. E. Dobbs, D. A. Gell, H. J. Grassl, and W. R. Skinner, “The high-resolution doppler imager on the Upper Atmosphere Research Satellite,” J. Geophys. Res. 98(D6), 10713–10723 (1993). [CrossRef]  

28. M. R. Swain, C. M. Bradford, G. J. Stacey, A. D. Bolatto, J. M. Jackson, M. L. Savage, and J. Davidson, “Design of the South Pole imaging Fabry-Perot interferometer (SPIFI),” Proc. SPIE 3354, 480–492 (1998). [CrossRef]  

29. Y. M. Eltagoury, Y. M. Sabry, and D. A. Khalil, “Novel Fourier transform infrared spectrometer architecture based on cascaded Fabry-Perot interferometers,” Proc. SPIE 9760, 97600L (2016). [CrossRef]  

30. I. S. El-Sayed, Y. M. Sabry, W. E. ElZeiny, N. Badra, and D. A. Khalil, “Transformation algorithm and analysis of the Fourier transform spectrometer based on cascaded Fabry-Perot interferometers,” Appl. Opt. 57(25), 7225–7231 (2018). [CrossRef]   [PubMed]  

31. J. M. Helbert, P. Laforie, and P. Miche, “Conditions of pressure scanning of a Fabry-Perot interferometer over a wide spectrum range,” Appl. Opt. 16(8), 2119–2126 (1977). [CrossRef]   [PubMed]  

32. B. Swinyard and M. Ferlet, “Cascaded interferometric imaging spectrometer,” Appl. Opt. 46(25), 6381–6390 (2007). [CrossRef]   [PubMed]  

33. K. D. Möller, “Wave-front-dividing array interferometers without moving parts for real-time spectroscopy from the IR to the UV,” Appl. Opt. 34(9), 1493–1501 (1995). [CrossRef]   [PubMed]  

34. E. V. Ivanov, “Static Fourier transform spectroscopy with enhanced resolving power,” J. Opt. A, Pure Appl. Opt. 2(6), 519–528 (2000). [CrossRef]  

35. A. Lacan, F. M. Bréon, A. Rosak, F. Brachet, L. Roucayrol, P. Etcheto, C. Casteras, and Y. Salaün, “A static Fourier transform spectrometer for atmospheric sounding: concept and experimental implementation,” Opt. Express 18(8), 8311–8331 (2010). [CrossRef]   [PubMed]  

36. Y. Zheng, J. Liang, and Z. Liang, “Design and fabrication of step mirrors used in space-modulated Fourier transform infrared spectrometer,” Opt. Express 21(1), 884–892 (2013). [CrossRef]   [PubMed]  

37. M. Born and E. Wolf, Principles of Optics (Cambridge University, 1999).

38. H. Nyquist, “Certain factors affecting telegraph speed,” Bell Syst. Tech. J. 3(2), 324–346 (1924). [CrossRef]  

39. C. E. Shannon, “Communication in the presence of noise,” Proc. IEEE 37(1), 10–21 (1949).

40. P. Jacquinot, “The luminosity of spectrometers with Prisms, Grating, or Fabry-Perot Etalons,” J. Opt. Soc. Am. 44(10), 761–765 (1954). [CrossRef]  

41. P. Jacquinot, “How the search for a throughput advantage led to Fourier transform spectroscopy,” Infrared Phys. 24, 99–101 (1984). [CrossRef]  

42. P. B. Fellgett, “Three concepts make a million points,” Infrared Phys. 24(2–3), 95–98 (1984). [CrossRef]  

43. P. B. Fellgett, “The nature and origin of multiplex Fourier spectrometry,” Notes Rec. R. Soc. 60(1), 91–93 (2006). [CrossRef]  

44. A. Barducci, D. Guzzi, C. Lastri, P. Marcoionni, V. Nardino, and I. Pippi, “Theoretical aspects of Fourier Transform Spectrometry and common path triangular interferometers,” Opt. Express 18(11), 11622–11649 (2010). [CrossRef]   [PubMed]  

45. A. Barducci, D. Guzzi, C. Lastri, V. Nardino, P. Marcoionni, and I. Pippi, “Radiometric and SNR properties of multiplex dispersive spectrometry,” Appl. Opt. 49(28), 5366–5373 (2010). [CrossRef]   [PubMed]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (7)

Fig. 1
Fig. 1 Optics of the UHRIS combining a FPI and a static SMI: (a) Basic optical layout and (b) Equivalent perspective view.
Fig. 2
Fig. 2 Resolving power of the UHRIS for the wavelength range from 0.8 µm to 2 µm when d 0 =2 cm and N=50.
Fig. 3
Fig. 3 Three interferograms produced by the SMI of the UHRIS with a spectral resolution 0.005 cm−1.
Fig. 4
Fig. 4 Spectrum obtained from Fourier transform of the three interferograms in Fig. 3.
Fig. 5
Fig. 5 Three interferograms produced by the SMI of the UHRIS with a spectral resolution 0.004 cm−1 (the second example).
Fig. 6
Fig. 6 Spectrum obtained from Fourier transform of the three interferograms in Fig. 5.
Fig. 7
Fig. 7 Resolving power of the UHRIS for the wavelength range from 0.8 µm to 2.5 µm when d 0 =2.5 cm and N=50 (the second example).

Tables (6)

Tables Icon

Table 1 Comparisons of the UHRIS with two other types of spectrometers with resolving power higher than 1,000,000 in near-infrared, short-wave infrared or mid-wave infrared region

Tables Icon

Table 2 Some key parameters of the UHRIS for the first example

Tables Icon

Table 3 Main parameters of the SMI in the UHRIS for the first example

Tables Icon

Table 4 Some parameters of both the UHRIS and a standard Michelson interferometer for the same spectral resolution 0.005 cm−1

Tables Icon

Table 5 Some key parameters of the UHRIS for the second example

Tables Icon

Table 6 Some parameters of both the UHRIS and a standard Michelson interferometer for the same spectral resolution 0.004 cm−1

Equations (18)

Equations on this page are rendered with MathJax. Learn more.

2d=mλ.
FS R σ = 1 2d .
FS R λ = λ 2 2d .
T FPI ( σ,d )= ( 1R ) 2 1+ R 2 2Rcos( 4πσd ) .
F r = π R 1R .
I UHRIS ( σ,d,x )= 0 B( σ ) T FPI ( σ,d )[ 1+cos( 2πσx ) ]dσ = 0 ( 1R ) 2 B( σ )[ 1+cos( 2πσx ) ] 1+ R 2 2Rcos( 4πσd ) dσ .
δ σ SMI = 1 2P d 0 .
δ σ UHRIS = FS R σ( 0 ) N = 1 2N d 0 .
δ σ UHRIS = P N δ σ SMI .
R UHRIS = λ λ λ FS R λ /N =mN= 2Nd λ .
Δ d k = k d 0 mN+k = k d 0 2 d 0 σ 0 N+k .
δ σ SMI = 1 2 x max .
x max =P d 0 .
χ< 1 2Δσ .
χ=2× h 1 .
Δσ 1 4 h 1 .
K= s 1 × s 2 .
K= P d 0 2 h 1 .
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.