Fast localization method of an anomaly in tissue based on differential optical density

Open Access Open Access

Abstract

The position of the source-detector (S-D) relative to an anomaly has an important influence on the detection effect in non-invasive near-infrared spectroscopy-based methods. In this study, a single-source multi-detector structure was designed in order to realize the rapid localization of anomalies within tissue. This method uses finite element analysis of the optical density distribution for different horizontal positions, depths and diameters of anomalies. The difference in optical density between the detectors was then calculated. The simulation results show that the horizontal position of the anomaly in the tissue can be quickly located according to the differential optical density difference curves formed by the multiple detectors. The Gaussian fitting feature of these curves shows strong correlation with the horizontal positions, depths and diameters of the anomaly. Through the differential optical density difference curves, rapid localization within the region of interest can be achieved. This method provides an important reference for sources and detectors location for tumor detection, brain function optical imaging and other fields using near infrared spectroscopy, and improves its detection accuracy.

© 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Tissue component detection methods based on near-infrared spectroscopy (NIRS) have drawn wide attention in recent years due to their rapidity and non-invasiveness. These methods have been applied in many fields such as tumor detection [1], brain imaging [2,3] and brain hematoma detection [4–7], and are gradually being applied to clinical detection. Based on multiple-source and multiple-detector patterns, using the tissue optical parameter reconstruction (Diffuse Optical Tomography, DOT) the three-dimensional distribution of the optical parameters of the tissue can be obtained in order to achieve the identification of tumors [8], hemoglobin distribution and other imaging tasks. However, these methods are computationally intensive and achieve only slow imaging speed, rendering them unsuitable for the rapid detection of anomalies. Britton Chance [9] proposed the differential Optical Density (△OD) method, which is based on the difference between the optical density of normal tissue and abnormal tissue. It has been fully used in brain tumors [10], tissue hemoglobin detection [11,12] and rapid cerebral hematoma detection [13,14] with good results. For example, the Infrascanner has been great clinical application in the hematoma detection . It's a simple, portable, accurate, and cost effective hematoma detection device for using in a variety of settings, including hospital emergency rooms and intensive care units. The exact placement of the source-detectors’ is the key to detecting whether an anomaly exists in the target effectively [15]. The relationship between source-detectors distribution and detection results has always been a research hotspot in this field. An accurate source-detector distribution can improve tissue detection accuracy of NIRS effectively [16–18].

In this study, we used a Single-Source Multi-Detector structure(SS-MD) and utilized Nirfast’s Optical Finite Element tool to construct an anomaly model with different horizontal positions, different depths and different diameters in uniform tissue [19]. We calculated the differential optical density on multiple detectors and plotted the differential optical density difference (ΔOD') curve between adjacent detectors. Gaussian curve fitting [20] was used to fit the differential optical density difference curve, and to obtain its characteristic parameters. The experimental results show that using the differential optical density difference curve, the characteristic parameters can be used to locate anomalies in uniform tissue quickly. This method provides an additional reference for the best placement of Source-Detectors positions to detect anomalies in tissues, and the accurate positions presented will improve the near-infrared spectroscopy detection effect.

2. The differential optical density difference

Figure 1 shows the differential optical density difference curve acquisition schematic. S and S' are incident sources, D1, D2,…, Dn and D1’, D2’,…, Dn’ are equally spaced photoelectric detectors, respectively. First,we used a fixed SS-MD distribution to detect the normal tissue area. Then we used n detectors to obtain intensity measurements recorded as I1, I2,…, In. Second, we used the same SS-MD distribution to detect the abnormal tissue area with the same optical properties background. Then we used n detectors to obtain intensity measurements recorded as I1’, I2’,…, In’. We then obtained each position’s optical density OD1, OD2,…, ODn and OD1’, OD2’,…, ODn’ according to the Eq. (1) and (2).

 figure: Fig. 1

Fig. 1 Differential optical density difference curve acquisition schematic.

Download Full Size | PDF

ODi=lg(I0Ii),(i=1,2...n)
ODi'=lg(I0Ii'),(i=1,2...n)

By dividing the optical density at symmetrical positions, we calculate the differential optical density.

ΔODi=ODiODi'=lgI0IilgI0Ii'=lgIiIi',(i=1,2...n)

Then, we subtract the differential optical density obtained by adjacent position detectors, to find the difference of the differential optical densities at each point.

