Optical properties of superconducting nanowire single-photon detectors

Open Access Open Access

Abstract

We measured the optical absorptance of superconducting nanowire single photon detectors. We found that 200-nm-pitch, 50%-fillfactor devices had an average absorptance of 21% for normally-incident front-illumination of 1.55-µm-wavelength light polarized parallel to the nanowires, and only 10% for perpendicularly-polarized light. We also measured devices with lower fill-factors and narrower wires that were five times more sensitive to parallel-polarized photons than perpendicular-polarized photons. We developed a numerical model that predicts the absorptance of our structures. We also used our measurements, coupled with measurements of device detection efficiencies, to determine the probability of photon detection after an absorption event. We found that, remarkably, absorbed parallel-polarized photons were more likely to result in detection events than perpendicular-polarized photons, and we present a hypothesis that qualitatively explains this result. Finally, we also determined the enhancement of device detection efficiency and absorptance due to the inclusion of an integrated optical cavity over a range of wavelengths (700-1700 nm) on a number of devices, and found good agreement with our numerical model.

©2008 Optical Society of America

1. Introduction

The short reset-time, low jitter, and broad wavelength response of superconducting nanowire single photon detectors (SNSPDs) makes them attractive candidates to replace other single photon detectors, including avalanche photodiodes [1], in applications such as free-space optical communications [2], quantum cryptography networks [3, 4], and quantum computation [5]. To fully take advantage of the boost in speed afforded by these properties, SNSPDs with high system detection efficiencies (SDE) are needed. But SNSPDs typically exhibit SDEs of only 0.2-10% for 1.55-µm-wavelength single-photons [6, 7, 8]. This limitation is currently believed to be due to poor coupling efficiency (η c); however, this may not be the entire story. To understand where the losses come from, we need to examine the factors that contribute to the SDE.

The SDE is a product of two lumped quantities, the coupling efficiency (η c) and the device detection efficiency (DDE). The quantity η c encapsulates all loss mechanisms encountered between the photon source and SNSPD, and is defined as the ratio of the number of photons that reach the active area to the number of photons emitted by the photon source. It is relatively straightforward to realize η c>0.9 through careful optical design that minimizes these losses. On the other hand, maximizing DDE is not as simple.

The DDE is the probability that a photon incident on the active area results in a voltage pulse. DDE depends on two quantities, the absorptance A, and the probability of resistive state formation due to an absorption event P R through

DDE=PRA.

While P R depends on the microscopic physics of the nanowires, A depends only on the optical properties of our fabricated structures and the incident field. A needs to be maximized to make efficient detectors; however, there are many ways an incident photon can remain unabsorbed. For example, the photon can pass through open gaps between the nanowires or be transmitted through the subwavelength thickness of NbN. In addition, a low value for P R will further suppress DDE (and therefore also SDE).

In this paper, we experimentally demonstrate that the optical polarization and changes in the geometry (pitch and fill-factor) of SNSPDs impact their absorptance and efficiency. For example, for devices with 200 nm pitch and 50% fill factor, we found that 21±1% of light incident from the front was absorbed for parallel polarization, while only 10±1% was absorbed for perpendicular polarization. The absorptance was reduced to 14±1% and 6±1.5% for parallel and perpendicular polarizations, respectively, for devices with the same pitch but a 25% fill-factor. We describe both a numerical model that predicts the absorptance of our structures, and experiments that directly measure absorptance. We found that the experimental data matches the predictions of our model to within the uncertainties in our knowledge of physical and optical parameters. We also measured the DDE for the same devices and compared it to the measured absorptance. We discovered, remarkably, that P R is different for photons polarized perpendicular (⊥) to the nanowires (P R,⊥) than for photons polarized parallel to the nanowires (P R,‖). We propose a model that qualitatively explains our observation of P R,‖P R,⊥.

Finally, we also made measurements to confirm that our numerical model and modeling parameters are accurate over a range of wavelengths. We measured the factor of enhancement of DDE due to the addition of an integrated optical cavity for single-photonwavelengths of 700-1700 nm, and found good agreement with our numerical model. As a result, we have shown that optical modelling can be used to design future SNSPDs in new regimes of operation.

