Aberration measurement and correction on a large field of view in fluorescence microscopy

Open Access Open Access

Abstract

The aberrations induced by the sample and/or by the sample holder limit the resolution of optical microscopes. Wavefront correction can be achieved using a deformable mirror with wavefront sensorless optimization algorithms but, despite the complexity of these systems, the level of correction is often limited to a small area in the field of view of the microscope. In this work, we present a plug and play module for aberration measurement and correction. The wavefront correction is performed through direct wavefront reconstruction using the spinning-pupil aberration measurement and controlling a deformable lens in closed loop. The lens corrects the aberrations in the center of the field of view, leaving residual aberrations at the margins, that are removed by anisoplanatic deconvolution. We present experimental results obtained in fluorescence microscopy, with a wide field and a light sheet fluorescence microscope. These results indicate that detection and correction over the full field of view can be achieved with a compact transmissive module placed in the detection path of the fluorescence microscope.

© 2021 Optical Society of America under the terms of the OSA Open Access Publishing Agreement

1. Introduction

Adaptive Optics is a technology used to improve image quality when the optical system is subject to phase aberrations. In optical microscopy the aberrations can be induced by imperfections and misalignments of the illumination or detection optics. This becomes particularly relevant when considering the continuous development of new microscopy configurations and measurement protocols, such as clearing methods [1,2] or customized sample holders [3,4]. On the other hand, the inhomogeneities in the refractive index of the sample itself, can induce aberrations and reduce the contrast of the formed image. In order to perform high resolution and super-resolution imaging, a good knowledge of these aberrations and their correction are required.

Recently, many techniques have been developed to compensate for optical aberrations in microscopy [5,6]. They either rely on the use of a deformable mirror or on the use of a liquid crystal spatial light modulator placed in the pupil, or in its image plane. An approach similar to astronomical adaptive optics with a closed loop control is not convenient in microscopy, because the samples might not have bright point sources for the wavefront sensor. For this reason, wavefront sensorless control has become a standard for AO in microscopy. Remarkable achievements have been obtained with two techniques: image sharpening [6,7] and pupil segmentation [8,9]. Image sharpening uses a merit function that displays a maximum when the aberrations are minimized. With this assumption, it is easy to apply some bias aberrations to estimate the wavefront error using an optimization algorithm and then to compensate for them. Pupil segmentation is based on a different approach: by means of a liquid crystal panel, the pupil is segmented, and the phase of each segment is tuned, maximizing the constructive interference, thus resulting in an increased signal intensity.

Although adaptive optics has demonstrated its benefits, several limitations are still hindering its widespread diffusion. The level of the correction depends on the sample structure and it is not measurable, together with the correction being limited to the isoplanatic patch, which is often smaller than the field of view. In addition, microscopes must be entirely redesigned to include adaptive optics, adding complexity and limiting its adoption in commercial systems.

Some of these problems have been recently approached, but a solution that deals with them together is not yet available. For example [1012], propose methods that strongly simplify the system using an adaptive lens, instead of a deformable mirror, combined with optimization algorithms. This has brought to promising results in microscopy and ophthalmic imaging [7], [13].

The problem of limited isoplanatic correction has been addressed with different techniques. Laslandes et al. [14], and Popovic et al. [15] apply multi-conjugate adaptive optics, using two deformable mirrors. Other approaches, such as anisoplanatic correction with liquid crystals, promise to be effective [16]. Furthermore, the dependence on the sample structure and contrast has been analyzed and solved adding high spatial frequencies in the illumination path [17].

In this paper, we present a method to achieve anisoplanatic aberration correction over the entire field of view of a widefield microscope. The method consists in a compact optical module, able to detect and compensate for aberrations. The detection is based on spinning-pupil aberration measurement (SPAM) that reconstructs the wavefront phase in the different locations of the field of view [18]. This device controls an adaptive lens based on multiple piezoelectric actuators [10], which is installed into the same chassis of the SPAM module in a compact device. The adaptive lens corrects the aberrations in one point in the field of view. The remaining residual aberrations, still present in the other regions of the field of view, are removed by deconvolution. This is possible because the corresponding point spread function is known, as the wavefront phase has been detected by SPAM over the entire sample. The pre-compensation with the deformable lens is of fundamental importance in this method to increase the signal to noise ratio and to improve the quality of the images obtained with the deconvolution reducing the generation of artifacts.