ΔOD'=ΔODi+lΔODi,(i=1,2...,n-l)

The multi-detector obtained ΔOD' values in turn are connected to obtain the differential optical density difference curve. This curve reflects the change rate of the optical density difference values at different S-D positions and its characteristics reflect the anomaly distribution within normal tissue.

3. Model foundation and results

In order to study the relationship between the variation trend of the differential optical density difference curve and the anomaly distribution in uniform tissue, this study used a two-dimensional uniform tissue model of 15cm*15cm based on the Nirfast optical finite element tool, in which μa = 0.01mm−1, and μs’ = 1 mm−1 [21]. The model featured three kinds of Single-Source Multi-Detectors (SS-MD) distribution and was used to compare the effects of different SS-MD positions on anomaly localization. As shown in Fig. 2, for SS-MD distribution A, the source was placed at the origin, while the 10 detectors were placed equidistantly from the origin from 2cm to 10cm. SS-MD distribution B was based on distribution A, moving horizontally 2 cm to the right. Distribution C was also based on the distribution A, this time moving horizontally 4 cm to the right.

 figure: Fig. 2

Fig. 2 Source (o) and detectors’ (x) location. The coordinate system established at model vertices was used to represent the position of the source and the detectors.

Download Full Size | PDF

3.1 Relationship between the anomaly’s horizontal position and ΔOD' simulation and results

While maintaining the same depth, the anomaly’s horizontal position was changed. A single anomaly was used for this experiment. The specific experimental parameters were the depth l = 10 mm, the diameter d = 30 mm, the absorption coefficient μa = 0.5 mm−1, and the scattering coefficient μs’ = 1 mm−1. Then, we set the abscissa value of the anomaly to 4.0 cm, 4.5 cm, 5.0 cm, 5.5 cm, 6.0 cm, 6.5 cm, 7.0 cm, 7.5 cm, and 8.0 cm, while keeping the other parameters unchanged. The ΔOD' values measured by each detector were calculated. We selected the horizontal position of 5.0 cm, 6.0 cm, 7.0 cm as an example. Figure 3(a), (b), (c) show the relationship between the SS-MD distribution’s source and detectors’ distance and ΔOD' values.

 figure: Fig. 3

Fig. 3 ΔOD' curve of anomaly at different horizontal positions.

Download Full Size | PDF

Figure 3 shows that at the different detectors’ positions different ΔOD' values were measured, with an obvious trend. This indicates that there is a specific relationship between the sensitivity of the ΔOD' value in the target detection zone and the source and detectors’ distance. The highest ΔOD' point in the curve indicates that the differential optical density has its largest change rate at this location. The corresponding source and the detectors’ distance reflects the anomaly’s lateral position.

3.2 Relationship between the anomaly depth and ΔOD' simulation and results

In this case, the anomaly’s depth was changed while maintaining the same horizontal position. The simulation parameters of the anomaly were d = 30mm, μa = 0.5 mm−1, μs’ = 1 mm−1 in this experiment. The anomaly was placed 6cm from the origin in the horizontal direction. For optical brain imaging applications, the average adult scalp skull thickness is usually 1.0-1.8cm, so the anomaly’s depth ranged from 1.0 to 1.8 cm at 0.1cm intervals. We then calculated the ΔOD' values measured by each detector. We selected the depths of 1.0cm, 1.2cm, 1.4cm, as examples. Figure 4(a), (b), (c) shows the relationship between the SS-MD distribution’s distance and ΔOD' values; in this case, the change trend is clear as well.

 figure: Fig. 4

Fig. 4 The ΔOD' curve of anomaly at different depths.

Download Full Size | PDF

This indicates that the ΔOD' value in the target detection zone was sensitive to the distance between the source and the detectors. Firstly, the distance of the anomaly to the origin in the horizontal direction was 6 cm. The maximum ΔOD' curve of the figure was also close to 6cm, thus confirming experiment 3.1’s conclusion. Meanwhile, Fig. 4 shows that the ΔOD' values were different when the depths of the anomaly were different. As the anomaly depth increases, the corresponding ΔOD' values decrease.

