Expand this Topic clickable element to expand a topic
Skip to content
Optica Publishing Group

Impact of vessel diameter and bandwidth of illumination in sidestream dark-field oximetry

Open Access Open Access

Abstract

We investigate the possibility of oxygen saturation estimation from images obtained by the sidestream dark-field (SDF) technique. The SDF technique is a method for microvascular imaging. In SDF imaging, light enters a tissue directly from illumination sources configured around a camera and then the camera captures the light scattered by the tissue. To advance the capability of the SDF imaging system, we develop a SDF oximetry method with LED illumination sources. In this paper, we evaluate some SDF oximetry methods from virtual SDF images obtained by the Monte Carlo photon propagation simulation. As a result, we verify that SDF imaging allows the estimation of oxygen saturation of the individual vessels from virtual images using the average extinction coefficients considering the bandwidth of the illumination and the effect of the integration of the camera.

© 2015 Optical Society of America

1. Introduction

Microcirculation refers to the blood circulation in those blood vessels having a diameter of about 100 μm. Microcirculation is found within organ tissues and exists throughout the body. The function of microcirculation is to exchange nutrients and oxygen delivered by erythrocytes for waste products of organ tissues. In other words, microcirculation occurs the region where arteries turn into veins.

Sidestream dark-field (SDF) imaging was developed as an imaging system to observe microcurculation directly [1]. The SDF system provides a clinically-applicable, portable, and hand-held imaging to study the microcirculation of tissue surfaces. A schematic diagram of SDF imaging is shown in Fig. 1. SDF imaging employs a light-guide equipped with concentrically placed light-emitting diodes (LEDs) to provide illumination. The lens system in the core of the light-guide is optically isolated from LEDs of the outer ring, preventing microcirculation images from contamination by tissue surface reflections. As for analysis of the microcirculation by the SDF technique, some physical quantities such as vessel length, vessel diameter, and velocity of red blood cells (RBCs) have been estimated by analyzing the SDF images of individual vessels [2].

 figure: Fig. 1

Fig. 1 Schematic diagram of the SDF imaging.

Download Full Size | PDF

Microcirculation imaging can be applied to diagnose the state of an organ. In digestive tract open surgery, surgeons typically diagnose the state of an organ as normal or ischemia on the basis of its visual color in order to decide whether to remove a part of it. The diagnosis relies on the surgeons’ visual impression and previous experiences. Therefore, a quantitative technique for evaluating the state of an organ is desirable. A macro technique with a hyperspectral camera has been suggested [3]. However, it is difficult for any macro techniques to specifically distinguish the boundary of the ischemia part. For a digestive organ, videoscope techniques such as laparoscopic and endoscopic surgery are commonly used recently. In this case, inaccurate color reproduction disturbs the diagnosis of surgeons. In contrast, a micro technique such as an observation of the microcirculation is more accurate to distinguish the boundary of the ischemia part, although it is a local observation. Therefore, the SDF imaging is useful to help in diagnosing the state of organs. Moreover, the SDF imaging can be applied to videoscope surgery because the SDF probe can be fabricated with a small diameter. However, for accurate diagnosis, monitoring of not only the flow of RBCs but also some other physical quantities are required. In particular, the most important item among the physical quantities of blood to evaluate the level of the state of an organ is oxygen saturation (SO2).

The measurement of SO2 from images, which is called image-based oximetry, has been actively studied in ophthalmology. The first retinal image-based oximetry experiment was reported by Hickam et al. [4]. They used broadband illumination provided by a dual-wavelength pair of illumination sources (peak wavelength: 510 and 640 nm) with the bandwidth of > 100 nm FWHM (full width at half maximum). Cohen et al. [5] reported results of an image-based experiment using illumination sources with dual-wavelengths of 470 and 515nm with a 20 nm bandwidth. Besides ophthalmology studies, there have been some to estimate SO2 of individual vessels. Pittman and Duling [6] reported transmission type SO2 measurements using a microscope. They used illumination sources with 546 and 555 nm with dual-wavelengths of a 10 nm bandwidth and considered the effect of light scattering. Styp-Rekowska et al. [7] also used a microscope and illumination sources from 500 to 598 nm wavelengths at 2 nm intervals and evaluated the influence of the number of wavelengths used, exposure time, and analysis area.

In the present study, we aim at establishing an estimation method of oxygen levels of the individual vessels from spectral images acquired with the SDF imaging system. For this purpose, we perform the Monte Carlo photon propagation simulation as the first approach in which we evaluate dependence of accuracy of the SDF oximetry on vessel diameters and wavelength-bandwidths of the illumination sources.

2. Methods and materials

2.1. Formula for oximetry

2.1.1. Lambert-Beer’s law for biological tissue

In this section, we discuss image-based oximetry by considering incident light of intensity Iin = Iin(λ) at wavelength λ. When the incident light enters into medium with a thickness d, the Lambert-Beer’s law gives the intensity Iout = Iout(λ) of the transmitted light through the medium as:

Iout=Iin10μa(λ)d,
where μa(λ) represents an absorption coefficient that can be written by μa(λ) = ε(λ)c, in which ε(λ) is an extinction coefficient and c is the concentration of the absorber in the medium. Using this expression, we define optical density D = D(λ) as:
D=log(IoutIin)=μa(λ)d.
We consider that the medium is blood and assume that the incident light is absorbed by only hemoglobin, then μa(λ) is written as:
μa(λ)=[sεHbO2(λ)+(1s)εHb(λ)]c,
where εHbO2(λ) and εHb(λ) represent the extinction coefficients of the oxygenated and the deoxygenated hemoglobin respectively, c represents the hemoglobin concentration, and s is the saturation. To be exact, the optical density D(λ) for real blood needs to be corrected with an additional term. However, we will use the above simple form in this subsection and modify it in the next subsection. To estimate the saturation s in the case that εHbO2(λ) and εHb(λ) are known, we solve the follow equations for s using different wavelengths λ1 and λ2:
D(λ1)=[sεHbO2(λ1)+(1s)εHb(λ1)]cd,D(λ2)=[sεHbO2(λ2)+(1s)εHb(λ2)]cd.
Next, we define the ratio of D(λ1) and D(λ2) as Φ = Φ(λ1, λ2), which is called the saturation parameter in reference [5] by:
Φ(λ1,λ2)=D(λ2)D(λ1).
From Eqs. (4) and (5), we obtain the saturation s as:
s=ΦεHb(λ1)εHb(λ2)Δλ2ΦΔλ1,
where Δλn := εHbO2(λn) − εHb(λn) for n = 1, 2. It should be noted that we do not have to know c and d to solve for s.

In the discussion above, we do not take into account the reflected light from the medium. In reality, however, there is a portion of light that returns from the surface of the medium in which light enters into. When a medium is highly scattering, the ratio of the reflected light is larger than the transmitted light. Therefore, the Lambert-Beer’s law can be applied to the reflected intensity of light in the case of a highly scattering medium [8]. A biological tissue is a highly scattering medium, hence the saturation s of a tissue can be estimated from Eq. (6) based on the Lambert-Beer’s law in principle, but there are some uncertainties such as light path length and the effect of scattering. Thus, many studies, as reviewed in [8], have suggested an appropriate way to correct them for each modality. In the case of reflected light, the optical density D = D(λ) of Eq. (2) can be written by using the reflectance R = R(λ):

D=logR.
As with the transmitted light, Eq. (6) can be applied to estimating s by measuring the reflectance.

2.1.2. SDF oximetry

We propose a SDF oximetry method that estimates the SO2 from SDF images of a vessel. The proposed method is constructed in an analogy to retinal oximetry methods. In this section, we show the SDF oximetry method using two spectral images. When applying retinal oximetry methods to SDF imaging, we must consider differences between the retinal oximetry and the SDF oximetry. As presented in Fig. 2, there is a difference in the light paths. In retinal oximetry (Fig. 2(a)), photons backscattered from the vessel mainly contribute to the detected intensity at the vessel part in a retinal fundus image [9]. In contrast, photons that pass once the vessel (single pass) mainly contribute to that in a SDF image (Fig. 2(b)). In the SDF technique, light enters a tissue from illumination sources configured around the camera and then the camera captures the light scattered by the tissue. Consequently, we apply retinal oxymetry methods to the SDF oximetry keeping in mind the above consideration. We consider the proposed SDF oximetry method. Figure 3 shows a concept of the proposed method. As shown in Figs. 3(a) and 3(b), we define the light that enters the body with intensity I0(λ) and returns outside without passing through any vessels as back reflection light and define its intensity as Iin(λ) = R(λ)I0(λ) where R(λ) denotes the total reflectance of layers behind the vessel near a surface, e.g. muscularis, submucosa, and mucosa. Next we assume the same amount of reflecting light illuminates the vessels. Therefore, the intensity of the transmitted light can be expressed in terms of I0 as in Eq. (1):

Iout=Iin(λ)10μa(λ)d=R(λ)I0(λ)10μa(λ)d,
where μa(λ) is the absorption coefficient defined in Eq. (3). Here we ignore the effect of photons that pass more than twice through the vessel (double pass) because all photons enter the peripheral region of the target vessel and most of them pass through the vessel once at most. In this case, the optical density D(λ) of blood can be written as:
D(λ)=μa(λ)d=[sεHbO2(λ)+(1s)εHb(λ)]cd.
As in the previous section, we define the saturation parameter by: Ψ(λ1, λ2) = D(λ2)/D(λ1). To determine s, we solve Eq. (9) for s with two wavelengths λ = λ1, λ2 and we obtain:
s=ΨεHb(λ1)εHb(λ2)Δλ2ΨΔλ1,
where Ψ = Ψ(λ1, λ2) and Δλn := εHbO2(λn) − εHb(λn) for n = 1, 2.

 figure: Fig. 2

Fig. 2 Light paths for retinal and SDF imaging. (a) For a retinal fundus camera. Three paths can be considered: incident light is backscattered from a blood vessel (backscattering); reflected or scattered light passes once through a vessel (single pass); the light passes twice through a vessel (double pass). (b) For a SDF imaging system. In contrast with a fundus camera, there is no backscattered light. The dashed line triangles denote the illumination area.

Download Full Size | PDF

 figure: Fig. 3

Fig. 3 A conceptual diagram of the SDF oximetry method. (a) Light with the intensity I0 enters into a tissue. (b) The incident light is reflected and/or scattered in each layer, then the intensity becomes RI0. The light with Iin = RI0 enters into a vessel and then goes through. Iout represents the intensity of the transmitted light.

Download Full Size | PDF

2.2. Average extinction coefficient

Light-emitting diodes (LEDs) are commonly used as illumination sources in SDF imaging [1, 10]. As discussed in the previous section, monochromatic light is required for the SO2 estimation. Thus, we have to take into account the effect of bandwidth of the LEDs since the bandwidth cannot be ignored. Moreover, a typical camera detects intensity of light by integrating over all wavelengths. Hence we also have to consider the effect of this integration on the SO2 estimation. In order to address these issues, we modified the extinction coefficients of hemoglobin by averaging over the wavelength in the SO2 estimation by referring to [5].