2. Results and discussion

2.1 Integrated wavefront aberration measurement and adaptive optics correction

The method consists in a device that incorporates both aberration detection and correction, placed in the proximity of a widefield microscope pupil (Fig. 1). From ray tracing simulation we saw that the AL location is close enough to the Exit Pupil plane so that we do not see any significant impact on the measurement and correction process [Section 5 Supplement 1]. The detection is performed using the spinning pupil aberration measurement (SPAM) device that, by moving an aperture inside the pupil and measuring the relative shift of each formed image, reconstructs the wavefront phase [18]. The correction is obtained using an adaptive lens (AL) made of ultrathin glass membranes bended by multiple piezoelectric actuators [10]. This lens is compact and can be placed directly in the detection path of the optical microscope which is, in this case, adjacent to the spinning sub-aperture module. The AL is controlled by the SPAM with a closed loop, as in typical adaptive optics applications [19] (see Methods section). The control system computes the actuators control voltages that minimize the wavefront deviation from flat using an integrative controller.

 figure: Fig. 1.

Fig. 1. Optical layout of the imaging system used in this experiment including the SPAM module in combination with an adaptive lens. The diagram shows the SPAM module in the full pupil position (c) and in the sub-pupil position (d). The microscope has been used in wide field (a) and light sheet microscopy (b) configurations, as illustrated in the insets.

For more detailed schematics on the lens and the SPAM system see [10] and [18]

Download Full Size | PDF

We tested this method in a wide field fluorescence microscope. The schematic of the system is shown in Fig. 1 (fluorescence filters are not shown). The module (18 actuators Adaptive Lens with 10mm pupil, AOL1810 Dynamic Optics srl and spinning pupil) is placed in proximity of the back aperture of the objective lens (Mitutoyo, NA=0.28, 10X, long working distance). The effective Numerical Aperture of the system is 0.24, limited by the lens diameter. We imaged fluorescence sample slides (Fluocell sample prepared slice #3), with a mouse kidney section labeled with Alexa Fluor 488 WGA, with a phase plate, made of 1mm thick flexible transparent polystyrene plate, placed in front of it to induce further controlled amount of aberrations. In the first experiment, the lens starts in flat position. We measured the wavefront, aberrated by the phase plate, in 9 × 9 positions (Fig. 2).

 figure: Fig. 2.

Fig. 2. Comparison between the uncorrected and the AL corrected image, with correction in different region. Each raw shows an image of the sample, the RMS error (wavelength = 530 nm) and the wavefront phase aberrations represented as interferograms in a 9 × 9 grid. (a) Aberrated image without AL. (b) Corrected with AL in the central field. (c) Corrected with AL in the right-top corner. The correction region is indicated by the blue square. Scale bar is 200 µm.

Download Full Size | PDF

The wavefront aberrations vary field to field with an average value of 0.78 waves, root mean squared (RMS), at 530 nm. In second place, we corrected the wavefront in a single point, at the center of the of the image. Then, we repeated the measurement correcting in the top-right corner. In both cases, the results (summarized in Table 1) indicate that the adaptive lenses can correct the aberration almost completely over the correction area. However, only a small portion in the nearby of the corrected area respects the Marechal criterion for a well corrected system (RMS error of the aberration < λ/14 waves RMS). Nevertheless, the whole image benefits from the wavefront correction in both experiments. The average aberration error on the whole image was 0.33 waves RMS (central point) and 0.3 waves RMS (top-right point), while the portion of image being well corrected is 3.7% and 8.6% respectively.

Tables Icon

Table 1. Comparison of the correction parameter for the wide field fluorescence microscope. The table reports the area (%) of the image below the Marechal criterion, the residual wavefront error for different corrected image segment (center and top-right, highlighted in blue in Fig. 2) and the average wavefront error on the whole image.

Therefore, we conclude that the adaptive lens, in combination with the SPAM, allows one to correct the aberration in a chosen region of the sample, but residual aberration remains (in this case, mainly defocus and astigmatism), away from the corrected location.

2.2 Aberrations measurement and correction in a light-sheet fluorescence microscope

Being able to measure and correct aberrations in a wide field configuration, the proposed device fits particularly well in the detection path of a Light Sheet Fluorescence Microscope (LSFM) [20]. This microscopy technique is based on the excitation of a single plane (light sheet) of a three-dimensional fluorescent sample. Widefield detection optics are used to collect light orthogonally to the excitation plane. The selective illumination of a single plane provides intrinsic optical sectioning capability, offering a powerful tool for the three-dimensional imaging of biological samples. LSFM is particularly well suited for imaging large, chemically cleared specimens, such as entire small animal’s organs and tissues. When imaging chemically cleared samples the mm to cm-sized, specimens are placed in a high refractive index medium. This refractive index is approximately n≈1.4 for samples cleared in gel-based media, such as Clarity [2], and n≈1.5 for samples cleared in solvents, such as Benzyl Benzoate (BABB) and dibenzyl-ether (DBE) [21]. In many LSFM configuration the sample is placed in a glass cuvette, since the solvents would damage standard liquid immersion objective-lenses.