3.3 Relationship between the anomaly size and ΔOD' simulation and results

For this experiment, the size of the anomaly was changed while maintaining the same horizontal position and depth. The simulation parameters of the anomaly were l = 30mm, μa = 0.5 mm−1, μs’ = 1 mm−1 in this experiment. The anomaly was placed 7.5cm from the origin in the horizontal direction. We changed the anomaly’s diameters to 1.0cm, 2.0cm, 3.0cm, 4.0cm and calculated the ΔOD' values measured by each detector. Figure 5 shows the relationship between the SS-MD distribution’s distance and ΔOD' values.

 figure: Fig. 5

Fig. 5 ΔOD' curves of anomaly at different sizes. (a) Result of ΔOD' at r = 0.5 cm at different SS-MD distribution positions. (b), (c), and (d) results of ΔOD' with different sizes in fixed SS-MD positions.

Download Full Size | PDF

Figure 5(a) shows the intensity measured by different SS-MD distributions when the anomaly was in a fixed position, confirming experiment 3.1’s conclusions. The curves show the relationship between the different anomaly sizes and their corresponding ΔOD' values under the same SS-MD distribution in Fig. 5(b), (c) and (d). As can be seen from the figure, the ΔOD' value trend of each anomaly is highly similar, and as the anomaly diameters increases, their detected ΔOD' also increases.

4. Analysis and discussion

The spectral curves obtained in the experiment are composed of discrete points. Let these discrete data points be (xi,yi). The goal of curve fitting is to find a function y = f (x) between the variables x and y. The Gaussian fitting spectral curve is based on the assumption that the original spectrum is formed by stacking a number of unimodal bands. The Gaussian function system is used as the basic function of the spectral curve, that is, y = f (x) is set as the Gaussian function system, where each Gaussian function is determined by three parameters: the peak height A, the peak position B. and the peak width C. The entire Gaussian function system can be written as:

f(x)=i=1nAiexp[(xBiCi)2]

In the actual fitting process, we do not require y = f (x) to pass through all the spectral points (xi, yi), and we only require that the fitting error (δ = f (x) - yi) on the point xi is the smallest according to a certain criterion. This study uses the least squares method to find the best curve fit. The peak information such as peak height, peak width and peak position form a new information matrix, which allows the extraction and simplification of the original spectral information, thus facilitating the result analysis.

ΔOD' data containing the anomaly position and size information were obtained during the simulation experiments 3.1, 3.2, and 3.3. In order to observe the variation of these discrete points more clearly, Fig. 6 used the Gaussian fitting method to extract the curve characteristic peak.

 figure: Fig. 6

Fig. 6 Eigenvalue fitting results. (a) Curve fit of the characteristic peak about the anomaly’s ΔOD' with different sizes. (b) Curve fit of the characteristic peak about the anomaly’s ΔOD' at different depths. (c) Curve fit of the characteristic peak about the anomaly’s ΔOD' at different positions.

Download Full Size | PDF

As can be seen from Fig. 6(a), the characteristic peak of ΔOD' curve fit had a good correlation, 98.67% when measured with the anomaly at different horizontal positions. With the anomaly moving from left to right, the curve fitting peak was more accurate and closer to the true value when the anomaly was located in the middle of the SS-MD distribution. When the anomaly was located on either side of the SS-MD distribution, the curve peak changed within the allowable error range. As can be seen from Fig. 6(b), the characteristic peak of the ΔOD' fitting curves had a good correlation, 94.35% when measured with the anomaly at different depths. With the increase of the anomaly’s depth, the peak of the ΔOD' curve fit tends to increase, and the maximum ΔOD' value can still be used to accurately represent the anomaly’s presence. As can be seen from Fig. 6(c), the characteristic peak size of the ΔOD' fitted curves remains at 7.5 ± 0.2 cm, measured using anomalies of different sizes. The peak correlation was around 69% within the allowable error range. With the increase of anomaly diameters, the peak of the ΔOD' curve fit tended to increase.

The data correlation of the Gaussian curve fitting shows that the curve characteristic peaks can be used as a basis for determining whether the anomaly is present or not. This method can extract the main features of the data curves quickly and greatly simplify the original data.

5. Conclusion

This study demonstrates a method for the rapid localization of anomalies in tissues. Different SS-MD distributions were used to detect tissue-derived anomalies in unknown locations, analyze the relation between source-detectors’ distance and ΔOD' values, and analyze ΔOD' trends in order to determine the anomaly’s position. The results show that ΔOD' represents the sensitivity of the system to anomalies. The higher the value of ΔOD', the higher the system’s sensitivity. Therefore, the proposed method can provide a reference for placing source-detector-based localization, such as tumor detection and brain function imaging. At the same time, the method can be used as an effective auxiliary means to quickly assess whether the sampled tissue contains anomalies, and also to further guide clinical screening tools.