First, we discuss the general case. As we mentioned earlier, the optical density of real blood is written with an additional term by:

D(λ)=μa(λ)d+B(λ),
where d is the diameter of a vessel, μa(λ) represents the absorption coefficient of hemoglobin. The additional term B(λ) represents a correction for the RBC geometry and the detector geometry, and all phenomena other than the absorption of hemoglobin such as scattering and the sieve effect [11]. The sieve effect is a discrepancy between spectral absorption of RBCs and that of hemoglobin itself. Its dominant wavelength range is approximately from 370 to 430 nm. Fortunately, since this wavelength range is far from that of LED used in this study (see the sub-subsection 2.3.2), this effect is negligible.

We consider light with the amplitude I0 = I(λ0) at the peak wavelength λ0 and the relative spectral intensity distribution F0(λ) of the light I0, i.e. the intensity of light represents I0F0(λ). A camera detects the intensity of the light transmitting through the vessel as:

I¯out(λ0)=ΩdλR(λ)S(λ)I0F0(λ)10[μa(λ)d+B(λ)],
where S(λ) represents the sensitivity of the camera, R(λ) represents the total reflectance of layers behind the vessel, and Ω represents the spectral sensitivity region of the camera. The intensity of light outside a vessel is written by:
I¯in(λ0)=ΩdλR(λ)S(λ)I0F0(λ).
Similarly to Eq. (2), we define the optical density (λ0) = −log[Īout(λ0)/Īin(λ0)] in consideration of the influence of the integral detection by the camera and the bandwidth of an illumination by:
D¯(λ0)=log[ΩdλR(λ)S(λ)F0(λ)10[μa(λ)d+B(λ)]ΩdλR(λ)S(λ)F0(λ)]=:[sε¯HbO2(λ0,d)+(1s)ε¯Hb(λ0,d)]cd+B(λ0),
where ε̄HbO2 and ε̄Hb represent average extinction coefficients. The average extinction coefficients are defined so that the optical density has the same form as Eq. (2). It should be noted that the average extinction coefficients are a function of not only the peak wavelength, but also the diameter of the vessel. Solving Eq. (14) with s = 1 and s = 0, we obtain:
ε¯HbO2(λ0,d)=1cd(log[ΩdλR(λ)S(λ)F0(λ)10[εHbO2(λ)cd+B(λ)]ΩdλR(λ)S(λ)F0(λ)]+B(λ0)),ε¯Hb(λ0,d)=1cd(log[ΩdλR(λ)S(λ)F0(λ)10[εHb(λ)cd+B(λ)]ΩdλR(λ)S(λ)F0(λ)]+B(λ0)),
respectively. Here μa(λ) is explicitly expressed as μa(λ) = εHbO2 (λ)c and μa(λ) = εHb(λ)c for s = 1 and s = 0, respectively. If light is monochromatic, i.e. F0(λ) = δ(λλ0) which is a Dirac delta function, then Eq. (15) becomes ε̄HbO2 (λ0) = εHbO2 (λ0) and ε̄Hb(λ0) = εHb(λ0).

Finally, we discuss the approximation made in the discussion above. Cohen and Laing [5] assumed that the dependences of the RBCs’ scattering part of B(λ) and the reflectance R(λ) on wavelength are much smaller than variations of S(λ), F(λ), and μa(λ), then they do not contribute at all to the integral. In addition to Cohen’s approximation, we assume that variation of B(λ) other than RBCs’ scattering is also smaller than that of S(λ), F(λ), and μa(λ), i.e. B(λ) ≃ B(λ0) in the wavelength range of LED used. Therefore, Eq. (15) can be rewritten to:

ε¯HbO2(λ0,d)=1cd(log[ΩdλS(λ)F0(λ)10εHbO2(λ)cdΩdλS(λ)F0(λ)]),ε¯Hb(λ0,d)=1cd(log[ΩdλS(λ)F0(λ)10εHb(λ)cdΩdλS(λ)F0(λ)]).
In this study, we assume that S(λ) is set to unity and F(λ) is set to a Gaussian function (see the next subsection) in order to simplify the calculation. The diameter d of a blood vessel is assumed to be consistent with the FWHM of the profile of a vessel image [12, 13]. Before estimating SO2, we measure d from images, and then evaluate Eq. (16) with known εHbO2 c and εHbc.

2.3. Setups of experiments

We performed a Monte Carlo photon propagation simulation to evaluate the SO2 estimation method. As a basic study, we investigated the trend of the estimated SO2 by changing values of SO2, vessel diameter, and bandwidth of illumination sources. Our procedure was as follows: first, we made a tissue model with a blood vessel and set blood with known SO2 into the vessel. Next, we simulated the photon propagation in the tissue and acquired virtual SDF images. Finally, we estimate SO2 from the virtual SDF images.