To demonstrate that light sheet imaging thought the cuvette is possible with standard, long-working distance microscope objective lenses, we used a LSFM microscope based on illumination with a cylindrical lens (see Methods section). The aberration correction module was placed in the detection arm of the microscope, right after the objective lens. Nonetheless, the light sheet thickness was large (the beam waist was c.a. 7 µm) so to have a uniform light sheet thickness over the entire field of view, with negligible illumination aberrations. We initially tested the module on fluorescence nano-beads (Estapor F1-XC 010). These polystyrene spheres have an average diameter of 160nm, largely below the detection diffraction limit, and two maxima fluorescence peaks at 525nm and 560nm. A beads solution in water was embedded in a 1.5% phytagel in water solution and then poured in a Fluorinated Ethylene Propylene (FEP) tube, which has a refractive index close to that of water. For imaging, the tube was inserted in an Optical Glass square cuvette filled with distilled water (thickness of 2cm). The system showed a strong spherical aberration induced by the interfaces between the air, the glass and the water contained in the cuvette. The aberration was stronger for off-axis fields, with a coma component. A summary of the wavefront measurements is shown in Table 2. The wavefront correction applied only in the central point led to a very small residual error (0.01 waves) while it left a non-negligible coma component in the image corners (0.23waved RMS). The beads profile, full width at half maximum (FWHM) was reduced from 1.79 µm to 1.35 µm showing a considerable improvement in resolution. The beads profile in the corner of the image decreases to 1.57 µm. This result confirms what observed in the widefield measurement of the microscope slide: the correction is effective in one point (at the center) but leaves significant aberrations at the margin of the image.

Tables Icon

Table 2. Wavefront data relative to the fluorescent nano-beads acquisition.

We repeated the experiment on a biological sample, a chemically cleared adult zebrafish brain immersed in DBE. This sample shows strong autofluorescence, primarily in the blood vessels. The results are shown in Fig. 3. The figure panels show the magnified details of two sample regions corresponding to the brain Optic Tectum (center) and to the Diencephalon (bottom-left).

 figure: Fig. 3.

Fig. 3. Light sheet microscopy acquisition of a chemically cleared adult zebrafish brain section, without and with AL correction. (a) Full image uncorrected (top-left triangle) vs corrected with the adaptive lens (bottom-right). (b) Wavefront aberration in the central region (blue in panel a). (c) Wavefront aberration at the left-bottom concern (red). (d-g) Magnified regions uncorrected (d, f) and corrected with the adaptive lens (e, g). Scalebar is 200 µm in a and 40 µm in b.

Download Full Size | PDF

For a fair comparison of the quality of the acquired images, we initially removed any adaptive lenses from the microscope (Fig. 3(a)). Then, we inserted the adaptive lenses for wavefront correction. As shown in Fig. 3(c)-(d), the correction in the central field leads to a clear contrast improvement over the overall image. The wavefront measurements (see Table 3) show that the aberration is almost completely corrected in the central field (0.05 waves RMS) while residual coma is present in the corners (0.27 waves RMS). Even if the fluorescence background is decreased in the entire field of view, improving the contrast of the acquired image, the resolution is not enhanced at the corners of the sample. An example is presented in Fig. 3(f)-(g) (showing a magnified detail of the red square in Fig. 3(a)), where the correction with a single adaptive optics element is not sufficient to improve the resolution at the margins of the sample, and some brain vessels appear even more blurred that in the uncorrected case. A further strategy is required, to extend the correction to the entire field of view.

Tables Icon

Table 3. Wavefront error data relative the zebrafish brain sample (Fig. 3), corrected in the central region.