This paper is organized as follows. In Section 2, we will describe a numerical model that predicts the absorptance of our structures. In Section 3, we will describe the setup and experiments conducted to measure absorptance on our devices. In Section 4, we will describe the experiments conducted to measure the wavelength dependence of the enhancement due to the addition of an optical cavity. We will not describe fabrication details or DDE measurement details as these were discussed in [9] and [10], respectively.

2. Optical model

We modeled the absorption process of photons by an SNSPD as a plane wave interacting with an infinite grating. These two approximations are justified because in our experiments, the photon was launched at normal incidence and had plane-wave phase-fronts when it reached the detectors, and because our beam spot was smaller than the test grating which will be described in Section 3.

While analytic methods, such as form-birefringence theory, have been used in the past to analyze grating structures such as wire-grid polarizers [11, 12], they break down for subwavelength grating thicknesses, in particular for electric-field polarized perpendicular to the wires in the grating (i.e. in the direction of grating periodicity)[13, 14]. Recently, a numerical technique (finite-element analysis) was used to approach this problem [15]; but because of either a difference in material parameters or simulation conditions, their findings were not consistent with our experimental or theoretical results.

 figure: Fig. 1.

Fig. 1. Infinite wire grid schematized in (a) is further reduced via symmetry to the unit cell shown in (b) in numerical modeling of the absorptance, A. (The schematic is not drawn to scale.) Plots of the calculated A as a function of fill-factor, f and pitch p are shown in (c) and (d) for parallel (‖) and perpendicular (⊥) electric field polarization. Inset in (d) shows how the calculated cross-sectional time-averaged electric field magnitude, |E⃗ cc′| varies with position x for ‖ and ⊥ polarizations across a f=12 , p=200nm structure. The NbN film thickness (t NbN) was 4 nm and incident electric field magnitude |E⃗ 0|=1. The NbN region extends from x=-50nm to x=+50nm.

Download Full Size | PDF

We used a different finite-element analysis software (Comsol Multiphysics v3.2b, EM module) than the one used in [15] to calculate the absorptance of the geometry shown in Fig. 1(a). We took advantage of symmetry by using the ‘In-plane Hybrid-Mode Waves’ implementation and defined the unit cell[16, 9, 10] shown in Fig. 1(b) with the desired NbN thickness (tNbN), pitch (p), wire width (w), fill-factor (f=w/p), and a 1.55-µm-wavelength. The thickness of NbNxOy was held constant at 2 nm because it was observed by transmission electron microscopy (TEM) to be the same thickness for two different NbN thicknesses [17]. The sapphire wafer thickness was not included because an anti-reflection coating was added to the sapphire-air interface in experiments. We applied periodic boundary conditions on the left (XY) and right (X′Y′) edges of the unit cell, scattering boundary conditions on the bottom (XX′) and top (YY′) edges. We applied either an out-of-plane (E ), or in-plane (E ) electric field on the top edge (front-illumination), which matches the experimental setup described in Sec. 3. The material parameters were defined in terms of complex refractive indices collected either from ellipsometry measurements [18] or the literature. We used n NbN=5.23-i5.82, nNbNxOy=2.28 , and n sapphire=1.75 [19]. We then applied a Lagrange (quadratic) mesh constrained to 0.25 nm in NbN and NbNxOy, and made to be denser at edge XY than X′Y′ to maintain accuracy when using periodic boundary conditions. Moving to a mesh with elements that were an order of magnitude smaller did not improve accuracy. We used a direct linear solver (UMFPACK) to solve for the spatial distribution of electric (E⃗) field, E⃗(x,y).

The absorptance A is proportional to the integral of the time-averaged electric field intensity in the NbN film. Because the electric field does not vary significantly over the thickness of the NbN, we can express A in terms of the cross-sectional electric field |E⃗ cc′(x)| in NbN (as pictured in Fig. 1(b)) through

A=p2p20tNbNQ(x,y)dxdyp2p2Iodx

where Q(x,y) is the time-average resistive dissipation in the nanowire, and I is the time-average incident Poynting power density given by

Io=12(εoμo)12Eo2

where |E⃗ | is the time-averaged incident electric field magnitude, ε is the permittivity of air, and µ is the permeability of air. For our wires which have typical t NbN of 4-6 nm, Q(x,y)≈Q(x) and is given by Ohm’s law to be

Q(x)=12ωIm[ε]Ecc(x)2.