Funding

National Natural Science Foundation of China (NSFC) (61705164).

Disclosures

The authors declare that there are no conflicts of interest related to this article.

References and links

1. Z. Tao, X. Dang, X. Huang, M. D. Muzumdar, E. S. Xu, N. M. Bardhan, H. Song, R. Qi, Y. Yu, T. Li, W. Wei, J. Wyckoff, M. J. Birrer, A. M. Belcher, and P. P. Ghoroghchian, “Early tumor detection afforded by in vivo imaging of near-infrared II fluorescence,” Biomaterials 134, 202–215 (2017). [CrossRef]   [PubMed]  

2. R. Nosrati, K. Vesely, T. A. Schweizer, and V. Toronov, “Event-related changes of the prefrontal cortex oxygen delivery and metabolism during driving measured by hyperspectral fNIRS,” Biomed. Opt. Express 7(4), 1323–1335 (2016). [CrossRef]   [PubMed]  

3. A. Kassab, J. Le Lan, P. Vannasing, and M. Sawan, “Functional near-infrared spectroscopy caps for brain activity monitoring: a review,” Appl. Opt. 54(3), 576–586 (2015). [CrossRef]  

4. P. Schober, S. M. Bossers, and L. A. Schwarte, “Intracranial Hematoma Detection by Near Infrared Spectroscopy in a Helicopter Emergency Medical Service: Practical Experience,” BioMed. Res. Int. 2017, 1–6 (2017). [CrossRef]  

5. P. A. Narayana, X. Yu, K. M. Hasan, E. A. Wilde, H. S. Levin, J. V. Hunter, E. R. Miller, V. K. S. Patel, C. S. Robertson, and J. J. McCarthy, “Multi-modal MRI of mild traumatic brain injury,” Neuroimage Clin. 7, 87–97 (2015). [CrossRef]   [PubMed]  

6. D. J. Davies, Z. Su, M. T. Clancy, S. J. Lucas, H. Dehghani, A. Logan, and A. Belli, “Near-Infrared Spectroscopy in the monitoring of adult traumatic brain injury: a review,” J. Neurotrauma 32(13), 933–941 (2015). [CrossRef]   [PubMed]  

7. R. Salonia, M. J. Bell, P. M. Kochanek, and R. P. Berger, “The utility of near infrared spectroscopy in detecting intracranial hemorrhage in children,” J. Neurotrauma 29(6), 1047–1053 (2012). [CrossRef]   [PubMed]  

8. M. Jermyn, H. Ghadyani, M. A. Mastanduno, W. Turner, S. C. Davis, H. Dehghani, and B. W. Pogue, “Fast segmentation and high-quality three-dimensional volume mesh creation from medical images for diffuse optical tomography,” J. Biomed. Opt. 18(8), 086007 (2013). [CrossRef]   [PubMed]  

9. C. S. Robertson, S. P. Gopinath, and B. Chance, “A new application for near-infrared spectroscopy: detection of delayed intracranial hematomas after head injury,” J. Neurotrauma 12(4), 591–600 (1995). [CrossRef]   [PubMed]  

10. J. B. Wu, C. Shi, G. C. Chu, Q. Xu, Y. Zhang, Q. Li, J. S. Yu, H. E. Zhau, and L. W. Chung, “Near-infrared fluorescence heptamethine carbocyanine dyes mediate imaging and targeted drug delivery for human brain tumor,” Biomaterials 67(10), 1–10 (2015). [CrossRef]   [PubMed]  

11. H. Tian, M. Li, Y. Wang, D. Sheng, J. Liu, and L. Zhang, “Optical wavelength selection for portable hemoglobin determination by near-infrared spectroscopy method,” Infrared Phys. Techn 86, 98–102 (2017).