2.3 Correction over an extended area

Altogether, the previous LSFM measurement suggest that the aberration contribution primarily comes from the refractive index mismatch, given by the cuvette. Its correction with a single adaptive optics element leaves a residual field curvature and coma, oriented towards the image center. This constitutes a general problem in the field of adaptive optics: the corrected area is limited to a small region, located around the correction point. Its size, called isoplanatic patch, depends on the volumetric aberration [16], [22,23]. In astronomy, this limitation has been overcome by correcting the wavefront with a series of deformable mirrors in the so called Multi Conjugate Adaptive Optics (MCAO) [21] configuration. However, its implementation in microscopy has been so far inconvenient, due to the difficulties in using more deformable mirrors together with the lack of a wavefront measurement system [14]. We have discussed in the previous section how, by using a single AL element in a microscope, we are able to compensate for low-order aberrations that are present over the entire imaging plane (typically, spherical aberration). However, residual aberrations remain in the lateral fields. This would require a sequential correction of each FOV region combined with a post processing stitching [24]. The sequential correction would increase the total correction time with a consequential increased risk of photobleaching or damage to the sample. On the other hand, in Section 4 of the Supplement 1, we demonstrated that a pure software correction does not provide good performance when aberrations increase substantially (Supplement 1). Additionally, the improvement provided by the AL always surpasses the quality of the RL deconvolution (Supplement 1). By mixing hardware and software, we find a balance between setup complexity and image quality, still retaining the ability to correct the images over the entire detection plane.

To this aim, we decided to put in place a multi-region anisoplanatic deconvolution [18]. We started from the measurement of the wavefront phase, which is provided by SPAM in multiple subregions over the entire image. In Fig. 4 the phases are shown in different areas (5 × 5 regions are presented), along with the measured RMS wavefront error. For each region, we calculated the 2D Fourier Transform of the wavefront and we took its squared value, to generate the corresponding point spread function (PSF). Figure 4(d) shows that the PSF has a typical comet-like shape with an elongation that radially increases around the correction region. The acquired image was then divided in tiles, deconvolved with the corresponding PSF (with a custom Richardson-Lucy iterative approach), and recomposed to form a single, higher resolution, image [18].

 figure: Fig. 4.

Fig. 4. (a) LSFM acquisition of a zebrafish brain with AL correction in the image centre. (b) Wavefront measured in 5X5 subregions of the image. (c) RMS wavefront error (d) PSF generated for each subregion.

Download Full Size | PDF

This deconvolution strategy successfully reconstructs the image in the entire field of view (Fig. 5) with improved resolution and contrast, allowing the identification of single micro-vessels. Taking into consideration the center of the image (Fig. 5(c)), already corrected by the adaptive lens, we observe that the deconvolution does not significantly vary the image quality in this region. The strength of the deconvolution is instead observable in the outer regions of the field of view (Fig. 5(b) and Fig. 5(d)), where the PSFs exhibit strongly asymmetric and elongated shapes.

 figure: Fig. 5.

Fig. 5. Comparison between hardware correction with adaptive lens (AL) and the combination of hardware correction with anisoplanatic deconvolution (ANI). AL corrected is shown in the top left triangles, AL corrected with ANI deconvolution is shown in the bottom right triangle. (a) Full image of a zebrafish brain plane (b) Magnified region in the top left corner. (c) Magnified region in the center. (d) Magnified region in the bottom right corner. Scalebar is 200 µm in a and 50 µm in b-d.

Download Full Size | PDF

Each PSF, in fact, represents how a single point-source would appear in that given region: deconvolving with a wrong PSF may lead to the production of artifact. As an intuitive example, let us imagine to deconvolve the signal emitted by a bead located at the extreme corner of the image (as in Fig. 4(d)) with the -wrong- symmetric kernel measured in the central part. This operation would enhance “features” of the bead-comet, but it will not remove its peculiar structure from the image, producing artifacts instead.

We report the presence of horizontal lines in the deconvolved image, particularly in Fig. 5(d). These are given by the known shadowing artifacts typical of light sheet microscopy, which have not been removed in this setup: light sheet is randomly attenuated along the propagation (light propagates from left to right in Fig. 5) by the absorbing and scattering sample, generating the lines. More importantly the visibility of the vessels and microvasculature is improved. We estimate the improvement in the image quality by calculating the contrast, here defined as the standard deviation of the intensity distribution of the image with respect to its mean. The original image has a contrast of 0.8 that, after isoplanatic deconvolution, increases to 0.83 Instead, with the anisoplanatic deconvolution, we achieve the increased value of 0.93. The image is also affected by a significant distortion that, at the corners of the object corners, reaches 1.0 µm (see Supplement 1). The anisoplanatic deconvolution effectively compensates for this deformation too (Supplement 1). It is worth noting that a standard isoplanatic deconvolution would not successfully correct the distortion introduced in the image (Supplement 1): the knowledge of the region-dependent aberrations and the anisoplanatic deconvolution play a central role in high contrast and high-resolution reconstruction.