Figure 1(c) shows the calculated dependence of A on f for constant values of p and electric field polarizations that were either parallel (‖) or perpendicular (⊥) to the periodicity of the nanowires, while Fig. 1(d) shows A as a function of p. Both figures show that A is invariant with pitch p when fill-factor f is kept constant, and that A can be maximized by narrowing the gaps between nanowires. We understand that A remains constant with varying p because the electric field intensity is continuous across the air-NbN boundaries as a result of tangential E⃗ cc′ continuity, as shown in Fig. 1(d) inset. Only the fraction of the electric field magnitude |E⃗ cc′| in NbN compared to air (i.e. f) changes A . On the other hand, decreasing p while keeping f constant decreases A . This effect is observed because there are more air-NbN interfaces per unit area for a grating with a smaller pitch than one with a larger pitch, and the boundary conditions dictate a lower |E⃗ cc′| in the NbN at each air-NbN interface.

The plots also illustrate an important limitation that A poses to the photon detection process: the maximum A (and therefore also DDE) can not exceed 30% with the configuration shown. There are two ways that A can be increased that have already been demonstrated: (1) by using back-illumination (i.e. through sapphire edge XX′) instead of front-illumination (i.e. through air edge YY′), and thereby reducing the index mismatch with NbN, A can be increased up to 45%; (2) by fabricating an optical cavity designed to intensify the field in the NbN nanowires [10]. An understanding of the impact of SNSPD geometry and optical polarization on A and P R will be helpful in finding other ways to improve A.

We will now describe how we determined the optical absorptance of the devices and how the results compared to the model. We will also compare our measurements of the absorptance to the DDE measurements of the same devices.

3. Absorptance measurements

Figure 2 shows a schematic of the optical apparatus we used to measure absorptance. Free-space optics were attached to a three-axis, computer-controlled stage (motorized actuator: Newport LTA-HS, motion controller: Newport ESP300). The stage could be affixed to a cryostat specially designed to allow free-space optical coupling to a cold sample, or be used without a cryostat for room-temperature measurements. We mounted a single-mode optical fiber (Corning SMF-28) to the stage that was illuminated by a 1.55-µm-wavelength diode laser (Thorlabs S1FC1550) and polarized using a state of polarization (SOP) locker (Thorlabs PL100S). The SOP locker was preferable to a manual polarizer because it provided mechanical isolation to our setup when we varied polarization and permitted automation of the data acquisition process. The polarized light was collimated and split equally into two arms. In one arm, a Ge-based sensor (Thorlabs S122A) connected to a power meter (Thorlabs PM100) was used to measure the incident light intensity, mainly for troubleshooting purposes. In the other arm, a long-working-distance microscope objective (Mitutoyo M Plan APO NIR 20X) was used to focus the light onto a device. Light reflected from the device was collected and measured using another power meter. The transmitted light was measured by a third power meter. We controlled the power meters, SOP locker, and motion controller using the Instrument Control Toolbox (ver. 2.4) for Matlab (ver. 7.2/R2006a) and custom software.

 figure: Fig. 2.

Fig. 2. Schematic of the optical setup used to measure the optical absorptance. A state of polarization (SOP) locker was used to set the input polarization of the incident 1.55-µm-wavelength light. Light was focused to a 4±0.5µm (FWHM) spot onto a device and the incident, reflected, and transmitted power were measured. The absorptance was calculated from these measurements and earlier calibrations discussed in the main text.

Download Full Size | PDF

It was necessary to fabricate devices which were specially designed to facilitate both absorptance and DDE measurements. A typical device is pictured in Figure 3 where the active area (3µm×3µm) of the SNSPD is centrally located within a large grating structure (30µm×10µm). The small active area facilitated DDE measurements because it enabled better uniformity in electrical response across the device. The larger grating structure, which has the same p and f as the smaller active area, facilitated measurements of A because it closely matched our simulation geometry. Thus, five groups of devices with identically sized large gratings and small active areas were fabricated with different (f, p[nm]) combinations of ( 12 , 200), ( 13 , 300), ( 14 ,400), ( 13 , 150), and ( 14 , 200) on the same chip.

 figure: Fig. 3.

Fig. 3. Scanning electron micrograph of a 50% fill-factor, 200-nm-pitch SNSPD is shown in (a). The central, photon-detecting region of the SNSPD structure outlined in (a) is magnified in (b). The extended parallel grating structure outside the active area was necessary for proximity-effect correction in the fabrication process, but was also useful in expanding the optically testable region of the device.