2.3.1. Tissue model

To construct a three-dimensional tissue model with a vessel, we used an open-source software iso2mesh [14,15] which is coded with Matlab (MathWorks, Natick, USA). The tissue model is constructed by referring to the result of hematoxylin and eosin (H&E) staining of a pig small intestine shown in Fig. 4(a). The simulated tissue model is shown in Figs. 4(b) and 4(c). As shown in Fig. 4(d), each of the layers simulate serosa, muscularis, submucosa, and mucosa from the top surface respectively. Here the model has a blood vessel for which the uppermost surface is located at 100 μm below the top surface in the serosa. We set the blood vessel diameter as 100, 200, and 300 μm. In this study, the optical properties of the blood vessel and blood were approximated by those of hemoglobin. The oxygenated saturation was varied by the amount of the oxygenated hemoglobin. We used these optical properties in the case of the hematocrit value H = 0.4 reported in the literature [16, 17]. The optical properties of blood are shown in Fig. 5(a). In the optical properties of tissue, we assumed that absorption and scattering of each layer were homogeneous, the saturation of tissue was fixed, and the absorption coefficient of each layer other than the serosa was equal. We used the value of mucosa in the study by Bashkatov et al. [18] as the absorption coefficient of tissue. The scattering coefficient is in proportion to the volume fraction of scatterer in a medium [19]. Accordingly, we assumed that the size of scatterer in each layer was equal. Since the submucosa includes an abundant number of blood vessels and tends to scatter the light, the ratio of the volume fraction of scatterer to that of the submucosa was set as 0.7 in the mucosa and 0.3 in the muscularis. The scattering coefficient of the submucosa was set to the value reference to the data of Bashkatov et al. [18]. Absorption coefficient and scattering coefficient of the submucosa are shown in Fig. 5(b). For the serosa, we set this layer as transparent, i.e. μa = 0 mm−1 and μs = 0 mm−1. The anisotropy factor of tissue was set as 0.9. This is because this value is approximately equal for any tissue.

 figure: Fig. 4

Fig. 4 The tissue model. (a) The cross section of a small intestine of a pig H&E stained. The layers, from the top surface, serosa, muscularis, submucosa, and mucosa. The inset on the right is an enlarged image that shows blood vessels are present in the serosa (blue arrows). (b) (c) The simulated tissue model which has a layered structure and a straight blood vessel at 0.1 mm below the top surface in the serosa. (b) Side view. (c) Overview. (d) A schematic diagram of the tissue model and the geometry of the simulation.

Download Full Size | PDF

 figure: Fig. 5

Fig. 5 Optical properties used in this study. (a) Optical properties of blood [16, 17]. Open and closed squares represent the absorption coefficient of oxygenated and de-oxygenated hemoglobin respectively, crosses represent the scattering coefficient, and circles represents the anisotropy factor. (b) Optical properties of the tissue [18]. Squares represent the absorption coefficient and crosses represent the scattering coefficient.

Download Full Size | PDF

2.3.2. Trial SDF device

We assumed the illumination geometry was similar to that in the SDF imaging device that we have developed. Figure 6 shows our trial device with LED chips (SMLV56RGB1W, ROHM Co., Ltd., Japan) and a charge-coupled device (CCD) camera. Six LED chips were configured around the exterior tip of a probe to provide uniform illumination. These LED chips could be switched from blue to green light. The peak wavelengths of each color were 470 nm and 527 nm and the bandwidths (FWHM) of each intensity distribution were 27.3 nm and 37.2 nm respectively. The reasons why we chose these wavelengths were that light with these wavelengths is well absorbed by hemoglobin and their penetration distance in a tissue is shorter than that of longer-wavelength light [20], so these wavelengths are suitable for observation of vessels near the surface. The incident angle of our device was 16°. Figure 4(d) shows the geometry of the simulation. A LED chip source was assumed as a set of point sources. In this simulation study, the number of point sources per LED chip was 100. We set the initial launch area to be the same as that of a LED chip of the trial device. The area of a LED chip was 2.8 × 3.1 mm2. To simulate an area source with a set of point sources, we generated position launching photons with the uniform probability distribution.

 figure: Fig. 6

Fig. 6 The trial SDF illumination device. (a) A schematic diagram of the trial SDF device. The outer diameter of this device was 13.5 mm. LED chips were used as illumination. (b) This LED chip could be switched from blue light at the peak wavelength of 470 nm to green light at 527 nm.

Download Full Size | PDF

We assumed that the spectral intensity distribution of a LED with a peak wavelength λ0 has a Gaussian function given by:

F0(λ)=Nexp[(λλ0)22σ2],
where N is the normalization coefficient and σ is the coefficient to control the bandwidth. We set bandwidth w as the FWHM of the Gaussian function, namely:
σ=w28ln2.
For w = 30 nm, σ is 12.74 nm.

2.3.3. Monte Carlo photon propagation simulation

We used a mesh-based Monte Carlo (MMC) approach [21, 22] developed based on references [2325] for a three-dimensional Monte Carlo code. The Monte Carlo simulation procedures are explained in detail in [2325].

For the simulation of light absorption in a tissue, we used the concept of the photon packet. A photon packet enters into a tissue with an initial weight W0 = 1, and then after s steps, its weight becomes Ws = Ws−1 exp[−μaΔ], where Ws−1 is the weight at the (s − 1)-th step, μa is the absorption coefficient of the tissue and Δ is the step distance as it propagates within the tissue. When a photon packet reaches the top surface, its weight is detected with respect to a sub-region in the top surface. We assumed that a virtual SDF image acquired by a CCD camera was the spatial intensity distribution of the average weight of the sub-region.

As setups for the simulation, we set the number of incident photons as 2.0 × 107 per LED chip. The number of a sub-regions was 100 × 100 and the area of a sub-region was 0.02 × 0.02 mm2.