3. Materials and methods

3.1 Adaptive optics

We used a Multi Actuator adaptive lens, similar to the one described in [10]. This lens relies on the use of ultrathin glass membranes bended by piezoelectric actuators. The lens is composed by two glass membranes with the inner space filled with a transparent liquid. It mounts two piezoelectric actuator rings, one on each face. Each ring is then segmented in 9 parts for a total of 18 singularly addressable actuators, so that it can generate aberrations up to the 4th order Zernike polynomials. The optical performances of the lens used in these experiments are reported in the Supplementary Text.

3.2 Wavefront measurement and closed loop control

The wavefront measurement system is composed by a spinning wheel that, rotating, scans the pupil with an iris of 2.5 mm. The scan happens on 4 radial distances, with the angular positions of the iris being designed to avoid overlaps. The total wavefront points are 18. The system automatically adjusts the exposure time proportionally to the ratio R of the area of the sub-apertures and of the whole pupil (R = 16). For the zebrafish samples, the exposure time with the fully open pupil was 40 ms and 640 ms for the sub-apertures. Hence, the total time for wavefront measurement was about 12 seconds. The measurement process is repeated 3 times during the correction close loop; hence the total correction time is about 36s. The spinning wheel is equipped with a large aperture used for image acquisition (Fig. 1) which only slightly limits the detection numerical aperture. The wavefront measurement was also used for the characterization of the deformable lenses necessary to operate the closed loop control. The characterization was performed poking each actuator individually and storing their wavefront deformation in a matrix (the so-called influence functions matrix). This operation was performed before the experiment using a phantom target in 9 positions of the field of view arranged in 3 × 3 equally spaced points. Then, in the correction process, we could select the positions used for the wavefront correction. It's worth noting that the aberration does depend on wavelength. Thus, the correction is valid only for a specific wavelength, otherwise the correction process must be run again.

3.3 Light-sheet fluorescence microscope

Light sheet fluorescence microscopy was performed on a custom system, similar that described in [25]. The system was based on a diode-pumped solid-state laser emitting at 473 nm, coupled in a single-mode optical fiber that delivered the light to the sample. The laser beam was collimated to a 2 mm waist and propagated through a cylindrical lens (f = 50 mm). The illumination power was 1.2 mW. A 10x, long working distance objective (Mitutoyo, NA = 0.28, WD = 34 mm) was used to collect the fluorescence signal. The objective, in combination with a tube lens and a long pass filter (FELH500, Thorlabs), formed the image on a CMOS camera (ORCA Flash 4.0, Hamamatsu), with a field of view of ca. 11mm×11mm. The sample was held in a glass cuvette of size 20mm x 20mm and it was moved with a 3-axis stage.

3.4 Anisoplanatic deconvolution

The SPAM permits to measure how the PSF changes within different regions of the image. To exploit this accurate information, we decided to address an anisoplanatic deconvolution problem [18]. To begin, each acquired image is divided into overlapping squared regions (tiles). For our purpose, we chose the tile dimension to be $W \times W$, where $W = 410px$. Each tile crops a portion of the image, and it is translated in steps of 82 pixels along both dimensions. By doing so, we obtain a total of $N \times N\textrm{\; }$tiles, with $N = 21$. The number of tiles should be equal to the number of PSFs that we are able to estimate. Thus, we interpolate the $5 \times 5$ PSF map to match the number of total tiles of the image, $21 \times 21$. Then, each tile is deconvolved by its corresponding PSF with a custom Richardson-Lucy (RL) algorithm, implemented in python language. In our case, the RL is written by using convolution functions from the PyTorch library. For each image reconstruction reported in the article, we run $50$ RL-iterations. Once done, we recompose the final image by placing each deconvolved tile in its appropriate location, removing an external canvas of $32\; px$ to avoid the import of boundary artifacts. To reassemble the image, the overlapping regions of the tiles where averaged by counting the number of tile contributions in each single pixel. This produces the reconstruction of the anisoplanatically deconvolved image. The whole RL correction described requires about $2\; s$ with an nVidia Titan RTX equipped with 24Gb of RAM.