Download Full Size | PDF

The procedure for collecting absorptance data was as follows. We first used the imaging capability of our system to determine the working distance used in our experiments. To find the working distance, we scanned the laser across a sapphire-gold edge and varied working distances between scans. We measured the reflected and transmitted light intensity as a function of position for each working distance. We fitted the measured intensity profiles to theoretical profiles that were calculated by convolving a unit step function with a Gaussian function, and were thus able to estimate the full-width half-max (FWHM) of the beam profile at different working distances. The minimum FWHM for our setup was 4±0.5µm. By operating at a working distance that corresponded to the minimum FWHM, we ensured that 98% of the laser intensity was incident on the large grating structure.

We then measured the reflected and transmitted power from gold and sapphire in the vicinity of the device. These measurements were repeated over polarizations spaced by 10° on the Poincaré sphere. We calibrated the measured reflected and transmitted power to the theoretical reflectance and transmittance (R, T) of gold [20, 21, 22] (0.98, 0) and sapphire (0.07, 0.93) calculated from their refractive index values at λ=1.55µm. This calibration procedure corrects for absorption or reflection from optical elements (e.g. optical windows). We then measured the reflected and transmitted power with the laser centered in the active area of the SNSPD, and calculated R, T, and A=1-R-T. We then repeated our measurements on 10-13 identical devices in each of the five groups of devices.

Figure 4 shows our measurements of A and A at room temperature. Room temperature measurements did not differ significantly from measurements of A made at 6 K, where the reflectance and transmittance were calibrated using the room temperature refractive index values of sapphire and gold at λ=1.55µm. We attempted to fit the model by using w and t NbN as free parameters, and assuming t NbN was between 4 nm and 6 nm. Our fit required wire widths that were systematically 10-15% larger than the nominal w used in our electron-beam-lithography pattern. There are many uncertainties in our model parameters that can contribute to this. One cause is that we may have inadvertently exposed the resist adjacent to the intended structure causing w to be larger than desired. We conducted scanning electron microscopy of our nanowires and found that w predicted by the model were within the uncertainty caused by increased secondary electron emission from the edges of the nanowires. However, there are other factors that can contribute to a fitting error. One factor is that the NbN refractive index we used in our simulations may not be accurate because the refractive index was measured on a thicker film (12 nm) than the film we used to fabricated our devices on (4-6 nm). In addition, the films were from different growth batches so may have had slight differences in optical properties. Another factor is that our simple model of infinite gratings may not be sufficient, i.e. a 3D model may be needed. Considering these uncertainties, the qualitative and quantitative fit of our measurements with the model was good.

 figure: Fig. 4.

Fig. 4. Statistical plot of the measured parallel (‖) and perpendicular (⊥) absorptance for devices with different pitch p and fill-factor f. Each data symbol represents measurements of 10-13 devices for the polarization that yielded the maximum (‖) or minimum (⊥) absorptance. The arrows indicate the calculated absorptance for structures with fitted wire widths (from left to right) w=104 nm, 108 nm, 114 nm, 58 nm, and 55 nm, NbN thickness t NbN=4.35 nm, and nominal values of pitch p.

Download Full Size | PDF

3.1. Comparison of device detection efficiency to absorptance

A comparison of device detection efficiency to absorptance can answer one important question about our device: is every absorbed photon actually detected? To answer this question, we measured both DDE (using the measurement apparatus described in [10] but with front-illumination instead of back-illumination) and A for the same devices. The two quantities are plotted against each other in Fig. 5 where, for reference, we have also plotted lines with constant slope P R. It should be noted that the value for DDE is measured with a relative accuracy of ±25% due to uncertainties in fiber output power calibration. Some features of this plot were as expected; for example, all of the devices had higher absorptance than device detection efficiency (P R<1). But there were other features that were unexpected. We will now discuss these other features and see how they yield an unexpected new insight into the microscopic physics of SNSPDs.

 figure: Fig. 5.

Fig. 5. Plot of device detection efficiency as a function of absorptance for the same devices. Dotted lines have a constant slope given by P R, the probability of resistive state formation.

Download Full Size | PDF

There are two remarkable features in the data plotted in Fig. 5. First, P R had a peak value of 0.85+0.15 -0.21 for parallel polarization. The natural question one can ask is what happens to absorbed photons that do not lead to detection events? One possible explanation is that the devices were in some way not biased at their “true” critical currents, i.e. that they were constricted in some way so that the absorption of photons does not cause a voltage pulse. Another explanation is that there is some other absorbing medium that was not accounted for. We aim to address this issue in future experiments.