12. M. N. Kim, B. L. Edlow, T. Durduran, S. Frangos, R. C. Mesquita, J. M. Levine, J. H. Greenberg, A. G. Yodh, and J. A. Detre, “Continuous optical monitoring of cerebral hemodynamics during head-of-bed manipulation in brain-injured adults,” Neurocrit. Care 20(3), 443–453 (2014). [CrossRef]   [PubMed]  

13. X. Dai, T. Zhang, H. Yang, J. Tang, P. R. Carney, and H. Jiang, “Fast non-invasive functional diffuse optical tomography for brain imaging,” J. Biophotonics 11, 3 (2017). [PubMed]  

14. J. J. Lee, D. J. Segar, J. F. Morrison, W. M. Mangham, S. Lee, and W. F. Asaad, “Subdural hematoma as a major determinant of short-term outcomes in traumatic brain injury,” J. Neurosurg. 128(1), 236–249 (2018). [CrossRef]   [PubMed]  

15. L. Di Sieno, H. Wabnitz, A. Pifferi, M. Mazurenka, Y. Hoshi, A. Dalla Mora, D. Contini, G. Boso, W. Becker, F. Martelli, A. Tosi, and R. Macdonald, “Characterization of a time-resolved non-contact scanning diffuse optical imaging system exploiting fast-gated single-photon avalanche diode detection,” Rev. Sci. Instrum. 87(3), 035118 (2016). [CrossRef]   [PubMed]  

16. J. Wang, D. Liu, J. Sun, Y. Zhang, Q. Sun, J. Ma, Y. Zheng, and H. Wang, “The study on increasing the equivalent SNR in the certain DOI by adjusting the SD separation in near-infrared brain imaging application,” Proc. SPIE 157, 101572P (2016). [CrossRef]  

17. D. Contini, L. Zucchelli, L. Spinelli, M. Caffini, R. Re, A. Pifferi, R. Cubeddu, and A. Torricelli, “Brain and muscle near infrared spectroscopy/imaging techniques,” J. Near Infrared Spectrosc. 20(1), 15–27 (2012). [CrossRef]  

18. C. C. Chuang, C. M. Chen, Y. S. Hsieh, T. C. Liu, and C. W. Sun, “Brain structure and spatial sensitivity profile assessing by near-infrared spectroscopy modeling based on 3D MRI data,” J. Biophotonics 6(3), 267–274 (2013). [CrossRef]   [PubMed]  

19. H. Dehghani, M. E. Eames, P. K. Yalavarthy, S. C. Davis, S. Srinivasan, C. M. Carpenter, B. W. Pogue, and K. D. Paulsen, “Near infrared optical tomography using NIRFAST: Algorithm for numerical model and image reconstruction,” Commun. Numer. Methods Eng. 25(6), 711–732 (2009). [CrossRef]   [PubMed]  

20. A. Hannukainen, L. Harhanen, N. Hyvönen, and H. Majander, “Edge-promoting reconstruction of absorption and diffusivity in optical tomography,” Inverse Probl. 32(1), 015008 (2016). [CrossRef]  

21. M. S. Wru00f3bel, A. P. Popov, A. V. Bykov, M. Kinnunen, and V. V. Tuchin, “Multi-layered tissue head phantoms for noninvasive optical diagnostics,” Journal of Innovation in Optical Health Science 8(03), 463- (2015).

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (6)

Fig. 1
Fig. 1 Differential optical density difference curve acquisition schematic.
Fig. 2
Fig. 2 Source (o) and detectors’ (x) location. The coordinate system established at model vertices was used to represent the position of the source and the detectors.
Fig. 3
Fig. 3 ΔOD' curve of anomaly at different horizontal positions.
Fig. 4
Fig. 4 The ΔOD' curve of anomaly at different depths.
Fig. 5
Fig. 5 ΔOD' curves of anomaly at different sizes. (a) Result of ΔOD' at r = 0.5 cm at different SS-MD distribution positions. (b), (c), and (d) results of ΔOD' with different sizes in fixed SS-MD positions.
Fig. 6
Fig. 6 Eigenvalue fitting results. (a) Curve fit of the characteristic peak about the anomaly’s ΔOD' with different sizes. (b) Curve fit of the characteristic peak about the anomaly’s ΔOD' at different depths. (c) Curve fit of the characteristic peak about the anomaly’s ΔOD' at different positions.

Equations (5)

Equations on this page are rendered with MathJax. Learn more.

O D i =lg( I 0 I i ),(i=1,2...n)
O D i ' =lg( I 0 I i ' ),(i=1,2...n)
ΔO D i =O D i O D i ' =lg I 0 I i lg I 0 I i ' =lg I i I i ' ,(i=1,2...n)
ΔO D ' =ΔO D i+l ΔO D i ,(i=1,2...,n-l)
f(x)= i=1 n A i exp[ ( x B i C i ) 2 ]
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.