3.5 Zebrafish brain

The adult zebrafish brain was prepared as follows (in accordance with the protocol n. 513/2018-PR, authorized on 4/7/2018 by the Italian Health Ministry). The fish was sacrificed by over-anaesthesia in a tricaine solution of 500 mg/l. When opercular movements stopped and all reflex reactivity was lost, the fish was rapidly placed in a cut made in a sponge within a Petri dish containing chilled tricaine solution. The brain was quickly and carefully removed and placed overnight in 4% paraformaldehyde on a shaker (at 4°C). The brain was subsequently processed as described in [26].

To reveal the origin of the moderate/high autofluorescence recorded at brain level we performed fixation by perfusion. The fish was placed in a tricaine solution (150 mg/l). When opercular movements stopped and all reflex reactivity was lost, the fish was placed in a sponge (the Petri dish contained 150 mg/l chilled tricaine solution). The heart was exposed, and a smoothed 30 Gauge needle inserted into the ventricle. A saline solution was flushed (1-2 ml, at 1 ml/min) and when fish gills whitened, a fixative solution (10 ml) was used to perfuse. The brain was then carefully removed and placed overnight in 2% paraformaldehyde on a shaker (at 4°C). The brain was subsequently processed as described above. Light sheet fluorescence microscopy revealed the absence of autofluorescence in the brain (not shown), indicating that autofluorescence derived from blood-containing brain vessels.

4. Conclusions

We have demonstrated that the use of deformable lenses in combination with anisoplanatic deconvolution increases the size of the corrected area on aberrated microscopy images. This allowed us to combine the deformable lens with the aberration detection system, based on a spinning sub-pupil, that enables wavefront measurement directly from the detected image. The SPAM system provides information about the aberrations in every point of the field of view. This is particularly important because it gives quantitative information about the aberration structure generated by the sample. Furthermore, in the presented cases, we evaluated the size of the isoplanatic patch after wavefront correction. It is worth noting that, with respect to all the literature relative to wavefront sensorless aberration correction in microscopy, we were able to measure the size of the well corrected patch in the case of the aberration correction with a single wavefront modulator. The results show that, even if the part of the image with diffraction limited features is extremely small (from 4% to 8%, Table 1), on average, the whole image benefits from the correction (from 0.78 waves RMS down to about 0.3 waves RMS, Table 1). The fact that the wavefront phase is measured over the entire field of view has another major impact. The residual aberration, outside the isoplanatic patch, can be quantified and corrected by deconvolution, providing high resolution reconstruction over the entire field of view. However, it is worth mentioning that, since the AL corrects aberrations over a localized region, repeating the hardware correction sequentially over the entire field of view could potentially improve the reconstructions. Unfortunately, this step necessitates acquiring multiple images, one for each corrected patch, leading to photobleaching and needs an ad-hoc image fusion scheme to recompose the enhanced image. Anisoplanatic deconvolution, instead, is much faster and can be applied directly on a single static image. Nevertheless, this is a significant aspect that we plan to investigate further in the follow-up studies. Importantly, our combined method of using hardware and computational correction leads to a strong improvement of the image quality and reduction of artifacts with respect to pure computational methods (see Supplement 1 “Deconvolution and hardware correction comparison”)

Finally, the measurement and correction of aberrations using the integrated SPAM and deformable lens module, can be of easy installation on many types of microscopes (Fig. 5 Supplement 1).

Funding

Office of Naval Research Global (GRANT12789919); H2020 Marie Skłodowska-Curie Actions (799230); Laserlab-Europe (871124).

Acknowledgments

CNR-IFN was supported by Office of Naval Research global (ONRG) and by Air Force Research Laboratory (AFRL) with grant GRANT12789919.

Politecnico di Milano has received funding from LASERLAB-EUROPE (grant agreement no. 871124, European Union’s Horizon 2020 research and innovation programme) and from H2020 Marie Skłodowska-Curie Actions (HI-PHRET project, 799230).

We are grateful to Dr Giulia Petrillo for helping in the zebrafish brain preparation.

Disclosures

S. Bonora acknowledges a financial interest in Dynamic Optics srl.

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Supplemental document

See Supplement 1 for supporting content.