A second remarkable feature of the data in Fig. 5 is that P R for parallel polarization (0.55+0.14 -0.14<P R,‖<0.85+0.15 -0.21) is different than P R for perpendicular polarization (0.35+0.09 -0.09<P R,⊥<0.75+0.19 -0.19). While we do not yet completely understand the origin of the disparity between P R,‖ and P R,⊥, we have developed a model that can form a starting point for further analysis.

The main assumption of this model is a position-dependent (but polarization-independent) probability density of resistive state formation, ψ(x), where x is the distance across the nanowire. The second element of this model is the time-averaged resistive dissipation Q(x) which we know to be both position-dependent and polarization-dependent. These two elements can be combined to give the device detection efficiency and absorptance for parallel and perpendicular polarization:

DDE=p2p20tNbNψ(x)Q(x)Iodxdy.

We can use this equation, Eq. (1), and a calculated Q(x) to find a ψ(x) that gives a different P R for parallel and perpendicular polarizations. Using our data, we found that ψ(x) needs to be larger at the edges of the nanowire than in the center to explain P R,‖>P R,⊥, however the five geometries we investigated did not provide enough resolution to fit ψ(x) to a specific shape at this point.

4. Verification of the model through measurement of the cavity enhancement factor

In the previous section, we verified that our model predicted the absorptance for our structures to within experimental uncertainties. But this verification was done only at a single wavelength of 1550 nm, mainly because all of our optical components, i.e., the laser, fiber, lenses, SOP locker, microscope objective, beam splitter, collimator, and photodiodes were optimized for that wavelength. In order to verify our modeling parameters for a range of wavelengths, we measured the cavity enhancement factor 𝓔 for SNSPDs over a range of wavelengths. The cavity enhancement factor 𝓔 is the ratio of the absorptance of an SNSPD with an integrated cavity to the absorptance without a cavity and depends on the refractive indices of the materials and the geometry. 𝓔 is also the factor by which the intensity of the electric field in NbN, and therefore Q(x), is increased due to the presence of a cavity. In view of Eq. (5), and because ψ(x) should not depend on the intensity, we can determine 𝓔 by measuring DDE for an SNSPD with a cavity and the DDE of an SNSPD without a cavity and taking the ratio of these two quantities for each wavelength.

We generated radiation spanning the range from 600-1700 nm using a supercontinuum source (Toptica photonics). This source consisted of a power-amplified, modelocked fiber laser with 100 fs pulses at 1550 nm, which could either be coupled into a highly-nonlinear step-index fiber (generating a frequency comb from ~600-1050 nm) or first doubled using a periodically-poled Lithium-niobate crystal and then coupled into a photonic crystal fiber (Blazed photonics) (generating a frequency comb from ~1150-1700 nm). We used a Pellin-Broca prism to select a wavelength band out of one or the other of these outputs (with a FWHM Δλ/λ~50) which we then coupled into an optical fiber (SMF-28, single mode from ~1200-1600 nm). The wavelength of the light output from the fiber could then be tuned by rotating the prism.

We used this tunable output to measure the enhancement 𝓔. The most obvious way to do this would simply be to make calibrated DDE measurements at all wavelengths before and after adding the cavity. However, we chose not to attempt this, since we had no way to accurately calibrate optical power levels at or near the single-photon level over such a wide wavelength range. To avoid the need for this calibration, we instead measured only detection efficiency ratios for pairs of detectors side-by-side. To do this, we fabricated a chip with 225 pairs of 3×3.3 µm detectors, where the two in each pair could be read out separately, and were spaced by only 100 µm. We measured critical currents and room-temperature resistances of all 450 detectors, and identified ~50 pairs for which these values were within 5% of each other. We then measured (calibrated) DDEs at 1550 nm for this subset of detectors, and further narrowed the experimental sample to the pairs having DDEs within 5% of each other. Then, we added optical cavities to one detector of each pair, and re-measured at 1550 nm. Although we commonly observe that DDE for a given device can vary between cooldowns (particularly if some additional processing has occurred), these variations are nearly always strongly correlated with a variation of the critical current (a reduced DDE correlates with a reduced critical current). So, we selected from our pairs of detectors only those where the critical currents remained the same to within 5% after the addition of cavities, and where the detector which did not have a cavity added had the same DDE as previously to within 5%. Lastly, we further reduced our experimental sample by selecting only the subset of devices which were relatively unconstricted [23]. In the present context, this corresponded to those devices with DDE at 1550 nm>17% for front illumination. Our final sample consisted of five detector pairs. For these ten detectors, we measured count rates at each of a sequence of wavelengths spanning the entire accessible range. By measuring both detectors in a pair in succession without adjusting the source in any way, we ensured that the optical power incident on the two detectors was nearly identical, and therefore that their count rates could be compared directly to give 𝓔. This precaution was important since the optical power obtainable in a given wavelength band was not repeatable, nor was the coupling into the SMF-28 optical fiber, which was not single-mode over a fraction of our wavelength range. Given the latter issue, we restricted ourselves to the wavelength range 700-1700 nm where the fiber output could be fairly well polarized (extinction ratio>20).