3. Results

3.1. Photon propagation simulation

Figure 7(a) shows a virtual SDF image simulated for the light at 470 nm with bandwidth w = 0 and the oxygenated blood vessel with diameter d = 100 μm. We can see the straight vessel in the image. Figure 7(b) shows the x-directional intensity profiles of the single pass, double pass, and no pass of light for the case of Fig. 7(a); there were obtained by averaging the two-dimensional distribution along the y-direction. This result indicates that the double pass light does not contribute to the virtual SDF image unlike for the single pass light. We measured the FWHM of the profiles from virtual SDF images, and then evaluated average extinction coefficients using Eq. (16). Figure 7(c) shows the average extinction coefficients results. Figure 7(d) shows the relation of intensities between incident and transmitted light for d = 100 μm. We used the weighted average of these intensities to evaluate the Lambert-Beer’s law below.

 figure: Fig. 7

Fig. 7 Results of the photon propagation simulation for a tissue model with the straight vessel. (a) The virtual SDF image (the gray level map of the intensity). (b) The profile of the virtual SDF image (blue line). The single pass, the double pass, and the no pass are depicted by red, magenta, and green, respectively. (c) Absorption coefficients with average extinction coefficients are plotted as a function of bandwidth of the illumination. ε̄(λ0) for λ0 = 470 nm and 527 nm are depicted by squares and triangles, respectively. Open and closed symbols represent SO2 s = 1 and s = 0, respectively. Diameter d values of 100 μm, 200 μm, and 300 μm are depicted by solid lines, dashed lines, and chain lines, respectively. (d) The histogram of the intensity Iin of light directly entering into the blood vessel and Iout of light passing through it when the model is (a).

Download Full Size | PDF

3.2. Three SO2 estimation methods and mean SO2 estimation error

We estimated SO2 using three methods: the first one was the Lambert-Beer’s law using weighted average values of the intensities of the incident and transmitted light of the blood vessel (Fig. 7(d)), which is called the pure Lambert-Beer’s law method (PLB); the second one was the image-based measurement with the extinction coefficients at the peak wavelength, which is called a method using the extinction coefficients at the peak wavelength (EPW); the last one was the imaged-based measurement with the average extinction coefficients defined in Eq. (16), which is called a method using the average extinction coefficients (AEC).

We used the SO2 estimation error (SEE) as an evaluation index of estimation accuracy. We modeled the estimated saturation s by a linear function to parameterize those trends:

s=αt+β,
where t (0 ≤ t ≤ 1) represents the true SO2 value and α and β are fitting coefficients. Those coefficients are obtained by the Levenberg-Marquadt method. If these coefficients are close to α = 1 and β = 0, the estimation is accurate. When the SEE Δs(t) is defined as an absolute value of the difference between the estimated curve Eq. (19) and the true curve s = t, we define the mean SEE Δs¯ as:
Δs¯:=1Mi=1MΔs(ti)=1Mi=1M|(α1)ti+β|,
where M represents the number of sample points over the SO2 range and ti represents the true SO2 value at the i-th sample point. In this study, we had 21 true SO2 values, that is, M = 21.

Figures 8, 9 and 10 show the results of the estimated value s and the estimated curve by Eq. (19) for d = 100 μm, 200 μm, and 300 μm, respectively. Three methods (PLB, EPW, and AEC) are shown together. We show the Δs¯ and standard deviation (SD) of each measurement in Table 1 and summarize the Δs¯ of these results in Fig. 11. In Fig. 11, the plots are fitted by a surface with bivariate d and w power series up to the second-order of both variables. Since extinction coefficients at the peak wavelength and average extinction coefficients match in w = 0, Δs¯ of EPW and AEC are equal. As a result, the tendency of the SEE of PLB (Fig. 11(a)) and AEC (Fig. 11(c)) are explicitly similar.

 figure: Fig. 8

Fig. 8 The SO2 estimation of d = 100 μm vessel. (a) The monochromatic light with no bandwidth (w = 0). (b) w = 10 nm. (c) w = 20 nm. (d) w = 30 nm. (e) w = 40 nm. (f) w = 50 nm.

Download Full Size | PDF

 figure: Fig. 9

Fig. 9 The SO2 estimation of d = 200 μm vessel. (a) The monochromatic light with no bandwidth (w = 0). (b) w = 10 nm. (c) w = 20 nm. (d) w = 30 nm. (e) w = 40 nm. (f) w = 50 nm.

Download Full Size | PDF

 figure: Fig. 10

Fig. 10 The SO2 estimation of d = 300 μm vessel. (a) The monochromatic light with no bandwidth (w = 0). (b) w = 10 nm. (c) w = 20 nm. (d) w = 30 nm. (e) w = 40 nm. (f) w = 50 nm.

Download Full Size | PDF

Tables Icon

Table 1. The mean SEE Δs¯±SD.

 figure: Fig. 11

Fig. 11 The mean of SEE Δs¯ surface with the contour interval color. (a) PLB. (b) EPW. (c) AEC.

Download Full Size | PDF

We focused on the changes in Δs¯ due to w. In the comparison between image-based methods, AEC is more accurate than EPW for all bandwidths in each d. As shown in Fig. 11(b) and Table 1, Δs¯ using EPW increases with increment of w. The changing rate of Δs¯ due to w decreases with an increase in d. On the other hand, as shown in Fig. 11(c) and Table 1, Δs¯ using AEC decreases with increment of w. Next, we focused on the changes in Δs¯ due to d of the vessel. From the results, AEC is more precise than EPW for all d values in each w. Δs¯ using EPW increases with increment of d. The changing rate of Δs¯ due to d decreases with an increase in w. By contrast, Δs¯ using AEC increases with increment of d.

4. Discussion and conclusion

We have investigated the validity of an SO2 estimation of individual vessels using two spectral SDF images obtained by a Monte Carlo photon propagation simulation where two image-based methods (EPW and AEC) were compared with the intensity method (PLB). As an experiment, we performed the simulation with illumination sources from w = 0 nm to 50 nm at 10 nm intervals and the tissue model with vessel diameter d = 100, 200, and 300 μm.