References

1. A. Ertürk and F. Bradke, “High-resolution imaging of entire organs by 3-dimensional imaging of solvent cleared organs (3DISCO),” Exp. Neurol. 242, 57–64 (2013). [CrossRef]  

2. K. Chung, J. Wallace, S. Y. Kim, S. Kalyanasundaram, A. S. Andalman, T. J. Davidson, J. J. Mirzabekov, K. A. Zalocusky, J. Mattis, A. K. Denisin, S. Pak, H. Bernstein, C. Ramakrishnan, L. Grosenick, V. Gradinaru, and K. Deisseroth, “Structural and molecular interrogation of intact biological systems,” Nature 497(7449), 332–337 (2013). [CrossRef]  

3. A. Kaufmann, M. Mickoleit, M. Weber, and J. Huisken, “Multilayer mounting enables long-term imaging of zebrafish development in a light sheet microscope,” Dev. 139(17), 3242–3247 (2012). [CrossRef]  

4. R. Tomer, M. Lovett-Barron, I. Kauvar, A. Andalman, V. M. Burns, S. Sankaran, L. Grosenick, M. Broxton, S. Yang, and K. Deisseroth, “SPED light sheet microscopy: fast mapping of biological system structure and function,” Cell 163(7), 1796–1806 (2015). [CrossRef]  

5. N. Ji, “Adaptive optical fluorescence microscopy,” Nat. Methods 14(4), 374–380 (2017). [CrossRef]  

6. M. J. Booth, “Adaptive optical microscopy: the ongoing quest for a perfect image,” Light Sci. Appl. 3(4), e165 (2014). [CrossRef]  

7. H. R. G. W. Verstraete, M. Heisler, M. J. Ju, D. Wahl, L. Bliek, J. Kalkman, S. Bonora, Y. Jian, M. Verhaegen, and M. V. Sarunic, “Wavefront sensorless adaptive optics OCT with the DONE algorithm for in vivo human retinal imaging [Invited],” Biomed. Opt. Express 8(4), 2261 (2017). [CrossRef]  

8. N. Ji, D. E. Milkie, and E. Betzig, “Adaptive optics via pupil segmentation for high-resolution imaging in biological tissues,” Nat. Methods 7(2), 141–147 (2010). [CrossRef]  

9. D. J. Wahl, C. Huang, S. Bonora, Y. Jian, and M. V. Sarunic, “Pupil segmentation adaptive optics for invivo mouse retinal fluorescence imaging,” Opt. Lett. 42(7), 1365–1368 (2017). [CrossRef]  

10. S. Bonora, Y. Jian, P. Zhang, A. Zam, E. N. Pugh, R. J. Zawadzki, and M. V. Sarunic, “Wavefront correction and high-resolution in vivo OCT imaging with an objective integrated multi-actuator adaptive lens,” Opt. Express 23(17), 21931 (2015). [CrossRef]  

11. P. Pozzi, M. Quintavalla, A. B. Wong, J. G. G. Borst, S. Bonora, and M. Verhaegen, “Plug-and-play adaptive optics for commercial laser scanning fluorescence microscopes based on an adaptive lens,” Opt. Lett. 45(13), 3585–3588 (2020). [CrossRef]  

12. K. Banerjee, P. Rajaeipour, Ç Ataman, and H. Zappe, “Optofluidic adaptive optics,” Appl. Opt. 57(22), 6338–6344 (2018). [CrossRef]  

13. M. Skorsetz, P. Artal, and J. M. Bueno, “Performance evaluation of a sensorless adaptive optics multiphoton microscope,” J. Microsc. 261(3), 249–258 (2016). [CrossRef]  

14. M. Laslandes, M. Salas, C. K. Hitzenberger, and M. Pircher, “Increasing the field of view of adaptive optics scanning laser ophthalmoscopy,” Biomed. Opt. Express 8(11), 4811 (2017). [CrossRef]  

15. Z. Popovic, J. Thaung, P. Knutsson, and M. Owner-Peterse, “Dual conjugate adaptive optics prototype for wide field high resolution retinal imaging,” in Adative Optics Progress (IntechOpen, 2012), 3–22.

16. P. Pozzi, C. Smith, E. Carroll, D. Wilding, O. Soloviev, M. Booth, G. Vdovin, and M. Verhaegen, “Anisoplanatic adaptive optics in parallelized laser scanning microscopy,” Opt. Express 28(10), 14222 (2020). [CrossRef]  

17. M. Žurauskas, I. M. Dobbie, R. M. Parton, M. A. Phillips, A. Göhler, I. Davis, and M. J. Booth, “IsoSense: frequency enhanced sensorless adaptive optics through structured illumination,” Optica 6(3), 370–379 (2019). [CrossRef]  

18. D. Ancora, T. Furieri, S. Bonora, and A. Bassi, “ Spinning pupil aberration measurement for anisoplanatic deconvolution,” to be published in Optics Letters.

19. R. K. Tyson, Introduction to Adaptive Optics (SPIE, 2009).

20. O. E. Olarte, J. Andilla, E. J. Gualda, and P. Loza-Alvarez, “Light-sheet microscopy: a tutorial,” Adv. Opt. Photonics (2018).