Figure 6 shows the enhancement measured in this manner, and a calculation of 𝓔 as a function of wavelength. We can see that the data has good experimental agreement with the calculation. Inset Fig. 6(a) shows the geometry that was used in the calculation while Fig. 6(b) shows a plot of the real and imaginary parts of refractive index for NbN, NbNxOy, and HSQ. Measurements of the refractive indices of NbN and NbNxOy were made at room temperature by J. A. Woolam, Inc. using spectroscopic ellipsometry on a 12-nm-thick NbN film deposited on a sapphire wafer. Measurements of refractive index of HSQ were also performed by J. A. Woolam, Inc. using spectroscopic ellipsometry. n and k for gold were taken from Ref. [21].

 figure: Fig. 6.

Fig. 6. Plot of enhancement factor, 𝓔 as a function of wavelength λ. The dotted line shows the calculation that was carried out on the unit cell shown in inset (a) using values for the refractive indices shown in inset (b).

Download Full Size | PDF

5. Conclusion

In this paper, we reported measurements of the absorptance of SNSPDs and showed how A changed with the fill-factor and pitch of SNSPDs and optical polarization. We confirmed that our numerical model made accurate predictions for a range of geometry, polarization, and wavelength. We found that 200-nm-pitch, 50% fill-factor devices had an average absorptance of 21±1% for light polarized parallel to the nanowires, and only 10±1% for perpendicularly-polarized light. This disparity in polarization-sensitive absorptivity was even more evident in lower fill-factor and narrowwire-width devices, where we measured that parallel-polarized photons were more than 5 times as likely to be absorbed over perpendicularly polarized photons. We also found that potentially, some absorbed photons do not result in detection events, and that this quantity is smaller for photons with an orthogonal polarization. These results present new challenges that need to be understood and overcome if higher efficiency devices will be possible.

Acknowledgments

The authors would like to thank Prof. Hank Smith for the use of his facilities and equipment, James Daley and Mark Mondol for technical assistance, and Ayman Abouraddy, Scott Hamilton, Rich Molnar, Prof. Steven Johnson, and Prof. Terry Orlando for helpful discussions. The authors would also like to thank Jeffrey Stern for useful discussions about using effective index theory for optical modeling and for pointing out the formula for n eff for parallel electric-field polarization. This work made use of MIT’s shared scanning-electron-beam-lithography facility in the Research Laboratory of Electronics (SEBL at RLE). This work was sponsored in part by the United States Air Force under Air Force Contract #FA8721-05-C-0002 and IARPA. Opinions, interpretations, recommendations and conclusions are those of the authors and are not necessarily endorsed by the United States Government.

References and links

1. K.-S. Hyun, Y. Paek, Y.-H. Kwon, I. Yun, and E.-H. Lee, “High-speed and highly reliable InP/InGaAs avalanche photodiode for optical communications,” in Quantum Sensing: Evolution and Revolution from Past to Future, M. Razeghi and G. J. Brown, eds., vol. 4999, pp. 130–137 (SPIE, San Jose, CA, USA, 2003).

2. E. A. Dauler, B. S. Robinson, A. J. Kerman, V. Anant, R. J. Barron, K. K. Berggren, D. O. Caplan, J. J. Carney, S. A. Hamilton, K. M. Rosfjord, M. L. Stevens, and J. K. Yang, “1.25 Gbit/s photon-counting optical communications using a two-element superconducting nanowire single photon detector,” in Proc. SPIE, W. Becker, ed., vol. 6372 of Advanced Photon Counting Techniques, p. 637212 (Oct 2006).