There were strong correspondences between PLB and the image-based methods. The SEE of AEC was especially similar to the SEE of PLB. In the EPW, the SEE was an increasing function with respect to d and w. On the other hand, in the PLB and AEC, the errors were not a monotone function. In addition, for all conditions, AEC was more accurate than EPW, hence using the average extinction coefficients of hemoglobin is effective for the SDF oximetry with narrowband illumination sources such as LEDs.

We considered that the SEE was caused by a deviation from Lambert-Beer’s law due to the scattering event in case of using the extinction coefficient of hemoglobin. In our study we assumed that, the scattering in a tissue and a vessel was more dominant than the absorption over the entire wavelength range and more than 440 nm, respectively. Therefore, its effect was not negligible. We considered that the its possible effects were the overestimation of intensity in the vessel and a positional displacement to reach the surface due to multiple scattering in the tissue. The displacement caused a loss or a gain of detected intensity. A correction of these effects is included in the term of B(λ) in Eq. (11). In our study, we assumed that B(λ) was constant compared to extinction coefficients of hemoglobin and the camera sensitivity, and then it vanishes in Eq. (16), i.e. we did not take into account the influence of the scattering and the other factors. This assumption is not very likely to be valid. Thus, we can correct these influence with the average extinction coefficients that include B(λ).

In addition to the above prediction, we considered that there was another cause of the SEE. Smith [26] pointed out that generally the SEE is due to the combination of the selected wavelengths in retinal vessel oximetry. Here this error is called the potential error (PE). The potential error occurs in the difference of the extinction coefficients of hemoglobin and de-oxygenated hemoglobin between selected wavelengths. Since average extinction coefficients vary in d and w, the PE also varies in a combination of d and w. Here, we evaluated the PE with the procedure proposed by Smith [26] with average extinction coefficients. Figure 12 shows the result with a combination of our calculated average extinction coefficients. We used mean FWHM of profiles as the diameter and ΔT1 = ΔT2 was set as 1.0 × 10−3 (see Eqs. (21) and in [26]). As a result, we verified that the tendency of our SEEs of AEC was consistent with the mean PEs due to changes in average extinction coefficients. From this result, the mean of SEE was greatly affected by the wavelength combination compared with the aforementioned scattering effect. When monochromatic light is irradiated to the model with vessel d = 300 μm, the mean PE takes the maximum value. In the case of w = 0, the mean PE occurs in the change of d rather than the difference of average extinction coefficients and increases as d increases. This is because average extinction coefficients of any d are equal.

 figure: Fig. 12

Fig. 12 The mean PE evaluated by Smith’s procedure [26]. The surface is a fitting curve with a bivariate power series until the second-order of both variables.

Download Full Size | PDF

As a future plan, we will construct a method for correcting the PE and deviation from Lambert-Beer’s law. One candidate correction method is to use a database. In the method, we obtain α and β for many combinations of d and w by a photon propagation simulation and estimate s from real SDF images. Then, we look up α and β in the corresponding condition in the database; finally, we predict the true SO2 value from t = (sβ)/α. This method can be used without knowing the effects of B(λ). We expect that this method is useful for the SO2 estimation of a vessel in microcirculation.

Acknowledgments

This research was supported by Technology Foundation, Kakenhi, the Grant-in-Aid for Scientific Research B25282151.

References and links

1. C. Ince, “The microcirculation is the motor of sepsis,” Critical care 9Suppl 4, S13–9 (2005). [CrossRef]   [PubMed]  

2. J. G. Dobbe, G. J. Streekstra, B. Atasever, R. Van Zijderveld, and C. Ince, “Measurement of functional microcirculatory geometry and velocity distributions using automated image analysis,” Med. Biol. Eng. Comput. 46, 659–670 (2008). [CrossRef]   [PubMed]  

3. H. Akbari, Y. Kosugi, K. Kojima, and N. Tanaka, “Detection and analysis of the intestinal ischemia using visible and invisible hyperspectral imaging,” IEEE Trans. Biomed. Eng. 57, 2011–2017 (2010). [CrossRef]   [PubMed]  

4. J. B. Hickam, R. Frayser, and J. C. Ross, “A study of retinal venous blood oxygen saturation in human subjects by photographic means,” Circulation 27, 375–385 (1963). [CrossRef]   [PubMed]  

5. A. J. Cohen and R. A. Laing, “Multiple scattering analysis of retinal blood oximetry,” IEEE Trans. Biomed. Eng. 23391–400 (1976). [CrossRef]   [PubMed]  

6. R. N. Pittman and B. R. Duling, “Measurement of percent oxyhemoglobin in the microvasculature,” J. Appl. Physiol. 38, 321–327 (1975). [PubMed]  

7. B. Styp-Rekowska, N. Disassa, B. Reglin, L. Ulm, H. Kuppe, T. Secomb, and A. Pries, “An imaging spectroscopy approach for measurement of oxygen saturation and hematocrit during intravital microscopy,” Microcirculation 14, 207–221 (2007). [CrossRef]   [PubMed]  

8. R. V. Maikala, “Modified Beer’s Law – historical perspectives and relevance in near-infrared monitoring of optical properties of human tissue,” Int. J. Indust. Ergo. 40, 125–134 (2010). [CrossRef]  

9. M. Hammer, S. Leistritz, L. Leistritz, and D. Schweitzer, “Light paths in retinal vessel oximetry,” IEEE Trans. Biomed. Eng. 48, 592–598 (2001). [CrossRef]   [PubMed]  

10. P. Goedhart, M. Khalilzada, R. Bezemer, J. Merza, and C. Ince, “Sidestream dark field (SDF) imaging: a novel stroboscopic LED ring-based imaging modality for clinical assessment of the microcirculation,” Opt. Express 15, 15101–15114 (2007). [CrossRef]   [PubMed]  

11. A. R. Pries, G. Kanzow, and P. Gaehtgens, “Microphotometric determination of hematocrit in small vessels,” Am. J. Physiol. 245, 167–177 (1983).