21. R. Ragazzoni, E. Marchetti, and G. Valente, “Adaptive-optics corrections available for the whole sky,” Nature 403(6765), 54–56 (2000). [CrossRef]  

22. J. Li, D. R. Beaulieu, H. Paudel, R. Barankov, T. G. Bifano, and J. Mertz, “Conjugate adaptive optics in widefield microscopy with an extended-source wavefront sensor,” Optica 2(8), 682 (2015). [CrossRef]  

23. J. Mertz, H. Paudel, and T. G. Bifano, “Field of view advantage of conjugate adaptive optics in microscopy applications,” Appl. Opt. 54(11), 3498–3506 (2015). [CrossRef]  

24. P. Rajaeipour, A. Dorn, K. Banerjee, H. Zappe, and Ç Ataman, “Extended field-of-view adaptive optics in microscopy via numerical field segmentation,” Appl. Opt. 59(12), 3784–3791 (2020). [CrossRef]  

25. L. Maiuri, A. Candeo, I. Sana, E. Ferrari, C. D’Andrea, G. Valentini, and A. Bassi, “Virtual unfolding of light sheet fluorescence microscopy dataset for quantitative analysis of the mouse intestine,” J. Biomed. Opt. 21, 056001 (2016). [CrossRef]  

26. A. Ertürk, K. Becker, N. Jährling, C. P. Mauch, C. D. Hojer, J. G. Egen, F. Hellal, F. Bradke, M. Sheng, and H. U. Dodt, “Three-dimensional imaging of solvent-cleared organs using 3DISCO,” Nat. Protoc. 7(11), 1983–1995 (2012). [CrossRef]  

Supplementary Material (1)

NameDescription
Supplement 1       Supplementary material

Data availability

Data underlying the results presented in this paper are not publicly available at this time but may be obtained from the authors upon reasonable request.

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (5)

Fig. 1.
Fig. 1. Optical layout of the imaging system used in this experiment including the SPAM module in combination with an adaptive lens. The diagram shows the SPAM module in the full pupil position (c) and in the sub-pupil position (d). The microscope has been used in wide field (a) and light sheet microscopy (b) configurations, as illustrated in the insets.
Fig. 2.
Fig. 2. Comparison between the uncorrected and the AL corrected image, with correction in different region. Each raw shows an image of the sample, the RMS error (wavelength = 530 nm) and the wavefront phase aberrations represented as interferograms in a 9 × 9 grid. (a) Aberrated image without AL. (b) Corrected with AL in the central field. (c) Corrected with AL in the right-top corner. The correction region is indicated by the blue square. Scale bar is 200 µm.
Fig. 3.
Fig. 3. Light sheet microscopy acquisition of a chemically cleared adult zebrafish brain section, without and with AL correction. (a) Full image uncorrected (top-left triangle) vs corrected with the adaptive lens (bottom-right). (b) Wavefront aberration in the central region (blue in panel a). (c) Wavefront aberration at the left-bottom concern (red). (d-g) Magnified regions uncorrected (d, f) and corrected with the adaptive lens (e, g). Scalebar is 200 µm in a and 40 µm in b.
Fig. 4.
Fig. 4. (a) LSFM acquisition of a zebrafish brain with AL correction in the image centre. (b) Wavefront measured in 5X5 subregions of the image. (c) RMS wavefront error (d) PSF generated for each subregion.
Fig. 5.
Fig. 5. Comparison between hardware correction with adaptive lens (AL) and the combination of hardware correction with anisoplanatic deconvolution (ANI). AL corrected is shown in the top left triangles, AL corrected with ANI deconvolution is shown in the bottom right triangle. (a) Full image of a zebrafish brain plane (b) Magnified region in the top left corner. (c) Magnified region in the center. (d) Magnified region in the bottom right corner. Scalebar is 200 µm in a and 50 µm in b-d.

Tables (3)

Tables Icon

Table 1. Comparison of the correction parameter for the wide field fluorescence microscope. The table reports the area (%) of the image below the Marechal criterion, the residual wavefront error for different corrected image segment (center and top-right, highlighted in blue in Fig. 2) and the average wavefront error on the whole image.

Tables Icon

Table 2. Wavefront data relative to the fluorescent nano-beads acquisition.

Tables Icon

Table 3. Wavefront error data relative the zebrafish brain sample (Fig. 3), corrected in the central region.