3. A. Beveratos, R. Brouri, T. Gacoin, A. Villing, J.-P. Poizat, and P. Grangier, “Single Photon Quantum Cryptography,” Phys. Rev. Lett. 89, 187,901 (2002). [CrossRef]  

4. H. Takesue, S. W. Nam, Q. Zhang, R. H. Hadfield, T. Honjo, K. Tamaki, and Y. Yamamoto, “Quantum key distribution over a 40-dB channel loss using superconducting single-photon detectors,” Nature Photonics 1, 343–348 (2007). [CrossRef]  

5. E. Knill, R. Laflamme, and G. J. Milburn, “A scheme for efficient quantum computation with linear optics,” Nature 409, 46–52 (2001). [CrossRef]   [PubMed]  

6. A. Verevkin, J. Zhang, R. Sobolewski, A. Lipatov, O. Okunev, G. Chulkova, A. Korneev, K. Smirnov, G. N. Gol’tsman, and A. Semenov, “Detection efficiency of large-active-area NbN single-photon superconducting detectors in the ultraviolet to near-infrared range,” Appl. Phys. Lett. 80, 4687–4689 (2002). [CrossRef]  

7. A. Korneev, P. Kouminov, V. Matvienko, G. Chulkova, K. Smirnov, B. Voronov, G. N. Gol’tsman, M. Currie, W. Lo, K. Wilsher, J. Zhang, W. Slysz, A. Pearlman, A. Verevkin, and R. Sobolewski, “Sensitivity and gigahertz counting performance of NbN superconducting single-photon detectors,” Appl. Phys. Lett. 84, 5338–5340 (2004). [CrossRef]  

8. R. H. Hadfield, M. J. Stevens, S. S. Gruber, A. J. Miller, R. E. Schwall, R. P. Mirin, and S. W. Nam, “Single photon source characterization with a superconducting single photon detector,” Opt. Express 13, 10,846–10,853 (2005). [CrossRef]  

9. J. K. W. Yang, E. Dauler, A. Ferri, A. Pearlman, A. Verevkin, G. Gol’tsman, B. Voronov, R. Sobolewski, W. E. Keicher, and K. K. Berggren, “Fabrication development for nanowire GHz-counting-rate single-photon detectors,” IEEE Trans. Appl. Supercond. 15, 626–630 (2005). [CrossRef]  

10. K. M. Rosfjord, J. K. W. Yang, E. A. Dauler, A. J. Kerman, V. Anant, B. M. Voronov, G. N. Gol’tsman, and K. K. Berggren, “Nanowire Single-photon detector with an integrated optical cavity and anti-reflection coating,” Opt. Express 14, 527–534 (2006). [CrossRef]   [PubMed]  

11. X. J. Yu and H. S. Kwok, “Optical wire-grid polarizers at oblique angles of incidence,” J. Appl. Phys. 93, 4407–4412 (2003). [CrossRef]  

12. P. Yeh, “New Optical-Model for Wire Grid Polarizers,” Opt. Commun. 26, 289–292 (1978). [CrossRef]  

13. For an electric field polarized parallel to the wires in a sub-wavelength grating, an accurate result for the A can be obtained with the Fresnel equations where an effective index neff=((1-f)n2air+fn2NbN)1/2 is used for the thin film consisting of NbN subwavelength gratings (nair=1, nNbN=5.23-i5.82, and f is the fill-factor). A simple effective index model that only depends on f will not work for perpendicular polarization since A depends on both f and p.

14. P. Lalanne and D. Lemercier-Lalanne, “Depth dependence of the effective properties of subwavelength gratings,” J. Opt. Soc. Am. A 14, 450–458 (1997). [CrossRef]  

15. Z. Yan, A. H. Majedi, and S. Safavi-Naeini, “Physical Modeling of Hot-Electron Superconducting Single-Photon Detectors,” IEEE Trans. Appl. Supercond. 17, 3789–3794 (2007). [CrossRef]  

16. While our fabrication process (see Refs. 9 and 10) leaves behind 10–40 nm of residual resist on top of the nanowires, we found that including the resist in our geometry did not affect our results.