12. J. Lowell, A. Hunter, D. Steel, A. Basu, R. Ryder, and R. L. Kennedy, “Measurement of retinal vessel widths from fundus images based on 2-D modeling,” IEEE Trans. Med. Imag. 23, 1196–1204 (2004). [CrossRef]  

13. L. Pedersen, M. Grunkin, B. Ersbøll, K. Madsen, M. Larsen, N. Christoffersen, and U. Skands, “Quantitative measurement of changes in retinal vessel diameter in ocular fundus images,” Patt. Recog. Lett. 21, 1215–1223 (2000). [CrossRef]  

14. Q. Fang and D. A. Boas, “Tetrahedral mesh generation from volumetric binary and grayscale images,” in “Biomedical Imaging: From Nano to Macro, 2009. ISBI’09. IEEE International Symposium on Biomedical Imaging 2009,” (IEEE, 2009), pp. 1142–1145.

15. Q. Fang, “iso2mesh,” http://iso2mesh.sourceforge.net/cgi-bin/index.cgi?Home

16. S. Prahl, “Optical absorption of hemoglobin,” http://omlc.ogi.edu/spectra/hemoglobin/index.html.

17. M. Friebel, K. Do, A. Hahn, G. Mu, D. Berlin, L. Medizin, and F. Universita, “Optical properties of circulating human blood in the wavelength range 400 – 2500 nm,” J. Biomed. Opt. 4, 36–46 (1999). [CrossRef]  

18. N. Bashkatov, E. Genina, V. I. Kochubey, and V. V. Tuchin, “Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm,” J. Phys. D: Appl. Phys. 38, 2543–2555 (2005). [CrossRef]  

19. D. Hidović-Rowe and E. Claridge, “Modelling and validation of spectral reflectance for the colon,” Phys. Med. Biol. 50, 1071–1093 (2005). [CrossRef]  

20. G. H. Weiss, R. Nossal, and R. F. Bonner, “Statistics of penetration depth of photons re-emitted from irradiated tissue,” J. Mod. Opt. 36, 349–359 (1989). [CrossRef]  

21. Q. Fang, “Mesh-based Monte Carlo method using fast ray-tracing in Plücker coordinates,” Biomed. Opt. Express 1, 165–175 (2010). [CrossRef]   [PubMed]  

22. Q. Fang, “Mesh-based Monte Carlo (MMC),” http://mcx.sourceforge.net/cgi-bin/index.cgi?MMC.

23. L. Wang, S. L. Jacques, and L. Zheng, “MCML – Monte Carlo modeling of light transport in multi-layered tissues,” Comput. Methods Programs Biomed. 47, 131–146 (1995). [CrossRef]   [PubMed]  

24. D. Boas, J. Culver, J. Stott, and A. Dunn, “Three dimensional Monte Carlo code for photon migration through complex heterogeneous media including the adult human head,” Opt. Express 10, 159–170 (2002). [CrossRef]   [PubMed]  

25. Q. Fang and D. A. Boas, “Monte Carlo simulation of photon migration in 3D turbid media accelerated by graphics processing units,” Opt. Express 17, 20178–20190 (2009). [CrossRef]   [PubMed]  

26. M. H. Smith, “Optimum wavelength combinations for retinal vessel oximetry,” Appl. Opt. 38, 258–267 (1999). [CrossRef]  

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (12)

Fig. 1
Fig. 1 Schematic diagram of the SDF imaging.
Fig. 2
Fig. 2 Light paths for retinal and SDF imaging. (a) For a retinal fundus camera. Three paths can be considered: incident light is backscattered from a blood vessel (backscattering); reflected or scattered light passes once through a vessel (single pass); the light passes twice through a vessel (double pass). (b) For a SDF imaging system. In contrast with a fundus camera, there is no backscattered light. The dashed line triangles denote the illumination area.
Fig. 3
Fig. 3 A conceptual diagram of the SDF oximetry method. (a) Light with the intensity I0 enters into a tissue. (b) The incident light is reflected and/or scattered in each layer, then the intensity becomes RI0. The light with Iin = RI0 enters into a vessel and then goes through. Iout represents the intensity of the transmitted light.
Fig. 4
Fig. 4 The tissue model. (a) The cross section of a small intestine of a pig H&E stained. The layers, from the top surface, serosa, muscularis, submucosa, and mucosa. The inset on the right is an enlarged image that shows blood vessels are present in the serosa (blue arrows). (b) (c) The simulated tissue model which has a layered structure and a straight blood vessel at 0.1 mm below the top surface in the serosa. (b) Side view. (c) Overview. (d) A schematic diagram of the tissue model and the geometry of the simulation.
Fig. 5
Fig. 5 Optical properties used in this study. (a) Optical properties of blood [16, 17]. Open and closed squares represent the absorption coefficient of oxygenated and de-oxygenated hemoglobin respectively, crosses represent the scattering coefficient, and circles represents the anisotropy factor. (b) Optical properties of the tissue [18]. Squares represent the absorption coefficient and crosses represent the scattering coefficient.
Fig. 6
Fig. 6 The trial SDF illumination device. (a) A schematic diagram of the trial SDF device. The outer diameter of this device was 13.5 mm. LED chips were used as illumination. (b) This LED chip could be switched from blue light at the peak wavelength of 470 nm to green light at 527 nm.
Fig. 7
Fig. 7 Results of the photon propagation simulation for a tissue model with the straight vessel. (a) The virtual SDF image (the gray level map of the intensity). (b) The profile of the virtual SDF image (blue line). The single pass, the double pass, and the no pass are depicted by red, magenta, and green, respectively. (c) Absorption coefficients with average extinction coefficients are plotted as a function of bandwidth of the illumination. ε̄(λ0) for λ0 = 470 nm and 527 nm are depicted by squares and triangles, respectively. Open and closed symbols represent SO2 s = 1 and s = 0, respectively. Diameter d values of 100 μm, 200 μm, and 300 μm are depicted by solid lines, dashed lines, and chain lines, respectively. (d) The histogram of the intensity Iin of light directly entering into the blood vessel and Iout of light passing through it when the model is (a).
Fig. 8
Fig. 8 The SO2 estimation of d = 100 μm vessel. (a) The monochromatic light with no bandwidth (w = 0). (b) w = 10 nm. (c) w = 20 nm. (d) w = 30 nm. (e) w = 40 nm. (f) w = 50 nm.
Fig. 9
Fig. 9 The SO2 estimation of d = 200 μm vessel. (a) The monochromatic light with no bandwidth (w = 0). (b) w = 10 nm. (c) w = 20 nm. (d) w = 30 nm. (e) w = 40 nm. (f) w = 50 nm.
Fig. 10
Fig. 10 The SO2 estimation of d = 300 μm vessel. (a) The monochromatic light with no bandwidth (w = 0). (b) w = 10 nm. (c) w = 20 nm. (d) w = 30 nm. (e) w = 40 nm. (f) w = 50 nm.
Fig. 11
Fig. 11 The mean of SEE Δ s ¯ surface with the contour interval color. (a) PLB. (b) EPW. (c) AEC.
Fig. 12
Fig. 12 The mean PE evaluated by Smith’s procedure [26]. The surface is a fitting curve with a bivariate power series until the second-order of both variables.

Tables (1)

Tables Icon

Table 1 The mean SEE Δ s ¯ ± SD.

Equations (20)

Equations on this page are rendered with MathJax. Learn more.

I out = I in 10 μ a ( λ ) d ,
D = log ( I out I in ) = μ a ( λ ) d .
μ a ( λ ) = [ s ε HbO 2 ( λ ) + ( 1 s ) ε Hb ( λ ) ] c ,
D ( λ 1 ) = [ s ε HbO 2 ( λ 1 ) + ( 1 s ) ε Hb ( λ 1 ) ] c d , D ( λ 2 ) = [ s ε HbO 2 ( λ 2 ) + ( 1 s ) ε Hb ( λ 2 ) ] c d .
Φ ( λ 1 , λ 2 ) = D ( λ 2 ) D ( λ 1 ) .
s = Φ ε Hb ( λ 1 ) ε Hb ( λ 2 ) Δ λ 2 Φ Δ λ 1 ,
D = log R .
I out = I in ( λ ) 10 μ a ( λ ) d = R ( λ ) I 0 ( λ ) 10 μ a ( λ ) d ,
D ( λ ) = μ a ( λ ) d = [ s ε HbO 2 ( λ ) + ( 1 s ) ε Hb ( λ ) ] c d .
s = Ψ ε Hb ( λ 1 ) ε Hb ( λ 2 ) Δ λ 2 Ψ Δ λ 1 ,
D ( λ ) = μ a ( λ ) d + B ( λ ) ,
I ¯ out ( λ 0 ) = Ω d λ R ( λ ) S ( λ ) I 0 F 0 ( λ ) 10 [ μ a ( λ ) d + B ( λ ) ] ,
I ¯ in ( λ 0 ) = Ω d λ R ( λ ) S ( λ ) I 0 F 0 ( λ ) .
D ¯ ( λ 0 ) = log [ Ω d λ R ( λ ) S ( λ ) F 0 ( λ ) 10 [ μ a ( λ ) d + B ( λ ) ] Ω d λ R ( λ ) S ( λ ) F 0 ( λ ) ] = : [ s ε ¯ HbO 2 ( λ 0 , d ) + ( 1 s ) ε ¯ Hb ( λ 0 , d ) ] c d + B ( λ 0 ) ,
ε ¯ HbO 2 ( λ 0 , d ) = 1 c d ( log [ Ω d λ R ( λ ) S ( λ ) F 0 ( λ ) 10 [ ε HbO 2 ( λ ) c d + B ( λ ) ] Ω d λ R ( λ ) S ( λ ) F 0 ( λ ) ] + B ( λ 0 ) ) , ε ¯ Hb ( λ 0 , d ) = 1 c d ( log [ Ω d λ R ( λ ) S ( λ ) F 0 ( λ ) 10 [ ε Hb ( λ ) c d + B ( λ ) ] Ω d λ R ( λ ) S ( λ ) F 0 ( λ ) ] + B ( λ 0 ) ) ,
ε ¯ HbO 2 ( λ 0 , d ) = 1 c d ( log [ Ω d λ S ( λ ) F 0 ( λ ) 10 ε HbO 2 ( λ ) c d Ω d λ S ( λ ) F 0 ( λ ) ] ) , ε ¯ Hb ( λ 0 , d ) = 1 c d ( log [ Ω d λ S ( λ ) F 0 ( λ ) 10 ε Hb ( λ ) c d Ω d λ S ( λ ) F 0 ( λ ) ] ) .
F 0 ( λ ) = N exp [ ( λ λ 0 ) 2 2 σ 2 ] ,
σ = w 2 8 ln 2 .
s = α t + β ,
Δ s ¯ : = 1 M i = 1 M Δ s ( t i ) = 1 M i = 1 M | ( α 1 ) t i + β | ,
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.