17. Energy dispersive X-ray (EDX) analysis was performed on two cross-section samples that confirmed the presence of oxide atop the NbN surface. There was also a contrast difference in the oxide layer from the NbN that was clearly visible in the TEM image. TEM imaging services were provided by Materials Analytical Services, Inc.

18. Measurements of the refractive indices of NbN and NbNxOy made at room temperature by J. A. Woolam, Inc. using spectroscopic ellipsometry on a 12-nm-thick NbN film deposited on a sapphire wafer.

19. I. H. Malitson, “Refraction and dispersion of synthetic sapphire,” J. Opt. Soc. Am. 52, 1377–1379 (1962). [CrossRef]  

20. The patterned gold film was approximately 100 nm thick. We used bulk values for the complex refractive index of gold found in Ref. 21. For a thin film thicker than 30 nm, the bulk refractive index can be used in this wavelength range (see Ref. 22).

21. E. D. Palik, Handbook of optical constants of solids, Academic Press handbook series (Academic Press, Orlando, 1985).

22. C. Reale, “Optical constants of vacuum deposited thin metal films in the near infrared,” Infrared Phys. 10, 173–181 (1970). [CrossRef]  

23. A. J. Kerman, E. A. Dauler, J. K. W. Yang, K. M. Rosfjord, V. Anant, K. K. Berggren, G. N. Gol’tsman, and B. M. Voronov, “Constriction-limited detection efficiency of superconducting nanowire single-photon detectors,” Appl. Phys. Lett. 90, 101,110 (2007). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (6)

Fig. 1.
Fig. 1. Infinite wire grid schematized in (a) is further reduced via symmetry to the unit cell shown in (b) in numerical modeling of the absorptance, A. (The schematic is not drawn to scale.) Plots of the calculated A as a function of fill-factor, f and pitch p are shown in (c) and (d) for parallel (‖) and perpendicular (⊥) electric field polarization. Inset in (d) shows how the calculated cross-sectional time-averaged electric field magnitude, |E⃗ cc′| varies with position x for ‖ and ⊥ polarizations across a f = 1 2 , p=200nm structure. The NbN film thickness (t NbN) was 4 nm and incident electric field magnitude |E⃗ 0|=1. The NbN region extends from x=-50nm to x=+50nm.
Fig. 2.
Fig. 2. Schematic of the optical setup used to measure the optical absorptance. A state of polarization (SOP) locker was used to set the input polarization of the incident 1.55-µm-wavelength light. Light was focused to a 4±0.5µm (FWHM) spot onto a device and the incident, reflected, and transmitted power were measured. The absorptance was calculated from these measurements and earlier calibrations discussed in the main text.
Fig. 3.
Fig. 3. Scanning electron micrograph of a 50% fill-factor, 200-nm-pitch SNSPD is shown in (a). The central, photon-detecting region of the SNSPD structure outlined in (a) is magnified in (b). The extended parallel grating structure outside the active area was necessary for proximity-effect correction in the fabrication process, but was also useful in expanding the optically testable region of the device.
Fig. 4.
Fig. 4. Statistical plot of the measured parallel (‖) and perpendicular (⊥) absorptance for devices with different pitch p and fill-factor f. Each data symbol represents measurements of 10-13 devices for the polarization that yielded the maximum (‖) or minimum (⊥) absorptance. The arrows indicate the calculated absorptance for structures with fitted wire widths (from left to right) w=104 nm, 108 nm, 114 nm, 58 nm, and 55 nm, NbN thickness t NbN=4.35 nm, and nominal values of pitch p.
Fig. 5.
Fig. 5. Plot of device detection efficiency as a function of absorptance for the same devices. Dotted lines have a constant slope given by P R, the probability of resistive state formation.
Fig. 6.
Fig. 6. Plot of enhancement factor, 𝓔 as a function of wavelength λ. The dotted line shows the calculation that was carried out on the unit cell shown in inset (a) using values for the refractive indices shown in inset (b).

Equations (5)

Equations on this page are rendered with MathJax. Learn more.

DDE = P R A .
A = p 2 p 2 0 t NbN Q ( x , y ) d x d y p 2 p 2 I o d x
I o = 1 2 ( ε o μ o ) 1 2 E o 2
Q ( x ) = 1 2 ω Im [ ε ] E cc ( x ) 2 .
DDE = p 2 p 2 0 t NbN ψ ( x ) Q ( x ) I o d x d y .